The Biocore9 Humeral Head Resurfacing Shoulder Component is intended for the reconstruction of painful and/or severely disabled shoulder joints resulting from osteoarthritis and rheumatoid arthritis. For proper function of this device, the humeral head and neck must be of sufficient bone stock to support the loads on it. Also, the presence of an intact or reconstructable rotator cuff is important for proper functioning and dislocation resistance. Porous-coated components are intended for cemented use only. This device is indicated for hemi-arthroplasty only.
Device Story
Non-constrained orthopedic implant; replaces articular surface of humeral head. Axisymmetric truncated shell with tapered central stem; manufactured from Ti-6Al-4V alloy with Titanium Nitride (TiN) ceramic coating. Internal surface features porous titanium bead coating for cemented fixation. Available in five sizes (40-56 mm diameters). Used by orthopedic surgeons in clinical settings for hemi-arthroplasty. Device articulates with natural glenoid; restores joint function in patients with osteoarthritis or rheumatoid arthritis. Benefits include pain relief and improved joint mobility.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical performance testing, including porous coating characterization (tensile strength, shear strength, porosity) and TiN coating evaluation (thickness, adhesion, abrasion resistance, and wear resistance).
Technological Characteristics
Material: Ti-6Al-4V alloy (ASTM F136-13) with Titanium Nitride (TiN) ceramic coating. Design: Axisymmetric truncated shell with tapered central stem. Fixation: Porous titanium bead coating (cemented use only). Sizes: 40-56 mm diameters. Energy source: None (mechanical implant).
Indications for Use
Indicated for patients with painful or severely disabled shoulder joints due to osteoarthritis or rheumatoid arthritis requiring hemi-arthroplasty. Requires sufficient humeral bone stock and intact or reconstructable rotator cuff. Contraindicated for cases without sufficient bone stock or rotator cuff integrity.
Regulatory Classification
Identification
A shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis is a device made of alloys, such as cobalt-chromium-molybdenum. It has an intramedullary stem and is intended to be implanted to replace the articular surface of the proximal end of the humerus and to be fixed without bone cement (§ 888.3027). This device is not intended for biological fixation.
Predicate Devices
Buechel-Pappas, Humeral Head Resurfacing Component (K992394)
Submission Summary (Full Text)
{0}------------------------------------------------
May 15, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Biocore9, LLC. Stephen J. Peoples Official Correspondent 9 Whippany Road. Bldg A1. Unit 12 Whippany, New Jersey 07981
Re: K193122
Trade/Device Name: Biocore9 Humeral Resurfacing System Regulation Number: 21 CFR 888.3690 Regulation Name: Shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis Regulatory Class: Class II Product Code: HSD Dated: November 12, 2019 Received: November 12, 2019
Dear Stephen J. Peoples:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: Michael Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193122
Device Name Biocore9 Humeral Resurfacing System
### Indications for Use (Describe)
The Biocore9 Humeral Head Resurfacing Shoulder Component is intended for the reconstruction of painful and/or severely disabled shoulder joints resulting from osteoarthritis. For proper function of this device, the humeral head and neck must be of sufficient bone stock to support the loads on it. Also, the presence of an intact or reconstructable rotator cuff is important for proper functioning and dislocation resistance. Porous-coated components are intended for cemented use only. This device is indicated for hemi-arthroplasty only.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| <div> <span> <b> </b> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <b> </b> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments reqarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the alphanumeric string "K193122". The text is displayed in a simple, sans-serif font. The color of the text is black, and it is set against a plain white background.
Image /page/3/Picture/2 description: The image shows the word "BIOCORES" in a bold, sans-serif font. The letters are a dark blue color. The "O" in "BIOCORES" is stylized with a circular design that has a gear-like pattern in the center.
Biocore9, LLC. Humeral Resurfacing Shoulder System Traditional 510(k)
## 510(k) SUMMARY
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the Humeral Resurfacing Shoulder.
| Submitted by: | Biocore9, LLC.<br>9 Whippany Road, Bldg A1, Unit 12.<br>Whippany NJ 07981 |
|------------------------------------|---------------------------------------------------------------------------|
| Contact Person: | Stephen J. Peoples, VMD, MS, FAOA |
| | Email: speoplesVMD@gmail.com |
| Proprietary Name: | Biocore9 Humeral Resurfacing<br>Shoulder |
| Common Name: | Humeral Resurfacing Head |
| Classification Name and Reference: | 21 CFR 888.3690, Class II |
| Device Product Code, Device Panel: | HSD |
| Predicate Device: | Buechel-Pappas, Humeral Head<br>Resurfacing Component 510(k)<br>#K992394 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "BIOCORE9" in a stylized, sans-serif font. The letters are a dark blue color. The "O" in "BIOCORE9" is replaced with a gear-like symbol, with a yellow center. The overall design appears to be a logo.
#### K193122 Page 2/2
# Device Description
The Biocore9 Humeral Head Resurfacing component is a non-constrained orthopedic implant intended to replace the articular surface of the existing humeral head in the patient's shoulder joint. It is intended for the reconstruction of painful and or severely disabled shoulder joints resulting from osteoarthritis or rheumatoid arthritis for patients who would be candidates for total shoulder procedure, whose proximal humerus has not been excessively damaged by disease or trauma and where damage is primarily associated with the articular surface of the humeral head. It is primarily intended to articulate with intact, natural, glenoid. Components are available in five sizes with available outside diameters from 40 mm to 56 mm in 4 mm increments.
The Biocore9 Humeral Resurfacing Head Component is manufactured from Ti-6Al-4V alloy (ASTM F136-13) with a Titanium Nitride (TiN) thin film ceramic coating. The Humeral Resurfacing component is axisymmetric with a truncated shell and a tapered central stem. The component consists of a thin walled spherical shell. The Humeral resurfacing Component's rim is anatomical with a rounded relief. The exterior of the spherical surface is intended to articulate with the Glenoid.
The internal surface of the spherical shell is covered with a porous coating consisting of titanium beads. The stem is not porous coated.
## Intended Use
The Biocore9 Humeral Head Resurfacing Shoulder Component is intended for the reconstruction of painful and/or severely disabled shoulder ioints resulting from osteoarthritis and rheumatoid arthritis. For proper function of this device, the humeral head and neck must be of sufficient bone stock to support the loads on it. Also, the presence of an intact or reconstructable rotator cuff is important for proper functioning and dislocation resistance. Porous-coated components are intended for cemented use only. This device is indicated for hemi-arthroplasty only.
## Non-Clinical Performance Testing
Performance testing was conducted in order to provide support in establishing substantial equivalence of technological characteristics.
Testing was conducted for the Porous coating: Tensile strength, Shear strength, Porosity Testing was conducted for the TiN Coating: Thickness, Adhesion Strength to Ti6Al4V substrate, Abrasion resistance, Wear resistance Mode 1
## Substantial Equivalence
Biocore9 Humeral Resurfacing Shoulder System is substantially equivalent to the Endotec - Buechel-Pappas Humeral Resurfacing Component - K992394, since there are no significant differences in form, fit, or functionality.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.