Arm Electronic Sphygmomanometer is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the arm of which the circumference includes 22 cm~32 cm. It is intended to be used in hospital environment or at home.
Device Story
Battery-powered automatic non-invasive blood pressure monitor; uses oscillometric technique to measure systolic/diastolic blood pressure and pulse rate. Device components: external cuff, analog circuit, MCU. Used in hospital or home by adults. User wraps cuff around upper arm; device performs automatic inflation/deflation; MCU processes pressure signals to calculate blood pressure and pulse rate. Results displayed on LCD; device includes data storage for review. Provides objective physiological data to assist healthcare providers or patients in monitoring cardiovascular status.
Clinical Evidence
Bench testing only. Evaluated for electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 80601-2-30), electromagnetic compatibility (IEC 60601-1-2, IEC 60601-1-6), performance (AAMI/ANSI/ISO 81060-2), battery safety (IEC 62133), biocompatibility (ISO 10993-5, ISO 10993-10), and usability (IEC 62366). Software verified and validated per FDA guidance.
Technological Characteristics
Battery-powered (AA alkaline or Li-ion) oscillometric blood pressure monitor. Components: inflatable cuff, analog circuitry, MCU. Dimensions/form factor: upper arm cuff (22-32 cm). Connectivity: standalone with LCD display and internal data storage. Sterilization: non-sterile. Software: embedded firmware for signal processing and measurement calculation.
Indications for Use
Indicated for adult individuals requiring non-invasive measurement of systolic and diastolic blood pressure and pulse rate via an upper arm cuff (22-32 cm circumference) in hospital or home settings.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 27, 2021
ShenZhen Changkun Technology Co., Ltd. % Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District Guangzhou, 510000 China
Re: K193387
Trade/Device Name: Arm Electronic Sphygmomanometer Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: December 27, 2020 Received: December 31, 2020
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193387
Device Name Arm Electronic Sphygmomanometer
#### Indications for Use (Describe)
Arm Electronic Sphygmomanometer is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the arm of which the circumference includes 22 cm-32 cm. It is intended to be used in hospital environment or at home.
Type of Use (Select one or both, as applicable)
| ☑ Own-The-Counter Use (21 CFR 201.66) |
|-------------------------------------------------|
| ☐ Prescription Use (Part 201 CFR 201 Subpart D) |
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Sponsor: | ShenZhen Changkun Technology CO., LTD. |
|-----------|------------------------------------------------------------------------------------------|
| Subject: | Arm Electronic Sphygmomanometer, Models: CK-A120,CK-A136,CK-A138,CK-A139,CK-A155,CK-A155 |
| Device: | A156, CK-A158, CK-A168 |
| Document: | FDA 510(k) Submission Report |
| Name: | |
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
## 1. Submitter's Information
510(k) Owner's Name: ShenZhen Changkun Technology CO., LTD. Establishment Registration Number: Applying Address: 801, 3 floor 4floor 6floor 7floor, B building, NO.69 ,zhenbi road, biling community, biling street, pingshan district, Shenzhen city, Guangdong, China Tel: +86-0755-29100487 Fax: +86-0755-23463480 Contact Person (including title): Steve Li (Manger Representative) E-mail: changkunkj@163.com
## Application Correspondent:
Contact Person: Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China Tel: +86 20 8266 2446 Email: regulatory@glomed-info.com
### 2. Subject Device Information
Type of 510(k): Traditional Common Name: Non-invasive blood pressure measurement systems Classification Name: System, Measurement, Blood-Pressure, Non-Invasive Trade Name:Arm Electronic Sphygmomanometer Model Name: CK-A120, CK-A136, CK-A138, CK-A155, CK-A156, CK-A156, CK-A158, CK-A168 Review Panel: Cardiovascular Product Code: DXN Regulation Number: 870.1130 Regulatory Class: 2
### 3. Predicate Device Information
| Sponsor | Shenzhen BSX Technology Electronics Co., Ltd. |
|--------------------------|-----------------------------------------------------------------------------------------------------|
| Device Name<br>and Model | Arm-type Electronic Blood Pressure Monitor<br>Model: BSX516, BSX523, BSX525, BSX583, BSX593, BSX595 |
| 510(k) Number | K183058 |
| Product Code | DXN |
{4}------------------------------------------------
ShenZhen Chanqkun Technoloqy CO., LTD. Sponsor: Subject Arm Electronic Sphygmomanometer, Models: CK-A120,CK-A136,CK-A138,CK-A139,CK-A155,CK-Device: A156. CK-A158.CK-A168 Document FDA 510(k) Submission Report Name:
| Regulation<br>Number | 870.1130 |
|----------------------|----------|
| Regulation<br>Class | II |
### 4. Device Description
The proposed devices are battery driven automatic non-invasive blood pressure Monitor. It is used for automatically conduct the inflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult at upper arm within its claimed range and accuracy by oscillometry technique. Devices are consisted of three main parts: external hardwares (such as cuff), analog circuit, and MCU. The devices have the data storage function in order for data reviewing, including the systolic pressure, diastolic pressure, pulse rate and measurement time.
