← Product Code [QKB](/productcode/QKB) · K193252

# Contour ProtegeAI (K193252)

_Mim Software, Inc. · QKB · Jul 2, 2020 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K193252

## Device Facts

- **Applicant:** Mim Software, Inc.
- **Product Code:** [QKB](/productcode/QKB.md)
- **Decision Date:** Jul 2, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device

## Indications for Use

Contour ProtégéAI is used by trained medical professionals as a tool to aid in the automated processing of digital medical images of modalities CT and MR, as supported by ACR/NEMA DICOM 3.0. Contour ProtégéAl assists in the following indications: The creation of contours using machine-learning algorithms for applications including, but not limited to, quantitative analysis, aiding adaptive therapy, transferring contours to radiation therapy treatment planning systems, and archiving contours for patient follow-up and management. Segmenting normal structures across a variety of CT anatomical locations. And segmenting normal structures of the prostate, seminal vesicles, and urethra within T2-weighted MR images. Contour ProtégéAI must be used in conjunction with MIM software to review and, if necessary, edit contours that were automatically generated by Contour ProtégéAI.

## Device Story

Contour ProtégéAI is a server-based software accessory to MIM software; it processes DICOM-compliant CT and MR images. It utilizes machine-learning algorithms to automatically generate anatomical contours for radiation therapy planning, adaptive therapy, and quantitative analysis. The device is operated by trained medical professionals (e.g., radiologists, oncologists, physicists). The system is deployed on a remote server using MIMcloud for data management. The healthcare provider reviews and edits the automatically generated contours within the MIM software environment. By automating the initial contouring process, the device aims to reduce the time required for manual segmentation and support clinical workflows.

## Clinical Evidence

Bench testing only. Performance evaluated using 286 CT images and 72 MR images from institutions disjoint from the training set. Non-inferiority testing compared mean Dice coefficients of Contour ProtégéAI against the predicate device. Results confirmed non-inferiority with a limit of 0.1 Dice, deemed clinically acceptable.

## Technological Characteristics

Server-based software application supporting Linux-based OS. Uses machine-learning algorithms for automated image segmentation. Compatible with ACR/NEMA DICOM 3.0 standards. Deployed via MIMcloud for data management and transfer. Operates on Windows, Mac, and Linux client systems.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- MIM on Linux ([K190379](/device/K190379.md))
- Workflow Box ([K181572](/device/K181572.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

MIM Software, Inc. % Ms. Lynn Hanigan Quality Assurance Director 25800 Science Park Drive, Suite 180 CLEVELAND OH 44122

Re: K193252

Trade/Device Name: Contour ProtégéAI Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: QKB Dated: June 1, 2020 Received: June 3, 2020

Dear Ms. Hanigan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193252

Device Name Contour ProtégéAI

#### Indications for Use (Describe)

Contour ProtégéAI is used by trained medical professionals as a tool to aid in the automated processing of digital medical images of modalities CT and MR, as supported by ACR/NEMA DICOM 3.0. Contour ProtégéAl assists in the following indications:

The creation of contours using machine-learning algorithms for applications including, but not limited to, quantitative analysis, aiding adaptive therapy, transferring contours to radiation therapy treatment planning systems, and archiving contours for patient follow-up and management.

Segmenting normal structures across a variety of CT anatomical locations.

And segmenting normal structures of the prostate, seminal vesicles, and urethra within T2-weighted MR images.

Contour ProtégéAI must be used in conjunction with MIM software to review and, if necessary, edit contours that were automatically generated by Contour ProtégéAI.

| Type of Use (Select one or both, as applicable)                                                                          |                                                                                                                         |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><div>☒</div><div>Prescription Use (Part 21 CFR 801 Subpart D)</div></div> | <div style="display:flex; align-items:center;"><div>☐</div><div>Over-The-Counter Use (21 CFR 801 Subpart C)</div></div> |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle cut out of the red square. To the right of the squares is the text "mim SOFTWARE" in black. The word "mim" is in a larger font than the word "SOFTWARE".

