Infrared Thermometer, Model Number CK-T1501, CK-T1502, CK-T1503
K193253 · Shenzhen Changkun Technology Co., Ltd. · FLL · Jun 17, 2020 · General Hospital
Device Facts
Record ID
K193253
Device Name
Infrared Thermometer, Model Number CK-T1501, CK-T1502, CK-T1503
Applicant
Shenzhen Changkun Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Jun 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Infrared Thermometer (model: CK-T1501, CK-T1502, CK-T1503) is a non-sterile, reusable, non-contact and handheld device. It can be used by consumers in homecare environment and doctors in clinic as reference. It is intended for measuring human body temperature of people over one month old by detecting infrared heat from the forehead.
Device Story
Handheld, battery-powered, non-contact infrared thermometer; measures human body temperature by detecting infrared energy emitted from the forehead at a distance of 3-5 cm. Device utilizes a thermopile sensor with an integrated thermistor for target readings and a secondary thermistor for ambient temperature compensation. Operates via user-activated measurement key; provides oral-equivalent temperature output on an LCD screen within 0.5 seconds. Features include temperature alarm points, temperature offset, warning tone switch, and 32-set memory storage. Used in homecare and clinical settings by consumers or healthcare providers. Output assists in clinical decision-making by providing rapid, non-invasive temperature assessment.
Clinical Evidence
Clinical performance evaluated in 240 subjects across three age groups: A1 (1-3 months), A2 (3 months-1 year), B (1-5 years), and C (>5 years). Testing conducted per ISO 80601-2-56 and ASTM E1965-98 (2016) standards. Results demonstrated compliance with clinical accuracy requirements for infrared thermometers.
Technological Characteristics
Handheld infrared thermometer; ABS plastic enclosure. Sensing principle: thermopile sensor with integrated thermistor for target and ambient temperature. Power: DC 3V (2 AA alkaline batteries). Connectivity: None. Sterilization: Non-sterile. Software: Embedded firmware for data processing and display. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-56, ASTM E1965-98.
Indications for Use
Indicated for measuring human body temperature in patients over one month old by detecting infrared heat from the forehead. Suitable for use by consumers in homecare environments and by clinicians in medical clinics.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Non Contact Infrared Forehead Thermometer (model: TVT-200, TVT-200 PLUS) (K170662)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 17, 2020
Shenzhen Changkun Technology Co., Ltd. % Cassie Lee Official Correspondent Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road Huangpu District Guangzhou, 510000 China
Re: K193253
Trade/Device Name: Infrared Thermometer, Models: CK-T1501, CK-T1502, CK-T1503 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: March 23, 2020 Received: May 18, 2020
Dear Ms. Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Tina Kiang, Ph.D. Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193253
Device Name
Infrared Thermometer (Models: CK-T1501, CK-T1502, CK-T1503)
Indications for Use (Describe)
Infrared Thermometer (model: CK-T1501, CK-T1503) is a non-sterile, reusable, non-contact and handheld device. It can be used by consumers in homecare environment and doctors in clinic as reference. It is intended for measuring human body temperature of people over one month old by detecting infrared heat from the forehead.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K193253
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
# 1. Date of the summary prepared: 2020-05-25
# 2. Submitter'sInformation
510(k) Owner's Name: Shenzhen Changkun Technology Co., Ltd. Establishment Registration Number : Applying Address: 801, 3 floor 4floor 6floor 7floor, B building, NO.69, zhenbi road, biling community, biling street, pingshan district, Shenzhen city, Guangdong, China Tel: +0755-29100487 Fax: +0755-29100487 Contact Person: Steve Li Email: changkunkj@163.com
# Application Correspondent:
Contact Person: Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China Tel: +86 20 8266 2446 Email: regulatory @glomed-info.com
# 3. Subject Device Information
Trade Name: Infrared Thermometer, models: CK-T1501, CK-T1502, CK-T1503 Common Name: Clinical electronic thermometer Classification name: Thermometer, electronic, clinical Review Panel: General Hospital Product Code: FLL Regulation Class: II Regulation Number: 21 CFR 880.2910
## 4. Predicate Device Information
{4}------------------------------------------------
| Sponsor: | Shenzhen Changkun Technology Co., Ltd. |
|-----------------|------------------------------------------------------------|
| Subject Device: | Infrared Thermometer, Models: CK-T1501, CK-T1502, CK-T1503 |
Sponsor: Intrinity Global Limited Trade Name: Non Contact Infrared Forehead Thermometer(model : TVT-200, TVT-200 PLUS) Common Name: Clinical electronic thermometer Classification Name: Thermometer, Electronic, Clinical 510(K) Number: K170662 Review Panel: General Hospital Product Code: FLL Regulation Number: 880.2910 Regulation Class: II
#### 5. Device Description
Infrared Thermometer (model: CK-T1501,CK-T1502,CK-T1503) is a hand-held, battery powered, infrared Thermometer that coverts a user's forehead temperature, using the infrared energy emitted in the area around the user's forehead to an oral equivalent temperature when meas ure from 3-5 cm of the subiect's forehead with no contact.
