K193306 · GE Medical Systems SCS · LLZ · Nov 17, 2020 · Radiology
Device Facts
Record ID
K193306
Device Name
PROView
Applicant
GE Medical Systems SCS
Product Code
LLZ · Radiology
Decision Date
Nov 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Prostate gland segmentation
Deep learning model
—
—
Database of MRI prostate exams
—
—
—
Indications for Use
PROView is a medical diagnostic software that is designed to provide easy processing, analysis, reviewing and communication of 3D reconstructed images and their relationship to originally acquired images from MR Scanning devices. The combination of acquired images, reconstructed images, annotations, and measurements performed by the clinician are intended to provide to the referring physician clinically relevant information that may aid in diagnosis and treatment planning.
Device Story
PROView is MR image analysis software used by clinicians/radiologists to review multi-parametric prostate exams. It takes T2-weighted, DWI, and DCE MR images as input. The device uses a deep learning algorithm to perform automatic 3D segmentation of the prostate gland, allowing for manual adjustment if needed. It calculates prostate volume and PSA density, maps lesions to sectors, and facilitates scoring per PI-RADS v2.1 guidelines. The software generates a comprehensive report containing measurements and images. This output is used by clinicians to aid in diagnosis and treatment planning for patients with known or suspected prostate lesions. The device operates as a clinical workflow tool for prostate cancer assessment.
Clinical Evidence
Bench testing only. No clinical trials were conducted. Validation of the deep learning-based automatic prostate segmentation algorithm was performed using a database of MRI prostate exams representative of clinical scenarios, protocols, and patient demographics. Results confirmed the algorithm met acceptance criteria and improved performance on volume accuracy.
Technological Characteristics
Software-based MR image analysis tool. Features deep learning-based automatic prostate segmentation algorithm for T2-weighted MRI acquisitions. Supports PI-RADS v2.1 reporting. Designed under 21 CFR 820 and ISO 13485 quality systems. Software level of concern is Moderate.
Indications for Use
Indicated for clinicians, radiologists, or physicians to review multi-parametric prostate MR images following PI-RADS guidelines. Used for assessment of prostate gland volume and findings analysis in male patients with known or suspected prostate lesions. Measurements and scoring are included in a report for referring physicians.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
November 17, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below the words "U.S. FOOD & DRUG".
GE Medical Systems SCS % Hong Cui Regulatory Affairs Director 3000 North Grandview Boulevard-HQ WAUKESHA WI 53188
Re: K193306
Trade/Device Name: PROView Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: October 21, 2020 Received: October 22, 2020
Dear Hong Cui:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193306
Device Name PROView
#### Indications for Use (Describe)
PROView is an aiding tool for the clinicians to review multi-parametric resonance (MR) inages following PI-RADS guidelines. It displays acquired and reformatted data for visualization and provides tools for assessment of the prostate gland volume and findings analysis in patients with known or suspected prostate lesions. Measurements and associated scoring are included in a report for communication to referring physicians. It is intended for use by professionals, such as clinicians, radiologists, or physicians. The clinician remains ultinately responsible for the final assessment and diagnosis based on state-of-the-art practices, clinical judgment and interpretation of prostate images or quantitative data.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three small white curved lines around the outside of the circle, evenly spaced. The logo is simple and recognizable.
GE Healthcare 510(k) Premarket Notification Submission K193306
### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | November 26, 2019 |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems SCS |
| | Establishment Registration Number - 9611343 |
| | 283 rue de la Miniere |
| | 78530 Buc, France |
| Primary Contact Person: | Hong Cui<br>Regulatory Affairs Director<br>GE Healthcare, (GE Medical Systems, LLC)<br>3000 N Grandview Blvd.,<br>Waukesha, WI - 53188<br>Phone: (262) 506-8238<br>Email: hong.cui@ge.com |
| Secondary Contact Person: | Elizabeth Mathew<br>Senior Regulatory Affairs Manager<br>Phone: (262) 424-7774<br>Email: Elizabeth.Mathew@ge.com |
| Proposed Device: | |
| Device Name: | PROView |
| Common/Usual Name: | MR image analysis software |
| Regulation number/ Product Code: | 21 CFR 892.2050 Picture archiving and communications system/<br>LLZ |
Classification: Class II
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined in a stylized script, enclosed within a blue circle. There are four stylized water droplets or flourishes evenly spaced around the perimeter of the circle, adding a decorative element to the design.
## GE Healthcare 510(k) Premarket Notification Submission
Predicate Device:
| Device Name: | <b>DynaCAD</b> |
|-------------------------------------|---------------------------------------------------------------------|
| 510(k) number: | K192200 cleared on Oct 9, 2019 |
| Regulation number/<br>Product Code: | 21 CFR 892.2050 Picture archiving and communication system /<br>LLZ |
| Classification: | Class II |
| Manufacturer: | Invivo Corporation, USA |
### Device Description/
Technology:
PROView offers a guided workflow for the review, assessment and reporting of multi-parametric MR prostate exams. From inputting clinical information, measuring prostate and lesion volume to scoring lesions to form a comprehensive MR report, PROView offers a simple workflow per PI-RADS™ v2.1 guideline.
