K193220 · May 5, 2020
AVIEW LCS
Coreline Soft Co., Ltd.
National Lung Screening Trial (NLST) cohort; Korean clinical chest CT datasets (COPD patients); PanCan clinical data; BCCA clinical data
Retrospective clinical data from the NLST and Korean clinical cohorts were used to validate the accuracy of the device's deep-learning-based lung/lobe segmentation and to support the clinical validity of the PANCAN malignancy risk calculation model.
Retrospective cohort; Algorithm validation; Deep learning; Clinical performance
AVIEW LCS is intended for the review and analysis and reporting of thoracic CT images for the purpose of characterizing nodules in the lung in a single study, or over the time course of several thoracic studies. Characterizations include nodule type, location of the nodule and measurements such as size (major axis), estimated effective diameter from the volume of the nodule, the volume of the nodule, Mean HU (the average value of the CT pixel inside the nodule in HU), Minimum HU, Max HU, mass (mass calculated from the CT pixel value), and volumetric measures (Solid Major, length of the longest diameter measured in 3D for a solid portion of the nodule. Solid 2nd Major: The length of the longest diameter of the solid part. measured in sections perpendicular to the solid portion of the nodule), VDT (Volume doubling time), and Lung-RADS (classification proposed to aid with findings). The system automatically performs the measurement, allowing lung nodules and measurements to be displayed and, also integrate with FDA certified Mevis CAD (Computer-aided detection) (K043617).
Device Story
AVIEW LCS is a software-based diagnostic tool for thoracic CT image analysis. It processes DICOM-formatted thoracic CT scans to perform semi-automatic nodule segmentation and measurement (size, volume, density, VDT). The device integrates with third-party CAD (Mevis Visia) for automated nodule detection. It calculates malignancy risk using the PANCAN model and categorizes findings according to Lung-RADS. The system supports longitudinal follow-up by automatically retrieving past data and matching nodules across studies. Used in clinical settings by radiologists/physicians, the software provides structured PDF reports and DICOM-compatible images. By standardizing measurements and risk assessment, it assists clinicians in diagnostic decision-making and lung cancer screening workflows.
Clinical Evidence
Bench testing only. No clinical trials were conducted. Performance was validated through software verification, unit testing, and system testing. Accuracy of nodule segmentation was verified against sphere phantoms (volume error <10%). Lung/lobe segmentation (deep learning) was validated using 192 COPD patient scans (expert qualitative assessment) and 80 NLST patient scans (Dice similarity coefficient). PANCAN risk model and Lung-RADS implementation were verified against established datasets and regulatory guidelines.
Technological Characteristics
Software-only device for PC/network deployment. Inputs: DICOM thoracic CT images. Features: Deep-learning based lung/lobe segmentation, semi-automatic nodule segmentation, PANCAN malignancy risk calculation, Lung-RADS categorization. Connectivity: DICOM 3.0 standard. Output: Structured PDF reports and DICOM files. No hardware components.
Indications for Use
Indicated for review, analysis, and reporting of thoracic CT images to characterize lung nodules in single or longitudinal studies. Intended for use by clinicians for nodule measurement, malignancy risk assessment, and Lung-RADS categorization.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Lung Nodule Assessment and Comparison Option (LNA) (K162484)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 5, 2020
Coreline Soft Co., Ltd. % Hye Yi Park Deputy General Manager Strategic Business Dept. 4, 5F (Yeonnam-dong), 49, World Cup buk-ro 6-gil, Mapo-gu Seoul. 03991 REPUBLIC OF KOREA
Re: K193220
Trade/Device Name: AVIEW LCS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ, JAK Dated: April 2, 2020 Received: April 6, 2020
Dear Hye Yi Park:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
### 510(k) Number (if known)
K193220
Device Name AVIEW LCS
### Indications for Use (Describe)
AVIEW LCS is intended for the review and analysis and reporting of thoracic CT images for the purpose of characterizing nodules in the lung in a single study, or over the time course of several thoracic studies. Characterizations include nodule type, location of the nodule and measurements such as size (major axis), estimated effective diameter from the volume of the nodule, the volume of the nodule, Mean HU (the average value of the CT pixel inside the nodule in HU), Minimum HU, Max HU, mass (mass calculated from the CT pixel value), and volumetric measures (Solid Major, length of the longest diameter measured in 3D for a solid portion of the nodule. Solid 2nd Major: The length of the longest diameter of the solid part. measured in sections perpendicular to the solid portion of the nodule), VDT (Volume doubling time), and Lung-RADS (classification proposed to aid with findings). The system automatically performs the measurement, allowing lung nodules and measurements to be displayed and, also integrate with FDA certified Mevis CAD (Computer-aided detection) (K043617).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K193220
# 510(k) Summary
#### SUBMITTER 1
Coreline Soft Co., Ltd. 4,5F (Yeonnam-dong), 49 World Cup buk-ro 6-gil, Mapo-gu, Seoul, 03991, Republic of Korea.
