Disposable high-speed air turbine handpiece is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
Device Story
Disposable high-speed air turbine handpiece (Model YCX-TOD-PP) is an air-driven dental instrument used by trained dental professionals in clinical settings. The device connects to a standard dental treatment unit via a 4-pin coupling, receiving drive air, exhaust air, spray air, and spray water. Air pressure drives an internal turbine assembly, which rotates a dental bur (ISO 1797-1 Type 3) to perform dental procedures such as cavity preparation and polishing. A water/air mist is delivered to the treatment site for cooling. The device is supplied non-sterile and must be autoclaved before use. Unlike reusable predicates, this device is intended for disposal after use, eliminating the need for lubrication. The output is the mechanical rotation of the bur, which allows the clinician to remove dental material and refine restorations, directly facilitating dental therapy and improving patient oral health.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing, including performance testing (ISO 14457, ISO 9168), biocompatibility (ISO 10993-1, ISO 10993-5, ISO 10993-10), and sterilization validation (AAMI/ANSI/ISO 17665-1, AAMI/ANSI/ISO 11737-1, AAMI/ANSI/ISO 11138-1, ANSI AAMI ST79, AAMI/ANSI/ISO 14161).
Technological Characteristics
Materials: Brass, PES, silicon rubber. Principle: Air-driven turbine. Connectivity: 4-pin pneumatic coupling. Dimensions: 142-145mm length, 20.5mm width. Sterilization: Moist heat (autoclave). Performance standards: ISO 14457, ISO 9168. Biocompatibility standards: ISO 10993-1, ISO 7405. Chuck type: Chuck tool.
Indications for Use
Indicated for dental patients requiring removal of carious material, excess filling material, cavity and crown preparation, finishing of tooth preparations and restorations, root canal preparations, and tooth polishing.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 11, 2020
Ningbo HPDOVE Dental Instruments Co., Ltd. % Charlie Mack Principal Engineer IRC 2950 E Lindrick Drive Chandler, AZ 85249
Re: K193264
Trade/Device Name: Disposable High Speed Air Turbine Handpiece Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class II Product Code: EFB Dated: May 20, 2020 Received: May 27, 2020
Dear Charlie Mack:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K193264
Device Name
Disposable High Speed Air Turbine Handpiece
Indications for Use (Describe)
Disposable high speed air turbine handpiece is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
Type of Use (Select one or both, as applicable)
| <img alt="Checked" src="checkbox-checked.png"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------------|
| <img alt="Unchecked" src="checkbox-unchecked.png"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K193264
Image /page/3/Picture/1 description: The image shows a logo with a stylized bird-like figure inside a globe. The globe is represented by a grid of lines, and the bird-like figure is dark and appears to be made of a single, continuous line. The logo is simple and abstract, with the bird-like figure being the focal point.
## NINGBO HPDOVE DENTAL INSTRUMENTS CO. LTD.
510(k) Summary (21 CFR §807.92)
## Submitter Information:
| Submitter Name: | NINGBO HPDOVE DENTAL INSTRUMENTS CO. LTD. |
|----------------------|--------------------------------------------------------------------------------|
| Address: | 588 Xiaolin Road, Xiaolin Town, Cixi City, Zhejiang<br>Province, P.R.C. 315321 |
| Telephone: | +86-574-63883911 |
| Fax: | +86-574-63519378 |
| Contact Person: | Mr. Liyue Huang<br>Manager |
| Email: | charliemack@irc-us.com |
| Date of Preparation: | July 20, 2020 |
{4}------------------------------------------------
### Subject Devices :
Trade/proprietary name: Disposable high-speed air turbine handpiece Model YCX-TOD-PP Common Name: Disposable high-speed air turbine handpiece Classification : 21CFR872.4200
#### Predicate Device:
| Trade Name: | High-Speed Handpieces and Accessories |
|--------------------|---------------------------------------|
| 510(k) Reference: | K152146 |
| Common Name: | Handpiece, air-powered, dental |
| Regulation Number: | 21CFR872.4200 |
| Regulatory Class: | Class II |
| Manufacturer: | Codent Technical Industry Co. Ltd. |
#### Purpose of Submission
This is a new traditional 510(K) submission of Disposable high-speed air turbine handpiece.
