The GoChek Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips. The GoChek Blood Glucose Monitoring System is intended for self-testing by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control programs. The GoChek Blood Glucose Monitoring System is intended for single-patient use and should not be shared. The GoChek Blood Glucose Monitoring System is for in vitro diagnostic use. The GoChek Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes. The GoChek Connect Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips. The GoChek Connect Blood Glucose Monitoring System is intended for self-testing by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control programs. The GoChek Connect Blood Glucose Monitoring System is intended for single-patient use and should not be shared. The GoChek Connect Blood Glucose Monitoring System is for in vitro diagnostic use. The GoChek Connect Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes.
Device Story
System uses amperometric electrochemical detection to measure glucose in fresh capillary whole blood (fingertip). Test strips contain glucose dehydrogenase (FAD-GDH) and mediator; reaction produces electrical current proportional to glucose concentration. Meter calculates and displays plasma-equivalent glucose. GoChek Connect model includes Bluetooth 4.0 for data transfer to Pancares App on Android smartphones; data transfer is read-only. Used by patients at home for diabetes management. Healthcare providers review stored results to assess therapy effectiveness. Benefits include convenient, quantitative glucose monitoring for diabetes control.
Clinical Evidence
Method comparison study conducted with lay users (naïve and non-naïve) comparing GoChek series to YSI Model 2300 STAT PLUS. Additional laboratory testing performed on samples across the 10-580 mg/dL range. Results: 95% of SMBG results within +/- 15% and 99% within +/- 20% of comparator across the entire measuring range. Bench testing included precision, linearity, interference, hematocrit effect, stability, and environmental/safety testing.
Technological Characteristics
Amperometric glucose monitoring system; FAD-GDH enzyme; 0.5 μL sample volume; 10-580 mg/dL reportable range. Connectivity: Bluetooth (Connect model). Power: Battery. Calibration: Automatic. Sterilization: N/A (single-use strips). Compliant with ISO 14971 risk management and CLSI EP06/EP07 standards.
Indications for Use
Indicated for self-testing by people with diabetes at home to monitor glucose levels in fresh capillary whole blood from fingertips. Not for use in healthcare settings, neonates, critically ill patients, or patients with shock, hypotension, impaired peripheral circulation, or severe dehydration.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
On Call Sharp Blood Glucose Monitoring System (k130284)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
k193340
B Applicant
MicroTech Medical, Inc.
C Proprietary and Established Names
GoChek Blood Glucose Monitoring System
GoChek Connect Blood Glucose Monitoring System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| NBW | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New Device
B Measurand:
Glucose in capillary whole blood obtained from the fingertip
C Type of Test:
Quantitative, amperometric measurement of glucose.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The GoChek Blood Glucose Monitoring System
The GoChek Blood Glucose Monitoring System comprises of the GoChek Blood Glucose Meter and GoChek Blood Glucose Test Strips. The GoChek Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips. The GoChek Blood Glucose Monitoring System is intended for self-testing by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control programs. The GoChek Blood Glucose Monitoring System is intended for single-patient use and should not be shared.
The GoChek Blood Glucose Monitoring System is for in vitro diagnostic use. The GoChek Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes.
The GoChek Connect Blood Glucose Monitoring System
The GoChek Connect Blood Glucose Monitoring System comprises of the GoChek Connect Blood Glucose Meter and GoChek Blood Glucose Test Strips. The GoChek Connect Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips. The GoChek Connect Blood Glucose Monitoring System is intended for self-testing by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control programs. The GoChek Connect Blood Glucose Monitoring System is intended for single-patient use and should not be shared.
The GoChek Connect Blood Glucose Monitoring System is for in vitro diagnostic use. The GoChek Connect Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes.
C Special Conditions for Use Statement(s):
- OTC - Over The Counter
- This device is not intended for use in healthcare or assisted-use settings such as hospitals, physician offices, or long-term care facilities because it has not been cleared by FDA for use in these settings, including for routine assisted testing or as part of glycemic control procedures. Use of this device on multiple patients may lead to transmission of Human Immunodeficiency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or other bloodborne pathogens.
