K193328 · Beijing Globalipl Development Co., Ltd. · ONF · Mar 25, 2020 · General, Plastic Surgery
Device Facts
Record ID
K193328
Device Name
IPL + Diode Laser Machine
Applicant
Beijing Globalipl Development Co., Ltd.
Product Code
ONF · General, Plastic Surgery
Decision Date
Mar 25, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IPL + Diode Laser machine (inclusive of the handpiece used to deliver pulsed-light energy) is indicated for use in surgical, and aesthetic applications in permanent hair reduction, reduction of benign pigmented lesions and benign vascular lesions, and (inclusive of the handpiece used to deliver diode laser energy)is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Device Story
Device (US419) combines Intense Pulsed Light (IPL) and 808nm Diode Laser for aesthetic/surgical hair removal and lesion treatment; operates via selective thermolysis (IPL) and selective photothermolysis (Diode Laser). Inputs: user-defined parameters via LCD control panel; outputs: pulsed light (480-1200nm) or 808nm laser energy delivered through sapphire-tipped handpiece. System includes main control module, constant current board, and cooling system (condenser, cold plate, fans). Operated by clinician in clinical setting; laser emission triggered by foot switch. Microcontroller monitors sensors (temperature, humidity, water flow) to ensure safety and cooling. Output affects target chromophores (melanin, oxyhemoglobin) to achieve hair reduction or lesion treatment; benefits patient through stable, long-term hair reduction.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety, laser safety, photobiological safety, and biocompatibility (ISO 10993-5, ISO 10993-10).
Technological Characteristics
System uses IPL (480-1200nm) and 808nm Diode Laser. Handpiece features sapphire crystal delivery. Cooling via water circulation, condenser, and cold plate. Electrical safety per IEC 60601-1, 60601-2-22, 60601-2-57. Laser safety per IEC 60825-1. Biocompatibility per ISO 10993-5/10. Microcontroller-based control system with sensor-driven parameter monitoring.
Indications for Use
Indicated for permanent hair reduction, reduction of benign pigmented lesions, and reduction of benign vascular lesions in patients of all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
March 25, 2020
Beijing Globalipl Development Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm. 912, Building #15, XiYueHui, No. 5. YiHe North Rd., FangShan District Beijing, 102401 Cn
Re: K193328
Trade/Device Name: IPL + Diode Laser Machine Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONF, GEX Dated: November 29, 2019 Received: December 2, 2019
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Mavadia-Shukla, PhD Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K193328
Device Name Trade Name: IPL + Diode Laser machine Model: US419
Type of Use (Select one or both, as applicable)
### Indications for Use (Describe)
The IPL + Diode Laser machine (inclusive of the handpiece used to deliver pulsed-light energy) is indicated for use in surgical, and aesthetic applications in permanent hair reduction of benign pigmented lesions and benign vascular lesions, and (inclusive of the handpiece used to deliver diode laser energy) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
> Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Tab #8 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K193328
- 1. Date of Preparation
02/28/2020
- Sponsor 2.
## Beijing Globalipl Development Co.,Ltd.
F-8 Qunyinghui Building, Jinyuan Rd.32, Daxing Economic Development Zone, Beijing, China 102628
Contact Person: Li Ye Position: Certification Department Manager Tel: +86-18865721581 Fax: +86-534-2756245 Email: Liye@globalipl.com
- 3. Submission Correspondent
Mr. Ray Wang
# Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, BeiJing, China 102401
Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: IPL + Diode Laser machine Common Name: Powered Light Based Non-Laser Surgical Instrument With Thermal Effect; Powered Laser Surgical Instrument Model(s): US419
Regulatory Information:
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology ;
Classification: II;
Product Code: ONF, GEX;
Regulation Number: 21 CFR 878.4810;
Review Panel: General & Plastic Surgery;
#### Indication For Use:
The IPL + Diode Laser machine (inclusive of the handpiece used to deliver pulsed-light energy) is indicated for use in surgical, and aesthetic applications in permanent hair reduction, reduction of benign pigmented lesions and benign vascular lesions, and (inclusive of the handpiece used to deliver diode laser energy)is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
#### ર. Device Description
The proposed device, IPL + Diode Laser machine (inclusive of the handpiece used to deliver pulsed-light energy) is indicated for use in surgical, and aesthetic applications in permanent hair removal, reduction of benign pigmented lesions and benign vascular lesions, and (inclusive of the handpiece used to deliver diode laser energy)is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.
