← Product Code [FDT](/productcode/FDT) · K193202

# EXALT Model D Single-Use Duodenoscope, EXALT Controller (K193202)

_Boston Scientific Corporation · FDT · Dec 13, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K193202

## Device Facts

- **Applicant:** Boston Scientific Corporation
- **Product Code:** [FDT](/productcode/FDT.md)
- **Decision Date:** Dec 13, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

EXALT Model D Single-Use Duodenoscope: The EXALT Model D Single-Use Duodenoscope is intended for use with a Boston Scientific endoscopic video imaging system, for endoscopy and endoscopic surgery within the duodenum. EXALT Controller: The EXALT Controller is intended for use with a Boston Scientific endoscopic diagnosis, treatment, and video observation.

## Device Story

EXALT Model D is sterile, single-use, flexible duodenoscope; EXALT Controller is endoscopic video imaging system. Duodenoscope captures video signals, provides illumination via integrated LEDs, enables 4-way steerable navigation, lens wash, insufflation, suction, and delivery of ancillary devices. Controller processes video signals from duodenoscope, outputs to monitor, and interfaces with external image capture systems. Used in clinical settings for ERCP procedures; operated by physicians. Single-use design eliminates reprocessing requirements, reducing cross-contamination risk compared to reusable endoscopes.

## Clinical Evidence

Clinical testing performed to confirm procedural performance of the EXALT Model D and EXALT Controller in endoscopic cholangiopancreatography (ERCP) procedures. Results demonstrate the system is safe and effective for its intended use.

## Technological Characteristics

Single-use, flexible duodenoscope; LED illumination integrated into scope tip; 4-way steerable navigation; video imaging system controller. Biocompatibility per ISO 10993-1. Electrical safety/EMC per AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-2, and IEC 60601-2-18.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- EVIS EXERA II Duodenovideoscope TJF Type Q180 ([K143153](/device/K143153.md))
- CV-170 Video System Center ([K122831](/device/K122831.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 13, 2019

Boston Scientific Corporation Carter Navarro Sr. Regulatory Affairs Manager 100 Boston Scientific Way Marlborough, MA 01752

Re: K193202

> Trade/Device Name: EXALT Model D Single-Use Duodenoscope EXALT Controller Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDT, FET Dated: November 19, 2019 Received: November 20, 2019

Dear Carter Navarro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K193202

Device Name

EXALT Model D Single Use Duodenoscope; EXALT Controller

Indications for Use (Describe)

EXALT Model D Single Use Duodenoscope:

The EXALT Model D Single-Use Duodenoscope is intended for use with a Boston Scientific endoscopic video imaging system, for endoscopy and endoscopic surgery within the duodenum.

EXALT Controller:

The EXALT Controller is intended for use with a Boston Scientific endoscopic diagnosis, treatment, and video observation.

**Type of Use (Select *one* or *both*, as applicable)**

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "K193202" followed by "Page 1 of 4". The text is in a simple, sans-serif font and is likely part of a document or report. The page number indicates that this is the first page of a four-page document. The alphanumeric code "K193202" could be a document identifier or reference number.

Boston Scientific Corporation 300 Boston Scientific Way Marlborough, MA 01752 (508) 382-4000 www.bostonscientific.com

### 510(k) Summary for EXALT Endoscopic Visualization System (EXALT Model D Single-Use Duodenoscope and EXALT Controller)

# 1. Submitter

Boston Scientific Corporation Endoscopy Division 100 Boston Scientific Way Marlborough, MA 01752

- Contact: Carter Navarro Fellow, Regulatory Affairs Phone: (508) 683-4793 E-mail: carter.navarro@bsci.com
Date Prepared: November 19, 2019

# 2. Device

| Trade Name:                | EXALT Model D Single-Use Duodenoscope                                  |
|----------------------------|------------------------------------------------------------------------|
| Common Name:               | Duodenoscope and accessories, flexible/rigid                           |
| Product Code:              | FDT                                                                    |
| Device Class:              | Class II                                                               |
| Panel:                     | Gastroenterology/Urology                                               |
| Classification Regulation: | 21 CFR 876.1500, Endoscope and accessories                             |
| Trade Name:                | EXALT Controller                                                       |
| Common Name:               | Endoscopic video imaging system/component,<br>gastroenterology-urology |
| Product Code:              | FET                                                                    |
| Device Class:              | Class II                                                               |
| Panel:                     | Gastroenterology/Urology                                               |
| Classification Regulation: | 21 CFR 876.1500, Endoscope and accessories                             |

Image /page/3/Picture/9 description: The image shows the logo for Boston Scientific. The logo is in blue and features the words "Boston Scientific" stacked on top of each other. The font is a serif font, and the overall design is simple and professional. The logo is likely used to represent the company in various marketing and branding materials.