- This meter can only be used to determine blood glucose levels using whole blood samples from fingertips.
- For in vitro diagnostic use.
- For single-patient use only.
K193340 - Page 2 of 12
{2}
- This system should not be used for the diagnosis of or screening for diabetes.
- This system is not for use on patients with abnormally low blood pressure or those who are in shock.
- Test results may be lower than the actual values or inaccurate if the individual is hypotensive, in a hyperglycemic-hyperosmolar state (with or without ketosis).
- This system is not for use with neonates.
- This system should not be used on critically ill patients.
- This system should not be used on patients with impaired peripheral circulation or severe dehydration.
- Altitudes above 10,000 feet may cause inaccurate results.
## D Special Instrument Requirements:
The GoChek Blood Glucose Meter
The GoChek Connect Blood Glucose Meter
## IV Device/System Characteristics:
### A Device Description:
GoChek/GoChek Connect Blood Glucose Monitoring System consists of a glucose meter (GoChek or GoChek Connect Blood Glucose Meter), GoChek Blood Glucose Test Strips (sold in vials of test strips and individually wrapped pouches of test strips; sold separately), carrying case, lancing device, batteries, and two levels of control solution (sold separately). The only difference between the GoChek and GoChek Connect meters is that the GoChek Connect meter has Bluetooth connectivity enabled.
### B Principle of Operation:
The GoChek and GoChek Connect Blood Glucose meters measure the glucose concentration in capillary whole blood. Blood is applied to the end tip of the test strip. The blood is then automatically absorbed into the reaction cell. The reaction takes place in the reaction cell. The enzyme used is a Flavin-Adenine-Dinucleotide-Dependent Glucose Dehydrogenase (FAD-GDH). FAD-GDH oxidizes glucose to gluconolactone (GDL) and a transient electrical current is formed during the reaction which is detected by the meter. The blood glucose concentration is then calculated based on the electrical current. The result is then shown on the meter display in mg/dL for the user. The meters are calibrated to display plasma equivalent results.
### C Instrument Description Information:
| Modes of Operation | Yes | No |
| --- | --- | --- |
| Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐ |
| Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☑ | ☐ |
K193340 - Page 3 of 12
{3}
| Modes of Operation | Yes | No |
| --- | --- | --- |
| **Software** | | |
| FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types. | ☑ | ☐ |
1. **Instrument Name:**
GoChek Blood Glucose Monitoring System
GoChek Connect Blood Glucose Monitoring System
2. **Specimen Identification:**
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
3. **Specimen Sampling and Handling:**
Samples are to be tested immediately upon collection.
4. **Calibration:**
The meter does not require calibration or coding by the user. The meter is automatically coded.
5. **Quality Control:**
The sponsor manufactures two (2) levels of GoChek control solutions. Each test strip vial is marked with a control solution range (CTRL1 and CTRL2). The user is instructed to compare the control results with the expected control ranges printed on the strip vial. Users are given instructions on when to conduct control solution testing and to call customer service if repeat testing with control material is out of range. The meter automatically detects that a control solution is being used. When a control solution test is performed, the test results will not be included in the 'day average' calculation.
