← Product Code [JAK](/productcode/JAK) · K193277

# SOMATOM On.site and On.scene (K193277)

_Siemens Medical Solutions USA, Inc. · JAK · Jul 22, 2020 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K193277

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [JAK](/productcode/JAK.md)
- **Decision Date:** Jul 22, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

This computed tomography system is intended to generate and process cross-sectional images by computer reconstruction of x-ray transmission data within a 25 cm field-of-view, primarily for the head and neck. The images delivered by SOMATOM On.site and On.scene can be used by a trained physician as an aid in diagnosis.

## Device Story

Mobile CT scanner system (SOMATOM On.scene) optionally mounted on motorized base (SOMATOM On.site); utilizes continuously rotating tube-detector fan beam technology. Inputs: X-ray transmission data. Outputs: DICOM-format cross-sectional images. Operated by trained staff in clinical settings (e.g., ICU, mobile stroke units) via tablet-based interface (Scan&GO). Features telescopic gantry design to minimize patient movement during scanning; hybrid air/water tube cooling; integrated radiation shielding curtains. Software (Somaris/10 syngo CT VA35A) manages patient data, scan control, image reconstruction, and archiving. Provides point-of-care imaging to reduce patient transport, improve access for immobile patients, and assist physician diagnosis and treatment planning.

## Clinical Evidence

Bench testing only. Included phantom tests (adult head, pediatric body/head) and volunteer human scans (brain with/without contrast, ankle, hand). Performance verified against standards (IEC 60601 series, NEMA XR-25/28). Usability evaluated via formative and summative testing. Board-certified radiologist confirmed high diagnostic quality of images.

## Technological Characteristics

32-slice CT system; 3kW generator; fan beam geometry; 25cm FOV. Materials/components: hybrid air/water X-ray tube cooling, telescopic gantry, motorized base. Connectivity: DICOM 3.1-3.20. Software: Somaris/10 syngo CT VA35A (command-based). Standards: IEC 60601-1, 60601-2-44, 60601-1-2, 62304, NEMA XR-25/28, ISO 14971.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- CereTom ([K051765](/device/K051765.md))
- SOMATOM go.Up ([K192061](/device/K192061.md))
- Mobilett Mira ([K111912](/device/K111912.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/10 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".

Siemens Medical Solutions USA, Inc. % M. Alaine Medio, RAC Regulatory Affairs Professional 810 Innovation Drive KNOXVILLE TN 37932

Re: K193277

Trade/Device Name: SOMATOM On.site and On.scene Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: July 2, 2020 Received: July 6, 2020

Dear M. Alaine Medio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

July 22, 2020

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K193277

Device Name SOMATOM On.site and On.scene

#### Indications for Use (Describe)

This computed tomography system is intended to generate and process cross-sectional images by computer reconstruction of x-ray transmission data within a 25 cm field-of-view, primarily for the head and neck.

The images delivered by SOMATOM On.scene can be used by a trained physician as an aid in diagnosis.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <div style="display:inline-block;"><b> <span style="font-size: 16px; vertical-align: middle;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</b></div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:inline-block;"><b> <span style="font-size: 16px; vertical-align: middle;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</b></div>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary – K193277

as required by 21 CFR Part 807.87(h)

## Identification of the Submitter

| Importer / Distributor:     | Siemens Medical Solutions USA, Inc.<br>40 Liberty Boulevard<br>Malvern, PA 19355<br>Establishment Registration Number: 2240869                                                         |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:               | Siemens Healthcare GmbH<br>Siemensstraße 1<br>D-91301 Forchheim, Germany<br>Establishment Registration Number: 3004977335                                                              |
| Submitter / Contact Person: | M. Alaine Medio, RAC<br>Regulatory Affairs Professional<br>Siemens Medical Solutions USA, Inc.<br>810 Innovation Drive<br>Knoxville, TN 37932<br>alaine.medio@siemens-healthineers.com |
| Telephone Number:           | (865)206-0337                                                                                                                                                                          |
| Fax Number:                 | (865)218-3019                                                                                                                                                                          |
| Alternative Contact:        | Tabitha Estes<br>Regulatory Affairs Professional                                                                                                                                       |
| Date of Submission:         | July 2, 2020                                                                                                                                                                           |

| Identification of the product |                                                      |
|-------------------------------|------------------------------------------------------|
| Device Proprietary Name:      | SOMATOM On.site and On.scene                         |
| Common Name:                  | Computed Tomography (CT) System                      |
| Classification Name:          | Computed Tomography X-Ray System per 21 CFR 892.1750 |
| Product Code:                 | JAK                                                  |
| Classification Panel:         | Radiology                                            |
| Device Class:                 | Class II                                             |