Phone: 82.2.517.7321 Fax: 82.2.571.7324
Contact Person: hyeyi. Park Date Prepared: 11.15.2019
#### DEVICE 2
Name of Device: AVIEW LCS Common or Usual Name: Image Processing Software Classification Name: System, image processing, radiological (21CFR 892.2050) Regulatory Class: II Product Code: LLZ, JAK
#### PREDICATE DEVICE 3
Lung Nodule Assessment and Comparison Option by Philips Medical System Nederland B.V. (K 162484)
Name of Device: Lung Nodule Assessment and Comparison Option (LNA) Common or Usual Name: Lung Nodule Assessment and Comparison Option (LNA) Classification Name: System, image processing, radiological (21CFR 892.1750) Regulatory Class: II Product Code: LLZ, JAK
This predicate has not been subject to a design-related recall
#### REFERENCE DEVICE 4
Lung Analysis Software by Vital Images, Inc. (K151283)
Name of Device: Lung Analysis Software Common or Usual Nmae: Radilolgical Image Processing Software Classification Name: System, Image Processing, Radiological Regulatory Class: II Product Code: JAK, LLZ
{4}------------------------------------------------
## AVIEW by Coreline Soft Co., Ltd. (K171199)
Name of Device: AVIEW Common or Usual Name: Image Processing Software Classification Name: System, image processing, radiological (21CFR 892.2050) Regulatory Class: II Product Code: LLZ
This reference device has not been subject to a design-related recall
#### DEVICEW DESCRIPTION 5
AVIEW LCS is intended for use as diagnostic patient imaging which is intended for the review and analysis of thoracic CT images. Provides following features as semi-automatic nodule measurement (segmentation), maximal plane measure, 3D measure and volumetric measures, automatic nodules detection by integration with 3th party CAD. Also provides cancer risk based on PANCAN risk model which calculates the malignancy score based on numerical or Boolean inputs. Follow up support with automated nodule matching and automatically categorize Lung-RADS score which is a quality assurance tool designed to standardize lung cancer screening CT reporting and management recommendations that is based on type, size, size change and other findings that is reported.
- -Nodule measurement
- Adding nodule by segmentation or by lines .
- Semi-automatic nodule measurement (segmentation) "
- . Maximal plane measure, 3D measure and volumetric measure.
- . Automatic large vessel removal.
- י Provides various features calculated per each nodule such as size, major(longest diameter measured in 2D/3D), minor (shortest diameter measured in 2D/3D), maximal plane, volume, mean HU, minimum HU, maximum HU for solid nodules and ratio of the longest axis for solid to non solid for paritla solid nodules.
- . Fully supporting Lung-RADS workflow: US Lung-RADS and KR Lung-RADS.
- . Nodule malignancy score (PANCAN model) calculation.
- . Importing from CAD results
- -Follow-up
- ' Automatic retrieving the past data
- י Follow-up support with nodule matching and comparison
- Automatic calculation of VDT (volume doubling time)
- Automatic nodule detection (CADe) -
- Seamless integration with Mevis Visia (FDA 510k Cleared) .
- -Lungs and lobes segmentation
- Better segmentation of lungs and lobes based on deep-learning algorithms.
- -Report
- PDF report generation .
