K193311 · Knox Medical Diagnostics, Inc. · BZH · Mar 25, 2020 · Anesthesiology
Device Facts
Record ID
K193311
Device Name
Aluna
Applicant
Knox Medical Diagnostics, Inc.
Product Code
BZH · Anesthesiology
Decision Date
Mar 25, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1860
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
Aluna is intended for monitoring FEV1 (Forced exhalation in the first second) and PEF (Peak Expired Flow Rate) for home use. The device is designed for children 5 years of age or older, adolescent and adult subjects. Additionally, the device may be used by clinicians for in-office monitoring.
Device Story
Aluna is a handheld peak flow meter; captures lung health data (FEV1, PEF) via Lilly-type pneumotachometer using differential pressure sensor across stainless-steel mesh. Device connects via Bluetooth 4.0 LE to mobile app (iOS); app provides user interface, data processing, display, and cloud storage. Includes gamification to incentivize correct spirometry technique. Used by patients at home or clinicians in-office. Output allows tracking of lung function trends; aids healthcare providers in monitoring respiratory conditions and clinical decision-making.
Clinical Evidence
Bench testing only. Performance verified per 2005 ATS/ERS Task Force standards for FEV1/PEF measurement accuracy and repeatability. Additional testing included biocompatibility (ISO 10993/18562), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), battery safety (IEC 62133-2), wireless coexistence (ANSI PC63.27), and usability (IEC 62366-1).
Technological Characteristics
Lilly-type pneumotachometer; differential pressure sensor; stainless-steel mesh; polymers/stainless steel construction. Dimensions: 90.3 x 43 x 115.5 mm; weight: 95g. Power: rechargeable lithium-ion battery. Connectivity: Bluetooth 4.0 LE. Non-sterile. Standards: ISO 10993, ISO 18562, IEC 62304, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 62133-2, ANSI PC63.27, IEC 62366-1.
Indications for Use
Indicated for monitoring FEV1 and PEF in children 5 years or older, adolescents, and adults for home or in-office clinical use.
Regulatory Classification
Identification
A peak-flow meter for spirometry is a device used to measure a patient's maximum ventilatory flow rate.
Predicate Devices
Smart One / MIR Medical International Research (K181666)
Wing Smart FEV1 and Peak Flow Meter / Sparo Inc. (K152276)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 25, 2020
Knox Medical Diagnostics, Inc. % Pierre Bounaud Senior Regulatory Specialist Acknowledge Regulatory Strategies 2251 San Diego Ave, Suite B-257 San Diego, California 92110
Re: K193311
Trade/Device Name: Aluna Regulation Number: 21 CFR 868.1860 Regulation Name: Peak-Flow Meter For Spirometry Regulatory Class: Class II Product Code: BZH Dated: February 26, 2020 Received: February 27, 2020
Dear Pierre Bounaud:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael Ryan Division Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193311
Device Name Aluna
Indications for Use (Describe)
Aluna is intended for monitoring FEV1 (Forced exhalation in the first second) and PEF (Peak Expired Flow Rate) for home use. The device is designed for children 5 years of age or older, adolescent and adult subjects. Additionally, the device may be used by clinicians for in-office monitoring.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <div> <span> ☑ Demonstration Use (§1.377-3(c), Subject to Exceptions) </span> </div> |
| <div> <span> ☑ On-The-Spot Testing (§1.377-3(d), Subject to Exceptions) </span> </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Aluna. The logo consists of a circle that is divided into two halves. The left half of the circle is a solid color, while the right half of the circle has two curved lines inside of it. Below the circle is the word "aluna" in lowercase letters.
