Naviswiss Hip Navigation System

K193094 · Naviswiss AG · OLO · Jun 10, 2020 · Neurology

Device Facts

Record IDK193094
Device NameNaviswiss Hip Navigation System
ApplicantNaviswiss AG
Product CodeOLO · Neurology
Decision DateJun 10, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

The Naviswiss Hip Navigation System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes to aid in the positioning of orthopedic implant system components where a reference to a rigid anatomical structure can be identified. The system is only compatible with acetabular cup impactors that have a parallel and straight segment. The system aids the surgeon in performing intra-operative measurements including measurements of leg length, offset, and cup inclination/anteversion. An example of a stereotaxic orthopedic surgical procedure includes Total Hip Arthroplasty: Lateral/Supine. The equipment is intended for use by trained surgeons in operating theaters.

Device Story

Image-free surgical navigation system; assists orthopedic surgeons during Total Hip Arthroplasty. System components: handheld navigation device (Naviswiss Hip M42), computer module, camera, and software (Naviswiss Platform M59). Inputs: infrared optical tracking of NAVItags attached to patient anatomy (femur) and surgical instruments. Operation: camera tracks tags to determine anatomical landmarks and instrument position; software guides surgeon through workflow; provides real-time intra-operative measurements of leg length, offset, and cup inclination/anteversion. Output: visual guidance and measurements displayed to surgeon. Used in operating theaters; surgeon-operated. Benefits: aids precise implant positioning relative to pre-operative planning.

Clinical Evidence

Evidence includes bench testing, phantom simulated use, cadaver wet labs, and an OUS prospective clinical study. Bench testing verified accuracy per ASTM F2554-18. Simulated use on metallic bone models and cadavers validated clinical accuracy and workflow requirements. OUS prospective study (n=35) evaluates agreement between intraoperative navigation data and post-operative CT measurements.

Technological Characteristics

Image-free infrared stereo camera navigation system. Components: handheld navigation unit, computer, and NAVItags. Connectivity: networked/integrated system. Software: application and tracking software. Sterilization: reusable instruments undergo cleaning/sterilization; navigation unit uses sterile drapes. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ASTM F2554-18 (positional accuracy).