510(k) Summary of Safety and Effectiveness (The following information is in conformance with 21 CFR 807.92)

# K193252

#### Submitter:

MIM Software Inc. 25800 Science Park Drive - Suite 180 Cleveland, OH 44122

| Phone:                 | 216-455-0600 |
|------------------------|--------------|
| Fax:                   | 216-455-0601 |
| Contact Person:        | Lynn Hanigan |
| Date Summary Prepared: | 06/23/2020   |

#### Device Name

Trade Name: Contour ProtégéAl Common Name: Medical Imaging Software Requlation Number / Product Code: 21 CFR 892.2050 Product Code QKB Classification Name: System, Imaging Processing, Radiological

#### Predicate Devices

| K190379 | MIM on Linux | MIM Software Inc.   |
|---------|--------------|---------------------|
| K181572 | Workflow Box | Mirada Medical Ltd. |

#### Intended Use

Contour ProtégéAl is an accessory to MIM software used for the contouring of anatomical structures in imaging data using machine-learning-based algorithms automatically.

Contour ProtégéAl must be used in conjunction with MIM software to review and, if necessary, edit results automatically generated by Contour ProtégéAl.

Contour ProtégéAl is not intended to detect or contour lesions automatically.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle cut out of the red square. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it.

### Indications for Use

Contour ProtégéAl is used by trained medical professionals as a tool to aid in the automated processing of digital medical images of modalities CT and MR, as supported by ACR/NEMA DICOM 3.0. Contour ProtégéAl assists in the following indications:

- . The creation of contours using machine-learning algorithms for applications including, but not limited to, quantitative analysis, aiding adaptive therapy, transferring contours to radiation therapy treatment planning systems, and archiving contours for patient follow-up and management.
- Segmenting normal structures across a variety of CT anatomical locations. .
- . And segmenting normal structures of the prostate, seminal vesicles, and urethra within T2-weighted MR images.

Contour ProtégéAl must be used in conjunction with MIM software to review and, if necessary, edit contours automatically generated by Contour ProtégéAl.

#### Device Description

Contour ProtégéAl is an accessory to MIM software that automatically creates contours on medical images through the use of machine-learning algorithms. It is designed for use in the processing of medical images and operates on Windows, Mac, and Linux computer systems. Contour ProtégéAl is deployed on a remote server using the MIMcloud service for data management and transfer.