V **Substantial Equivalence Information:**
A **Predicate Device Name(s):**
On Call Sharp Blood Glucose Monitoring System
B **Predicate 510(k) Number(s):**
k130284
K193340 - Page 4 of 12
{4}
K193340 - Page 5 of 12
# C Comparison with Predicate(s):
| Device & Predicate Device(s): | K193340 | K130284 |
| --- | --- | --- |
| Device Trade Name | GoChek Blood Glucose Monitoring System
GoChek Connect Blood Glucose Monitoring System | On Call Sharp Blood Glucose Monitoring System |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | For the quantitative measurement of glucose in fresh capillary whole blood from the fingertips as an aid in monitoring the effectiveness of diabetes control programs. | Same |
| Assay Method | Electrochemical | Same |
| Detection Method | Amperometry | Same |
| Enzyme | Glucose Dehydrogenase (FAD-GDH) | Same |
| General Device Characteristic Differences | | |
| Sample Type | Fresh capillary whole blood from the fingertips. | Fresh capillary whole blood from fingertips, forearm and palm |
| Hematocrit Range | 10-70% | 25-70% |
| Operating Temperature | 41-113°F (5-45°C) | 50-113°F (10-45°C) |
# VI Standards/Guidance Documents Referenced:
- ISO 14971:2007: Medical devices - Application of risk management to medical devices
- Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use: Guidance for Industry and Food and Drug Administration Staff, October 11th, 2016
- CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
- CLSI EP07-A2: Interference Testing in Clinical Chemistry; Approved Guideline Second Edition
{5}
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
Unless otherwise noted, the analytical studies detailed below were performed with the GoChek Blood Glucose Monitoring System, which is identical to the GoChek Connect Blood Glucose Monitoring System with the exception of the presence of Bluetooth in the GoChek Connect System. Therefore, the analytical studies performed with the GoChek Blood Glucose Monitoring System are representative of the performance for both systems in this 510(k) submission.
1. Precision/Reproducibility:
Within-Run Precision
The sponsor performed a repeatability study (within day) precision study using venous whole blood adjusted to five different glucose concentration ranges (30-50, 51-110, 111-150, 151-250, and 251-400 mg/dL). Each glucose concentration was analyzed in replicates of 10 with three test strips lots, and 10 meters, for a total of 100 tests per glucose level. Results are summarized in the table below:
| Glucose Level | Average | SD | CV |
| --- | --- | --- | --- |
| 1 | 35.3 | 1.88 | 5.3% |
| 2 | 95.8 | 2.81 | 2.9% |
| 3 | 128.3 | 3.74 | 2.9% |
| 4 | 221.4 | 6.31 | 2.8% |
| 5 | 382.9 | 10.05 | 2.6% |
Intermediate Precision
Intermediate (day-to-day) precision was performed using five ranges of glucose control solution concentrations (30-50, 51-110, 111-150, 151-250, and 251-400 mg/dL). Testing was performed for 10 days with three test strip lots. For each level, on each day, 10 meters were used for testing, with one replicate collected per meter for a total of 10 replicates per day for each control solution level, for a total of 100 tests per control solution level. Results are summarized below:
| Glucose Level | Average | SD | CV |
| --- | --- | --- | --- |
| 1 | 37.5 | 1.61 | 4.3% |
| 2 | 77.5 | 2.27 | 2.9% |
| 3 | 128.0 | 3.01 | 2.4% |
| 4 | 193.0 | 3.71 | 1.9% |
| 5 | 346.7 | 7.07 | 2.0% |
2. Linearity:
Linearity testing was evaluated with test strips from three lots and 11 mixed pools of venous whole blood samples with glucose concentrations of 7.5, 25.2, 43.7, 72.6, 95.5, 152.5, 199.5, 252.0, 300.0, 405.5, 511.5, and 578.0 mg/dL as measured with the YSI Model 2300 STAT
K193340 - Page 6 of 12
{6}
PLUS glucose analyzer. Each level was measured in replicates of four and the values obtained on the GoChek Blood Glucose Monitoring System were compared to the comparator. Linear regression analysis of the results yielded the following:
| Lot | Slope | y-intercept | R2 |
| --- | --- | --- | --- |
| 1 | 0.9960 | -4.4024 | 0.9997 |
| 2 | 0.9743 | -0.9445 | 0.9997 |
| 3 | 0.9949 | -3.1076 | 0.9996 |
The results of the study support the sponsor's claimed glucose measuring range of 10 - 580 mg/dL. If a sample is less than 10 mg/dL, the result is flagged by the meter as LO. If the result is greater than 580 mg/dL, the result is flagged by the meter as HI. This low and high functions were validated and demonstrated to function as intended.
