Evis Exera III Duodenovideoscope Olympus TJF-Q190V

K193182 · Olympus Medical Systems Corp. · FDT · Jan 17, 2020 · Gastroenterology, Urology

Device Facts

Record IDK193182
Device NameEvis Exera III Duodenovideoscope Olympus TJF-Q190V
ApplicantOlympus Medical Systems Corp.
Product CodeFDT · Gastroenterology, Urology
Decision DateJan 17, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.

Device Story

The TJF-Q190V is a flexible duodenovideoscope used for endoscopic procedures in the duodenum. It features an insertion section with an internal instrument channel and a distal forceps elevator to guide EndoTherapy accessories. The device captures visual data via a distal CCD chip, which converts incident light into electrical signals for processing by an external Olympus video system center and light source (supporting white light and Narrow Band Imaging). It is operated by physicians in clinical settings. The device incorporates a single-use distal cover (MAJ-2315) to protect the distal end and forceps elevator, and utilizes a specialized flushing adapter (MAJ-2319) to facilitate reprocessing. Output is displayed on a monitor for real-time visualization, enabling diagnostic assessment and therapeutic intervention. The device benefits patients by allowing minimally invasive access to the duodenum for surgical or diagnostic tasks.

Clinical Evidence

No clinical or animal studies were performed. Substantial equivalence is supported by bench testing, including sterilization/shelf-life validation (ASTM F1980-16), reprocessing validation, biocompatibility (ISO 10993-1), software verification/validation, electrical safety (AAMI ANSI ES60601-1), EMC (IEC 60601-1-2), and mechanical/thermal performance testing.

Technological Characteristics

Flexible duodenovideoscope with CCD imaging sensor. Features include a single-use distal cover (MAJ-2315) and flushing adapter (MAJ-2319). Complies with AAMI ANSI ES60601-1, IEC 60601-2-18, and IEC 60601-1-2. Biocompatibility per ISO 10993-1. Connectivity via proprietary Olympus video system center/light source. Sterilization/reprocessing validated per FDA guidance.