{4}------------------------------------------------

## Marketed Devices to which Equivalence is claimed

#### Primary Predicate:

| Device Proprietary Name: | CereTom             |
|--------------------------|---------------------|
| Manufacturer:            | Samsung NeuroLogica |
| Product Code:            | JAK                 |
| Device Class:            | Class II            |
| 510(k) Number:           | K051765             |

| Predicate Devices:       |                         |
|--------------------------|-------------------------|
| Device Proprietary Name: | SOMATOM go.Up           |
| Manufacturer:            | Siemens Healthcare GmbH |
| Product Code:            | JAK                     |
| Device Class:            | Class II                |
| 510(k) Number:           | K192061                 |

| Device Proprietary Name: | Mobilett Mira           |
|--------------------------|-------------------------|
| Manufacturer:            | Siemens Healthcare GmbH |
| Product Code:            | IZL                     |
| Device Class:            | Class II                |
| 510(k) Number:           | K111912                 |

Recall Information: All applicable recalls are considered and addressed as part of the design control process

#### Device Description

The Siemens SOMATOM On scanners are comprised of a Computed Tomography (CT) Scanner System (SOMATOM On.scene) which can be mounted on an optional motorized base (SOMATOM On.site). The CT scanner features one continuously rotating tube-detector system that functions according to the fan beam principle. The system software is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation.

The SOMATOM On scanners produce CT images in DICOM format, which can be used by trained staff for post-processing applications commercially distributed by Siemens and other vendors as an aid in diagnosis and treatment preparation. The computer system included in the CT Scanner is able to run optional post processing applications.

The software version for the SOMATOM On scanner system is Somaris/10 syngo CT VA35A, is a command-based program used for patient management, X-ray scan control, image reconstruction, and image archive/evaluation. The software platform SOMARIS/10 syngo CT VA35A is

{5}------------------------------------------------

designed to provide a plugin interface to integrate potential advanced post processing tasks, tools, or extendable functionalities.

As with the primary predicate device, the SOMATOM On. Scanners will be available in a 32 row, 32 slice configuration.

## Indications for Use

This computed tomography system is intended to generate and process cross-sectional images by computer reconstruction of x-ray transmission data within a 25cm field-of-view, primarily for the head and neck.

The images delivered by SOMATOM On.site and On.scene can be used by a trained physician as an aid in diagnosis.

## Comparison of Technological Characteristics with the Predicate Device

The SOMATOM On.site and On.scene scanners provide the same technological characteristics in terms of materials, energy source, and control mechanisms when compared to the predicate devices. The software and hardware components of this scanner have been modified or improved in comparison to the predicate devices to support enhanced device functionality and to reduce the potential for scattered radiation.

The SOMATOM On Scanners are similar to the CereTom mobile CT scanner (Neurologica, K051765) which is the primary predicate, and the Go.Platform CT scanners (Siemens K192061). Different features of the two systems are incorporated into the SOMATOM On platform:

- . CereTom mobile CT Scanner (Neurologica K051765) - Primary predicate Similar to the SOMATOM On scanners that are the subject of this application, the CereTom CT scanner is a mobile scanner with a similar bore size that is used in the same manner as to the SOMATOM On scanners.
- SOMATOM Go. Platform Scanners (Siemens K192061) - Predicate Siemens is incorporating multiple features of the Siemens SOMATOM Go scanners into the SOMATOM On platform, including the software that runs the CT Systems, as well as the Gantry rotation mechanics and tablet based touch screen.

