Datasets from several large institutions including CT images and their associated segmentations.
—
286 images from a different and disjoint set of institutions from the training data.
—
MR prostate, seminal vesicle, and urethra segmentation
Neural networks
Non-inferiority limit of 0.1 Dice
Non-inferior to predicate
Datasets from several large institutions including MR images and their associated segmentations.
—
72 images from a different and disjoint set of institutions from the training data.
—
Indications for Use
Contour ProtégéAI is used by trained medical professionals as a tool to aid in the automated processing of digital medical images of modalities CT and MR, as supported by ACR/NEMA DICOM 3.0. Contour ProtégéAl assists in the following indications: The creation of contours using machine-learning algorithms for applications including, but not limited to, quantitative analysis, aiding adaptive therapy, transferring contours to radiation therapy treatment planning systems, and archiving contours for patient follow-up and management. Segmenting normal structures across a variety of CT anatomical locations. And segmenting normal structures of the prostate, seminal vesicles, and urethra within T2-weighted MR images. Contour ProtégéAI must be used in conjunction with MIM software to review and, if necessary, edit contours that were automatically generated by Contour ProtégéAI.
Device Story
Contour ProtégéAI is a server-based software accessory to MIM software; it processes DICOM-compliant CT and MR images. It utilizes machine-learning algorithms to automatically generate anatomical contours for radiation therapy planning, adaptive therapy, and quantitative analysis. The device is operated by trained medical professionals (e.g., radiologists, oncologists, physicists). The system is deployed on a remote server using MIMcloud for data management. The healthcare provider reviews and edits the automatically generated contours within the MIM software environment. By automating the initial contouring process, the device aims to reduce the time required for manual segmentation and support clinical workflows.
Clinical Evidence
Bench testing only. Performance evaluated using 286 CT images and 72 MR images from institutions disjoint from the training set. Non-inferiority testing compared mean Dice coefficients of Contour ProtégéAI against the predicate device. Results confirmed non-inferiority with a limit of 0.1 Dice, deemed clinically acceptable.
Technological Characteristics
Server-based software application supporting Linux-based OS. Uses machine-learning algorithms for automated image segmentation. Compatible with ACR/NEMA DICOM 3.0 standards. Deployed via MIMcloud for data management and transfer. Operates on Windows, Mac, and Linux client systems.
Indications for Use
Indicated for trained medical professionals to aid in automated segmentation of normal anatomical structures in CT images and normal structures (prostate, seminal vesicles, urethra) in T2-weighted MR images. Not intended for lesion detection or contouring.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
MIM Software, Inc. % Ms. Lynn Hanigan Quality Assurance Director 25800 Science Park Drive, Suite 180 CLEVELAND OH 44122
Re: K193252
Trade/Device Name: Contour ProtégéAI Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: QKB Dated: June 1, 2020 Received: June 3, 2020
Dear Ms. Hanigan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193252
Device Name Contour ProtégéAI
#### Indications for Use (Describe)
Contour ProtégéAI is used by trained medical professionals as a tool to aid in the automated processing of digital medical images of modalities CT and MR, as supported by ACR/NEMA DICOM 3.0. Contour ProtégéAl assists in the following indications:
The creation of contours using machine-learning algorithms for applications including, but not limited to, quantitative analysis, aiding adaptive therapy, transferring contours to radiation therapy treatment planning systems, and archiving contours for patient follow-up and management.
Segmenting normal structures across a variety of CT anatomical locations.
And segmenting normal structures of the prostate, seminal vesicles, and urethra within T2-weighted MR images.
Contour ProtégéAI must be used in conjunction with MIM software to review and, if necessary, edit contours that were automatically generated by Contour ProtégéAI.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><div>☒</div><div>Prescription Use (Part 21 CFR 801 Subpart D)</div></div> | <div style="display:flex; align-items:center;"><div>☐</div><div>Over-The-Counter Use (21 CFR 801 Subpart C)</div></div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle cut out of the red square. To the right of the squares is the text "mim SOFTWARE" in black. The word "mim" is in a larger font than the word "SOFTWARE".