# 3. Analytical Specificity/Interference:
Interference testing was conducted using venous whole blood adjusted to a hematocrit of $42\% \pm 2\%$ and glucose was adjusted to the following levels: $50 - 70 \mathrm{mg} / \mathrm{dL}$ , $110 - 130 \mathrm{mg} / \mathrm{dL}$ and $250 - 270 \mathrm{mg} / \mathrm{dL}$ . Control blood samples were prepared by adding solvent with no interferant. Control and test samples were measured using 10 strips on each of 2 meters using 3 strip lots. The highest tested concentrations at which no significant interference was observed (defined by the sponsor as average $\%$ bias $\leq \pm 10\%$ ) are presented in the following table:
| Substances | Highest concentration testing with no significant interference |
| --- | --- |
| Acetaminophen | 20 mg/dL |
| Ascorbic Acid | 3 mg/dL |
| Cholesterol | 500 mg/dL |
| Conjugated-Bilirubin | 50 mg/dL |
| Creatinine | 15 mg/dL |
| Dopamine | 20 mg/dL |
| EDTA | 200 mg/dL |
| Galactose | 100 mg/dL |
| Gentisic acid | 10 mg/dL |
| Glutathione Reduced | 92 mg/dL |
| Hemoglobin | 2000 mg/dL |
| Heparin Sodium | 80000 u/L |
| Ibuprofen | 50 mg/dL |
| Icodestrin | 1094.4 mg/dL |
| Isomalt | 0.09 mg/dL |
| L-Dopa (Levo-Dopa) | 3 mg/dL |
| Lactitol | 0.09 mg/dL |
| Maltitol | 0.09 mg/dL |
| Maltose | 9000 mg/dL |
| Mannitol | 0.09 mg/dL |
| Methyl Dopa | 1.5 mg/dL |
K193340 - Page 7 of 12
{7}
| Substances | Highest concentration testing with no significant interference |
| --- | --- |
| Paralidoxime Idodine (PAM) | 80 mg/dL |
| Salicylic Acid | 60 mg/dL |
| Sodium | 414 mg/dL |
| Sorbitol | 0.09 mg/dL |
| Tolazamide | 10 mg/dL |
| Tolbutamide | 64 mg/dL |
| Triglycerides | 3000 mg/dL |
| Uric Acid | 24 mg/dL |
| Xylitol | 0.09 mg/dL |
| Xylose | 40 mg/dL |
The sponsor includes the following in the labeling for the GoChek and the GoChek Connect Blood Glucose Monitoring Systems:
- If you are taking Vitamin C (ascorbic acid, blood concentrations >3 mg/dL) at doses higher than recommended, it may interfere with your glucose meter and cause you to get inaccurate results with this system.
- Do not use this system during or shortly after receiving xylose absorption therapy; xylose may case inaccurate blood glucose results.
4. Assay Reportable Range:
Assay reportable range, 10 - 580 mg/dL.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The system is traceable to NIST SRM 917b glucose reference material. A method comparison was performed using the candidate device and YSI Model 2300 STAT PLUS glucose analyzer as the comparator method.
Test Strip Stability:
Test strip (vials and individually wrapped pouches of strips) stability was assessed using accelerated and real time stability studies. Protocols and acceptance criteria were reviewed and found acceptable to support the labeling claims that vialed test strips are stable for 6 months after first being opened and vialed and individually wrapped pouches are stable for 3 years when stored between of 35.6°F to 95°F (2°C-35°C) and 10-90% relative humidity.
6. Detection Limit:
The claimed reportable range for this assay is 10-580 mg/dL.
7. Assay Cut-Off:
Not applicable.
K193340 - Page 8 of 12
{8}
8. Accuracy (Instrument):
Not applicable.
9. Carry-Over:
Not applicable. The device uses single-use strips.
## B Comparison Studies:
1. Method Comparison with Predicate Device:
See lay user study below, section VII.C3, for system accuracy in the hands of the intended user.