lt uses a thermopile sensor with integrated thermistor for the target reading and a thermistor mounted in the head of the thermometer for ambient temperature readings.
It composed by a measuring sensor, PCB, 4 buttons, a LCD and an enclosure. The functions of the three models are similar. The functions of "Temperature Alarm Point", "Temperature Offset", and "Warning Tone Switch" can be set. When measuring body temperature, users need to measure in body mode from 3-5 cm from their forehead. Press the measuring key, after 0.5 second with the sound of "beep", the measurement is completed and the temperature is displayed on the LCD screen. Without any operation, it will close automatically in 15 sec. The No. and stored value of the stored data are displayed at the bottom of the LCD screen. Press "+" / " -" to view the previous or next stored data, it can store 32 sets of measurements.
User contact components are the enclosure and keys. The materials of the components are ABS plastic.
## 6. Intended Use / Indications for Use
Infrared Thermometer (model: CK-T1501, CK-T1502, CK-T1503) is a non-sterile, reusable, non-contact and handheld device. It can be used by consumers in homecare environment and doctors in clinic as reference. It is intended for measuring human body temperature of people over one month old by detecting infrared heat from the forehead.
#### 7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of Infrared Thermometer is substantially equivalent to the predicate devices quoted above.
The differences between the subject device and predicate devices do not raise new issues of
{5}------------------------------------------------
safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Company | Shenzhen Changkun<br>Technology Co., Ltd. | Intrinity Global Limited | -- |
| Trade Name | Infrared Thermometer | Non Contact Infrared Forehead<br>Thermometer | -- |
| Classification<br>Name | Clinical Electronic Thermometer | Clinical Electronic Thermometer | Same |
| 510(k)<br>Number | K193253 | K170662 | -- |
| Product Code | FLL | FLL | Same |
| Thermometer<br>Type | Infrared Forehead Thermometer | Infrared Forehead Thermometer | Same |
| Intended Use<br>/ Indications<br>for Use | Infrared Thermometer (model:<br>CK-T1501, CK-T1502,<br>CK-T1503) is a non-sterile,<br>reusable, non-contact and<br>handheld device. It can be used<br>by consumers in homecare<br>environment and doctors in clinic<br>as reference. It is intended for<br>measuring human body<br>temperature of people over one<br>month old by detecting infrared<br>heat from the forehead. | Non Contact Infrared Forehead<br>Thermometer is a non-sterile,<br>reusable, handheld device. It<br>can be used by consumers in<br>homecare environment and<br>doctors in clinic as reference. It<br>is intended for measuring<br>human body temperature of all<br>ranges of people by detecting<br>infrared heat from the forehead. | Similar<br>Note 3 |
| Display | LCD Digital Display | LCD Digital Display | Same |
| Measurement<br>method | Infrared radiation detection | Infrared radiation detection | Same |
| Measurement<br>mode | Forehead measure mode | Forehead measure mode | Same |
| Measuring<br>range | 32°C~42.5°C (89.6°F~108.5°F) | 32°C to 43°C (89.6°Fto109.4°F) | Similar<br>Note 1 |
| Display<br>resolution | 0.1°C/0.1°F | 0.1°C/0.1°F | Same |
| C/F<br>switchable | YES | YES | Same |
| Measuring | 32°C~34.9°C ±0.3°C/ | +0.2°C (0.4°F) | Similar |
| Elements of | Subject Device | Predicate Device | Verdict |
| Comparison | | | |
| accuracy | 89.6 °F~94.8°F ±0.5°F<br>35°C~42°C ±0.2°C/<br>95.0°F~107.6°F±0.4°F<br>42.1°C~42.5°C ±0.3°C/<br>107.8°F~108.5°F ±0.5°F | | Note 1 |
| Measurement<br>distance | 3-5cm | 1cm | Similar<br>Note 1 |
| Display | LCD | LCD | Same |
| Memory | 32 sets. | 16 sets. | Similar<br>Note 2 |
| Power source | DC 3V<br>(2 of AA alkaline<br>batteries) | DC 3V<br>(2 of AA alkaline<br>batteries ) | Same |
| Operating<br>condition | 10°C ~ 40°C;<br>15% ~ 85%RH;<br>80kPa~106kPa | 15°C~40°C (59°F~ 104°F);<br>≤95% RH | Similar<br>Note 2 |
| Patient<br>contact<br>materials | ABS | ABS with colorants (pink, grey,<br>orange and purple), Glass and<br>Metal | Same |
