Carto Octaray Mapping Catheter with Trueref Technology
Applicant
Biosense Webster, Inc.
Product Code
MTD · Cardiovascular
Decision Date
Jul 31, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
The CARTO® OCTARAY™ Mapping Catheter with TRUEref™ Technology is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The catheter provides location information only when used with compatible versions of the CARTO® 3 EP Navigation System.
Device Story
Multi-electrode diagnostic catheter; flower-shaped distal probe with 8 spines and 48 electrodes; includes central unipolar reference electrode. Used in cardiac electrophysiology labs by physicians. Input: electrical signals from heart endocardium. Output: electrograms and location data transmitted to CARTO 3 EP Navigation System. Principle: catheter contacts cardiac tissue to record/stimulate; navigation system tracks position. Output used by physicians to identify arrhythmia substrates and guide ablation procedures. Benefits: enables precise mapping of complex cardiac structures, facilitating treatment of arrhythmias.
Clinical Evidence
Prospective, single-arm, non-randomized, multi-center study (N=31) across three arrhythmia subgroups (VT, AT/PAF, PsAF). Primary endpoints: successful pre-ablation mapping without secondary devices (effectiveness) and no serious adverse events within 7 days (safety). 82.1% of per-protocol subjects achieved pre-ablation mapping success; 100% of clinically indicated mapping completed without secondary devices. No device-related serious adverse events reported.
Technological Characteristics
Flower-shaped distal tip with 8 spines, 48 ring electrodes, and 1 central unipolar reference electrode. Uni-directional piston-like deflection mechanism. Compatible with CARTO 3 EP Navigation System for location tracking. Materials and construction similar to predicate. Sterilization, packaging, and shelf-life validated.
Indications for Use
Indicated for multiple electrode electrophysiological mapping (recording or stimulation) of cardiac structures in atrial and ventricular regions. Patient population includes adults requiring mapping for ischemic ventricular tachycardia, scar-related atrial tachycardia, or persistent atrial fibrillation.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it.
July 31, 2020
Biosense Webster, Inc. Richard Lauhead Sr. Program Lead, RA 33 Technology Dr Irvine. California 92618
Re: K193237
Trade/Device Name: Carto Octaray Mapping Catheter with TRUEref Technology Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: MTD Dated: June 29, 2020 Received: July 1, 2020
Dear Richard Lauhead:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Mark Fellman Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use 1
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. | |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|--|
| 510(k) Number (if known) | K193237 | |
| Device Name | CARTO® OCTARAY™ MAPPING CATHETER WITH TRUEREF™ TECHNOLOGY | |
#### Indications for Use (Describe)
The CARTO® OCTARAY™ Mapping Catheter with TRUErcf™ Technology is indicated for multiple electrode clectrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The catheter provides location information only when used with compatible versions of the CARTO® 3 EP Navigation System.
| Type of Use (Select one or both, as applicable) | <span> [X] Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
#### 1. 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
| Date Summary<br>Prepared | 25 June 2020 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------|
| Applicant | Biosense Webster, Inc.<br>33 Technology Drive<br>Irvine, CA 92618<br>Establishment Registration Number: 9044811 |
| Official<br>Correspondent | Richard Lauhead<br>Senior Program Lead, Regulatory Affairs<br>Telephone: (949) 789-8583<br>Fax: (949) 450-6886 |
| Trade Name | CARTO® OCTARAY™ MAPPING CATHETER WITH<br>TRUEREF™ TECHNOLOGY |
| Common Name | Electrophysiology Catheter |
| Classification Name | Electrode recording Catheter |
| Device<br>Classification | Class II, 21 CFR 870.1220<br>Product Code: MTD |
| Model Numbers | D-1609-01-S, D-1609-02-S, D-1609-03-S, D-1609-04-S,<br>D-1609-05-S, D-1609-06-S |
| Predicate device | PENTARAY® Nav eco Mapping Catheter (K123837) |
## Substantially Equivalent To:
The Biosense Webster Inc. CARTO® OCTARA Y™ Mapping Catheter with TRUEref™ Technology is substantially equivalent to the Biosense Webster Inc. PENTARAY® Nav eco Mapping Catheter [510(k) K123837 cleared March 05, 2013]. Like the predicate device, the CARTO® OCTARAY™ Mapping Catheter with TRUEref™ Technology features a piston like deflection mechanism, allowing for uni-directional deflection of the shaft. The proposed catheter also has similar deflectable curve offerings and a similar flower shaped distal tip. The original intended use of the predicate device as a diagnostic catheter designed to facilitate mapping of structures within the heart remains the same in the proposed device.
