← Product Code [FOZ](/productcode/FOZ) · K193278

# MedSource CathMED IV Catheter (K193278)

_Medsource Labs · FOZ · Aug 7, 2020 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K193278

## Device Facts

- **Applicant:** Medsource Labs
- **Product Code:** [FOZ](/productcode/FOZ.md)
- **Decision Date:** Aug 7, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The MedSource CathMED IV Catheter is indicated to sample blood or administer fluids intravenously. The MedSource CathMED IV Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.

## Device Story

Intravascular catheter for blood sampling or fluid administration; used in clinical settings by healthcare providers. Device consists of a sharp needle with an over-the-needle catheter, protective cover, hub, and gauge chamber. Operation: clinician inserts needle/catheter into vein; catheter advanced off needle; needle removed; catheter remains in vein for short-term access. Provides venous access for therapy or sampling. No needle-stick prevention feature present; labeling includes warning regarding absence of safety mechanism. Available in 14G to 24G sizes.

## Clinical Evidence

Bench testing only. Conformance to ISO 10555-1, ISO 10555-5, ISO 594-1, ISO 594-2, and USP <788> standards. Biocompatibility established via reference to predicate device data. Sterilization validated per ISO 11135:2014.

## Technological Characteristics

Materials: Polyurethane/PTFE catheter, stainless steel needle, K-Resin hub/body, Polyacetal (POM) catheter holder. Design: Over-the-needle peripheral catheter with Luer taper fittings. Sterilization: Terminal Ethylene Oxide (EtO). No integrated software or electronics.

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- MedSource IV Safety Catheter ([K131555](/device/K131555.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 7, 2020

MedSource Labs Emilie Andrews Quality Technician 8600 Shelby Court Chanhassen, MN 55317

Re: K193278

Trade/Device Name: MedSource CathMED IV Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: July 13, 2020 Received: July 13, 2020

Dear Emilie Andrews:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Payal Patel Assistant Director, General Hospital DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K193278

Device Name MedSource CathMED IV Catheter

#### Indications for Use (Describe)

The MedSource CathMed IV Catheter is indicated to sample blood or administer fluids intravenously. The MedSource CathMED IV Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D)  | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510 K-Summary- K193278

#### Date of Prepared: August 6, 2020

### A. Submitter:

MedSource International, LLC 8600 Shelby Court Chanhassen, MN 55317

#### B. Contact Person:

Emilie Andrews, Quality and Regulatory Affairs 8600 Shelby Court Chanhassen. MN 55317 Phone: 952-472-0131

#### C. Proposed Device:

Product Code: FOZ Trade Name: MedSource CathMED IV Catheter Common Name: Intravascular catheter Classification Name: Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days Regulatory Reference: 21 CFR §880.5200 Review Panel: General Hospital Classification: Class II

#### D. Predicate:

510(K): K131555 Product Code: FOZ Trade Name: MedSource IV Safety Catheter Common Name: Intravascular catheter Classification Name: Catheter, Intravascular, Therapeutic, Short-Term Less Than 30Days Regulatory Reference: 21 CFR §880.5200 Review Panel: General Hospital Classification: Class II

#### E. Indications for Use:

The MedSource CathMED IV Catheter is indicated to sample blood or administer fluids intravenously. The MedSource CathMED IV Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.

{4}------------------------------------------------

## F. Device Description:

The MedSource CathMED IV Catheter is indicated to sample blood or administer fluids intravenously. The MedSource CathMED IV Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.

Each MedSource CathMED IV Catheter is comprised of the following components: protective needle cover, color-coded catheter hub, radiopaque catheter tube, Teflon catheter holder, needle guide, medical grade stainless steel needle hub, porous plug and gauge chamber. The MedSource CathMED IV Catheter components combine for an ergonomic design.

The device inserts a catheter intravenously by inserting a sharp needle with an over-theneedle catheter. Once the needle and catheter have been passed into a vein the catheter can be advance off the needle is the needle is removed while the catheter remains in the vein for short term use to sample blood or administer fluids.