PROView Processes data from a single date.
The PROView workflow includes:
- Prostate volume extracted from automatic organ segmentation
- PSA Density
- Lesion(s) mapping to sectors and measurement ●
- Scoring of T2-weighted, diffusion weighted imaging (DWI) and, when ● applicable, dynamic contrast enhanced (DCE) acquisitions.
- Automatically generated report with all measurements and images ●
Prostate volume can be automatically calculated by defining the contours of the prostate gland with the use of a deep learning algorithm, or through a manual method. Users can cancel or switch to manual prostate gland volume definition if the automatic prostate gland segmentation fails or provides unsatisfactory results
#### Intended Use:
PROView is a medical diagnostic software that is designed to provide easy processing, analysis, reviewing and communication of 3D reconstructed images and their relationship to originally acquired images from MR Scanning devices. The combination of acquired images, reconstructed images, annotations, and measurements performed by the clinician are intended to provide to the referring physician clinically relevant information that may aid in diagnosis and treatment planning.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. There are decorative, wave-like elements surrounding the circle, giving the logo a classic and recognizable appearance.
## GE Healthcare 510(k) Premarket Notification Submission
Indication for Use: PROView is an aiding tool for the clinicians to review multi-parametric prostate magnetic resonance (MR) images following PI-RADS guidelines. It displays acquired and reformatted data for visualization and provides tools for assessment of the prostate gland volume and findings analysis in patients with known or suspected prostate lesions. Measurements and associated scoring are included in a report for communication to referring physicians. It is intended for use by professionals, such as clinicians, radiologists, or physicians. The clinician remains ultimately responsible for the final assessment and diagnosis based on state-of-the-art practices, clinical judgment and interpretation of prostate images or quantitative data. Technological Characteristic: The goal of the new PROView software algorithm is to provide an automatic segmentation of the prostate on MRI T2 weighted acquisitions. It is a routine anatomical acquisition (as opposed to functional MRI), routinely done by the clinicians for prostate cancer assessment, as it is part of PIRADS guidelines. The algorithm provides a fully automatic segmentation of the prostate, based on a deep learning model. Comparison: The below comparison identifies the similarities and differences of the proposed
PROView to the DynaCAD Prostate module of the predicate device DynaCAD (K192200) to which substantial equivalency is claimed.:
| Specification | DynaCAD<br>K192200<br>(DynaCAD Prostaet) | Proposed Device:<br>PROView |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Targeted clinical<br>condition | Male patient with suspected<br>or known prostate lesions | Male patient with suspected<br>or known prostate lesions |
| Anatomy | Prostate | Prostate |
| Imaging<br>modality | MRI | MRI |
| Gland<br>segmentation | Automatically performs a<br>3D segmentation of the<br>gland. Users can alter or<br>make adjustments to the<br>segmented results in all<br>three planes. The resulting<br>segmentation reports<br>overall gland volume and<br>sets the stage for UroNav<br>MR/US guided fusion | The software provides a fully<br>automatic segmentation of<br>the prostate, based on a deep<br>learning model. Contour of<br>the prostate gland can be<br>adjusted by the user. Prostate<br>volume is extracted from<br>automatic gland<br>segmentation. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. There are three water droplet shapes surrounding the circle. The logo is simple and recognizable, representing the company's brand identity.
# GE Healthcare
510(k) Premarket Notification Submission
| | biopsy | |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Gland volume | The resulting segmentation<br>reports overall gland<br>volume and sets the stage<br>for UroNav MR/US guided<br>fusion biopsy | Prostate volume is extracted<br>from automatic gland<br>segmentation after validation<br>of the contour by the user. |
| Segmentation<br>algorithm type | Model-based automatic<br>prostate gland segmentation | Automatic segmentation of<br>the prostate based on a deep<br>learning model. |
| Standardized<br>report | Following PI-RADS™ v2 | Following PI-RADS™ v2.1 |
### Determination of Substantial Equivalence:
PROView has successfully completed the required design control testing per GE's quality system. PROView was designed and will be manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485. The following quality assurance measures have been applied to the development of the device:
- Risk Analysis ●
- . Requirements Reviews
- Design Reviews ●
- Performance testing (Verification) ●
- Safety testing (Verification)
- Algorithm Qualification (Validation) ●
Engineering has validated PROView algorithm's capability of automatic segmentation based on deep learning technique by using a database of MRI prostate exams. This database of exams is considered as representative of the clinical scenarios where PROView is intended to be used, with consideration of the different protocols, practices and ethnics factors. The results and feedback concluded that the algorithm meets the acceptance criteria and improves performance on volume accuracy.
The substantial equivalence was also based on software documentation for a "Moderate" level of concern device.
#### Conclusion:
GE Healthcare considers PROView to be as safe, as effective, and performance is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.