Nipro Dry Complete Dry Acid Concentrate for Hemodialysis
K193155 · Nipro Renal Solutions USA, Corp. · KPO · Nov 25, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K193155
Device Name
Nipro Dry Complete Dry Acid Concentrate for Hemodialysis
Applicant
Nipro Renal Solutions USA, Corp.
Product Code
KPO · Gastroenterology, Urology
Decision Date
Nov 25, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For DA-100 series: This acid concentrate product is formulated for use in acute and chronic hemodialysis and to be used in conjunction with MedicaLyteTM bicarbonate, or an equivalent bicarbonate labeled for use in acute and chronic hemodialysis and having the same composition as MedicaLyteTM bicarbonate, in a 36.83X three-stream artificial kidney (hemodialysis) machine. -Or- For DA-200 series: This acid concentrate product is formulated for use in acute and to be used in conjunction with Medical yteTM bicarbonate, or an equivalent bicarbonate labeled for use in acute and chronic hemodialysis and having the same composition as MedicalyteTM bicarbonate, in a 45X three-stream artificial kidney (hemodialysis) machine. -Or- For DA-300 series: This acid concentrate product is formulated for use in acute and to be used in conjunction with MedicaLyteTM bicarbonate, or an equivalent bicarbonate labeled for use in acute and chronic hemodialysis and having the same composition as MedicaLyteTM bicarbonate, in a 35X three-stream artificial kidney (hemodialysis) machine.
Device Story
Nipro Dry Complete consists of dry acid concentrate (USP grade sodium chloride, magnesium chloride, calcium chloride, potassium chloride, dextrose, acetic acid). Used in clinical settings for acute and chronic hemodialysis. Device components are proportioned with bicarbonate concentrate and water by a three-stream hemodialysis machine to create the final dialysis solution. This solution circulates through a hemodialyzer, separated from patient blood by semi-permeable membranes to remove waste products and toxins. Healthcare providers (clinicians/nurses) operate the hemodialysis machine; the concentrate is a consumable component. Benefits include effective removal of metabolic waste and toxins in renal failure patients.
Clinical Evidence
Bench testing only. Performance testing conducted on representative batches across all three series (DA-100, DA-200, DA-300) to verify chemical concentration accuracy per ANSI/AAMI/ISO 13958:2014. Endotoxin analysis confirmed levels below 0.5 EU/mL. Transportation and package integrity testing performed per ASTM D4169-16 and ISTA Procedure 3A:2018. Biocompatibility evaluated per ISO 10993-1.
Technological Characteristics
Dry acid concentrate containing USP grade NaCl, MgCl2, CaCl2, KCl, dextrose, and acetic acid. Packaged in HDPE bottles and LDPE bags. Proportioning ratios: 36.83X, 45X, or 35X. Non-sterile. Complies with ANSI/AAMI/ISO 13958:2014 for hemodialysis concentrates. Packaging integrity tested per ISO 2859-1:2011.
Indications for Use
Indicated for patients with acute or chronic renal failure requiring hemodialysis. Used in conjunction with bicarbonate concentrate and water in a three-stream proportioning hemodialysis machine.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services - USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 25, 2020
Nipro Renal Solutions Corporation USA % David Marcus Associate McNees Wallace & Nurick LLC 21 East State Street, Suite 1700 Columbus, OH 43215
Re: K193155
Trade/Device Name: Nipro Dry Complete Dry Acid Concentrate for Hemodialysis Regulation Number: 21 CFR 876.5820 Regulation Name: Hemodialysis System and Accessories Regulatory Class: II Product Code: KPO Dated: October 29, 2020 Received: October 30, 2020
Dear David Marcus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K193155
### Device Name
Nipro Dry Complete™ Dry Acid Concentrate for Hemodialysis
Indications for Use (Describe) For DA-100 series:
This acid concentrate product is formulated for use in acute and chronic hemodialysis and to be used in conjunction with MedicaLyteTM bicarbonate, or an equivalent bicarbonate labeled for use in acute and chronic hemodialysis and having the same composition as MedicaLyteTM bicarbonate, in a 36.83X three-stream artificial kidney (hemodialysis) machine.
-Or-
For DA-200 series:
This acid concentrate product is formulated for use in acute and to be used in conjunction with Medical yteTM bicarbonate, or an equivalent bicarbonate labeled for use in acute and chronic hemodialysis and having the same composition as MedicalyteTM bicarbonate, in a 45X three-stream artificial kidney (hemodialysis) machine.
-Or-
For DA-300 series:
This acid concentrate product is formulated for use in acute and to be used in conjunction with MedicaLyteTM bicarbonate, or an equivalent bicarbonate labeled for use in acute and chronic hemodialysis and having the same composition as MedicaLyteTM bicarbonate, in a 35X three-stream artificial kidney (hemodialysis) machine.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY UNDER 21 CFR 807.92
This 510(k) summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content in this 510(k) summary has been provided in conformance with 21 CFR Part 807.92.
