K193158 · Global Medikit Limited · FOZ · Nov 24, 2020 · General Hospital
Device Facts
Record ID
K193158
Device Name
ACENT™ Central Venous Catheter
Applicant
Global Medikit Limited
Product Code
FOZ · General Hospital
Decision Date
Nov 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ACENT™ Central Venous Catheter is an intravascular catheter, designed for infusion of fluid to the central venous system and/or for pressure measurements. - for short-term (less than 30 days) infusion therapy or parenteral nutrition - for continuous or intermittent monitoring of the central venous pressure, - for blood sampling.
Device Story
ACENT™ Central Venous Catheter is a polyurethane, radiopaque, single/double/triple-lumen catheter used for central venous access. Inserted percutaneously via Seldinger technique into a large vein and threaded to the SVC-RA junction. Used in clinical settings for short-term infusion therapy, parenteral nutrition, central venous pressure monitoring, and blood sampling. Features depth markings and soft distal tip to minimize trauma. Radiopacity allows for X-ray verification of placement. Output is fluid delivery or pressure data for clinical monitoring. Benefits include reliable vascular access for critical care and nutrition.
Clinical Evidence
No clinical data. Bench testing only. Testing included ISO 10555-1/3 (catheter requirements), ISO 594-1/2 (luer fittings), USP <788> (particulate), and comprehensive biocompatibility (ISO 10993-1) including cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation, hemocompatibility, and pyrogenicity. Sterilization validated per ISO 11135; packaging/shelf-life validated per ISO 11607-1 and ASTM 1980.
Technological Characteristics
Polyurethane construction; radiopaque; single, double, or triple lumen configurations (4F-8F). Percutaneous insertion via Seldinger technique. Ethylene oxide sterilization (SAL 10^-6). Complies with ISO 10555-1/3, ISO 594-1/2, ISO 10993-1, ISO 11135, ISO 11607-1, and ASTM 1980.
Indications for Use
Indicated for patients requiring short-term (less than 30 days) central venous access for infusion therapy, parenteral nutrition, continuous or intermittent central venous pressure monitoring, or blood sampling.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
November 24, 2020
Global Medikit Limited % Atonu Dutta, CEO Meditech Consulting 102, Platinum Avenue, Vikram Society, Gotri Road Vadodara, Gujarat India 390021
Re: K193158
Trade/Device Name: ACENT™ Central Venous Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: October 17, 2020 Received: October 30, 2020
Dear Atonu Dutta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Payal Patel
Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193158
Device Name ACENTTM Central Venous Catheter
Indications for Use (Describe)
ACENT™ Central Venous Catheter is an intravascular catheter, designed for infusion of fluid to the central venous system and/or for pressure measurements.
- for short-term (less than 30 days) infusion therapy or parenteral nutrition
- for continuous or intermittent monitoring of the central venous pressure,
- for blood sampling.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 12px;"> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span style="font-size: 12px;"> <span> <span style="border: 1px solid black; display: inline-block; width: 13px; height: 13px; vertical-align: middle;"></span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) Summary
## K193158
#### l. Submitter
M/s. Global Medikit Limited
Khasra No. 323 (MI) Camp Road, Selaqui Deharadun, Uttrakhand, India
Tel. No: +91 11 27667889.
E-mail : regulatory@globalmedikit.in
Web site: www.globalmedikit.in
Contact Person: Mr. Mahendra Singh
Date of Prepare: 24th November 2020
## Application Correspondent:
C/o: Atonu Dutta, CEO
Meditech Consulting
102 Platinum Avenue, Vikram Society, Gotri Road, Vadodara, Gujarat INDIA 390021
Tel: (91) 990-5805571
E-mail: medtect.requlatory@gmail.com
#### II. Device
| Trade Name | ACENT™ Central Venous Catheter |
|-------------------|-------------------------------------------------------------------------|
| Common Name | Catheter, Intravascular, Therapeutic, Short-term<br>(less than 30 days) |
| Device Class | II |
| Review Panel | General Hospital |
| Product Code | FOZ |
| Regulation Number | 21 CFR 880.5200 |
| Regulation Name | Intravascular Catheter |
### III. Predicate Device
| Trade Name | : | Able Central Venous Catheter |
|-------------|---|---------------------------------------------------|
| Manufacture | : | Foshan Nanhai Bai He Medical Technology Co., Ltd. |
| 510(K) No. | : | K070451 |
### IV. Device Description
ACENT™ catheters are polyurethane, radiopaque single, double, or triple lumen catheters. The catheter is available in sizes that vary from 4 French through 8 French and lengths 8 cm through 20 cm. Each lumen extends from the vicinity of the distal tip to the main (bifurcation or junction) hub, where it branches via the connector into dedicated extension tubes for infusion of fluids. Each of the extension line has a slide clamp and are labeled to provide identification of the lumen size and location. The catheter body has depth markings, measured in cm from the catheter tips. The extension line marked "distal" is used for device placement with a compatible
{4}------------------------------------------------
guide wire.
