The surgical patties and surgical strips are indicated for the use including brain and other tissues of the central nervous system, during surgery.
Device Story
Codman Surgical Patties and Strips are single-use, sterile, radiopaque devices used during neurosurgery to protect brain and central nervous system tissues. Patties consist of cottonoid material, a string, and an X-ray detectable monofilament; strips consist of cottonoid material and an X-ray detectable monofilament. The device is used by surgeons in an operating room setting. The radiopaque monofilament allows for visualization under X-ray to ensure the device is not left in the surgical site. The subject device utilizes a non-phthalate PVC resin for the monofilament, replacing the phthalate-containing PVC used in the predicate. The device is provided sterile (ETO) and is intended to be placed by the surgeon to provide a protective barrier or absorb fluids during procedures. It does not involve active electronic components or software.
Clinical Evidence
No clinical studies were performed. Substantial equivalence was established through bench testing, including visual inspection, functional/mechanical testing, radiopacity evaluation per ASTM F640-12, sterilization validation per ISO 11135:2014, shelf-life testing, and comprehensive biocompatibility testing (cytotoxicity, irritation, sensitization, systemic toxicity, and hemolysis) per ISO 10993 standards.
Technological Characteristics
Materials: Cottonoid, polyester string, non-phthalate PVC monofilament. Radiopacity: ASTM F640-12. Sterilization: Ethylene Oxide (ETO) to 10^-6 SAL per ISO 11135:2014. Form factor: Surgical patties and strips. No energy source or software.
Indications for Use
Indicated for use in the protection of brain and other central nervous system tissues during surgery. No specific age or gender restrictions provided.
Regulatory Classification
Identification
A neurosurgical paddie is a pad used during surgery to protect nervous tissue, absorb fluids, or stop bleeding.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 20, 2020
Integra LifeSciences Corporation Alyssa Woodcock Regulatory Affairs Product Manager 11 Cabot Boulevard Mansfield, Massachusetts 02048
## Re: K193346
Trade/Device Name: Codman Surgical Pattie & Strip Regulation Number: 21 CFR 882.4700 Regulation Name: Neurosurgical Paddie Regulatory Class: Class II Product Code: HBA Dated: October 19, 2020 Received: October 21, 2020
# Dear Alyssa Woodcock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193346
Device Name Codman Surgical Patties and Strips
Indications for Use (Describe)
The surgical patties and surgical strips are indicated for the use including brain and other tissues of the central nervous system, during surgery.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Codman® Surgical Pattie & Strip
# 510(k) Summary
| Date: | November 20, 2020 | |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Submission<br>number: | K193346 | |
| I. Submitter | Integra LifeSciences Corporation<br>11 Cabot Boulevard<br>Mansfield, MA 02048<br><br>Contact: Alyssa Woodcock<br>Phone: (508) 615-7426<br><br>Establishment Registation: 3014334038 | |
| II. Device | Device Proprietary Name | Codman® Surgical Pattie and Strip |
| | Common Name | Pattie and Strip |
| | Classification Name | Neurosurgical Paddie (21 CFR 882.4700) |
| | Regulatory Classification | II |
| | Product Code | HBA |
| | Review Panel | Neurology |
| III. Predicate<br>Device | The predicate device for this submission is the Codman® Surgical Pattie<br>(K880402), which was cleared on March 2, 1988. | |
| IV. Device<br>Description | The Codman Surgical Patties and Strips are single use, sterile devices indicated<br>for the use in protection of tissue, including brain and other tissues of the central<br>nervous system, during surgery.<br><br>The Surgical Patties are comprised of 3 main components: Cottonoid Material,<br>String and X-ray detectable monofilament.<br><br>The Strips are comprised of 2 main parts: Cottonoid Material and X-ray<br>detectable monofilament. | |
| V. Indications<br>for Use | The Surgical Patties and Surgical Strips are indicated for the use in protection of<br>tissue, including brain and other tissues of the central nervous system, during<br>surgery. | |
{4}------------------------------------------------
VI. The proposed Codman Surgical Patties & Strips are identical to the currently Comparison to marketed Codman Surgical Patties and Strips (K880402) with the exception that Predicate this submission proposes to use a non-phthalate PVC Resin instead of PVC Device Resin containing phthalates for the x-ray detectable monofilament affixed to the Surgical Patties and Strips.
> The indications for use, design, principle of operation, manufacturing process, clinical utility, packaging, and sterilization will remain unchanged and identical to the predicate device.