IPL are intense pulsed light system which delivers intense pulsed light at a wavelength ranging from 480mm-1200nm. Intense Pulsed Light (IPL) systems work on the principles of selective thermolysis. That is, causing thermal damage to target chromophores by using light of appropriate wavelength in pulses that exceeds the chromophores' thermal relaxation time but sparing normal skin by limiting the pulse width below the thermal relaxation time for skin.
IPL are different from lasers in that they deliver many wavelengths in each pulse of light instead of just one wavelength. Generally, IPL enhances penetration without using excessive energy levels and enables targeting of specific chromophores.
{5}------------------------------------------------
Diode Laser machine is based on the principles of selective photothermolysis (SPTL): a combination of the appropriate laser wavelength, pulse duration, and fluence can obtain optimal effect on a targeted tissue with minimal effect on surrounding tissue.
Wavelengths of 808 nanometers (nm) are widely used in hair removal treatment, it is selectively absorbed by melanin in the hair shaft, damaging the follicular epithelium, and further causes hair loss, while the competing chromophores (oxyhemoglobin and water) absorb less energy at these wavelengths.
The user could power on the device and use the key switch to start the system, and then set the output parameter, when the setting is done, the user could use the handpiece to aim the area to be treated by 808 nm laser delivering , the laser emitting could be triggered by foot switch controlled by user.
#### Function module description
#### a. Control Panel
The module uses the microcontroller as the heart, utilizes the LCD screen to display all prompt information and the system state information to complete the human-machine interaction function, and realizes the device parameters settings and accurate control of the output IPL/laser energy by the operator.
#### b. Main Control Module
The module uses the microcontroller as the heart, receives the IPL/laser energy parameters and work instructions from the control panel and detects the footswitch state; Utilizes the sensors of temperature, humidity, liquid level and flow to detect the parameters such as temperature, humidity and water flow during system working, and according to the detected values to calculate the dew point temperature; Controls and detects the work state of constant current board module as well as the temperature and humidity control system; Uploads the state data and alarm information of water circulation system, cooling system, handpiece module and constant current board module during system working.
#### c. Constant current board module
The module uses the high-power MOS as the heart, receives the IPL/laser energy parameters from the main control module, supplies the IPL/laser with constant drive current which corresponding to the received IPL/laser energy parameters to drive the IPL/laser to emit light. The module also has the detection function of over-current, over-voltage, over-temperature and handpiece state, and uploads the detected data to the control module.
#### d. Temperature and humidity control system
The system mainly includes the condenser, cold plate, water circulation subsystem and fans. The microcontroller of main control module according to the temperature, humidity parameter detected by the sensors to control the working state of the condenser, cold plate and cooling fan to meet the temperature and humidity requirements during the IPL/laser working.
{6}------------------------------------------------
### e. Handpiece module
Handpiece module is the heart of the device, which is the execution unit of the device and completes the IPL/laser emission function. The module is mainly composed of IPL/laser, sapphire, temperature and humidity sensor, data storage chips, cooling components and water flow path. The IPL/laser emits light to output energy, temperature and humidity sensors detects the temperature and humidity parameters during handpiece working, the cooling components and water flow path take away the heat of the IPL/laser to prevent it from being damaged caused by over-temperature, so prolongs the service life of the IPL/laser.