The intended use and fundamental scientific technology for the SOMATOM On. remains unchanged from the predicate devices.

The hardware components of the subject device have been modified as compared to the predicate devices to include:

- . a gantry translation system
- . a Control Device (Scan&GO) user interface (e.g. user interface via mobile tablet software application),
- modified gantry mechanics, and
- an optional motorized base.

{6}------------------------------------------------

- Hybrid air / water X-Ray tube cooling system ●
- 3kW maximum power Generator
- Gantry mounts at stretcher height in mobile stroke unit
- Method for "complete-angle" shielding of scattered x-ray radiation
- Pre-aligned patient holder attached to scanner system (to reduce human-error in patient alignment resulting in poor scan quality)
- Telescopic gantry design to provide no movement of patient vs scanner during the scan sequence.

The software is based on the commercially available SOMARIS/10 syngo CT VA30A software (K192061). The syngo CT VA35A software supports features that are designed to adapt to a mobile workflow and accommodate the scanner specific hardware/reconstruction.

Any differences in technological characteristics do not raise different questions of safety and effectiveness. Siemens believes that the subject device is substantially equivalent to the predicate devices. Testing and validation have been provided. Test results show that the subject device, the SOMATOM On., is comparable to the predicate devices in terms of technological characteristics and safety and effectiveness, and therefore are substantially equivalent to the predicate devices.

Testing provided and described below support marketing claims associated with the following:

- . Point of care imaging benefits, including reducing in-house transports of ICU patients, faster access to neuro CT imaging for immobile patients,
- Telescopic gantry design benefits for scatter radiation reduction
- Radiation safety and protection, including enabling "apron free scanning" when used with both radiation shields employed
- . lmage quality, including low image noise and quality consistent with SOMATOM CT scanners
- Streamline of the workflow and allowing physicians to determine appropriate facility for transfer when used in ambulatory situations
- . Usability claims, including those associated with the Graphical User Interface, patient positioning, single operator system handling and reduction of physical burden on staff associated with patient transport.
- . Improved workflow in ICU as well as radiation department

### Performance Data

### Non-Clinical Testing

Non-clinical testing (integration and functional) including phantom tests and volunteer human scans were conducted for the SOMATOM On. during product development. The modifications described in this Premarket Notification were supported with verification testing. Siemens claims conformance to the following performance standards: ISO 14791, NEMA XR-29, IEC 61223-2-6, IEC 61223-3-5, IEC 62304, NEMA XR-25, and DICOM 3.1-3.20.

Electrical Safety and Electromagnetic Compatibility (EMC) testing were conducted on the SOMATOM On. scanners in accordance with the following standards: IEC 60601-1, 60601-2-44, and 60601-1-2 (class A).

{7}------------------------------------------------

Radiation safety testing was performed in two configurations – with and without the front protection curtains in use. Both configurations indicate that the radiation is under published limits, and the protection curtain in place reduces the radiation significantly.

Usability testing was performed with both scanner configurations, On.site and On.scene. This testing included formative and Summative evaluations. No new use errors, hazards or hazardous situations were identified. There were not any impacts to the existing residual risks identified.

## Verification and Validation

Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. The Risk Analysis was completed, and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation supports the claims of substantial equivalence.

Siemens Healthcare conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Cybersecurity information in accordance with guidance document "Content of Premarket Submissions for Management of Cybersecurity Medical Devices issues on October 2, 2014" is included within this submission.

Additionally, Siemens conforms to the requirements for Radio Frequency Wireless Technology as defined in FDA guidance document "Radio Frequency Wireless Technology in Medical Devices, Guidance for Industry and Food and Drug Administration Staff, issued on August 14, 2013" by adhering to the EMC and risk based verification and validation requirements in design, testing, and labeling of the wireless remote control components of the subject devices.

### Imaging Studies

Phantom scans were performed with phantoms representing adult heads, pediatric bodies and pediatric heads. Additionally, volunteer scans were performed including brain scans with and without contrast, ankle and hand scans. All imaging results were as expected and were determined by a board-certified radiologist to be of high diagnostic quality.