#### A. Submitter's Information
| (1) Name: | McNees Wallace & Nurick LLC |
|-------------------------------------------------|-------------------------------------------------------|
| (2) Address: | 21 East State St., Suite 1700, Columbus, OH 43215 USA |
| (3) Phone: | 614-719-2856 |
| (4) Fax: | 614-469-4653 |
| (5) Contact Person: | David M. Marcus, J.D., Ph.D. |
| (6) Preparation Date: | November 11, 2019 |
| (7) Revised Preparation Date: | October 29, 2020 |
| On Behalf of Applicant Entity (owner of 510(k)) | |
| (1) Applicant Name: | Nipro Renal Solutions Corporation USA |
|----------------------------|-------------------------------------------------|
| (2) Applicant Address: | 509 Fishing Creek Rd., Lewisberry, PA 17339 USA |
| (3) Applicant Phone: | 717-938-8391 |
| (4) Applicant Fax: | 717-938-3957 |
| (5) Establishment Reg. No: | 1528807 |
### B. Device Name
| (1) Trade Name: | Nipro Dry Complete™ Dry Acid Concentrate for Hemodialysis |
|---------------------|---------------------------------------------------------------------|
| (2) Common Name: | Dialysis Concentrate for Hemodialysis (Dry Acid) |
| (3) Classification: | Class II, per 21 CFR 876.5820 (Hemodialysis system and accessories) |
| (4) Product Code: | KPO: Gastroenterology/Urology |
### C. Legally Marketed Predicate Devices
Nipro Renal Solutions Corporation USA MedicaPure™ Liquid Acid Concentrate (K901471) Diasol, Inc. Dryasol Acid Concentrate Mix (K993212) Nipro Renal Solutions Corporation USA MedicaLyte™ Bicarbonate Powder (K131202)
{4}------------------------------------------------
## 510(K) SUMMARY UNDER 21 CFR 807.92
#### D. Device Description
The Nipro Dry Complete™ dry acid concentrate devices for hemodialysis are comprised of USP grade sodium chloride, USP grade magnesium chloride, USP grade calcium chloride, USP grade potassium chloride, USP grade dextrose, and USP grade acetic acid. These products may be used in conventional, commercially available hemodialysis machines or monitors as one of the necessary components of three-component hemodialysis solution, and are formulated for 36.83X (DA-100 series), 45X (DA-200 series), and 35X (DA-300 series) proportioning dialysis machines. The hemodialysis concentrate solutions presented in this 510(k) premarket notification are intended to be used in an appropriately proportioned three-stream hemodialysis machines (36.83X, 45X, or 35X as specified on the device label) in which bicarbonate concentrate is proportioned into one stream, an acid solution prepared from the Nipro Dry Complete™ dry acid concentrate device is proportioned into a second stream, and water is proportioned into the third stream of the hemodialysis machine proportioning system. These three streams are then mixed by the hemodialysis machine to prepare the final proportioned hemodialysis solution. The final hemodialysis solution is separated from the patient's blood by semi-permeable membranes which permit the passage of waste products and toxins contained in the patient's blood circulating through the hemodialyzer, and such waste products and toxins pass through the semi-permeable membranes into the hemodialysis solution and exit the hemodialyzer with the hemodialysis solution. By such treatment, waste products and toxins are removed from the patient's blood during acute and end-stage renal failure.
#### E. Indications for Use
### For DA-100 series:
This acid concentrate product is formulated for use in acute and chronic hemodialysis and to be used in conjunction with MedicaLyte™ bicarbonate, or an equivalent bicarbonate labeled for use in acute and chronic hemodialysis and having the same composition as MedicaLyte™ bicarbonate, in a 36.83X three-stream artificial kidney (hemodialysis) machine.
### -01-
## For DA-200 series:
This acid concentrate product is formulated for use in acute and chronic hemodialysis and to be used in conjunction with MedicaLyte™ bicarbonate, or an equivalent bicarbonate labeled for
{5}------------------------------------------------
# 510(K) SUMMARY UNDER 21 CFR 807.92
use in acute and chronic hemodialysis and having the same composition as MedicaLyte™ bicarbonate, in a 45X three-stream artificial kidney (hemodialysis) machine.
-01-
## For DA-300 series:
This acid concentrate product is formulated for use in acute and chronic hemodialysis and to be used in conjunction with MedicaLyte™ bicarbonate, or an equivalent bicarbonate labeled for use in acute and chronic hemodialysis and having the same composition as MedicaLyte™ bicarbonate, in a 35X three-stream artificial kidney (hemodialysis) machine.