### Device Description
The disposable high-speed air turbine handpiece is an air-driven dental handpiece for the use by a trained professional in the field of general dentistry. The device is an ergonomically shaped air-powered handpiece that is disposable. The device features a 4-pin (drive air, exhaust air, spray air and spray water holes) connector for air and water input and return. A water/air mist cools the head and bur. Dental burs (not part of this 510(k)) according to ISO 1797-1 type 3 could be inserted into the chuck system of the turbine.
The disposable high-speed air turbine dental handpiece is an instrument widely applied in oral therapy. It is composed of interrelated parts, working as a complete set. The disposable high-speed air turbine dental handpiece is an air-driven handpiece which supplied with water and air through the pipe and the coupling of a dental treatment unit. The device contains air and water pipes inside which supply air for driving the turbine assembly rotating and cooling water for work site as well as supply air and water for creating spray. The air-driven turbine assembly provides rotation to
{5}------------------------------------------------
the bur for removing carious material, excess filling material, cavity, and is used for crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
The handpiece is supplied non-sterile and must be cleaned and sterilized before use.
#### Indication for use
Disposable high-speed air turbine handpiece is intended for removing carious material, excess filling material, cavity, and crown preparation finishing tooth preparations and restorations, root canal preparations and polishing teeth.
### Comparison with the predicate device:
Ningbo HPDOVE Dental Instruments Co. Ltd. believes that the Disposable high-speed air turbine handpiece is substantially equivalent to the Codent Technical Industry Co. Ltd. High-speed Handpieces and Accessories (K152146).
{6}------------------------------------------------
The differences noted between the submitted Disposable high-speed air turbine handpiece and the predicate High-Speed Handpieces and Accessories are based on the use. The Disposable High Speed Air Turbine Handpiece, YCX-TOD-PP has the same use as the predicate and used in the same manner. Both the predicate and the submitted Ningbo HPDOVE Dental Instruments Co. Ltd Disposable High-Speed Air Turbine Handpiece, YCX-TOD-PP come non-sterile and must be sterilized before use. Please refer to the following pages for specific difference details.
{7}------------------------------------------------
# Comparison to Predicate Devices
| Characteristics | Subject Device | Predicate Device |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name and model | Disposable high-speed air turbine<br>handpiece | High-speed Handpieces and Accessories |
| Model | YCX-TOD-PP | A4M4 |
| 510K Applicant | NINGBO HPDOVE DENTAL<br>INSTRUMENTS CO., LTD. | Codent Technical Industry Co., Ltd. |
| 510(K) Number | Pending | K152146 |
| Regulation<br>Number | CFR872.4200 | CFR872.4200 |
| Product Code | EFB | EFB |
| Classification<br>Name | Air-powered dental handpiece | Air-powered dental handpiece |
| OTC or<br>Prescription | Prescription Use | Prescription Use |
| Medical Specialty | Dental | Dental |
| Indication for Use | Disposable high-speed air turbine<br>handpiece is intended for removing<br>carious material, excess filling material,<br>cavity, and crown preparation, finishing<br>tooth preparations and restorations, root<br>canal preparations, and polishing teeth. | High-speed Handpieces and Accessories<br>are intended for removing carious<br>material, excess filling material, cavity,<br>and crown preparation, finishing tooth<br>preparations and restorations, root canal<br>preparations and polishing teeth. |
| Disposable | Disposable | Re-useable |
| Principles of<br>Operation | It is a handheld device that connects to<br>a standard turbine connection on a<br>dental operative unit and delivers a<br>mixture of water and air to a treatment<br>site. | It is a handheld device that connects to a<br>standard turbine connection on a dental<br>operative unit and delivers a mixture of<br>water and air to a treatment site. |
| Operational Mode | Air-powered | Air-powered |