| Electric<br>Safety<br>and EMC | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-56 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-56 | Same |
{6}------------------------------------------------
Sponsor: Shenzhen Changkun Technology Co., Ltd. Infrared Thermometer, Models: CK-T1501, CK-T1502, CK-T1503 Subject Device:
# Comparison in Detail(s):
## Note 1:
The "measuring range", "Measurement distance" and "Measuring accuracy" of the subject device is similar with predicate device, both of them meet the requirement of safety and essential performance standard ISO 80601-2-56. The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
## Note 2:
The "Memory" and "Operating condition" of subject device is similar with predicate device, the software verification and validation test met the requirements. The performance testing shows that the subject device complies with performance standard.
The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
{7}------------------------------------------------
## Note 3:
Although the "Intended Use / Indications for Use" of subject device is a little different from the predicate devices, but both of subject device and predicate device are meet the clinical accuracy requirements of the standards ISO 80601-2-56 and ASTM E1965-98. Based on the performance evaluation, the differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
## 8. Summary of Non-Clinical Testing
8.1 Non-clinical testing was conducted to verify that the subject devicesmet all design specifications, demonstrated safety based on current industry standards, and to demonstrate substantial equivalence to the predicate. The following tests were performed: A. Electromagnetic Compatibility, Electrical Safety, and Battery Safety:
The subject devices were tested in compliance with the following:
- � ANSI/AAMIES60601-1 Medicalelectricalequipment-Part 1: General requirements for basic safety and essential performance
- � IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- � IEC 60601-1-11 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- � ISO 80601-2-56 First Edition 2009-10-01, Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
#### B. Biocompatibility:
- � Patient contactingcomponents were subjected to biocompatibility testing in compliance with ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within
{8}------------------------------------------------
a risk management process, including cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10) and irritation (ISO 10993-10).
C. Software Verification: Software documentation was provided in accordance with FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued May 11, 2005.
# D. Performance Testing:
- ◆ ISO 80601-2-56 First Edition 2009-10-01, Medicalelectrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinicalthermometers for body temperature measurement
- ◆ ASTME1965-98 (2016): Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
# 8.2 Discussion of Clinical TestsPerformed
The clinical performance test protocol and data analysis is conducted as the requirement of ASTM E1965-98 (2009). The test report showed the clinical performance of the subject device complied with the requirement of ASTM E1965-98 (2016).
The clinical tests evaluated 240 of subjects. and the thermometer was evaluated in three age groups including subgroup A1 and A2: A1 – one month up to three month, A2 - three months to one year; B - older than one years and younger than five years; and C - older than five years old. The clinical performance test protocol and data analysis were conducted in accordance with the requirement of ISO 80601-2-56 and ASTM E1965-98 (2016). The test report showed the clinical performance of the subject devices complied with the requirement of ISO 80601-2-56 and ASTM E1965-98 (2016).
## 9. Conclusion:
Based on the performance testing, comparison and analysis in this submission, the subject device Infrared Thermometer (model: CK-T1501, CK-T1502, CK-T1503) is substantially equivalent to the predicate device K 170662.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.