510(k) Summary of Safety and Effectiveness (The following information is in conformance with 21 CFR 807.92)
# K193252
#### Submitter:
MIM Software Inc. 25800 Science Park Drive - Suite 180 Cleveland, OH 44122
| Phone: | 216-455-0600 |
|------------------------|--------------|
| Fax: | 216-455-0601 |
| Contact Person: | Lynn Hanigan |
| Date Summary Prepared: | 06/23/2020 |
#### Device Name
Trade Name: Contour ProtégéAl Common Name: Medical Imaging Software Requlation Number / Product Code: 21 CFR 892.2050 Product Code QKB Classification Name: System, Imaging Processing, Radiological
#### Predicate Devices
| K190379 | MIM on Linux | MIM Software Inc. |
|---------|--------------|---------------------|
| K181572 | Workflow Box | Mirada Medical Ltd. |
#### Intended Use
Contour ProtégéAl is an accessory to MIM software used for the contouring of anatomical structures in imaging data using machine-learning-based algorithms automatically.
Contour ProtégéAl must be used in conjunction with MIM software to review and, if necessary, edit results automatically generated by Contour ProtégéAl.
Contour ProtégéAl is not intended to detect or contour lesions automatically.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle cut out of the red square. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it.
### Indications for Use
Contour ProtégéAl is used by trained medical professionals as a tool to aid in the automated processing of digital medical images of modalities CT and MR, as supported by ACR/NEMA DICOM 3.0. Contour ProtégéAl assists in the following indications:
- . The creation of contours using machine-learning algorithms for applications including, but not limited to, quantitative analysis, aiding adaptive therapy, transferring contours to radiation therapy treatment planning systems, and archiving contours for patient follow-up and management.
- Segmenting normal structures across a variety of CT anatomical locations. .
- . And segmenting normal structures of the prostate, seminal vesicles, and urethra within T2-weighted MR images.
Contour ProtégéAl must be used in conjunction with MIM software to review and, if necessary, edit contours automatically generated by Contour ProtégéAl.
#### Device Description
Contour ProtégéAl is an accessory to MIM software that automatically creates contours on medical images through the use of machine-learning algorithms. It is designed for use in the processing of medical images and operates on Windows, Mac, and Linux computer systems. Contour ProtégéAl is deployed on a remote server using the MIMcloud service for data management and transfer.
# Substantial Equivalence
| ITEM | MIM Software Inc.<br>Contour ProtégéAl<br>(K193252) | MIM Software Inc.<br>MIM on Linux<br>(K190379) | Mirada Medical Ltd.<br>Workflow Box<br>(K181572) |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clearance Date | TBD | 03-19-2019 | 07-10-2018 |
| Intended Use | Contour ProtégéAl is an<br>accessory to MIM software<br>used for the contouring of<br>anatomical structures in<br>imaging data using machine-<br>learning-based algorithms<br>automatically. | MIM software is intended for<br>trained medical professionals<br>including, but not limited to,<br>radiologists, oncologists,<br>physicians, medical<br>technologists, dosimetrists and<br>physicists. | Workflow Box is a system<br>designed to allow users to route<br>DICOM-compliant data to and<br>from automated processing<br>components.<br>Workflow Box includes<br>processing components for<br>automatically contouring<br>imaging data using deformable |
| ITEM | MIM Software Inc. | MIM Software Inc. | Mirada Medical Ltd. |
| | Contour ProtégéAl | MIM on Linux | Workflow Box |
| | (K193252) | (K190379) | (K181572) |
| | Contour ProtégéAl must be<br>used in conjunction with MIM<br>software to review and, if<br>necessary, edit results<br>automatically generated by<br>Contour ProtégéAl.<br>Contour ProtégéAl is not<br>intended to detect or contour<br>lesions automatically. | MIM is a medical image and<br>information management<br>system that is intended to<br>receive, transmit, store,<br>retrieve, display, print and<br>process digital medical images,<br>as well as create, display and<br>print reports from those images.<br>The medical modalities of these<br>medical imaging systems<br>include, but are not limited to,<br>CT, MRI, CR, DX, MG, US,<br>SPECT, PET and XA as<br>supported by ACR/NEMA<br>DICOM 3.0.<br>MIM provides the user with the<br>means to display, register and<br>fuse medical images from<br>multiple modalities.