# Substantial Equivalence

| ITEM                                                               | MIM Software Inc.<br>Contour ProtégéAl<br>(K193252)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | MIM Software Inc.<br>MIM on Linux<br>(K190379)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Mirada Medical Ltd.<br>Workflow Box<br>(K181572)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clearance Date                                                     | TBD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 03-19-2019                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 07-10-2018                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Intended Use                                                       | Contour ProtégéAl is an<br>accessory to MIM software<br>used for the contouring of<br>anatomical structures in<br>imaging data using machine-<br>learning-based algorithms<br>automatically.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | MIM software is intended for<br>trained medical professionals<br>including, but not limited to,<br>radiologists, oncologists,<br>physicians, medical<br>technologists, dosimetrists and<br>physicists.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Workflow Box is a system<br>designed to allow users to route<br>DICOM-compliant data to and<br>from automated processing<br>components.<br>Workflow Box includes<br>processing components for<br>automatically contouring<br>imaging data using deformable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| ITEM                                                               | MIM Software Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | MIM Software Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Mirada Medical Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                    | Contour ProtégéAl                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | MIM on Linux                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Workflow Box                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                                    | (K193252)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | (K190379)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | (K181572)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                                    | Contour ProtégéAl must be<br>used in conjunction with MIM<br>software to review and, if<br>necessary, edit results<br>automatically generated by<br>Contour ProtégéAl.<br>Contour ProtégéAl is not<br>intended to detect or contour<br>lesions automatically.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | MIM is a medical image and<br>information management<br>system that is intended to<br>receive, transmit, store,<br>retrieve, display, print and<br>process digital medical images,<br>as well as create, display and<br>print reports from those images.<br>The medical modalities of these<br>medical imaging systems<br>include, but are not limited to,<br>CT, MRI, CR, DX, MG, US,<br>SPECT, PET and XA as<br>supported by ACR/NEMA<br>DICOM 3.0.<br>MIM provides the user with the<br>means to display, register and<br>fuse medical images from<br>multiple modalities.<br>Additionally, it evaluates<br>cardiac left ventricular function<br>and perfusion, including left<br>ventricular end-diastolic<br>volume, end-systolic volume,<br>and ejection fraction.<br>The Region of Interest (ROI)<br>feature reduces the time<br>necessary for the user to define<br>objects in medical image<br>volumes by providing an initial<br>definition of object contours.<br>The objects include, but are not<br>limited to, tumors and normal<br>tissues.<br>MIM provides tools to quickly<br>create, transform, and modify<br>contours for applications<br>including, but not limited to,<br>quantitative analysis, aiding<br>adaptive therapy, transferring<br>contours to radiation therapy<br>treatment planning systems<br>and archiving contours for | image registration and machine<br>learning based algorithms.<br>Workflow Box must be used in<br>conjunction with appropriate<br>software to review and edit<br>results generated automatically<br>by Workflow Box components,<br>for example image visualization<br>software must be used to<br>facilitate the review and edit of<br>contours generated by<br>Workflow Box component<br>applications.<br>Workflow Box is not intended to<br>automatically detect lesions.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| ITEM                                                               | MIM Software Inc.<br>Contour ProtégéAl<br>(K193252)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | MIM Software Inc.<br>MIM on Linux<br>(K190379)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Mirada Medical Ltd.<br>Workflow Box<br>(K181572)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | patient follow-up and<br>management.<br>MIM aids in the assessment of<br>PET/SPECT brain scans. It<br>provides automated<br>quantitative and statistical<br>analysis by automatically<br>registering PET/SPECT brain<br>scans to a standard template<br>and comparing intensity values<br>to a reference database or to<br>other PET/SPECT scans on a<br>voxel by voxel basis, within<br>stereotactic surface projections<br>or standardized regions of<br>interest.<br>MIM allows the dose<br>distribution of an implant to be<br>individually shaped for each<br>patient and is a general-<br>purpose brachytherapy<br>planning system used for<br>prospective and confirmation<br>dose calculations for patients<br>undergoing a course of<br>brachytherapy using permanent<br>implants of various<br>radioisotopes (not including<br>radioactive microspheres).<br>MIM allows voxel-based dose<br>calculations for patients who<br>have been administered<br>radioisotopes or radioactive<br>microspheres.                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Indications for Use                                                | Contour ProtégéAl is used by<br>trained medical professionals<br>as a tool to aid in the automated<br>processing of digital medical<br>images of modalities CT and<br>MR, as supported by<br>ACR/NEMA DICOM 3.0.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | MIM software is used by trained<br>medical professionals as a tool<br>to aid in evaluation and<br>information management of<br>digital medical images. The<br>medical image modalities<br>include, but are not limited to,<br>CT, MRI, CR, DX, MG, US,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Workflow Box is a software<br>system designed to allow users<br>to route DICOM-compliant data<br>to and from automated<br>processing components.<br>Supported modalities include<br>CT, MR, RTSTRUCT.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| ITEM                                                               | MIM Software Inc.<br>Contour ProtégéAl<br>(K193252)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | MIM Software Inc.<br>MIM on Linux<br>(K190379)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Mirada Medical Ltd.