| Head Type | Standard | Standard |
| Angle of Shaft | 90° | 90° |
| Speed Range in<br>rpms | 300,000 rpm to 360,000 rpm | 360,000 rpm |
| Drive Air Pressure | 200 kPa - 245 kPa | 245 kPa- 275 kPa |
| | 2.09 kg/cm² - 2.5 kg/cm² | 2.5 kg/cm²- 2.8 kg/cm² |
| Water Spray | Single | Single |
{8}------------------------------------------------
| Characteristics | Subject Device | Predicate Device |
|-----------------------------|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Dental Bur Size | ISO 1797-1 Type3 / Ø1.59~1.60mm | ISO 1797-1 Type3 / Ø1.59~1.60mm |
| Dimensions(mm) | Length: 142~145<br>Width: 20.5 | Length: 84~117<br>Width: 19~21 |
| How Supplied | Non-sterile; Subject is sterilized Prior to use. | Non-sterile; Subject is sterilized before each use |
| Cleaning | Wipe the surface of the device with a<br>clean, soft brush or cotton ball with 75%<br>alcohol | Use the cleaning wire to clean the water<br>spray hole. |
| Applicable<br>Sterilization | Moist heat - Autoclave sterilization | Moist heat - Autoclave sterilization |
| Material<br>Composition | Brass, PES, silicon rubber | Stainless steel, Brass, Aluminum,<br>Titanium |
| Coupling Pin | 4-pin(drive air, exhaust air, spray air and<br>spray water holes) | 4-pin(drive air, exhaust air, spray air and<br>spray water holes) |
| Fiber Optics | Without light | Without light |
| Accessories | Without coupling | Without coupling |
| Chunk Design | Chuck Tool | Push Button |
| Bur Extraction<br>Force | 22-45 N | 22-45 N |
| Lubrication | Lubricant is not required | Lubricant is required (Codent spray<br>nozzle) |
| Performance | Complies with ISO 14457<br>Complies with ISO 9168 | Complies with ISO 14457<br>Complies with ISO 9168 |
| Biocompatibility | Complies with ISO10993-1 & ISO 7405 | Complies with ISO10993-1 & ISO 7405 |
{9}------------------------------------------------
### Safety and Performance Data :
To establish substantial equivalence to the identified predicate devices, tests were completed as defined below to the subject devices, Disposable high-speed air turbine handpiece. The results of the testing demonstrate that the device complies with the applicable standards requirements, and the device is substantially equivalent to the predicate device.
#### Non-Clinical Study:
#### Performance:
- ISO 14457 First edition 2012-09-15 Dentistry Handpieces and motors -
- ISO 9168 Third edition 2009-07-15 Dentistry Hose connectors for air-driven dental handpieces
#### Biocompatibility:
- ISO10993-1:2009/AC2010 -
- ISO 10993-5: 2009 -
- -ISO 10993-10: 2010
#### Sterility:
- AAMI/ANSI/ISO 17665-1:2006. Sterilization of health care products -- Moist heat --- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices (Sterility)
- -AAMI/ANSI/ISO 11737-1:2018, Sterilization of health care products --Microbiological methods -- Part 1: Determination of a population of microorganisms on products.
- AAMI/ANSI/ISO 11138-1:2017, Sterilization of health care products -- Biological indicators -- Part 1: General requirements
- -ANSI AAMI ST79:2017, Comprehensive guide to steam sterilization and sterility assurance in health care facilities
- -AAMI/ANSI/ISO 14161: 2009, Sterilization of health care products — Biological indicators - Guidance for the selection, use and interpretation of results
#### Clinical Study:
No clinical studies were performed. l
{10}------------------------------------------------
#### Package and Shelf Life:
- The Disposable high-speed air turbine handpiece (YCX-TOD-PP ) is not subject to shelf life, as the device does not contain any sterile or degradable elements.
#### Conclusion:
The differences between the subject device and predicate device do not raise issues of safety and effectiveness based on the indication for use, technological characteristics, and performance testing. The subject device complies with the same applicable standards as the predicate device.
In accordance with the Federal Food, Druq and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, the subject device. Disposable high-speed air turbine handpiece, model YCX-TOD-PP is safe and effective and substantially equivalent to predicate devices as described herein.
END
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.