2. Matrix Comparison:
Not applicable. The device is only intended for use with fresh capillary whole blood from a fingerstick.
## C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
**Lay-user Performance study:**
To assess the performance of the GoChek Connect Blood Glucose Monitoring System in the hands of the intended users, the sponsor performed studies with 360 English speaking participants. The participants were responsible for obtaining their own fingertip capillary sample and performing a blood glucose test using only the instructions from the product labeling. A total of 360 meters and 20 test strip vials, each containing 25 test strips, from 3 test strip lots were used. The glucose concentrations across all samples ranged from 42.2 to $420.1\mathrm{mg / dL}$, which includes 34 native samples $< 80\mathrm{mg / dL}$ and 37 samples $>250\mathrm{mg / dL}$ as measured by the comparator method. Results were analyzed by comparing the blood glucose results obtained by the lay users with the GoChek Connect Blood Glucose Monitoring System against results obtained with a laboratory-based comparator method (YSI Model 2300 STAT PLUS glucose analyzer). The glucose concentrations in the samples ranged from $42.2 - 420.1\mathrm{mg / dL}$, as measured by the YSI Model 2300 STAT PLUS glucose analyzer. Results are summarized in the tables below:
K193340 - Page 9 of 12
{9}
K193340 - Page 10 of 12
| For glucose concentrations < 75 mg/dL: | | | |
| --- | --- | --- | --- |
| Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL | Within ± 20 mg/dL |
| 21/25 (84.0%) | 25/25 (100%) | 25/25 (100%) | 25/25 (100%) |
| For glucose concentrations > 75 mg/dL: | | | |
| --- | --- | --- | --- |
| Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% |
| 165/335 (49.3%) | 269/335 (80.3%) | 329/335 (98.2%) | 335/335 (100%) |
| Combined glucose concentrations across the measuring range: | | | |
| --- | --- | --- | --- |
| Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% |
| 176/360 (48.9%) | 289/360 (80.3%) | 353/360 (98.1%) | 360/360 (100%) |
**Accuracy at Extreme Glucose Study:**
An additional accuracy study was performed by testing 240 capillary fingerstick whole blood samples containing glucose concentrations from 12.5 to 584.3 mg/dL. Samples were altered by spiking or glycolysis to obtain 120 samples with glucose <80 mg/dL, ranging from 12.5 to 79 mg/dL, and 120 samples with glucose concentrations >250 mg/dL, ranging from 251 to 584.3 mg/dL. Samples obtained using the GoChek Connect Blood Glucose Monitoring System were compared to the results obtained on the comparator method (YSI 2300 STAT PLUS). Results are summarized in the tables below:
| For glucose concentrations < 80 mg/dL: | | | |
| --- | --- | --- | --- |
| Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% |
| 40/120 (39.2%) | 92/120 (76.7%) | 117/120 (97.5%) | 120/120 (100%) |
| For glucose concentrations >250 mg/dL: | | | |
| --- | --- | --- | --- |
| Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% |
| 46/120 (38.3%) | 92/120 (76.7%) | 117/120 (97.5%) | 119/120 (99.2%) |
**Readability:**
The user manual and inserts were assessed for readability using the Flesch-Kincaid's assessment and it was observed that the overall readability level was as that of an 8th grader.
**D Clinical Cut-Off:**
Not Applicable
**E Expected Values/Reference Range:**
The following expected glucose values for people without diabetes are included in the device labeling:
| Time | Range, mg/dL | Range, mmol/L |
| --- | --- | --- |
| Fasting and Before Meals | 70 – 100 | 3.9 – 5.6 |
| 2 Hours After Meal | Less than 140 | Less than 7.8 |
{10}
Source: Standards of Medical Care in Diabetes - 2020, Diabetes Care 2020 Jan; 43 (Supplement 1): S1-S2.