Indications for Use

Indicated for patients undergoing orthopedic surgery, such as Total Hip Arthroplasty, where stereotactic surgery is considered safe and effective. Intended for use by trained surgeons in operating theaters.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 10, 2020 Naviswiss AG % Viky Verna Co-Founder & VP (US Office) confinis 15807 Glacier Ct Potomac, Maryland 20878 Re: K193094 Trade/Device Name: Naviswiss Hip Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: June 5, 2020 Received: June 8, 2020 Dear Viky Verna: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K193094 Device Name Naviswiss Hip Navigation System #### Indications for Use (Describe) The Naviswiss Hip Navigation System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes to aid in the positioning of orthopedic implant system components where a reference to a rigid anatomical structure can be identified. The system is only compatible with acetabular cup impactors that have a parallel and straight segment. The system aids the surgeon in performing intra-operative measurements of leg length, offset, and cup inclination/anteversion. An example of a stereotaxic orthopedic surgical procedure includes Total Hip Arthroplasty: Lateral/Supine. The equipment is intended for use by trained surgeons in operating theaters. | Type of Use (Select one or both, as applicable) | |-----------------------------------------------------------------------------------------------------| | <label><input checked="true" type="checkbox"/> Remediation Use (Part 31, GER 901 Subpart D)</label> | | <label><input type="checkbox"/> On-Site Corrective Use (31 GER 903)</label> | > Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 5. 510(k) Summary In accordance with 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. ## Applicant: Naviswiss AG Submitter name: Jan Stifter Responsible person: Jan Stifter Phone: +41 61 761 85 37 E-mail: jan.stifter@naviswiss.eu ### Official Correspondent: | Contact person: | Viky Verna, MS BME, MS Pharm, RAC | |-----------------|-----------------------------------| | Phone: | 786-525-9811 | | Email: | Viky.verna@confinis.com | | Date prepared: | 4 November 2019 | #### Device Name: Naviswiss Hip Navigation System | Proprietary name: | Naviswiss Hip Navigation System | |----------------------|-----------------------------------| | 510(k) number: | To be assigned | | Common name: | Orthopedic Stereotaxic Instrument | | Classification name: | Orthopedic Stereotaxic Instrument | | Product code: | OLO | #### Predicate Device: Substantial Equivalence is claimed with the device, K162364 "intellijoint HIP Generation 2A System", manufactured by Intellijoint Surgical Inc. on the basis of equivalent intended use / indications for use, technological characteristics and principle of operation. #### Device Description: The Naviswiss Hip Navigation System is an image-free surgical navigation system intended to assist the orthopedic surgeon during the implantation of an artificial joint (hip). The device consists of a handheld navigation device, used to register the patient's anatomy into a software platform. Subsequently the navigation system helps the surgeon maneuver the surgical instruments with precision. Thus, the surgeon can position the implant in accordance to pre-operative planning. The Naviswiss Hip Navigation System is comprised of two major sub-systems: the Naviswiss Hip (M42) and Naviswiss Platform (M59). {4}------------------------------------------------ The Naviswiss Hip (M42) is composed of: - . NAVItags used for reference and registration - . Ancillary instruments - . Application Software The Naviswiss Platform (M59) is composed of: - . Computer - Camera ● - . Software The M42 employs the base computing, power supply, and programing in the M59 Platform. The NAVItags allow the surgeon to determine anatomical landmarks or navigate an instrument. The camera utilizes application and tracking software to guide the surgeon through a workflow process for the surgery. The camera is covered by a sterile, single use drape for use of the navigation device in a sterile environment. #### Indications for Use: The Naviswiss Hip Navigation System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes to aid in the positioning of orthopedic implant system components where a reference to a rigid anatomical structure can be identified. The system is only compatible with acetabular cup impactors that have a parallel and straight segment. The system aids the surgeon in performing intra-operative measurements including measurements of leg length, offset, and cup inclination/anteversion. An example of a stereotaxic orthopedic surgical procedure includes Total Hip Arthroplasty: Lateral/Supine. The equipment is intended for use by trained surgeons in operating theaters. #### Comparison of Technological Characteristics: The substantial equivalence of the Naviswiss Hip Navigation System to the predicate is shown by similarity in intended use, indications for use and performance. {5}------------------------------------------------ | Characteristic | Subject Device | Predicate Device | |---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Naviswiss Hip Navigation System | intellijoint HIP Generation 2A System | | 510(k) # | to be assigned | K162364 | | Intended use | The Naviswiss Hip Navigation System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes to aid in the positioning of orthopedic implant system components where a reference to a rigid anatomical structure can be identified. The system is only compatible with acetabular cup impactors that have a parallel and straight segment. The system aids the surgeon in performing intra-operative measurements including measurements of leg length, offset, and cup inclination/anteversion.<br><br>An example of a stereotaxic orthopedic surgical procedure includes Total Hip Arthroplasty: Lateral/Supine.<br><br>The equipment is intended for use by trained surgeons in operating theaters. | The intellijoint HIP Generation 2A System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified. The system is only compatible with straight acetabular cup impactors. The intellijoint HIP Generation 2A System is indicated for patients undergoing orthopaedic surgery, and where the use of stereotactic surgery is considered safe and effective. The system aids the surgeon in performing intra-operative measurements including measurements of limb position, joint center-of-rotation, and implant component positioning.