{5}------------------------------------------------
- . It saves or sends the pdf report and captured images in DICOM files.
- . It provides structured report including following items such as nodule location and, also input finding on nodules.
- Reports are generated using the results of all nodules nodules detected so far (Lung RADS) .
- -Save Result
- . It saves the results in internal format
#### INDICATIONS FOR USE 6
AVIEW LCS is intended for the review and analysis and reporting of thoracic CT images for the purpose of characterizing nodules in the lung in a single study, or over the time course of several thoracic studies. Characterizations include nodule type, location of the nodule and measurements such as size (major axis), estimated effective diameter from the volume of the volume of the nodule, Mean HU (the average value of the CT pixel inside the nodule in HU), Minimum HU, Max (mass (mass calculated from the CT pixel value), and volumetric measures (Solid Major; length of the longest diameter measured in 3D for a solid portion of the nodule. Solid 2™ Major: The length of the longest diameter of the solid part, measured in sections perpendicular to the solid portion of the nodule), VDT (Volume doubling time), and Lung-RADS (classification proposed to aid with findings). The system automatically performs the measurement, allowing lung nodules and measurements to be displayed and, also integrate with FDA certified Mevis CAD (Computer-aided detection) (K043617).
## COMPARISION OF TECHNOLOGICAL CHARACTERISTICS WITH 7 THE PREDICATE DEVCIE
AVIEW LCS has the same intended use and the principle of operation, and also has similar features to the predicate devices. Lung Nodule Assessment and Comparison Option (LNA) (K162484)
There might be slight differences in features and menu, but these differences between the predicate device and the proposed device are not so significant since they do not raise any new or potential safety risks to the user or patient and questions of safety or effectiveness. Based on the results of software validation and verification tests, we conclude that the proposed device is substantially equivalent to the predicate devices.
| Characteristic | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|---------------------------------------------|---------------------------------------------|
| Device Name | AVIEW LCS | Lung Nodule<br>Assessment and<br>Comparison Option<br>(LNA) | Lung Analysis<br>Software | AVIEW |
| Classification<br>Name | System, image<br>Processing<br>Radiological | System, image<br>Processing<br>Radiological | System, image<br>Processing<br>Radiological | System, image<br>Processing<br>Radiological |
| Regulatory<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | 21 CFR 892.2050 | 21 CFR 892.2050 |
| Product Code | LLZ, JAK | LLZ, JAK | LLZ, JAK | LLZ |
| Review Panel | Radiology | Radiology | Radiology | Radiology |
| 510k Number | - | K162484 | K151283 | K171199 |
| Indications<br>for use | AVIEW LCS<br>AVIEW LCS is intended for the review and analysis and reporting of thoracic CT images for the<br>purpose of characterizing nodules in the lung in a single study, or over the time course of several | | | |
{6}------------------------------------------------
| thoracic studies. Characterizations include nodule type, location of the nodule and measurements<br>such as size (major axis, minor axis), estimated effective diameter from the volume of the nodule,<br>the volume of the nodule, Mean HU (the average value of the CT pixel inside the nodule in HU),<br>Minimum HU, Max HU, mass (mass calculated from the CT pixel value), and volumetric measures<br>(Solid Major; length of the longest diameter measured in 3D for a solid portion of the nodule. Solid<br>2nd Major: The length of the longest diameter of the solid part, measured in sections perpendicular<br>to the Major axis of the solid portion of the nodule), VDT (Volume doubling time), and Lung-RADS<br>(classification proposed to aid with findings). The system automatically performs the measurement,<br>allowing lung nodules and measurements to be displayed and, also integrate with FDA certified<br>Mevis CAD (Computer-aided detection) (K043617).<br>Lung Nodule Assessment and Comparison Option (LNA)<br>The Lung Nodule Assessment and Comparison Option is intended for use as a diagnostic patient-<br>imaging tool. It is intended for the review and analysis of thoracic CT images, providing quantitative<br>and characterizing information about nodules in the lung in a single study, or over the time course<br>of several thoracic studies. Characterizations include diameter, volume and volume over time. The<br>system automatically performs the measurements, allowing lung nodules and measurements to be<br>displayed.