<br>Workflow Box<br>(K181572)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                    | Contour ProtégéAl assists in<br>the following indications:<br>The creation of<br>contours using<br>machine-learning<br>algorithms for<br>applications including,<br>but not limited to,<br>quantitative analysis,<br>aiding adaptive<br>therapy, transferring<br>contours to radiation<br>therapy treatment<br>planning systems, and<br>archiving contours for<br>patient follow-up and<br>management. Segmenting normal<br>structures across a<br>variety of CT<br>anatomical locations. And segmenting<br>normal structures of<br>the prostate, seminal<br>vesicles, and urethra<br>within T2-weighted MR<br>images.              Contour ProtégéAl must be<br>used in conjunction with MIM<br>software to review and, if<br>necessary, edit contours that<br>were automatically generated<br>by Contour ProtégéAl. | SPECT, PET and XA as<br>supported by ACR/NEMA<br>DICOM 3.0. MIM assists in the<br>following indications:<br>Receive, transmit, store,<br>retrieve, display, print, and<br>process medical images<br>and DICOM objects. Create, display and print<br>reports from medical<br>images. Registration, fusion<br>display, and review of<br>medical images for<br>diagnosis, treatment<br>evaluation, and treatment<br>planning. Evaluation of cardiac left<br>ventricular function and<br>perfusion, including left<br>ventricular end-diastolic<br>volume, end-systolic<br>volume, and ejection<br>fraction. Localization and definition<br>of objects such as tumors<br>and normal tissues in<br>medical images. Creation, transformation,<br>and modification of<br>contours for applications<br>including, but not limited<br>to, quantitative analysis,<br>aiding adaptive therapy,<br>transferring contours to<br>radiation therapy<br>treatment planning<br>systems, and archiving<br>contours for patient follow-<br>up and management. Quantitative and statistical<br>analysis of PET/SPECT<br>brain scans by comparing<br>to other registered<br>PET/SPECT brain scans.                                                                                                                                                                               | Workflow Box includes<br>processing components for<br>automatically contouring<br>imaging data using deformable<br>image registration to support<br>atlas-based contouring, re-<br>contouring of the same patient<br>and machine learning based<br>contouring.<br><br>Workflow Box is a data routing<br>and image processing tool<br>which automatically applies<br>contours to data which is sent<br>to one or more of the included<br>image processing workflows.<br>Contours generated by<br>Workflow Box may be used as<br>an input to clinical workflows<br>including, but not limited to,<br>radiation therapy treatment<br>planning.<br><br>Workflow Box must be used in<br>conjunction with appropriate<br>software to review and edit<br>results generated automatically<br>by Workflow Box components,<br>for example image visualization<br>software must be used to<br>facilitate the review and edit of<br>contours generated by<br>Workflow Box component<br>applications.<br><br>Workflow Box is intended to be<br>used by trained medical<br>professionals.<br><br>Workflow Box is not intended to<br>automatically detect lesions. |
| ITEM                                                               | MIM Software Inc.<br>Contour ProtégéAl<br>(K193252)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | MIM Software Inc.<br>MIM on Linux<br>(K190379)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Mirada Medical Ltd.<br>Workflow Box<br>(K181572)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Planning and evaluation of permanent implant<br>brachytherapy procedures (not including radioactive<br>microspheres). Calculating absorbed radiation dose as a result<br>of administering a radionuclide.              When using device clinically, the user should only use FDA<br>approved radiopharmaceuticals. If using with unapproved ones, this<br>device should only be used for research purposes.<br>Lossy compressed mammographic images and digitized film screen images<br>must not be reviewed for primary image interpretations. Images that are printed to film<br>must be printed using an FDA-approved printer for the diagnosis of digital<br>mammography images.<br>Mammographic images must be viewed on a display system that has been cleared by the<br>FDA for the diagnosis of digital mammography images. The software is not to be used for                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Modalities                                                         | CT and MR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | CT, MR, CR, DX, MG, US,<br>NM, PET, XA, and other<br>DICOM modalities                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | CT and MR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| ITEM                                                               | MIM Software Inc.<br>Contour ProtégéAl<br>(K193252)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | MIM Software Inc.<br>MIM on Linux<br>(K190379)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Mirada Medical Ltd.<br>Workflow Box<br>(K181572)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Atlas-based<br>Contour<br>Segmentation                             | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Automatically<br>Contour Imaging<br>Data Using<br>Machine-Learning | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Operating Platform                                                 | Server-based application<br>supporting<br>Linux-based OS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Microsoft® Windows,<br>Apple® OS,<br>Linux-based OS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Server based application<br>supporting<br>Microsoft Windows 10 (64-bit)<br>and<br>Microsoft Windows Server<br>2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is clean and modern, with a simple color palette.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is simple and modern, and the colors are eye-catching.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is simple and modern, and the colors are eye-catching.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is simple and modern, and the colors are eye-catching.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is clean and modern, with a simple color palette.

Contour ProtégéAl is substantially equivalent to a combination of the predicate devices MIM on Linux (K190379) and Mirada Workflow Box (K181572).

# Testing and Performance Data

The neural networks used in the Contour ProtégéAl device were trained on datasets from several large institutions. These datasets included CT images and MR images and their associated segmentations.

For testing, 286 images were used to evaluate the neural network models that segmented CT images, while 72 images were used to evaluate for the MR segmentation network. In all cases, the test images were gathered from a different and disjoint set of institutions from the training data. Both Contour ProtégéAI and the MIM predicate device were used to automatically segmented the independent test sets to show substantial equivalence.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font…

---

**Source:** [https://fda-staging.innolitics.com/device/K193252](https://fda-staging.innolitics.com/device/K193252)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com