The distal tip is soft to minimize patient trauma during insertion. The device is radiopaque to allow verification of location in the patient. Central venous catheters are inserted into a large vein and threaded into the central venous system.
## V. Indications for Use
ACENT™ Central Venous Catheter is an intravascular catheter, designed for infusion of fluid to the central venous system and/or for pressure measurements.
- for short-term (less than 30 days) infusion therapy or parenteral nutrition
- for continuous or intermittent monitoring of the central venous pressure,
- for blood sampling.
## VI. Substantial Equivalence
The ACENT™ Central Venous Catheter is substantially equivalent to marketed predicate devices with respect to intended use and technological characteristics. The ACENT™ Central Venous Catheter operates on the same principle as predicate device. Comparison Chart of ACENT™ Central Venous Catheter System and Predicate devices is provided below:
| Characteristics | Subject Device | Predicate Device | Change |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | ACENT™<br>Central Venous<br>Catheter | Able Central Venous<br>Catheter | - |
| 510 (K) | K193158 | K070451 | - |
| Manufacturer | Global Medikit Ltd. | Foshan Nanhai Bai He<br>Medical Technology Co.,<br>Ltd | - |
| Common Name | Central Venous<br>Catheter | Central Venous Catheter | No Change |
| Regulation<br>Name | Intravascular Catheter | Intravascular Catheter | No Change |
| Class | II | II | No Change |
| Product Code | FOZ | FOZ | No Change |
| Regulation<br>Number | 21 CFR 880.5200 | 21 CFR 880.5200 | No Change |
| Use duration | Short Term<br>(Less than 30<br>days) | Short Term (Less<br>than 30 days) | No Change |
| Single Patient<br>Use | Yes | Yes | No Change |
| Patient<br>Contact | External communicating device<br>circulating blood contact<br>prolonged use (Short term) | External communicating<br>device circulating blood<br>contact prolonged use<br>(Short term) | No Change |
| Indications for<br>Use | ACENT™ Central Venous<br>Catheter is an intravascular<br>catheter, designed for infusion<br>of fluid to the central venous<br>system and/or for pressure<br>measurements.<br>for short-term (less than<br>30 days) infusion therapy<br>or parenteral nutrition for continuous or<br>intermittent monitoring of<br>the central venous<br>pressure, for blood sampling. | The Central Venous Catheters Indications are<br>including:<br>14G, 16G, 18G, and 20G in<br>single-lumen catheters;<br>4F, 5F, 7F, and 8F in double-<br>lumen catheters;<br>are intended for vascular<br>access infusion<br>and withdrawal of blood,<br>blood products, and fluids,<br>plasma pheresis,<br>hyperalimentation, central<br>venous blood sampling and<br>continuous and intermittent<br>drug infusion. | same.<br>In subject device<br>indication is for<br>general infusion<br>therapy and<br>parenteral<br>nutrition<br>whereas the<br>predicate<br>includes plasma<br>pheresis, and<br>infusion of blood<br>and blood<br>products. The<br>intended use of<br>the subject<br>device is within<br>the scope of the<br>intended use of<br>the predicate. |
| Catheter<br>Placement<br>method | Percutaneous via Seldinger<br>technique | Percutaneous via Seldinger<br>technique | No Change |
| Number of<br>Catheter Lumens | 1 to 3 | 1 to 3 | No Change |
| Catheter size | | | |
| Single Lumen | 4fr, 5fr, 6fr, 7fr (14G,16G,18G) | 3fr,4fr,5fr,6fr,7fr<br>(14G,16G,18G,20G) | Sizes are similar<br>except that the<br>Predicate has an<br>additional 3fr<br>(20G) size.<br>Sizes of subject<br>device are within<br>the range of the<br>predicate. The<br>difference will<br>not affect safety<br>and<br>effectiveness of<br>the subject<br>device<br>performance<br>testing is<br>conducted on<br>worst case |
| | | | |
| Double Lumen | 5fr, 5.5fr, 7fr, 7.5fr | 4fr,5fr,7fr,8fr | The subject<br>device has<br>additional sizes<br>(5Fr and 7.5 Fr)<br>that are within<br>the range of the<br>predicate device<br>sizes (4fr and<br>8fr). The<br>additional sizes<br>will not affect<br>safety and<br>effectiveness of<br>the subject<br>device because<br>performance<br>testing is<br>conducted on<br>worst case<br>sizes. |
| Triple Lumen | 5fr, 5.5fr, 7fr, 7.5fr, 8fr | 5.5fr,7fr | The subject<br>device has<br>additional sizes,<br>5fr,7.5fr, &8fr.<br>The additional<br>sizes will not<br>affect safety and<br>effectiveness of<br>the subject<br>device because<br>performance<br>testing is<br>conducted on<br>worst case sizes. |
