The T2 ICF System is intended for internal bone fixation.
Device Story
T2 ICF System is a cannulated intramedullary nail system for internal fixation and stabilization of foot instabilities and reconstructions. Device consists of sterile-packed titanium alloy (Ti6Al4V ELI) intramedullary nails with pre-loaded compression screws and endcaps, plus non-sterile targeting instrumentation. Used in conjunction with existing VariAx 2 bone screws for fixation. Operated by surgeons in clinical/OR settings. Nail design allows compression of bony fragments to facilitate healing. Benefits include stabilization of foot fractures and fusions.
Clinical Evidence
No clinical testing conducted. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993), bacterial endotoxin testing (ANSI/AAMI ST72), MR compatibility (ASTM F2182, F2052, F2213, F2119), dynamic cantilever bending, static axial load to failure, and targeting device accuracy.
Technological Characteristics
Intramedullary nail system; material: titanium alloy (Ti6Al4V ELI) per ASTM F136; cannulated design with pre-loaded compression screw; includes non-sterile targeting instrumentation; used with VariAx 2 bone screws; MR conditional.
Indications for Use
Indicated for internal bone fixation for patients requiring treatment for neuropathic osteoarthropathy (Charcot), fracture fixation, osteotomies, non-unions, mal-unions, and fusions.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
March 20, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the full name. The full name of the agency is "U.S. Food & Drug Administration".
Stryker GmbH Cindy Leon Staff Specialist, Regulatory Affairs 325 Corporate Drive Mahwah, New Jersey 07430
Re: K193366
Trade/Device Name: T2 ICF Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: February 19, 2020 Received: February 20, 2020
Dear Cindy Leon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193366
Device Name T2 ICF System
Indications for Use (Describe)
The T2 ICF System is indicated for internal bone fixation of the following conditions and procedures:
- · Neuropathic osteoarthropathy (Charcot)
- · Fracture fixation
- · Osteotomies
- · Non-unions
- · Mal-unions
- · Fusions
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments rearding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Proprietary Name: | T2 ICF System |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Smooth or Threaded Metallic Bone Fixation Fastener<br>(21 CFR 888.3040) |
| Regulation Description: | Smooth or Threaded Metallic Bone Fixation Fastener<br>(21 CFR 888.3040) |
| Regulation Number: | 21 CFR 888.3040 |
| Product Code: | HWC |
| Device Class: | Class II |
| Sponsor: | Stryker GMBH<br>Bohnackerweg 1<br>2545 Selzach / Switzerland |
| Contact Person: | Cindy Leon<br>Staff Specialist, Regulatory Affairs<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: 201-831-5374<br>Fax: 201-831-6500 |
| Date Prepared: | December 03, 2019 |
| Primary Predicate: | Synthes (USA) 6.5mm Midfoot Fusion Bolt: K081071 |
| Additional Predicate: | Wright Medical Technology, Inc. SALVATION Midfoot Nail:<br>K180024 |
510(k) Summary
#### Description
This Traditional 510(k) premarket notification is being supplied to the U.S. FDA to gain clearance to market the T2 ICF System. The T2 ICF System is a cannulated intramedullary nail system intended for internal fixation and stabilization of various foot instabilities and reconstructions. The T2 ICF system includes sterile packed implants (intramedullary nails and endcaps) available in various diameters and lengths to better accommodate patient anatomy as well as non-sterile instrumentation (targeting devices). The intramedullary nails are assembled with a pre-loaded compression screw, which allows for the nail to compress the bony fragments. The sterile packed intramedullary nails (including the preloaded compression screw) and endcaps are made of titanium alloy (Ti6Al4V ELI) as per ASTM F136.
Additionally, the T2 ICF System will be used with an existing subset of VariAx 2 bone screws (previously cleared under K191412) to achieve fixation of the nail. The use of this subset of the
{4}------------------------------------------------
VariAx 2 bone screws with the T2 ICF System does not alter the intended use, indications for use, or overall function of the VariAx 2 bone screws.
Lastly, the T2 ICF System will be used with several existing Class I 510(k) exempt devices for various orthopedic purposes.
# Intended Use
The T2 ICF System is intended for internal bone fixation.
# Indications for Use
The T2 ICF System is indicated for internal bone fixation of the following conditions and procedures:
- · Neuropathic osteoarthropathy (Charcot)
- · Fracture fixation
- · Osteotomies
- · Non-unions
- Mal-unions
- Fusions
## Summary of Technologies
A comparison of the systems demonstrated that the subject T2 ICF System is substantially equivalent to the previously cleared predicate Synthes (USA) 6.5mm Midfoot Fusion Bolt (K081071) and the Wright Medical Technology, Inc. SALVATION Midfoot Nail (K180024) regarding their intended use, indications for use, technological characteristics (design features, material and performance), as well as underlying operational principles.
Overall, the subject device and the predicate devices share many of the same technological characteristics including:
- Intramedullary nailing system intended to provide fracture fixation and alignment of the foot
- Nail design (available in various size configurations) and material "(titanium alloy – Ti6Al4V ELI per ASTM F136)
- * Synthes (USA) 6.5mm Midfoot Fusion Bolt offered in both titanium alloy and stainless steel variants
The following technological differences do exist between the subject and predicate devices:
- . Fixation technology
- Nail diameter size offerings
These differences have been addressed through performance testing, which demonstrates that the subject device is as safe and effective as the predicate devices.
## Non-Clinical Testing
The following bench testing was conducted to determine substantial equivalence between the subject device and the primary predicate Synthes (USA) 6.5mm Midfoot device:
{5}------------------------------------------------
- l . Biocompatibility evaluation per ISO 10993-1 and FDA Guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical device – Part 1: Evaluation and testing within a risk management process". Biocompatibility testing per ISO 10993-5, ISO 10993-12, ISO 10993-18.
- 2. Bacterial endotoxin testing (BET) as specified in ANSI/AAMI ST72 was used for pyrogenicity testing to achieve an Endotoxin limit of <20 EU/Device.
- 3. MR Compatibility:
- Radio frequency (RF)-Induced heating per ASTM F2182 a.
- Magnetically Induced Displacement Force per ASTM F2052 b.
- Magnetically Induced Torque per ASTM F2213 C.
- d. MR Image Artifacts per ASTM F2119
- 4. Dynamic Cantilever Bending Test
- Static Axial Load to Failure Test 5.
- 6. Targeting Device Accuracy Testing
The results of the testing demonstrated that the subject device is substantially equivalent to the marketed predicate device.
# Clinical Testing
No clinical testing of the T2 ICF System has been conducted.
## Conclusion
The subject T2 ICF System is substantially equivalent to the primary predicate Synthes (USA) 6.5mm Midfoot Fusion Bolt (K081071) and the additional predicate Wright Medical Technology, Inc. SALVATION Midfoot Nail (K180024) identified as part of this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.