## 510(k) Summary
### DATE PREPARED
March 25, 2020
#### MANUFACTURER AND 510(k) OWNER
Aluna 345 California Street, Suite 600, San Francisco, CA 94104, USA Telephone: Official Contact: Charvi Shetty, CEO
### REPRESENTATIVE/CONSULTANT
| Pierre Bounaud, Ph.D. |
|----------------------------------------|
| Allison C. Komiyama, Ph.D., R.A.C. |
| Acknowledge Regulatory Strategies, LLC |
| Telephone: +1 (619) 458-9547 |
| Email: pbounaud@acknowledge-rs.com |
| akomiyama@acknowledge-rs.com |
| Website : www.Acknowledge-RS.com |
#### DEVICE CLASSIFICATION
| Proprietary Name/Trade Name: | Aluna |
|------------------------------|-------------------------------------------------------------------------------------|
| Common Name: | Peak-flow meter |
| Regulation Number: | 21 CFR 868.1860 |
| Class: | II |
| Product Code: | BZH |
| Premarket Review: | OHT1/ENT, Sleep Disordered Breathing, Respiratory and<br>Anesthesia Devices (DHT1C) |
| Review Panel: | Anesthesiology |
#### PREDICATE DEVICE IDENTIFICATION
Aluna is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|--------------------------------------------------|-------------------|
| K181666 | Smart One / MIR Medical International Research | ✓ |
| K152276 | Wing Smart FEV1 and Peak Flow Meter / Sparo Inc. | |
### DEVICE DESCRIPTION
Aluna is a small hand-held peak flow meter that captures lung health information and makes it available through a mobile application. Aluna is intended for monitoring FEV1 (Forced Expiratory Volume in one second) and PEF (Peak Expiratory Flow) for over-the-counter use. Aluna is designed for use by pediatric to adult users. Aluna is not recommended for children under 5 years of age. Additionally, the device may be used by clinicians for in-office monitoring.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo for a company called "aluna". The logo consists of the word "aluna" in lowercase letters, with a stylized circle above it. The circle is divided into three sections by two curved lines, creating a crescent moon shape on the right side. The logo is simple and modern, with a clean font and a light blue color scheme.
### INDICATIONS FOR USE
Aluna is intended for monitoring FEV1 (Forced exhalation in the first second) and PEF (Peak Expired Flow Rate) for home use. The device is designed for children 5 years of age or older, adolescent and adult subjects. Additionally, the device may be used by clinicians for in-office monitoring.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Aluna is substantially equivalent to the predicate device based on the information summarized here:
The subject device has a similar design, dimensions and uses similar materials as the devices cleared in K181666 and in K152276. The subject device has the same intended use and similar technological characteristics to the devices cleared in K181666 and in K152276.
Technological differences of the subject device compared to the predicate device cleared in K181666 include:
- . rechargeable lithium ion battery
- . data storage on a cloud server
- . game function
These technological differences have undergone testing to ensure the device is as safe and effective as the predicate.
The game function in the subject device is similar to the game function in the predicate device cleared in K152276. The role of the game function is to encourage users to take readings and to perform correct spirometry measurements with the use of game rewards as incentives.
Substantial equivalence comparison between subject devices cleared in K181666 and in K152276 is summarized in the table below.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Aluna. The logo consists of a circle with a vertical line through the center and two curved lines on the right side, resembling a crescent moon. Below the circle is the word "aluna" in lowercase letters and a light blue color.
| | Subject Device | Predicate Device | Predicate Device |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Aluna | Smart One<br>(K181666) | Wing Smart FEV1 and<br>Peak Flow Meter<br>(K152276) |
| Indications<br>for Use | Aluna is intended for<br>monitoring FEV1 (Forced<br>exhalation in the first<br>second) and PEF (Peak<br>Expired Flow Rate) for<br>home use. The device is<br>designed for pediatric to<br>adult users. Aluna is not<br>recommended for<br>children under 5 years of<br>age. Additionally, the<br>device may be used by<br>clinicians for in-office<br>monitoring. | Smart One is intended<br>for home use by<br>patients to monitor PEF<br>(Peak Expiratory Flow)<br>and FEV1 (Forced<br>Expiratory Volume in<br>one second). The<br>device is designed for<br>children greater than<br>five years of age,<br>adolescent and adult<br>subjects. | Wing is intended for<br>monitoring FEV1<br>(Forced exhalation in<br>the first second) and<br>PEF (Peak Expired Flow<br>Rate) for home use.<br>The device is designed<br>for pediatric to adult<br>users. Wing is not<br>recommended for<br>children under 5 years<br>of age. |