{4}------------------------------------------------

# 3. Predicate Devices

| Trade Name:                | EVIS EXERA II Duodenovideoscope TJF Type Q180                          |
|----------------------------|------------------------------------------------------------------------|
| Manufacturer:              | Olympus Medical Systems Corp.                                          |
| Clearance Number:          | K143153                                                                |
| Common Name:               | Duodenoscope and accessories, flexible/rigid                           |
| Product Code:              | FDT                                                                    |
| Device Class and Panel:    | Class II, Gastroenterology/Urology                                     |
| Classification Regulation: | 21 CFR 876.1500, Endoscope and accessories                             |
| Trade Name:                | CV-170 Video System Center                                             |
| Manufacturer:              | Olympus Medical Systems Corp.                                          |
| Clearance Number:          | K122831                                                                |
| Common Name:               | Endoscopic video imaging system/component,<br>gastroenterology-urology |
| Product Code:              | FET                                                                    |
| Device Class and Panel:    | Class II, Gastroenterology/Urology                                     |
| Classification Regulation: | 21 CFR 876.1500, Endoscope and accessories                             |

# 4. Device Description

The EXALT Model D Single-Use Duodenoscope is a sterile, single-use, flexible duodenoscope used to examine the duodenum and perform various procedures within the duodenum including endoscopy and endoscopic surgery. When connected to an EXALT Controller, the EXALT Model D Single-Use Duodenoscope provides imaging and illumination, 4-way steerable navigation, lens wash, insufflation, suction, facilitates image capture initiation, and allows the delivery of various ancillary devices.

The EXALT Controller is an endoscopic video imaging system that receives video signals from the EXALT Model D Single-Use Duodenoscope, processes the video signals, outputs video images to a video monitor, and outputs electrical signals that interface with external image capture systems. The EXALT Controller also controls the light transmitted by the tip of the EXALT Model D Single-Use Duodenoscope to illuminate the area of interest within the anatomy.

### 5. Indications for Use

The EXALT Model D Single-Use Duodenoscope is intended for use with a Boston Scientific endoscopic video imaging system, for endoscopy and endoscopic surgery within the duodenum.

The EXALT Controller is intended for use with a Boston Scientific endoscope for endoscopic diagnosis, treatment, and video observation.

{5}------------------------------------------------

### 6. Technological Characteristics

A direct comparison of key characteristics demonstrates that the EXALT Model D Single-Use Duodenoscope and EXALT Controller are substantially equivalent to their respective predicate devices in terms of intended use, technological characteristics, and performance characteristics.

The EXALT Model D Single-Use Duodenoscope and the EVIS EXERA II Duodenovideoscope TJF Type Q180V share similar mechanical and optical characteristics, including working length, diameters, articulation angles, flow rates, resolution, direction of view, and field of view. The EXALT Model D Single-Use Duodenoscope is sterile, single-use, and not intended for reprocessing, whereas the EVIS EXERA II Duodenovideoscope TJF Type Q180V is intended to be reprocessed.

The EXALT Controller and the Olympus CV-170 Video System Center share similar video formats and outputs and illumination technology, though the light-emitting diodes in the EXALT Endoscopic Visualization System are located in the EXALT Model D Single-Use Duodenoscope rather than the EXALT Controller; whereas the light sources in Olympus endoscopic systems (including the CV-170 Video System Center) are located in the capital equipment.

### 7. Performance Data

Non-clinical testing was successfully performed on the proposed EXALT Model D Single-Use Duodenoscope and EXALT Controller

Performance testing (bench) was successfully completed to establish substantial equivalence between the proposed EXALT Model D Single-Use Duodenoscope and EXALT Controller and the predicate devices. This testing included the following:

- Working Length
- Insertion Portion Diameter ●
- Articulation Angles ●
- Flow Rates
- Reliability ●
- Resolution ●
- Direction of View
- Field of View
- Light Output .

{6}------------------------------------------------

Biocompatibility of the EXALT Model D Single-Use Duodenoscope was evaluated in accordance with ISO 10993-1. Electrical safety and electromagnetic compatibility of the EXALT Model D Single-Use Duodenoscope and EXALT Controller were evaluated in accordance with AAMI / ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-2, and IEC 60601-2-18.

Clinical testing was performed to confirm procedural performance of the EXALT Model D Single-Use Duodenoscope and EXALT Controller in endoscopic cholangiopancreatography (ERCP) procedures.

# 8. Conclusion

The results of non-clinical and clinical testing demonstrate that the EXALT Model D Single-Use Duodenoscope and EXALT Controller are considered safe and effective for their intended uses. Boston Scientific has demonstrated that the proposed devices are substantially equivalent to the currently marketed predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K193202](https://fda-staging.innolitics.com/device/K193202)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