<br>Additionally, it evaluates<br>cardiac left ventricular function<br>and perfusion, including left<br>ventricular end-diastolic<br>volume, end-systolic volume,<br>and ejection fraction.<br>The Region of Interest (ROI)<br>feature reduces the time<br>necessary for the user to define<br>objects in medical image<br>volumes by providing an initial<br>definition of object contours.<br>The objects include, but are not<br>limited to, tumors and normal<br>tissues.<br>MIM provides tools to quickly<br>create, transform, and modify<br>contours for applications<br>including, but not limited to,<br>quantitative analysis, aiding<br>adaptive therapy, transferring<br>contours to radiation therapy<br>treatment planning systems<br>and archiving contours for | image registration and machine<br>learning based algorithms.<br>Workflow Box must be used in<br>conjunction with appropriate<br>software to review and edit<br>results generated automatically<br>by Workflow Box components,<br>for example image visualization<br>software must be used to<br>facilitate the review and edit of<br>contours generated by<br>Workflow Box component<br>applications.<br>Workflow Box is not intended to<br>automatically detect lesions. |
| ITEM | MIM Software Inc.<br>Contour ProtégéAl<br>(K193252) | MIM Software Inc.<br>MIM on Linux<br>(K190379) | Mirada Medical Ltd.<br>Workflow Box<br>(K181572) |
| | | patient follow-up and<br>management.<br>MIM aids in the assessment of<br>PET/SPECT brain scans. It<br>provides automated<br>quantitative and statistical<br>analysis by automatically<br>registering PET/SPECT brain<br>scans to a standard template<br>and comparing intensity values<br>to a reference database or to<br>other PET/SPECT scans on a<br>voxel by voxel basis, within<br>stereotactic surface projections<br>or standardized regions of<br>interest.<br>MIM allows the dose<br>distribution of an implant to be<br>individually shaped for each<br>patient and is a general-<br>purpose brachytherapy<br>planning system used for<br>prospective and confirmation<br>dose calculations for patients<br>undergoing a course of<br>brachytherapy using permanent<br>implants of various<br>radioisotopes (not including<br>radioactive microspheres).<br>MIM allows voxel-based dose<br>calculations for patients who<br>have been administered<br>radioisotopes or radioactive<br>microspheres. | |
| Indications for Use | Contour ProtégéAl is used by<br>trained medical professionals<br>as a tool to aid in the automated<br>processing of digital medical<br>images of modalities CT and<br>MR, as supported by<br>ACR/NEMA DICOM 3.0. | MIM software is used by trained<br>medical professionals as a tool<br>to aid in evaluation and<br>information management of<br>digital medical images. The<br>medical image modalities<br>include, but are not limited to,<br>CT, MRI, CR, DX, MG, US, | Workflow Box is a software<br>system designed to allow users<br>to route DICOM-compliant data<br>to and from automated<br>processing components.<br>Supported modalities include<br>CT, MR, RTSTRUCT. |
| ITEM | MIM Software Inc.<br>Contour ProtégéAl<br>(K193252) | MIM Software Inc.<br>MIM on Linux<br>(K190379) | Mirada Medical Ltd.<br>Workflow Box<br>(K181572) |
| | Contour ProtégéAl assists in<br>the following indications:<br>The creation of<br>contours using<br>machine-learning<br>algorithms for<br>applications including,<br>but not limited to,<br>quantitative analysis,<br>aiding adaptive<br>therapy, transferring<br>contours to radiation<br>therapy treatment<br>planning systems, and<br>archiving contours for<br>patient follow-up and<br>management. Segmenting normal<br>structures across a<br>variety of CT<br>anatomical locations. And segmenting<br>normal structures of<br>the prostate, seminal<br>vesicles, and urethra<br>within T2-weighted MR<br>images. Contour ProtégéAl must be<br>used in conjunction with MIM<br>software to review and, if<br>necessary, edit contours that<br>were automatically generated<br>by Contour ProtégéAl. | SPECT, PET and XA as<br>supported by ACR/NEMA<br>DICOM 3.0. MIM assists in the<br>following indications:<br>Receive, transmit, store,<br>retrieve, display, print, and<br>process medical images<br>and DICOM objects. Create, display and print<br>reports from medical<br>images. Registration, fusion<br>display, and review of<br>medical images for<br>diagnosis, treatment<br>evaluation, and treatment<br>planning. Evaluation of cardiac left<br>ventricular function and<br>perfusion, including left<br>ventricular end-diastolic<br>volume, end-systolic<br>volume, and ejection<br>fraction. Localization and definition<br>of objects such as tumors<br>and normal tissues in<br>medical