{7}------------------------------------------------
- 6. Identification of Predicate Device Predicate Device 1: 510(k) Number: K161286 Product Name: IPL Therapy Machine Manufacturer: BEIJING ADSS DEVELOPMENT CO., LTD.
Predicate Device 2: 510(k) Number: K192516 Product Name: Diode Laser 808nm Manufacturer: Beijing Superlaser Technology Co., Ltd.
- 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > IEC 60601-1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
- > IEC 60601-2-22 Edition 3.1 2012-10, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment.
- > IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification and requirements.
- > IEC60601-2-57:2011, Medical electrical equipment -- Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- > IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems
- IEC 60601-1-2:2014 , Medical electrical equipment Part 1-2: General requirements for basic A safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- > ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)
- > ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. (Biocompatibility)
- Clinical Test Conclusion 8.
No clinical study is included in this submission.
{8}------------------------------------------------
# 9. Substantially Equivalent (SE) Comparison
| Item | Proposed Device | Predicate Device 1<br>K161286 | Predicate Device 2<br>K192516 | Remark |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code | ONF;GEX | ONF | GEX | SE |
| Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | SE |
| Intended Use | The IPL + Diode Laser machine (inclusive of the handpiece used to deliver pulsed-light energy) is indicated for use in surgical, and aesthetic applications in permanent hair reduction, reduction of benign pigmented lesions and benign vascular lesions, and (inclusive of the handpiece used to deliver diode laser energy)is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. | The VE2000 device is indicated for use in surgical, aesthetic applications in permanent hair reduction, reduction of benign pigmented lesions and benign vascular lesions. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen | The Diode Laser 808nm is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. | SE |
| Configuration | Main Unit<br>Handpiece<br>Foot Control | Main Unit<br>Handpiece<br>Foot Control | Main Unit<br>Handpiece<br>Foot Control | SE |
| ITEM | Proposed Device | Predicate Device1<br>IPL Therapy Machine(K161286) | Remark | |
| Light source | Intense pulsed light | Intense pulsed light | SE | |
| Wavelength | 480-1200 nm<br>590-1200 nm<br>640-1200 nm | 480nm-1200nm<br>590nm-1200nm<br>640nm-1200nm | SE | |
| Deliver system | Sapphire | Sapphire | SE | |
| Energy density | 1~40J/cm2 | 1-50J/cm² | Analysis 1 | |
| Max. Power | 2200VA | 2200W | SE | |
| permanent hair reduction | | | | |
| Wavelength Range (nm) | 640-1200 | 640-1200 | SE | |
| Energy Range (J/cm²) | 11-40 | 10-44 | Analysis 1 | |
| Pulse Width (ms) | 16-32 | 3-14 | Analysis 1 | |
| Pulse Delay (ms) | 5-14 | 16-32 | Analysis 1 | |
| Spot Size (mm²) | 15*50 | 12*30 | Analysis 1 | |
| pigmented lesions | | | | |
| Wavelength Range (nm) | 480-1200 | 480-1200 | SE | |
| Energy Range (J/cm²) | 10-40 | 12-44 | Analysis 1 | |
| Pulse Width (ms) | 16-32 | 3-9 | Analysis 1 | |
| Pulse Delay (ms) | 3-8 | 16-32 | Analysis 1 | |
| Spot Size (mm²) | 15*50 | 12*30 | Analysis 1 | |
| vascular lesions | | | | |
| Wavelength Range (nm) | 590-1200 | 590-1200 | SE | |
| Energy Range (J/cm²) | 10-40 | 10-42 | Analysis 1 | |
| Pulse Width (ms) | 16-32 | 3-8 | Analysis 1 | |
| Pulse Delay (ms) | 5-9 | 16-32 | Analysis 1 | |
| Spot Size (mm²) | 15*50 | 12*30 | Analysis 1 | |
# Table 7-1 General Comparison
{9}------------------------------------------------
Table 7-2 Performance Comparison(IPL)
## Analysis 1
The proposed device is different in energy density, energy range, pulse width and pulse delay from the predicate device. Buy the IPL system of the proposed device has passed the IEC60601-2-57 and IEC62471 test, the IPL performance of the proposed device can be guaranteed. By complying with non-clinical test conducted, the proposed device is determined to be substantially equivalency with predicate device.