The proposed devices are intended to be used in medical facilities or at home and provided non-sterile.
#### 5. Intended Use / Indications for Use
Arm Electronic Sphygmomanometer is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the arm of which the circumference includes 22 cm~32 cm. It is intended to be used in hospital environment or at home.
#### 6. Test Summary
Arm Electronic Sphygmomanometer has been evaluated the safety and performance by lab bench testing as following:
- � Electrical safety test according to IEC 60601-1, IEC 60601-1-11 and IEC 80601-2-30 standards
- Electromagnetic compatibility test according to IEC 60601-1-2 standard �
- � Electromagnetic compatibility test according to IEC 60601-1-6 standard
- � Performance according to AAMI / ANSI / ISO 81060-2 standard
- � Lithium battery report in accordance with IEC62133
- � Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standards
- � Usability test according to IEC62366 standard
- � Software verification and validation test according to the requirements of the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
#### 7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of Arm Electronic Sphygmomanometer is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Remark |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Remark |
| Company | ShenZhen Changkun Technology<br>CO., LTD. | Shenzhen BSX Technology<br>Electronics Co., Ltd. | -- |
| Product Name | Arm Electronic<br>Sphygmomanometer | Arm-type Electronic Blood Pressure<br>Monitor | -- |
| Model Name | CK-A120, CK-A136, CK-A138, CK-<br>A139, CK-A155, CK-A156, CK-<br>A158,CK-A168 | BSX516, BSX523,<br>BSX525, BSX583, BSX593 and<br>BSX595 | -- |
| Classification<br>Name | systems,measurement,blood<br>pressure, Non-invasive | systems,measurement,blood<br>pressure, Non-invasive | SE |
| 510(k) Number | Applying | K183058 | -- |
| Product Code | DXN | DXN | SE |
| | Intended Use and Indications for Use | | |
| Intended Use /<br>Indications for<br>Use | Arm Electronic<br>Sphygmomanometer is a non-<br>invasive blood pressure<br>measurement system intended to<br>measure the diastolic and systolic<br>blood pressures and pulse rate of<br>an adult individual by using a non-<br>invasive oscillometric technique in<br>which an inflatable CUFF is<br>wrapped around the arm of which<br>the circumference includes 22<br>cm~32 cm. It is intended to be<br>used in hospital environment or at<br>home. | The blood pressure monitor is a<br>non-invasive blood pressure<br>measurement system intended to<br>measure the diastolic and systolic<br>blood pressures and pulse rate of<br>an adult individual by using a non-<br>invasive oscillometric technique in<br>which an inflatable CUFF is<br>wrapped around the arm of which<br>the circumference includes 22<br>cm~32 cm. It is intended to be used<br>in hospital environment or at home. | SE |
| Electrical requirement | | | |
| Power Supply | 4x1.5V AA Alkaline Battery (CK-<br>A138, CK-A139)<br>4x1.5V AAA Alkaline Battery (CK-<br>A120, CK-A136)<br>3.7V 400mAh Li-ion Battery (CK-<br>A155, CK-A156, CK-A158, CK-<br>A168) | 4x1.5V AAA Alkaline Battery<br>(BSX516, BSX525, BSX583,<br>BSX593 and BSX595)<br>3.7V 400mAh Li-ion Battery<br>(BSX523) | SE<br>Note 1 |