images. Creation, transformation,<br>and modification of<br>contours for applications<br>including, but not limited<br>to, quantitative analysis,<br>aiding adaptive therapy,<br>transferring contours to<br>radiation therapy<br>treatment planning<br>systems, and archiving<br>contours for patient follow-<br>up and management. Quantitative and statistical<br>analysis of PET/SPECT<br>brain scans by comparing<br>to other registered<br>PET/SPECT brain scans. | Workflow Box includes<br>processing components for<br>automatically contouring<br>imaging data using deformable<br>image registration to support<br>atlas-based contouring, re-<br>contouring of the same patient<br>and machine learning based<br>contouring.<br><br>Workflow Box is a data routing<br>and image processing tool<br>which automatically applies<br>contours to data which is sent<br>to one or more of the included<br>image processing workflows.<br>Contours generated by<br>Workflow Box may be used as<br>an input to clinical workflows<br>including, but not limited to,<br>radiation therapy treatment<br>planning.<br><br>Workflow Box must be used in<br>conjunction with appropriate<br>software to review and edit<br>results generated automatically<br>by Workflow Box components,<br>for example image visualization<br>software must be used to<br>facilitate the review and edit of<br>contours generated by<br>Workflow Box component<br>applications.<br><br>Workflow Box is intended to be<br>used by trained medical<br>professionals.<br><br>Workflow Box is not intended to<br>automatically detect lesions. |
| ITEM | MIM Software Inc.<br>Contour ProtégéAl<br>(K193252) | MIM Software Inc.<br>MIM on Linux<br>(K190379) | Mirada Medical Ltd.<br>Workflow Box<br>(K181572) |
| | | Planning and evaluation of permanent implant<br>brachytherapy procedures (not including radioactive<br>microspheres). Calculating absorbed radiation dose as a result<br>of administering a radionuclide. When using device clinically, the user should only use FDA<br>approved radiopharmaceuticals. If using with unapproved ones, this<br>device should only be used for research purposes.<br>Lossy compressed mammographic images and digitized film screen images<br>must not be reviewed for primary image interpretations. Images that are printed to film<br>must be printed using an FDA-approved printer for the diagnosis of digital<br>mammography images.<br>Mammographic images must be viewed on a display system that has been cleared by the<br>FDA for the diagnosis of digital mammography images. The software is not to be used for | |
| Modalities | CT and MR | CT, MR, CR, DX, MG, US,<br>NM, PET, XA, and other<br>DICOM modalities | CT and MR |
| ITEM | MIM Software Inc.<br>Contour ProtégéAl<br>(K193252) | MIM Software Inc.<br>MIM on Linux<br>(K190379) | Mirada Medical Ltd.<br>Workflow Box<br>(K181572) |
| Atlas-based<br>Contour<br>Segmentation | No | Yes | Yes |
| Automatically<br>Contour Imaging<br>Data Using<br>Machine-Learning | Yes | No | Yes |
| Operating Platform | Server-based application<br>supporting<br>Linux-based OS | Microsoft® Windows,<br>Apple® OS,<br>Linux-based OS | Server based application<br>supporting<br>Microsoft Windows 10 (64-bit)<br>and<br>Microsoft Windows Server<br>2016 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is clean and modern, with a simple color palette.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is simple and modern, and the colors are eye-catching.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is simple and modern, and the colors are eye-catching.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is simple and modern, and the colors are eye-catching.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is clean and modern, with a simple color palette.
Contour ProtégéAl is substantially equivalent to a combination of the predicate devices MIM on Linux (K190379) and Mirada Workflow Box (K181572).
# Testing and Performance Data
The neural networks used in the Contour ProtégéAl device were trained on datasets from several large institutions. These datasets included CT images and MR images and their associated segmentations.
For testing, 286 images were used to evaluate the neural network models that segmented CT images, while 72 images were used to evaluate for the MR segmentation network. In all cases, the test images were gathered from a different and disjoint set of institutions from the training data. Both Contour ProtégéAI and the MIM predicate device were used to automatically segmented the independent test sets to show substantial equivalence.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.