<br><br>Example orthopedic surgical procedures include, but are not limited to: Total Hip Arthroplasty<br>Minimally Invasive Hip Arthroplasty | | Technology /<br>Scientific principle | Image-free surgical navigation system using infrared stereo camera<br><br>Identical to predicate<br><br>Identical to predicate | Uses infrared optical technology and integrated microelectronics<br><br>intellijoint HIP® posterior and anterior applications provide real-time, intraoperative measurements to assist orthopaedic surgeons in accurate implant positioning for cup position, leg length, offset and hip center of rotation.<br><br>Measure changes in leg length, offset and hip center, as well as native acetabular position and cup position (anteversion and inclination) | | Device Descrip-<br>tion | Naviswiss Hip Navigation System is<br>an image-free surgical navigation<br>system, consists of a handheld navi-<br>gation device and a computer mod-<br>ule | 3D mini-optical navigation system,<br>surgeon controlled surgical guidance<br>tool | | Cleaning and<br>Sterilization<br>Shelf-Life | The reprocessing of reusable instru-<br>ments consists of preparation, pre-<br>cleaning, and automated cleaning<br>followed by sterilization. It was veri-<br>fied that tags are correct after the in-<br>itial shelf life validation. The naviga-<br>tion unit is draped and used in the<br>sterile area. | The work station is situated outside<br>of the sterile field, in view of the sur-<br>geon, while the camera and other<br>equipment remain within the sterile<br>field. | | Operating Prin-<br>ciple | The M42 employs the base compu-<br>ting, power supply, and programing<br>in the M59 Platform. The NAVItags<br>attach various objects (e.g. the fe-<br>mur, surgical instruments, etc.) for<br>either determining anatomical land-<br>marks or navigating an instrument.<br>The camera utilizes software to<br>guide the surgeon through a work-<br>flow process for the surgery. | System generates positional meas-<br>urements between a camera<br>(mounted on the iliaccrest) and a<br>tracker than can be mounted on vari-<br>ous objects (e.g. the femur, the ace-<br>tabular cup impactor, etc.) during<br>surgery to measure their positions.<br>The camera detects the tracker<br>within its field of view and communi-<br>cates with a workstation, which exe-<br>cutes the application software to<br>guide the device workflow and dis-<br>play measurements to the surgeon. | | Performance | All performance tests were passed<br>successfully | All performance tests were passed<br>successfully | {6}------------------------------------------------ The technological principle for both the subject and predicate devices is to provide intra-operative measurements during orthopedic surgery. It is based on the use of infrared optical technology to provide real-time, intraoperative data on leg length, offset, and cup position. The subject and selected predicate devices are based on similar technological elements. The system generates positional measurements between a camera and a tracker that can be mounted on different objects. The camera uses the application software and the tracking software to guide the surgeon through a workflow process for the surgery. #### Summary of Testing: The following tests were performed to demonstrate the substantial equivalence of the Naviswiss Navigation Hip System to the intellijoint HIP Generation 2A System. #### Biocompatibility Testing {7}------------------------------------------------ The biocompatibility evaluation for the Naviswiss Hip Navigation System was conducted in accordance with ISO 10993-1. #### Reprocessing Validation A reprocessing validation was conducted for the reusable instruments. #### EMC Testing Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the Naviswiss Hip Navigation System. The system complies with the IEC 60601-1, standard for safety and the IEC 60601-1-2 standard for EMC. #### Software Verification and Validation Testing Software verification and validation testing were conducted to ensure functional requirements were met and the system performs as intended. Algorithms and measurement calculations were also verified in these tests. All requirements and specifications were met. ### System Accuracy The Naviswiss Hip Navigation System's accuracy was verified according to the methodology in ASTM F2554-18 - Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems. Testing simulated normal conditions, and a variety of worst-case use scenarios and realistic tracking disturbances. All requirements were met. #### Benchtop Accuracy The Naviswiss Hip Navigation System's accuracy was verified using calibrated test fixtures. All requirements were met. #### Anatomical Phantom Simulated Use and Clinical Accuracy Simulated use testing was performed on a metallic bone simulator by orthopedic surgeons in THA procedures following a typical workflow. The test validated that the Naviswiss Hip Navigation System satisfies user needs, intended use, and clinical accuracy requirements. This was assessed by comparing the measurements obtained in the simulated use with known values. #### Cadaver Simulated Use Simulated use testing was performed in multiple cadaver wet labs to validate the Naviswiss Hip Navigation System satisfies clinical use requirements and performed as intended on human specimans when used in an OR environment by trained surgeons. #### OUS Prospective Clinical Study {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for "confinis project mgmt & consulting". The word "confinis" is written in a combination of red and gray letters. Below the word "confinis" is the phrase "project mgmt & consulting" in a smaller, gray font. To the left of the word "confinis" are four squares, one of which is filled in with red. Naviswiss is conducting a prospective clinical study in Australia. The study was designed to evaluate 35 patients undergoing elective THA surgery. The primary outcome is the agreement between intraoperative component positioning data captured by the Naviswiss Hip Navigation System compared with post-operative CT measurements. #### Conclusion: Based on equivalence of intended use / indications for use, technological characteristics and operational principle the applicant concludes, that substantial equivalence between the new and the predicate device has been demonstrated and that the new device, Naviswiss Hip Navigation System, is substantially equivalent to the legally marketed predicate device, intellijoint HIP Generation 2A System (K162364).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...