| Catheter Length | | | |
| Single Lumen | 15,16 & 20 cm | 15,16 & 20 cm | No Change |
| Double Lumen | 15,16 & 20 cm | 15,16 & 20 cm | No Change |
| Triple Lumen | 8,12,13, 15,16 & 20cm | 8,12,13, 15,16 & 20cm | No Change |
| Catheter<br>Tip Target<br>anatomy | SVC-RA junction | SVC-RA junction | No Change |
| Catheter Tip<br>Location<br>Confirmation<br>Method | X-Ray | X-Ray | No Change |
| Catheter Distal<br>End<br>Configuration | Straight and tapered | Straight and tapered | No Change |
| Catheter Ink<br>Markings | yes | yes | No Change |
| Catheter Shaft<br>Material | Polyurethane | Polyurethane | No Change |
| Catheter Tip &<br>Shaft Material | Polyurethane | Polyurethane | No Change |
| Lumen Shape | Round, Crescent | Round, Crescent | No Change |
| Packaging | Plastic Tray with Tyvek Lid stock | Plastic Tray with Tyvek Lid stock | No Change |
| Sterility Assurance<br>Level (SAL) | $10^{-6}$ | $10^{-6}$ | No Change |
| Sterilization | Ethylene Oxide | Ethylene Oxide | No Change |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
## VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## Bench Performance
The performance testing of ACENT™ Central Venous Catheter is carried out on sterilized production lot and per the following standard:
- ISO 10555 -1, "Intravascular catheters Sterile and single-use catheters Part 1: . General requirements"
- ISO 10555-3, "Intravascular catheters Sterile and single-use catheters Part 3: . Central venous catheters"
- ISO 594-1, "Conical fittings with a 6% (Luer) taper for syringes, needles and certain . other medical equipment - Part 1: General requirements"
- ISO 594-2, Conical fittings with a 6% (Luer) taper for syringes, needles and certain . other medical equipment - Part 2: Lock fittings"
- ISO 14971, "Medical devices Applications of risk management to medical devices" .
- Particulate testing per USP <788> ●
## Biocompatibility
The biocompatibility evaluation for ACENT™ Central Venous Catheter was conducted in accordance with the guidance document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and ISO 10993-1, "Biological evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process" as recognized by FDA.
As per ISO 10993-1, ACENT™ Central Venous catheter is an external communicating device with prolonged circulating blood contact. Accordingly, the following biological evaluation tests were conducted:
- Cytotoxicity
- Sensitization
- Irritation / Intracutaneous reactivity ●
- Acute Systemic Toxicity
- Sub Chronic Toxicity Study
- Genotoxicity
- Implantation study
- Hemocompatibility
- Pyrogenicity (Material and Bacterial)
## Sterilization
The sterilization process has been validated per AAMI/ANSI/ISO 11135, ", Sterilization of Health-Care Products - Ethylene Oxide - Requirements for the Development, Validation and Routine Control of a Sterilization Process for Medical Devices" and demonstrated a sterility assurance level of 10°. Ethylene oxide residuals were tested and met ISO 10993-7, "Biological
{8}------------------------------------------------
Evaluation Of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals" requirements.
## Packaging & Shelf life
Following packaging and shelf life study was conducted to ensure package integrity throughout the shelf life. The shelf life is established for 2 years.
- · Packaging validation as per ISO 11607-1, "Packaging for terminally Sterilized Medical Devices - Part 1: Packaging For Terminally Sterilized Medical Devices - Part 1: Requirements for Materials, Sterile Barrier Systems And Packaging Systems"
- Shelf life validation as per ASTM 1980, "Standard guide for Accelerated Aging of Sterile . Barrier Systems for Medical Devices" & ISO 11607-1, "Packaging for terminally Sterilized Medical Devices - Part 1: Packaging For Terminally Sterilized Medical Devices - Part 1: Requirements for Materials, Sterile Barrier Systems And Packaging Systems"
## VIII. Conclusion
ACENT™ Central Venous Catheter is substantially equivalent to the predicate device and presents no new questions of safety or effectiveness when compared in design, technological characteristics, performance testing, and intended use to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.