| Performance specification | | | |
| Measurement Site | Upper Arm | Upper Arm | SE |
| Measuring range | Pressure: 0mmHg~299 mmHg:<br>Pulse: 40 bpm~199 bpm | Pressure: 0mmHg~299 mmHg<br>Pulse: 40 bpm ~180 bpm | SE<br>Note |
| Pressure<br>resolution | 1 mmHg or 0.1kPa | 1 mmHg or 0.1kPa | SE |
| Measuring<br>accuracy | Pressure: ± 3mmHg<br>Pulse: ±5% | Pressure: ± 3mmHg<br>Pulse: ±5% | SE |
| Cuff<br>Circumference | 22~32cm | 22~32cm | SE |
| Inflation and | Automatic | Automatic | SE |
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Remark |
| Deflation | | | SE |
| Measuring<br>Method | Non-invasive Oscillometric | Non-invasive Oscillometric | SE |
| Patient Population | Adult | Adult | SE |
| Display | LCD | LCD | SE |
| Operation<br>condition | Temperature: +5 to +40 °C,<br>Humidity: 15 to 90%<br>Atmospheric Pressure: 80 kPa~106 kPa | Temperature: +5 to +40 °C,<br>Humidity: 15 to 85%<br>Atmospheric Pressure: 80 kPa~106 kPa | SE<br>Note 2 |
| Transport/storage<br>environment | Temperature: -25°C to +70°C,<br>Humidity: ≤93%R.H.<br>Atmospheric Pressure: 70 kPa~106 kPa | Temperature: -20°C to +55 °C,<br>Humidity: 10 to 93%<br>Atmospheric Pressure: 70 kPa~106 kPa | SE<br>Note 2 |
| Safety | IEC 60601-1<br>IEC 60601-1-11<br>IEC 80601-2-30 | IEC 60601-1<br>IEC 60601-1-11<br>IEC 80601-2-30 | SE |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | SE |
| Biocompatibility | ISO 10993 | ISO 10993 | SE |
{5}------------------------------------------------
Sponsor: ShenZhen Changkun Technology CO., LTD. Arm Electronic Sphygmomanometer, Models: CK-A120,CK-A136,CK-A138,CK-A139,CK-A155,CK-Subject Device: A156, CK-A158,CK-A168 Document FDA 510(k) Submission Report Name:
{6}------------------------------------------------
ShenZhen Chanqkun Technoloqy CO., LTD. Sponsor: Subject Arm Electronic Sphygmomanometer, Models: CK-A120,CK-A136,CK-A138,CK-A139,CK-A155,CK-Device: A156, CK-A158,CK-A168 FDA 510(k) Submission Report Document Name:
## Comparison
### Note 1:
Although the "Power Supply" of the subject device is a little different from the predicate device, they all complied with the requirements of safety standards as IEC 60601-1, IEC 60601-1-2, IEC 80601-2-30 required. So this difference will not raise any safety or effectiveness issue,
### Note 2:
Although "Measuring scope", "Operation condition" and "Transport/storage environment" of the subject device are a little different from the predicate devices, but the difference of subject device is very similar with the K183058. So these parameters' differences will not raise any safety or effectiveness issue.
### Final Conclusion :
The subject device Non-invasive Sphygmomanometer (Arm Electronic Sphygmomanometer) (Model: CK-A120, CK-A136, CK-A138, CK-A139, CK-A155, CK-A158, CK-A168) has all features of the predicate device. The few differences do not affect the safety and effectiveness of the subject device. Thus, the subject device is substantially equivalent to the predicate device K183058.
### 8. Date of the summary prepared: December 27, 2020
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.