The MedSource CathMED IV Catheter gauges are color coded for positive identification and range from 14G to 24G

ORANGE = 14G GREY = 16G GREEN = 18G PINK = 20G BLUE = 22G YELLOW = 24G

## G. SUMMARY OF COMPARISON TABLE OF PROPOSED DEVICE and PREDICATE

The proposed device is substantially equivalent to the predicate device the MedSource Safety Catheter (K131555).

| Comparison<br>Point                                 | Submission Device<br>MedSource CathMED IV<br>Catheter                                                                                                                                                                                                                                                                                                                | Predicate Device<br>MedSource IV Safety<br>Catheter (K131555)                                                                                                                                                                                                                                                                                                | Discussion | Comments                                                                                                                                                                                                                                                                    |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technology                                          |                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                              |            |                                                                                                                                                                                                                                                                             |
| Indications for<br>Use                              | The MedSource<br>CathMED IV Catheter is<br>indicated to sample blood<br>or administer fluids<br>intravenously. The<br>MedSource IV Catheter<br>may be used for any<br>patient population with<br>consideration given to<br>adequacy for the vascular<br>anatomy appropriateness<br>for the solution being<br>administered and                                        | The MedSource IV Safety<br>Catheter is indicated to<br>sample blood or administer<br>fluids<br>intravenously. The<br>MedSource IV Safety<br>Catheter may be used for<br>any patient population<br>with consideration given to<br>adequacy for the vascular<br>anatomy appropriateness<br>for the<br>solution being                                           | Same       |                                                                                                                                                                                                                                                                             |
|                                                     | duration of the therapy.                                                                                                                                                                                                                                                                                                                                             | administered and duration<br>of the therapy.                                                                                                                                                                                                                                                                                                                 |            |                                                                                                                                                                                                                                                                             |
| Intended Use                                        | The MedSource<br>CathMED IV Catheter is<br>indicated to sample blood<br>or administer fluids<br>intravenously. The<br>MedSource CathMED IV<br>Catheter may be used for<br>any patient population<br>with consideration given<br>to adequacy for the<br>vascular anatomy<br>appropriateness for the<br>solution being<br>administered and<br>duration of the therapy. | The MedSource IV Safety<br>Catheter is indicated to<br>sample blood or administer<br>fluids intravenously. The<br>MedSource IV Safety<br>Catheter may be used for<br>any patient population with<br>consideration given to<br>adequacy for the vascular<br>anatomy appropriateness<br>for the solution being<br>administered and duration<br>of the therapy. | Same       |                                                                                                                                                                                                                                                                             |
| Distal End<br>configuration                         | Beveled                                                                                                                                                                                                                                                                                                                                                              | Beveled                                                                                                                                                                                                                                                                                                                                                      | Same       |                                                                                                                                                                                                                                                                             |
| Proximal End<br>configuration                       | Cupper                                                                                                                                                                                                                                                                                                                                                               | Cupper                                                                                                                                                                                                                                                                                                                                                       | Same       |                                                                                                                                                                                                                                                                             |
| Needle Stick<br>Prevention<br>Feature               | Not present                                                                                                                                                                                                                                                                                                                                                          | Push-button needle<br>shielding                                                                                                                                                                                                                                                                                                                              | Different  | Absent in the<br>submission<br>device.<br>A warning was<br>added to the<br>labeling of this<br>product that a<br>Needle Stick<br>Prevention<br>Feature is not<br>present in this<br>device.<br>This does not<br>affect the safety<br>and<br>effectiveness of<br>the device. |
| Catheter<br>Length<br>Catheter O.D.<br>Needle Gauge | Gauge Length OD<br>mm mm                                                                                                                                                                                                                                                                                                                                             | Gauge Length OD<br>mm mm                                                                                                                                                                                                                                                                                                                                     | Different  | one new Gauge<br>Length "20G<br>length 33 mm<br>OD 1.1 mm"                                                                                                                                                                                                                  |
|                                                     | 24G                                                                                                                                                                                                                                                                                                                                                                  | 19                                                                                                                                                                                                                                                                                                                                                           | 0.7        |                                                                                                                                                                                                                                                                             |
|                                                     | 24G                                                                                                                                                                                                                                                                                                                                                                  | 19                                                                                                                                                                                                                                                                                                                                                           | 0.7        |                                                                                                                                                                                                                                                                             |
|                                                     | 22G                                                                                                                                                                                                                                                                                                                                                                  | 25                                                                                                                                                                                                                                                                                                                                                           | 0.9        |                                                                                                                                                                                                                                                                             |
|                                                     | 22G                                                                                                                                                                                                                                                                                                                                                                  | 25                                                                                                                                                                                                                                                                                                                                                           | 0.9        |                                                                                                                                                                                                                                                                             |
|                                                     | 20G 33 1.1<br>20G 25 1.1<br>18G 45 1.3<br>16G 45 1.8<br>14G 45 2.1                                                                                                                                                                                                                                                                                                   | 20G 25 1.1<br>18G 25 1.3<br>16G 30 1.7<br>14G 45 2.1                                                                                                                                                                                                                                                                                                         | Same       | added to the<br>submission<br>device.<br>This was added<br>to expand the<br>size availability<br>for patient care.<br>This does not<br>affect the safety<br>and<br>effectiveness of<br>the device.                                                                          |