| Product<br>Code /<br>Regulation<br>Number | BZH / 21 CFR 868.1860 | BZH / 21 CFR 868.1860 | BZH / 21 CFR 868.1860 |
| Regulation<br>Description | Peak-flow meter for<br>spirometry | Peak-flow meter for<br>spirometry | Peak-flow meter for<br>spirometry |
| Rx/OTC | OTC | OTC | OTC |
| Mode of<br>operation | Aluna is a Lilly type<br>pneumotachometer that<br>uses a differential<br>pressure sensor to<br>measure the pressure<br>difference across the<br>stainless-steel mesh. The<br>resulting pressure drop is<br>used to calculate FEV1<br>and PEF. | The device is equipped<br>with a plastic<br>mouthpiece connected<br>to a turbine flow meter<br>based on the infrared<br>interruption principle.<br>The device detects the<br>signals generated by<br>the turbine and<br>measures the<br>exhalation flow. At the<br>end of the expiration,<br>the device calculates<br>the PEF and FEV1. | The Wing uses an<br>acoustic transducer to<br>measure frequency of<br>oscillation caused by<br>the airstream to<br>calculate FEV1 and PEF |
| Dimensions | 90.3 mm x 43 mm x<br>115.5 mm | 49 mm x 109 mm x 21<br>mm | 3.2 x 2.4 x 1.5 inches<br>(81 x 61 x 37mm) |
| Weight | 95 g | 60.7 g | 70g |
| FEV1 Max | 8 L | 10 L | 0.99 L |
| | Subject Device | Predicate Device | Predicate Device |
| | Aluna | Smart One<br>(K181666) | Wing Smart FEV1 and<br>Peak Flow Meter<br>(K152276) |
| FEV1<br>accuracy | $\pm$ 3% or 0.05 L whichever<br>is greater | $\pm$ 3% or 0.1 L whichever<br>is greater | $\pm$ 5% or 0.1 L whichever<br>is greater |
| PEF max | 14 L/s | 16 L/s | 15 L/s |
| PEF accuracy | $\pm$ 10% or 0.3 L/s<br>whichever is greater | $\pm$ 10% or 0.4 L/s<br>whichever is greater | $\pm$ 12% or 0.417 L/s<br>whichever is greater |
| Materials | Polymers, stainless steel | Plastic body | Plastic body |
| Transmission<br>Method | Bluetooth 4.0 LE | Bluetooth 4.0 LE | Audio cable |
| Power<br>Source | Rechargeable battery | 2x1.5V AAA batteries | Smartphone battery |
| Smartphone<br>compatibility | iPhone (iOS 11 or later) | iPhone, Android<br>devices | iPhone and iPad (iOS 9<br>or later) |
| Sterilization | Non-sterile | Non-sterile | Non-sterile |
| App<br>functions | • Primary user interface<br>• Data acquisition and<br>processing<br>• Data display<br>• Data transmission to<br>server | • Primary user<br>interface<br>• Data acquisition and<br>processing<br>• Data display and<br>storage | • Primary user<br>interface<br>• Data acquisition and<br>processing<br>• Data display<br>• Data transmission to<br>server |
| Game<br>function | Yes | No | Yes |
| Data Storage | Cloud server | Phone | Cloud server |
| Data sharing | Yes, via email | Yes, via email | Yes, via email |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a logo with the word "aluna" in lowercase letters. Above the word is a circular symbol that resembles a stylized moon or a partially shaded circle. The logo has a clean and modern design, with a light and airy feel.
### SUMMARY OF NON-CLINICAL TESTING
The following tests were performed to demonstrate safety based on current industry standards:
- Biocompatibility testing per ISO 10993-1, ANSI/AAMI/ISO 10993-5, ISO 10993-10, ISO . 18562-1, ISO 18562-2, and ISO 18562-3
- . Software testing per IEC 62304
- . Electrical safety testing per IEC 60601-1, IEC 60601-1-6, and IEC 60601-1-11
- EMC testing per IEC 60601-1-2
- Battery testing per IEC 62133-2
- Wireless coexistence testing per ANSI PC63.27
- Non-clinical performance bench testing including: ●
- o FEV1/PEF measurement accuracy and repeatability per 2005 American Thoracic Society document ATS/ERS Task Force: Standardization of Lung Function Testing
- o Ambient temperature and ambient pressure testing
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Aluna. The logo consists of a circle divided into three sections, with the left section being a half-circle and the other two sections being quarter-circles. Below the circle is the word "aluna" in lowercase letters.
- Usability testing per IEC 62366-1
The results of these tests indicate that Aluna is substantially equivalent to the predicate device.
#### CONCLUSION
Based on the testing performed, including biocompatibility, software testing, electrical safety and EMC testing, battery testing, wireless coexistence testing, and non-clinical performance bench testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar indications for use, technological characteristics, and performance characteristics for the proposed Aluna are assessed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.