{10}------------------------------------------------
| Item | Proposed Device | Predicate Device 2<br>Diode Laser 808nm<br>K192516 | Remark |
|----------------------|--------------------------|----------------------------------------------------|------------|
| Laser Type | Diode Laser | Diode Laser | SE |
| Laser Classification | Class IV | Class IV | SE |
| Laser Wavelength | 808 nm | 808 nm | SE |
| Spot Size | 13*13mm² | 1.2 cm² | Analysis 2 |
| Fluence | 1~69J/c m² | 1-70J/ cm² | Analysis 2 |
| Frequency | SHR:1-10Hz<br>LHR:1-3 Hz | 1-20 Hz | Analysis 2 |
| | Pulse Duration | 5~400ms | 5~400ms |
| | Table 7-3 Performance Comparison (Diode Laser ) | | | |
|--|-------------------------------------------------|--|--|--|
|--|-------------------------------------------------|--|--|--|
## Analysis 2
The proposed device is different in Spot Size, Fluence, Frequency and Pulse Duration from the predicate device. Buy the Diode Laser system of the proposed device has passed the IEC60601-2-22 and IEC60825-1 test, the Diode Laser performance of the proposed device can be guaranteed. By complying with non-clinical test conducted, the proposed device is determined to be substantially equivalency with predicate device.
{11}------------------------------------------------
| | | Predicate Device1 | Predicate Device 2 | |
|------------------------------------------------|-----------------------------------------------------------------|---------------------------------|-----------------------------------------------|------------|
| Item | Proposed Device | IPL Therapy | Diode Laser 808nm | Remark |
| | | Machine (K161286) | (K192516) | |
| Patient Contact Materials and Biocompatibility | | | | |
| Patient<br>Contact<br>Materials | IPL + Diode Laser<br>handpiece | IPL handpiece | Diode Laser handpiece | SE |
| Cytotoxicity | No Cytotoxicity | No Cytotoxicity | No Cytotoxicity | |
| Sensitization | No evidence of<br>sensitization | No evidence of<br>sensitization | No evidence of<br>sensitization | SE |
| Irritation | No evidence of<br>irritation | No evidence of<br>irritation | No evidence of<br>irritation | |
| EMC, Electrical and Laser Safety | | | | |
| Electrical<br>Safety | Comply with IEC<br>60601-1, IEC<br>60601-2-22,<br>IEC60601-2-57 | Comply with IEC<br>60601-1 | Comply with IEC<br>60601-1, IEC<br>60601-2-22 | SE |
| EMC | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | SE |
| Laser Safety | Comply with<br>IEC 60601-2-22,<br>IEC 60825 ,<br>IEC60601-2-57 | Comply with<br>IEC60601-2-57 | Comply with IEC<br>60601-2-22, IEC<br>60825 | SE |
| Power Supply | AC110V,<br>50/60Hz | AC 110V 50Hz | AC 110V/60Hz or<br>250V 50Hz | SE |
| Dimension | 620x610x1290mm | 430x 590 x 1360 mm | 598x440x1093mm | Analysis 3 |
| Weight | 78kg | 45kg | 60Kg | Analysis 3 |
| | | Table 7-4 Safety Comparison |
|--|---|-----------------------------|
| | ้ | |
## Analysis 3
The proposed device is different in Dimension and Weight from the predicate device. but the proposed device has passed the IEC 60601-1-2, IEC 60601-2-22, IEC60601-2-57, IEC 60825 test, we believe these differences will not affect the effectiveness and safety compared with the predicate device., the proposed device is determined to be substantially equivalency with predicate device.
{12}------------------------------------------------
## 10. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.