| Sterilization                                       | Terminal Ethylene Oxide<br>(EtO gas)                                                                                                                                                                                                                                                                                                                                 | Terminal Ethylene Oxide<br>(EtO gas)                                                                                                                                                                                                                                                                                                                         | Same       |                                                                                                                                                                                                                                                                             |
|                                                     | Materials                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                              |            |                                                                                                                                                                                                                                                                             |
| Catheter                                            | Polyurethane (PUR)<br>/ (PTFE)<br>Polytetrafluoroethylene                                                                                                                                                                                                                                                                                                            | Polyurethane (PUR)<br>/ (PTFE)<br>Polytetrafluoroethylene                                                                                                                                                                                                                                                                                                    | Same       |                                                                                                                                                                                                                                                                             |
| Needle                                              | Stainless Steel                                                                                                                                                                                                                                                                                                                                                      | Stainless Steel                                                                                                                                                                                                                                                                                                                                              | Same       |                                                                                                                                                                                                                                                                             |
| Catheter Body                                       | K-Resin                                                                                                                                                                                                                                                                                                                                                              | K-Resin                                                                                                                                                                                                                                                                                                                                                      | Same       |                                                                                                                                                                                                                                                                             |
| Catheter<br>Holder                                  | Polyacetal (POM)                                                                                                                                                                                                                                                                                                                                                     | Polyacetal (POM)                                                                                                                                                                                                                                                                                                                                             | Same       |                                                                                                                                                                                                                                                                             |
| Needle Hub                                          | K-Resin                                                                                                                                                                                                                                                                                                                                                              | K-Resin                                                                                                                                                                                                                                                                                                                                                      | Same       |                                                                                                                                                                                                                                                                             |
| Activation<br>Lever                                 | Absent                                                                                                                                                                                                                                                                                                                                                               | Polyacetal (POM)                                                                                                                                                                                                                                                                                                                                             | Different  | Absent in the<br>submission<br>device.<br>A warning was<br>added to the<br>labeling of this<br>product that a<br>Needle Stick<br>Prevention<br>Feature is not<br>present in this<br>device.<br>This does not<br>affect the safety<br>and<br>effectiveness of                |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                              |            | the device.                                                                                                                                                                                                                                                                 |
| Flashback<br>Chamber                                | PUR/PTFE                                                                                                                                                                                                                                                                                                                                                             | PUR/PTFE                                                                                                                                                                                                                                                                                                                                                     | Same       |                                                                                                                                                                                                                                                                             |
| Hydrophobic<br>Filter                               | Stainless Steel                                                                                                                                                                                                                                                                                                                                                      | Stainless Steel                                                                                                                                                                                                                                                                                                                                              | Same       |                                                                                                                                                                                                                                                                             |
| Materials                                           | K-Resin                                                                                                                                                                                                                                                                                                                                                              | K-Resin                                                                                                                                                                                                                                                                                                                                                      | Same       |                                                                                                                                                                                                                                                                             |
| Sterilization<br>method                             | EtO                                                                                                                                                                                                                                                                                                                                                                  | EtO                                                                                                                                                                                                                                                                                                                                                          | Same       |                                                                                                                                                                                                                                                                             |

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

## H. Performance Data:

## I. Performance Testing:

- ISO 10555-5: Intravascular catheters Sterile and single-use catheters Part 5: ● Over-needle peripheral catheters
- ISO 10555-1 Intravascular catheters Sterile and single-use catheters Part 1: ● General requirements
- ISO 594-1Conical fittings with a 6% (Luer) taper for syringes, needles and certain ● other medical equipment. - Part 1: General requirements
- . ISO 594-2 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings

## II. Biocompatibility

There were no changes to the material that impacted the biocompatibility of the device. A signed statement of conformance was included in place of testing as the testing from the predicate device was referenced.

- . USP <788> Particulate Matter in Injections
## III. Sterility

- ISO 11135:2014 Sterilization of health-care products Ethylene oxide -. Requirements for the development, validation and routine control of a sterilization process for medical devices
### I. Conclusion:

In summary, based on the indication for use, technological characteristics and performance testing, the differences between the MedSource CathMED IV Catheter and the predicate did not raise concern for safety and effectiveness. Therefore, we conclude that the subject device is substantially equivalent to the legally marketed predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K193278](https://fda-staging.innolitics.com/device/K193278)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).