> Accelerated shelf-life testing has been completed to for 2 years and will be updated to 5 years once testing has been completed. The current marketed Codman Surgical Patties and Strips (K880402) has a 5-year shelf-life; therefore, the proposed Surgical Patties and Strips will be tested to a 5-year shelf-life.
| Comparison of the Predicate and Subject Device | | | |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Device:<br>Codman Sugical Patties<br>& Strips<br>(K880402) | Subject Device:<br>Codman Patties &<br>Strips<br>(This Submission) | Difference and<br>Justification |
| FDA Product<br>Code | HBA | Same as predicate | No difference |
| Classification | Class II - 21 CFR<br>882.4700 | Same as predicate | No difference |
| Classification<br>Name | Neurosurgical Paddie | Same as predicate | No difference |
| Intended Use | The surgical patties and<br>surgical strips are<br>indicated for the use in<br>protection of tissue,<br>including brain and other<br>tissues of the central<br>nervous system, during<br>surgery. | Same as predicate | No difference |
| Single Use | Yes | Same as predicate | No difference |
| Radiopaque | Yes | Same as predicate | No difference |
| Material | Pattie/Strip: Cottonoid<br>String: braided, textured<br>Polyester<br>X-ray Monofilament:<br>Phthalate containing PVC<br>material for x-ray<br>detection | Same materials as<br>predicate except a<br>similar PVC material<br>without phthalate will<br>be used for the x-ray<br>monofilament. | The testing (i.e.<br>Visual Inspection,<br>Functional,<br>Mechanical, Shelf<br>Life, Sterility and<br>Biocompatibility<br>testing) performed<br>verified that the<br>performance and<br>safety of the<br>proposed device is<br>substantially<br>equivalent to that of<br>the predicate device<br>and does not<br>introduce any new<br>questions of safety<br>and effectiveness. |
| Sterile | Yes, ETO Sterilized | Same as predicate | No difference |
| Sterility<br>Assurance<br>Level (SAL) | 10-6 | Same as predicate | No difference |
{5}------------------------------------------------
| Comparison of the Predicate and Subject Device | | | |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Device:<br>Codman Sugical Patties<br>& Strips<br>(K880402) | Subject Device:<br>Codman Patties &<br>Strips<br>(This Submission) | Difference and<br>Justification |
| Packaging | <b>Patties:</b><br>Packaging using Form<br>Fill Seal to create Sterile<br>Barrier, placed in a<br>Pouch with Tyvek Lid,<br>(Pouch – plastic film, Lid<br>– Tyvek) and then placed<br>in a Carton and Labeled<br>(Carton - Single walled<br>cardboard)<br><br><b>Strips:</b><br>Packaged using Ceratek<br>Pouch Sealer to create<br>Sterile Barrier, placed in<br>a Pouch with Tyvek for<br>sterilization (Pouch –<br>Plastic Film and Tyvek)<br>and then placed in Carton<br>and Labeled (Carton –<br>Single walled cardboard) | Same as predicate | No difference |
| Shelf Life | 5 years | 2 years | The 2-year shelf life<br>does not raise new<br>questions of safety<br>and effectiveness<br>because testing was<br>performed per FDA<br>recognized standard,<br>ASTM F640 and the<br>results met the<br>acceptance criteria. |
VII. There were no changes made that affect the Codman Surgical Patties and Strips Performance indications for use, principle of operation, manufacturing process, clinical utility, Data packaging, shelf life and sterilization. The difference between the predicate and proposed device is the use of a non-phthalate PVC Resin instead of PVC Resin that contains phthalates for the construction of the x-ray detectable monofilament. In addition, the proposed device has been tested to 2-year shelf life and testing is ongoing to 5-year shelf life to match the currently marketed Codman Surgical Patties and Strips shelf life of 5 years. Below is a summary of the testing performed to support the substantial equivalence between the proposed and predicate devices:
| Test | Test Method Summary | Results |
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Visual Inspection<br>(Patties) | Evaluated the following pattie specifications:<br>1. String and X-ray isolation<br>2. X-ray weld strength<br>3. X-ray flash | All samples passed the acceptance criteria demonstrating that the proposed patties meet the same performance specification as the predicate device. This supports that |
| | | |
| Visual Inspection<br>(Strips) | Evaluated the following parameters of the strips:<br>1. X-ray weld strength<br>2. X-ray flash | All samples passed the acceptance criteria demonstrating that the proposed strips meet the same specifications as the predicate strips. This supports that the performance of the proposed device with a non-phthalate PVC Resin is substantially equivalent to that of the predicate device. |
| Radiopacity<br>(Patties and<br>Strips) | Evaluated the visibility of the finished good<br>under X-ray per FDA recognized standard,<br>ASTM F640-12 | All samples passed the qualitative acceptance criteria. This supports that the performance of the proposed device with a non-phthalate PVC Resin is substantially equivalent to that of the predicate device. |