<br>Toshiba's Lung Analysis Software<br>The separately licensed Lung Analysis option is intended for the review and analysis of thoracic CT<br>images for the purpose of characterizing nodules in the lung in a single study, or over the time<br>course of several thoracic studies. Characterization include diameter, volume and volume over time.<br>The system automatically performs the measurement, allowing lung nodules and measurement to<br>be displayed.<br>AVIEW | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AVIEW provides CT values for pulmonary tissue from CT thoracic datasets. This software can be<br>used to support the physician quantitatively in the diagnosis. Follow-up evaluation and<br>documentation of CT lung tissue images by providing image segmentation of sub-structures in the<br>left and right lung (e.g., the five lobes and airway), volumetric and structural analysis, density<br>evaluations and reporting tools. AVIEW is also used to store, transfer, inquire and display CT data<br>sets. AVIEW is not meant for primary image Interpretation in mammography. | | | | |
| Platform | IBM-compatible PC or<br>PC network | IBM-compatible PC or<br>PC network | IBM-compatible PC or PC network | |
| User Interface | Monitor, Mouse,<br>Keyboard | Monitor, Mouse,<br>Keyboard | Monitor, Mouse,<br>Keyboard | |
| Image Input<br>Sources | Images can be scanned,<br>loaded from card<br>readers, or imported<br>from a radiographic<br>imaging device | Images can be scanned,<br>loaded from card<br>readers, or imported<br>from a radiographic<br>imaging device | Images can be scanned, loaded from<br>card readers, or imported<br>from a radiographic<br>imaging device | |
| Image format | DICOM | DICOM | DICOM | |
| Intended<br>body part | Chest | Chest | Chest | |
| Type of scans | Thoracic CT images | Thoracic CT images | Thoracic CT images | |
| General<br>Description | AVIEW LCS is intended for use as diagnostic patient imaging which is intended for the review and<br>analysis of thoracic CT images. Provides following features as semi-automatic nodule measurement<br>(segmentation), maximal plane measure, 3D measure and volumetric measures, automatic nodules<br>detection by integration with 3rd party CAD. Also provides cancer risk based on PANCAN risk<br>model which calculates the malignancy score based on numerical or Boolean inputs. Follow up<br>support with automated nodule matching and automatically categorize Lung-RADS score which is<br>a quality assurance tool designed to standardize lung cancer screening CT reporting and | | | |
| Key<br>Functions | | | | |
| | management recommendations that is based on type, size, size change and other findings that is<br>reported. | | | |
| | Lung Nodule Assessment and Comparison Option (LNA) | | | |
| | The Lung Nodule Assessment and Comparison Option application is intended for use as a<br>diagnostic patient imaging tool. It is intended for the review and analysis of thoracic CT images,<br>providing quantitative and characterizing information about nodules in the lung in a single study,<br>or over the time course of several thoracic studies. The system automatically performs the<br>measurements, allowing lung nodules and measurements to be displayed. The user interface and<br>automated tools help to determine growth patterns and compose comparative reviews. The Lung | | | |
| | Nodule Assessment and Comparison Option application requires the user to identify a nodule and<br>to determine the type of nodule in order to use the appropriate characterization tool. Lung Nodule<br>Assessment and Comparison Option may be utilized in both diagnostic and screening evaluations<br>supporting Low Dose CT Lung Cancer Screening | | | |
| | Lung Analysis Software | | | |
| | Lung Analysis aids in measuring and characterizing lung nodules. The interface and automated tools<br>help to efficiently determine growth patterns and compose comparative reviews. Lung Analysis is<br>intended for the review and analysis of thoracic CT images for the purpose of characterizing nodules<br>in the lung in a single study, or over the time course of several thoracic studies. Characterizations<br>include diameter, volume and volume over time. The system automatically performs the | | | |
| | measurements, allowing lung nodules and measurements to be displayed. The Lung Analysis<br>Software requires the user to identify a nodule and to determine whether it is a GGO or solid nodule<br>in order to use the appropriate characterization too. | | | |
| | AVIEW | | | |
| | The AVIEW is a software product which can be installed on a PC. It sh…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.