#### F. Technological Characteristics:
Comparing the proposed Dry Complete™ dry acid concentrate devices to the MedicaPure™ Liquid Acid Concentrate (K901471) predicate devices utilize the same chemical compositions and produce the same acid solutions for hemodialysis. The devices also utilize substantially equivalent high density polyethylene bottles, which are made by the same manufacturing process with the same resin by the same manufacturer (CKS Packaging located at 943 Trollingwood Rd., Haw River, North Carolina 27258). The only difference between the bottles of the MedicaPure™ Liquid Acid Concentrate predicate devices (K901471) and the bottles of the proposed Dry Complete™ dry acid concentrate devices is the size of the bottles, which does not present any new hazard to the patients. Thus, there are no significant differences, and a finding of substantial equivalence is appropriate.
Comparing the proposed Nipro Dry Complete™ dry acid concentrate devices to the Diasol, Inc. Dryasol Acid Concentrate Mix (K993212) predicate devices utilize the same chemical compositions and produce the same acid concentrates for thirty-three of the seventy-three proposed devices (with the other proposed devices being substantiated by K901471 as described above). There are no significant differences, and therefore a finding of substantial equivalence is appropriate.
Comparing the proposed Dry Complete™ dry acid concentrate devices to the MedicaLyte™ Bicarbonate Powder (K131202) predicate devices, the powder components of the devices are packaged in identical low density polyethylene bags (same material composition and processing) from the same manufacturer (Elkay Plastics Co., Inc. located at 6000 Sheila Street,
Tab 5: Page 3
{6}------------------------------------------------
## 510(K) SUMMARY UNDER 21 CFR 807.92
Commerce, California 90040). There are no significant differences, and therefore a finding of substantial equivalence is appropriate.
# Performance Testing:
Performance testing was conducted on three batches each of DA-100, DA-104, DA-106, DA-113, DA-115, and DA-118 for 36.83X proportioning systems; DA-210, DA-211, DA-213, DA-229, and DA-234 for 45X proportioning systems; and DA-303, DA-308, and DA-314 for 35X proportioning systems. These formulations encompass the entire range of the chemical components present in the devices for each proportioning system and were thus determined to be representative of the proposed formulations. Testing was accomplished by formulating the dialysis liquid concentrate pursuant to the instructions on the device labels and measuring the concentrations of all components to ensure that following the label instructions generates the liquid concentrate specified on the label. All testing was conducted pursuant to ANSI/AAMI/ISO 13958:2014, specifically with reference to sections 4.1.2.2 and 4.1.2.1. All samples of all batches of each device tested were within acceptable limits as set by ANSI/AAMI/ISO 13958:2014. Therefore, the performance of these devices is considered to be satisfactory.
Transportation testing per ASTM D4169-16 has been completed, including environmental conditioning, per ISTA Procedure 3A:2018, simulated distribution cycle testing, and inner package integrity testing. The results indicate that the proposed devices can be shipped and transported so that they may be received and used in a safe and effective way.
Biocompatibility was evaluated for the low density polyethylene bags and high density polyethylene bottles in accordance with Use of International Standard ISO 19003-1 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process."
## Endotoxin Analysis on Device Contents:
Endotoxin analysis was performed on three batches each of DA-100, DA-104, DA-106, DA-113, DA-115, and DA-118 for 36.83X proportioning systems; DA-210, DA-211, DA-213, DA-229, and DA-234 for 45X proportioning systems; and DA-303, DA-308, and DA-314 for 35X proportioning systems. These formulations encompass the entire range of the chemical components present in the devices for each proportioning system and were thus determined to be representative of the proposed formulations. Testing was performed in
{7}------------------------------------------------
# 510(K) SUMMARY UNDER 21 CFR 807.92
accordance with ANSVAAMI/ISO 13958:2014 Section 4.1.2. under which the limit for endotoxins is 0.5 endotoxin units/mL or less. The standard dialysis fluid generated from the devices was well within the acceptable limits.
# Sterilization and Shelf Life
The devices are not provided in sterilized form, and the devices are non-sterile when used. Stability testing was conducted on aged samples.
All chemistry and bioburden stability testing on aged product was performed to conform to standards as laid out in ANSI/AAMI/ISO 13958:2014, Concentrates for haemodialysis and related therapies. Additionally, the aged product tested well within endotoxin specifications for safe release of this product. Results from package integrity testing were analyzed and accepted based on Acceptance Quality Limit (AQL) criteria developed using ISO 2859-1:2011, Sampling Procedures for Inspection by Attributes, Part 1: Sampling Schemes Indexed by Acceptance Quality Limit (AQL) for Lot-by-Lot Inspection.
#### G. Conclusion
The proposed Nipro Dry Complete™ dry acid concentrate devices are believed to be substantially equivalent in design, materials, and indications for use to the commercially available Nipro Renal Solutions Corporation USA MedicaPure™ Liquid Acid Concentrate (K901471) predicate devices, Diasol, Inc. Dryasol Acid Concentrate Mix (K993212) predicate devices, and Nipro Renal Solutions Corporation USA MedicaLyte™ Bicarbonate Powder (K131202) predicate devices, and are therefore expected to be as safe, as effective, and perform as well as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.