## Description of the Device Subject to Premarket Notification:
The Biosense Webster CARTO® OCTARA Y™ Mapping Catheter with TRUEref™ Technology is a multi-electrode diagnostic catheter designed to facilitate
{4}------------------------------------------------
electrophysiological mapping of all structures in the heart (recording and stimulation). The catheter's distal end is a flower-shaped probe with 8 spines that radiate from the center. Each spine has 6 ring electrodes that are used for stimulation and recording within the heart. The catheter also includes an electrode located at the confluence of the spines that can be used as a close internal unipolar reference. The flower is available in a 30mm or a 40 mm diameter and several ring spacing configurations to achieve optimal mapping and contact with various cardiac structures.
## Indications for Use:
The CARTO® OCTARAY™ Mapping Catheter with TRUEref™ Technology is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The catheter provides location information only when used with compatible versions of the CARTO® 3 EP Navigation System.
## Technical Characteristics:
The CARTO® OCTARAY™ Mapping Catheter with TRUEref™ design uses similar technology, has similar intended use, functions, materials and method of operation as the predicate PENTARAY® Nav eco Mapping Catheter. The CARTO® OCTARAY™ Mapping Catheter with TRUEref™ Technology has a flower shaped tip containing 48 electrodes arranged across 8 spines and includes a close uni-polar reference electrode at the confluence of the spines. Otherwise, there are no special technical aspects of the ability of this catheter to detect electrical signals from heart endocardium and transmit this information to the CARTO® 3 EP Navigation system and/or recording equipment for display, analysis, and interpretation in detection of various heart arrhythmias. The differences in the number of spines and electrodes between the CARTO® OCTARA Y™ Mapping Catheter with TRUEref™ Technology and the PENTARAY® Nav eco Mapping Catheter have been evaluated through bench, animal and biocompatibility testing. Bench, clinical and animal testing demonstrated that the subject device is substantially equivalent to the predicate device. Results from the bench, animal, biocompatibility, and clinical data did not result in new questions with regards to safety and effectiveness of the device.
#### Performance Data:
The CARTO® OCTARAY™ Mapping Catheter with TRUEref™ Technology underwent bench testing, animal testing, and Clinical testing to demonstrate substantial equivalence.
Testing included Mechanical integrity, electrical leakage, deflection, device functionality, simulated use, biocompatibility, electrical properties, visualization and navigation, sterilization, packaging, shelf life, Animal testing to assess device safety, device maneuverability and signal quality, and clinical testing to assess device effectiveness and safety.
The catheter passed all predetermined acceptance criteria.
{5}------------------------------------------------
The Clinical study was a prospective, single arm, non-randomized, multi-center study. The study intended to enroll 30 subjects with a target of equal numbers of subjects (n=10) among three study arrhythmia subgroups: ischemic ventricular tachycardia, scar related AT (including PAF re-do/repeat procedures), and PsAF. The Study was performed outside of the United States.
The Primary objectives include meeting an effectiveness endpoint of completing preablation mapping and clinically indicated mapping without the need for another device and a safety endpoint of having no serious adverse events within 7 days of the study procedure.
The study population included subjects scheduled to have a clinically-indicated catheter mapping and ablation procedure for management of an arrhythmia in the following 3 subgroups: Ischemic ventricular tachycardia (VT) n=31, Scar-related atrial tachycardia (AT; includes atypical atrial flutter) resulting from previous paroxysmal atrial fibrillation (PAF) ablation procedures, mitral valve repair procedures, or PAF "re-do" procedures n=30, and Persistent Atrial Fibrillation (PsAF) n=30.
Table A summarizes the Subject enrollment.