## F Other Supportive Instrument Performance Characteristics Data:
### 1. Altitude Study
To evaluate the effect of altitude on the GoChek Blood Glucose Monitoring System, a study was conducted using venous whole blood samples adjusted with four concentration levels of glucose ranging from 40-500 mg/dL. Samples were tested at two different altitude conditions at sea level (65 feet) and 11,410 feet. Blood samples were tested using 10 test strips at each altitude condition for each glucose concentration. Eighty test strips from 3 lots were used. The results obtained on the candidate device were compared with the results obtained with the comparator method, YSI 2300 STAT PLUS glucose analyzer. The results demonstrated acceptable bias to the comparator to support the claim in the labeling that glucose measurement performance of the GoChek Blood Glucose Monitoring System is maintained at altitudes up to 10,000 feet.
### 2. Hematocrit
To evaluate the effect of hematocrit on the GoChek Blood Glucose Monitoring System, venous blood samples were adjusted to hematocrit levels of 10%, 15%, 20%, 25%, 30%, 35%, 40%, 45%, 50%, 55%, 60%, 65%, 70%. Each hematocrit level was adjusted to achieve five glucose concentration ranges (30-50 mg/dL, 51-110 mg/dL, 111-150 mg/dL, 151-250 mg/dL, and 251-400 mg/dL). Blood samples were tested using the GoChek meter with test strips from three lots and the mean glucose values were compared to the results obtained on the comparator method, YSI 2300 STAT PLUS glucose analyzer. The percent biases relative to the comparator method demonstrated were acceptable to support the claimed hematocrit range of 10 to 70%.
### 3. Operating Conditions
The effect of temperature and relative humidity on the GoChek Blood Glucose Monitoring System was evaluated using venous whole blood samples adjusted to approximately four glucose concentration ranges (40-50, 100-120, 250-300, and 500-550 mg/dL). Testing was conducted under the following temperature and relative humidity (RH) combinations: 5°C at 10% relative humidity (RH), 5°C at 90% RH, 21°C at 10% RH, 21°C at 90% RH, 21°C at 45% RH, 45°C at 10% RH and 45°C at 90% RH. Test strips from three lots were used to test the blood samples at each glucose concentration in 10 replicates of 10. Values were compared to results from the comparator method, YSI 2300 STAT PLUS glucose analyzer. The study results support the operating conditions claim of 41-113°F (5-45°C) with relative humidity of 10% to 90%.
### 4. Sample volume Study
The sponsor performed a study to support the claimed minimum sample volume of 0.5 μL for the GoChek Blood Glucose Monitoring System. Venous whole blood samples
K193340 - Page 11 of 12
{11}
with three ranges of glucose concentrations (50-65, 100-120, 200-250 mg/dL) were tested at five sample volumes (0.3, 0.4, 0.5, 0.6, and 0.7 μL). Each sample was tested with 10 test strips from 3 lots. Values obtained were compared to the comparator method, YSI 2300 STAT PLUS. The meter has an error message that is displayed if insufficient blood is added to the test strip. This feature was validated and was shown to function as intended. The study results support the claimed sample volume of 0.5 μL.
## 5. Flex Studies
The following additional flex studies were performed with the GoChek Blood Glucose Monitoring System: sample perturbation, sample outside measuring range, intermittent sampling, drop, vibration, and shipping testing were completed. The testing performed demonstrated that the device is robust under these conditions.
## 6. Infection Control Studies
The two device systems, GoChek and GoChek Connect Blood Glucose Monitoring System, are intended for single-patient use only. Disinfection efficacy studies were performed on the materials comprising the meter by an outside commercial testing laboratory demonstrating complete inactivation of hepatitis B virus (HBV) with the chosen disinfectant, CaviWipes (EPA Registration # 46781-8). Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or external materials of the meter after 7,665 cleanings and disinfection steps with the CaviWipes wipes. The robustness studies were designed to simulate 3 years of single-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
## 7. Electromagnetic Compatibility (EMC) Testing
The sponsor provided documentation certifying that acceptable electrical safety and EMC testing had been performed and the system was found to be compliant.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K193340 - Page 12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.