Indications for Use

Indicated for endoscopy and endoscopic surgery within the duodenum in patients requiring diagnostic or therapeutic procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 17, 2020 Olympus Medical Systems Corp. % Sheri Musgnung Regulatory Affairs Manager Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610 Re: K193182 Trade/Device Name: Evis Exera III Duodenovideoscope Olympus TJF-Q190V Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDT, NWB, OCX Dated: November 15, 2019 Received: November 18, 2019 Dear Sheri Musgnung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shanil P. Haugen, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K193182 Device Name EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V Indications for Use (Describe) This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum. Type of Use (Select one or both, as applicable) | <div> <span></span> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> | |-----------------------------------------------------------------------------------------| | <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly shadowed, giving them a three-dimensional appearance. Below the word, there is a gradient of yellow and orange, creating a subtle background effect. # 5.1 General Information - OLYMPUS MEDICAL SYSTEMS CORP. ■ Applicant: 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 Establishment Registration No: 8010047 - Official Correspondent: Sheri L. Musgnung Olympus Surgical Technologies America 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 Fax: 484-896-7128 Email: sheri.musgnung@olympus.com - . Manufacturing site: Aizu Olympus Co., Ltd., 500 Muranishi, Niidera, Monden-machi, Aizuwakamatsu-shi, Fukushima 965-8520, Japan # 5.2 Device Identification | ■ Device Name | - EVIS EXERA III DUODENOVIDEOSCOPE<br>OLYMPUS TJF-Q190V | |---------------------|-----------------------------------------------------------------------------------------------------| | ■ Model Name | TJF-Q190V | | ■ Common Name | Duodenoscope and accessories | | ■ Regulation Number | 876.1500 | | ■ Regulation Name | Endoscope and Accessories | | ■ Regulatory Class | II | | ■ Product Code | FDT; Duodenoscope, Accessories, Flexible/Rigid<br>NWB; Endoscope, accessories, narrow band spectrum | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. The word is slightly angled upwards from left to right. There is a registered trademark symbol to the right of the word. Traditional 510(k) Notification DUODENOVIDEOSCOPE TJF-Q190V OCX; Endoscopic Irrigation/Suction System ■ Classification Panel Gastroenterology/Urology # 5.3 PREDICATE DEVICE | Table 1 Predicate device on TJF-Q190V | | | |---------------------------------------|----------------------------------|------------| | Device name | 510(k) Submitter | 510(k) No. | | EVIS EXERA II<br>DUODENOVIDEOSCOPE | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K143153 | | OLYMPUS TJF TYPE Q180V | | | # 5.4 DEVICE DESCRIPTION # EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V The TJF-Q190V has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum. The TJF-Q190V is compatible with Olympus system "Video System Center OLYMPUS CV-190 and XENON LIGHT SOURCE OLYMPUS CLV-190 (K112680)". The subject device consists of flexible insertion section, control section and endoscope connector section with equipped CCD chip which delivers images. The light from the light source travels through the light guide to the light guide lens at the distal end. The light source can offer both the white light for the normal observation and the narrow band imaging (NBI). The CCD chip transduces the incident light from the objective lens to electrical signal. The video processor transduces electrical signal to video signal. There is an instrument channel entirely inside of the flexible insertion section. Endo Therapy accessories can be inserted through the instrument channel. A forceps elevator is located at the distal end of the insertion to elevate EndoTherapy accessories for endoscopic treatment. The TJF-Q190V consists of a single-use distal cover, MAJ-2315 which has been designed to be attached to OLYMPUS TJF-O190V to cover the distal end of the insertion tube and around the forceps elevator. MAJ-2315 is to be discarded after clinical use. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. A registered trademark symbol is located in the upper right corner of the word. The following new reprocessing accessory has also been designed for use with TJF-Q190V: # DISTAL END FLUSHING ADAPTER MAJ-2319 The MAJ-2319 was designed to flush the distal end of the endoscope with reprocessing fluids. The MAJ-2319 can be attached to the distal end of the endoscope during the manual cleaning and disinfection process to flush the distal end with reprocessing fluids. The reprocessing fluid is flushed through the MAJ-2319 to the distal end using a syringe. #### Indications for Use 5.5 This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum. #### 5.6 Comparison of Technological Characteristics Although there are some major differences including those described below, the technological characteristics of the TJF-Q190V are functionally equivalent to the predicate device, TJF-Q180V. 1) New removable SINGLE USE DISTAL COVER MAJ-2315 is introduced to the subject device compared to fixed distal cover of the predicate device. Design modifications were made around the forceps elevator to optimize the movement of the forceps elevator. 2) The reprocessing process of the subject device is modified from the predicate device due to the introduction of SINGLE USE DISTAL COVER MAJ-2315. This includes the use of a DISTAL END FLUSHING ADAPTER MAJ-2319. # 5.7 PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination. # 1) Sterilization/Shelf life testing Sterilization/shelf life testing for the MAJ-2315 were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are sans-serif and appear to be a standard font. Below the word "OLYMPUS" is a thick, horizontal, yellow bar. The bar is slightly shorter than the length of the word "OLYMPUS" above it. "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile". Accelerated aging test for the MAJ-2315 was conducted in accordance with ASTM F1980-16, the standard guide for accelerated aging of sterile barrier systems for medical devices. The real-time aging test for three-years will be performed to demonstrate longer stability and support the results of the accelerated aging test. # 2) Reprocessing validation testing Reprocessing instruction and reprocessing method validation testing for the TJF-Q190V were conducted and documentation was provided as recommended by Guidance for Industry and Food and Drug Administration Staff, "Reprocessing Medical Devices in Health Care Setting: Validation Methods and Labeling". # 3) Biocompatibility testing Biocompatibility testing for the TJF-Q190V and MAJ-2315 were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests: - Cytotoxicity Study Using the Colony Assay । - Intracutaneous Study in Rabbits । - Guinea Pig Maximization Sensitization Test । # 4) Software verification and validation testing Software verification and validation testing for the TJF-Q190V were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". # 5) Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the TJF-Q190V. The system complies with the AAMI ANSI ES60601-1: 2005/(R)2012 and A1:2012 and IEC 60601-2-18: Edition 3.0 2009-08 standards for safety and the IEC 60601-1-2: Edition 4: 2014-02 standards for EMC. # 6) Performance testing - Bench Bench testing for the TJF-Q190V and its accessories as listed below was conducted to ensure that the subject device performs as intended and meet design specifications. Device performance assessed the design requirements, and included process verification, design verification, and design validation. {7}------------------------------------------------ - Thermal Safety test - Mechanical durability test - Performance testing for MAJ-2315 - Photobiological safety test - Accidental Physical Impact testing on distal tip # 7) Performance testing - Animal No animal study was performed to demonstrate substantial equivalence. # 8) Performance testing - Clinical No clinical study was performed to demonstrate substantial equivalence. # 9) Risk analysis Risk analysis for the TJF-Q190V was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2007 and the human factors validation was conducted in accordance with the FDA Guidance, "Applying Human Factors and Usability Engineering to Medical Devices". The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment. # 5.8 CONCLUSIONS Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate devices, the TJF-Q190V raise no new issue of safety and effectiveness and are substantially equivalent to the predicate devices in terms of safety, efficacy and performance.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...