Table B summarizes subject demographic information
| | All<br>Enrolled<br>(N=31)<br>n (%) | VT<br>(n=11)<br>n (%) | AT/PAF<br>(n=10)<br>n (%) | PSAF<br>(n=10)<br>n (%) |
|---------------------------------------------------------|------------------------------------|-----------------------|---------------------------|-------------------------|
| Disposition | | | | |
| Enrolled Subjects | 31<br>(100.0%) | 11<br>(100.0%) | 10<br>(100.0%) | 10<br>(100.0%) |
| No study catheter<br>inserted | 1<br>(3.2%)* | 0 (0.0%) | 0 (0.0%) | 1<br>(10.0%)* |
| Safety Population | 30<br>(96.8%) | 11<br>(100.0%) | 10<br>(100.0%) | 9<br>(90.0%) |
| Catheter inserted<br>but no mapping | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| Mapping<br>performed but not<br>for study<br>arrhythmia | 2<br>(6.5%) † | 1<br>(9.1%) † | 0 (0.0%) | 1<br>(10.0%) † |
| Per Protocol<br>Population | 28<br>(90.3%) | 10<br>(90.9%) | 10<br>(100.0%) | 8<br>(80.0%) |
#### Table A - Subject Enrollment and Accountability (Enrolled Subjects, N=31)
* Subject 632-003 did not meet eligibility criteria and was withdrawn prior to a study procedure
† Subjects 847-001 (PsAF) & 847-003 (VT) had mapping performed despite not having study arrhythmia
{6}------------------------------------------------
| Disposition | All<br>Arrhythmias<br>(n=31) | OCTARAY<br>FIM<br>AT/PAF<br>(n=10) | OCTARAY<br>FIM<br>PsAF<br>(n=10) | OCTARAY<br>FIM<br>VT<br>(n=11) |
|---------------------------|------------------------------|------------------------------------|----------------------------------|--------------------------------|
| Age at Consent | | | | |
| Mean ± SD (n) | 67.8 ± 8.52<br>(31) | 64.5 ± 9.23<br>(10) | 68.1 ± 7.75<br>(10) | 70.5 ± 8.25<br>(11) |
| Median (Q1/Q3) | 66.0<br>(63.0,74.0) | 65.5<br>(57.0,73.0) | 68.0<br>(63.0,72.0) | 69.0<br>(64.0,79.0) |
| Min / Max | 51.0 / 83.0 | 51.0 / 76.0 | 54.0 / 79.0 | 57.0 / 83.0 |
| Sex, n/N (%) | | | | |
| Male | 27 (87.1%) | 9 (90.0%) | 7 (70.0%) | 11<br>(100.0%) |
| Female* | 4 (12.9%) | 1 (10.0%) | 3 (30.0%) | 0 (0.0%) |
| Race, n/N (%) | | | | |
| Black/African<br>American | 1 (3.2%) | 0 (0.0%) | 1 (10.0%) | 0 (0.0%) |
| White | 25 (80.6%) | 8 (80.0%) | 9 (90.0%) | 8 (72.7%) |
| Race not reported | 5(16.1%) | 2(20.0%) | 0(0.0%) | 3(27.3%) |
| Ethnicity, n/N (%) | | | | |
| Not Hispanic or<br>Latino | 9 (90.0%) | 9 (90.0%) | 10 (100.0%) | 7 (63.6%) |
| Not Reported | 5(16.1%) | 1(10.0%) | 0(0.0%) | 4(36.4%) |
Table B - Study Demographic information (Enrolled Subjects, N=31)
# Study Results:
The study met Primary endpoints for both safety and effectiveness.
No Serious Adverse events were reported resulting from the use of the study catheter and all physicians were able to perform pre-ablation mapping with the subject device.
In the Per-Protocol Population, 82.1% (23/28) had pre-ablation mapping requirements successfully performed without the need of another catheter.
- . Three (3) VT subjects did not have LAT mapping performed for reasons of patient safety. It is the position of the Sponsor that these were acceptable minor protocol deviations for patient safety and that the associated endpoint failures do not reflect a
{7}------------------------------------------------
deficiency of the catheter's ability to map but rather a unique unforeseen scenario. Pre-ablation voltage mapping was successfully performed for these subjects.
- . One (1) VT subject did not have LAT mapping completed due to the operator's inability to maintain stable contact with the area of interest; the Sponsor has made note of this operator's feedback.
- . One (1) PsAF subject that had CARTOFINDER mapping performed out of sequence was due to human error and was not affected by the catheter design or protocol prescribed workflow.
Clinically indicated mapping was started and finished in 100.0% (15/15) without the need for another device.
The study had 17 adverse events occur across 12 subjects and all were classified as unrelated to the study device. The Adverse Events reported include: Heart Failure, thoracic pain, perifissural nodule, Flu, constipation (two separate subjects), bladder infection, fever, hematuria, AV block 2:1, Dizziness, shoulder bruises, sore throat, non cardiac chest aches, shortness of breath, headaches, and allergic reaction.
# Basis for Determination of Substantial Equivalence:
The CARTO® OCTARAY™ Mapping Catheter with TRUEref™ Technology is substantially equivalent to its currently cleared predicate. PENTARA Y Nav eco Mapping Catheter, based on the successful completion of nonclinical bench testing, pre-clinical animal studies, a clinical study, as well as the technological comparison exhibiting similar principles of design, operation, and indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.