← Product Code [DXO](/productcode/DXO) · K193279

# Comet II Pressure Guidewire (K193279)

_Boston Scientific Corporation · DXO · Dec 23, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K193279

## Device Facts

- **Applicant:** Boston Scientific Corporation
- **Product Code:** [DXO](/productcode/DXO.md)
- **Decision Date:** Dec 23, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2870
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Comet II Pressure Guidewire measures the blood pressure gradient across coronary and peripheral lesions during endovascular procedures. The Comet II Pressure Guidewire may also be used as a coronary or peripheral guidewire for interventional treatments.

## Device Story

Comet II is a 0.014" coronary/peripheral guidewire with an integrated distal optical pressure sensor; used to direct catheters and measure blood pressure gradients across lesions. Operated by physicians during endovascular procedures. Guidewire connects via detachable cable to a Boston Scientific signal-processing module (e.g., iLab Polaris Multi-Modality Guidance System). System displays aortic pressure, distal pressure, and calculated fractional flow reserve (FFR). Radiopaque, shapeable tip aids placement; torque device included. Provides real-time hemodynamic data to assist clinical decision-making regarding lesion severity and treatment strategy.

## Clinical Evidence

No clinical data; substantial equivalence demonstrated via non-clinical bench testing.

## Technological Characteristics

0.014" guidewire; optical pressure sensor located ~3 cm from distal tip; radiopaque, shapeable tip; detachable optical cable connector; torque device; single-use; sterilized.

## Regulatory Identification

A catheter tip pressure transducer is a device incorporated into the distal end of a catheter. When placed in the bloodstream, its mechanical or electrical properties change in relation to changes in blood pressure. These changes are transmitted to accessory equipment for processing.

## Predicate Devices

- Comet™ Pressure Guidewire ([K151610](/device/K151610.md))

## Reference Devices

- iLab Polaris™ Multi-Modality Guidance System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

December 23, 2019

Boston Scientific Corporation Kimberly Berg Sr. Regulatory Affairs Specialist Three Scimed Place Maple Grove, Minnesota 55311-1566

Re: K193279

Trade/Device Name: Comet™ II Pressure Guidewire Regulation Number: 21 CFR 870.2870 Regulation Name: Catheter Tip Pressure Transducer Regulatory Class: Class II Product Code: DXO, DQX Dated: November 25, 2019 Received: November 27, 2019

Dear Kimberly Berg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K193279

Device Name Comet™ II Pressure Guidewire

Indications for Use (Describe)

The Comet II Pressure Guidewire is indicated to direct a catheter through a blood vessel and measure physiological parameters in the coronary and peripheral blood vessels.

Type of Use (Select one or both, as applicable)

| <div><span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|------------------------------------------------------------------------------------------------|
| <div><span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div>  |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Boston Scientific. The words "Boston Scientific" are written in a serif font. The color of the text is blue. 

# 510(k) Summary per 21 CFR §807.92

| Sponsor                         | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA                                           |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name and<br>Information | Kimberly Berg<br>Three Scimed Place<br>Maple Grove, MN 55311-1566<br>Phone: 763-494-2701<br>Fax: 763-494-2981<br>e-mail: Kimberly.Berg@bsci.com |
| Prepared by                     | Kimberly Berg<br>November 25, 2019                                                                                                              |
| Proprietary Name                | Comet™ II Pressure Guidewire                                                                                                                    |
| Common Name                     | Transducer, Pressure, Catheter Tip                                                                                                              |
| Product Code                    | DXO, DQX                                                                                                                                        |
| Classification                  | Class II, 21 CFR Part 870. 2870                                                                                                                 |
| Predicate Device                | Comet™ Pressure Guidewire<br>K151610 October 6, 2019                                                                                            |

{4}------------------------------------------------

300 Boston Scientific Way Marlborough, MA 01752-1234 508-683-4000 www.bostonscientific.com

### Device Description

Comet™ II is a coronary and peripheral quidewire with a pressure sensor mounted approximately 3 cm proximally to the distal end of the straight tip. The optical pressure sensor is integrated within the distal region of a 0.014" quidewire to measure blood pressure gradient changes across coronary and peripheral lesions during endovascular procedures. The tip is radiopaque to help with guidewire placement and shapeable by the user. The guidewire is attached to an optical cable through a detachable connector. The optical cable has a connector on the proximal end for interface with an ancillary Boston Scientific signal processing module. When connected to the signal-processing module, a Boston Scientific iLab Polaris™ Multi-Modality Guidance System displays various physiological parameters including aortic pressure, distal pressure and a fractional flow reserve (FFR) value calculated from the ratio of distal pressure over aortic pressure. A torque device is preloaded onto the proximal end of the wire. The pressure guidewire, optical cable, and torque device are all connected in the single use sterilized packaging. The intended use of Comet II remains within the scope of the predicate intended use.

## Indications for Use / Intended Use

#### Intended Use

The Comet II Pressure Guidewire measures the blood pressure gradient across coronary and peripheral lesions during endovascular procedures. The Comet II Pressure Guidewire may also be used as a coronary or peripheral guidewire for interventional treatments.

#### Indications for Use

The Comet II Pressure Guidewire is indicated to direct a catheter through a blood vessel and measure physiological parameters in the coronary and peripheral blood vessels.

#### Comparison of Technological Characteristics

The Comet II Pressure Guidewire incorporates substantially equivalent design, packaging, fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate, Comet Pressure Guidewire, K151610.

#### Non-clinical Performance Data

Determination of substantial equivalence is based on an assessment of non-clinical performance bench testing data.

#### Bench Testing

Bench testing was performed to evaluate physical integrity, functionality, and performance of the Comet II Pressure Guidewire. Performance criteria includes dimensional requirements, pressure gradient measurement (sensor) requirements, and interface with compatible devices.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Boston Scientific. The words "Boston Scientific" are written in a serif font. The color of the text is blue. 

300 Boston Scientific Way Marlborough, MA 01752-1234 508-683-4000 www.bostonscientific.com

## Clinical Testing

Performance testing from clinical studies is not required to demonstrate substantial equivalence of the Comet II.

#### Conclusion

Based on the indications for use, technological characteristics, and performance testing the Comet II Pressure Guidewire has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the Comet Pressure Guidewire, K151610.

---

**Source:** [https://fda-staging.innolitics.com/device/K193279](https://fda-staging.innolitics.com/device/K193279)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