### General Safety and Effectiveness Concerns

The device labeling contains instructions for use as well as necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a system related Risk analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing according to the Risk Management process. In order to minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.

{8}------------------------------------------------

| Standard                    | Version                    | Content                                                                                                                                                                                              | FDA<br>Recognition<br>Number |
|-----------------------------|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| ES60601-1                   | :2005 +<br>A1:2012         | Medical electrical equipment - part 1:<br>general requirements for basic safety and<br>essential performance                                                                                         | 19-4                         |
| IEC 60601-1-2               | :2014                      | Medical Electrical Equipment -- Part 1-2:<br>General Requirements For Basic Safety And<br>Essential Performance -- Collateral Standard:<br>Electromagnetic Disturbances --<br>Requirements And Tests | 19-8                         |
| IEC 60601-1-3               | :2013                      | Collateral standard: Radiation protection in<br>diagnostic X-ray equipment                                                                                                                           | 12-269                       |
| IEC 60601-1-6               | :2010 +A1:2013             | Collateral Standard: Usability                                                                                                                                                                       | 5-89                         |
| IEC 60601-2-28              | :2017                      | Particular measurements for the safety of X-<br>ray source assemblies and X-ray tube<br>assemblies for medical diagnostics.                                                                          | 12-309                       |
| IEC 60601-2-44              | :2009 +A1:2012<br>+A2:2-16 | Medical Electrical Equipment - Part 2-44:<br>Particular Requirements for the Basic Safety<br>and Essential Performance of X-Ray<br>Equipment for Computed Tomography                                 | 12-302                       |
| IEC 62366-1                 | :2015                      | Medical devices -- Application of usability<br>engineering to medical devices                                                                                                                        | 5-114                        |
| IEC 61223-2-6               | :2006                      | Constancy Tests - Imaging performance of<br>computed tomography X-ray equipment                                                                                                                      | 12-226                       |
| IEC 61223-3-5               | :2004                      | Acceptance Tests – Imaging performance of<br>computer tomography X-ray equipment                                                                                                                     | 12-270                       |
| IEC 60825-1                 | :2007                      | Safety of Laser Products – Part 1: Equipment<br>Classification and Requirements [Including:<br>Technical Corrigendum 1 (2008),<br>Interpretation Sheet 1 (2007), Interpretation<br>Sheet 2 (2007)]   | 12-273                       |
| NEMA Standard<br>PS3.1-3.20 | :2016                      | Digital Imaging and Communications in<br>Medicine (DICOM)                                                                                                                                            | 12-300                       |
| NEMA XR-25                  | :2010                      | Computed Tomography Dose Check                                                                                                                                                                       | 12-225                       |
| NEMA XR-28                  | :2013                      | Supplemental Requirements for User<br>Information and System Function related to<br>Dose in CT                                                                                                       | 12-287                       |
| IEC 62304                   | :2006+A1:2015              | Medical Device Software - Software Life<br>Cycle Processes                                                                                                                                           | 13-79                        |
| ISO 14971                   | :2007                      | Application of Risk Management to Medical<br>Devices                                                                                                                                                 | 5-40                         |

### Summary

The features described in this premarket notification are supported with verification and validation testing, dosimetry and imaging performance, and analysis of phantom images to assess device and feature performance during product development. The risk analysis was completed, and risk control implemented to mitigate identified hazards. The test results show that all the software specifications have met the acceptance criteria. Verification testing of the device was found acceptable to support the claim of substantial equivalence.

{9}------------------------------------------------

#### Conclusions

The predicate devices were cleared based on the results of non-clinical testing including verification and validation, phantom tests, and supportive literature. The subject device is also tested using the same methods as used for the predicate devices. The non-clinical data supports the safety of the device and the hardware and software verification and validation demonstrates that the SOMATOM On. Scanners should perform as intended in the specified use conditions. The data included in this submission demonstrates that the SOMATOM On.site and SOMATOM On.scene scanners perform comparably to the predicate devices currently marketed for the same intended use. Since both devices were tested using the same methods, Siemens believes that the data generated supports a finding of substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K193277](https://fda-staging.innolitics.com/device/K193277)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