| Sterilization<br>Validation<br>(Cottonoid and<br>X-ray Strip) | Testing was conducted to achieve a 10-6 sterility assurance level using the overkill process per FDA recognized standard, ISO 11135:2014 with a reference product load that included "worst case" product. | The test results met the EO/ECH residual and bioburden acceptance criteria and achieved a sterility assurance level of 10-6. The acceptance criteria is the same as the predicate device, supporting substantial equivalence between the proposed and predicate devices. |
| Shelf-Life (2-<br>years) | The following test methods were used to evaluate the product after 2-years accelerated aging:<br>• X-ray shall not proture<br>• ASTM F640<br>• X-ray shall exhibit adequate adherence<br>• X-ray shall not overlap string. | All product tested passed the acceptance criteria. The proposed device has been tested to 2-year shelf life which is within the current shelf life claim for the predicate device. |
| Biocompatibility Testing | | |
| Cytotoxicity<br>Testing | Determine the potential of a test article extract to cause cytotoxicity per ISO 10993-5 Tests for in vitro cytotoxicity (Less than or equal to grade 2 reactivity on the 0 to 4 reactivity scale) | All product tested passed the acceptance criteria demonstrating that the device is biocompatible and therefore is substantially equivalent to the predicate device. |
| ISO<br>Intracutaneous<br>Study in Rabbits | Evaluate the local dermal irritation of a test article extract following intracutaneous injection in rabbits per ISO 10993-10 Tests for irritation and skin sensitization (Difference between test article extract overall mean score and control group score is 1.0 or less) | All product tested passed the acceptance criteria demonstrating that the device is biocompatible and therefore is substantially equivalent to the predicate device. |
| USP Rabbit<br>Pyrogen Study,<br>Material<br>Mediated | Determine whether an extract of the test article induced a pyrogenic response following intravenous injection in rabbits per ISO 10993-11 Tests for Systemic Toxicity (No single animal shows a rise of 0.5 °C above its baseline temperature) | All product tested passed the acceptance criteria demonstrating that the device is biocompatible and therefore is substantially equivalent to the predicate device. |
| ISO Acute<br>Systemic Toxicity<br>Study in Mice | Evaluate the acute systemic toxicity of a test article extract following injection in mice per ISO 10993-11 Tests for Systemic Toxicity (No single animal treated with the individual test extract exhibits a significantly greater reaction | All product tested passed the acceptance criteria demonstrating that the device is biocompatible and therefore is substantially equivalent to the predicate device. |
| | | |
| <b>ISO Guinea Pig<br/>Maximization<br/>Sensitization Test</b> | than the control animals: no more than two<br>animal deaths; no abnormal behavior such as<br>convulsions or prostration occurs in two or more<br>animals; no final body weight loss greater than<br>10% occurred in three or more animals) | All product tested passed the<br>acceptance criteria demonstrating<br>that the device is biocompatible and<br>therefore is substantially equivalent<br>to the predicate device. |
| Evaluate the potential of the test article to cause | | |
| delayed dermal contact sensitization in the | | |
| guinea pig maximization test per ISO 10993-10 | | |
| Tests for irritation and skin sensitization (A<br>grade of 0 on a patch test reaction scale of 0-3) | | |
| <b>ASTM Hemolysis<br/>Study</b> | Evaluate the potential to cause hemolysis. This<br>study was conducted based on ASTM F756 and<br>ISO 10993-4 (Acceptance criteria of hemolytic<br>index for the test article extract to be 0.1% and<br>non-hemolytic.) | All product tested passed the<br>acceptance criteria demonstrating<br>that the device is biocompatible and<br>therefore is substantially equivalent<br>to the predicate device. |
{6}------------------------------------------------
{7}------------------------------------------------
Codman® Surgical Pattie & Strip
### Bench Testing
Visual Inspection, Functional Testing (Radiopacity Testing), Mechanical Testing (Weld Strength), Shelf Life testing, Sterility testing and Biocompatibility testing were performed to verify that the performance of the proposed device with a non-phthalate PVC Resin is substantially equivalent to that of the current device, which uses PVC Resin containing phthalates).
#### Animal Studies
No animal studies were performed as appropriate verification and validation of the subject device was achieved based on the comparison to the predicate device and from the results of the bench, sterilization, shelf-life and biocompatibility testing.
#### Clinical Studies
No clinical studies were performed as appropriate verification and validation of the subject device was achieved based on the comparison to the predicate device and from the results of the bench, sterilization, shelf-life and biocompatibility testing.
Based upon the intended use, design, operating principles, patient/user VIII. Conclusion interface, comparison to the predicate device, and testing conducted, it is concluded that the subject device, Codman Surgical Pattie and Strip is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.