K193082 · Merit Medical Systems, Inc. · KRA · Aug 7, 2020 · Cardiovascular
Device Facts
Record ID
K193082
Device Name
Fountain ValveTip Infusion Catheter
Applicant
Merit Medical Systems, Inc.
Product Code
KRA · Cardiovascular
Decision Date
Aug 7, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1210
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fountain ValveTip Infusion System is intended to administer infusions of various therapeutic solutions into the peripheral vasculature of a patient.
Device Story
Fountain ValveTip Infusion Catheter; peripheral infusion catheter for therapeutic solutions. Features distal infusion holes and radiopaque marker bands for positioning. Incorporates distal tip valve to seal catheter, facilitating flow and spray distribution; allows tracking over 0.035" guide wire. Eliminates need for occluding guide wire used in predicate. Available in 4 and 5 French sizes; lengths 45-135 cm. Used by clinicians in clinical settings for targeted delivery of fluids into peripheral vasculature. Output is controlled infusion of therapeutic agents; assists in localized treatment. Benefits patient via precise fluid delivery and simplified procedural workflow.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included biocompatibility (ISO 10993 series), sterilization validation (ISO 11135-1), and mechanical/functional bench testing (ISO 10555-1/3, ISO 594-1/2). Tests included leak testing, tensile force, trackability, tip stiffness, and valve performance (penetration/drag force). All results met predetermined acceptance criteria.
Technological Characteristics
Intravascular infusion catheter; 4-5 French; 45-135 cm length. Materials evaluated per ISO 10993-1. Features distal valve for flow control and 0.035" guide wire compatibility. Radiopaque marker bands. Sterilized via Ethylene Oxide (ISO 11135-1). Conforms to ISO 10555-1/3 for intravascular catheters and ISO 594-1/2 for Luer fittings. Mechanical testing per ASTM F2096, F1929, F640, and D4169.
Indications for Use
Indicated for administration of therapeutic solutions into the peripheral vasculature of patients.
Regulatory Classification
Identification
A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.
Predicate Devices
Fountain Infusion Catheter and Occluding Guide Wire (K974067)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 7, 2020
Merit Medical Systems, Inc Spriha Pandey Regulatory Affairs Specialist I 1600 West Merit Parkway South Jordan, Utah 84095
Re: K193082
Trade/Device Name: Fountain ValveTip Infusion Catheter Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: July 9, 2020 Received: July 10, 2020
Dear Spriha Pandey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193082
Device Name Fountain ValveTip Infusion Catheter
Indications for Use (Describe)
The Fountain ValveTip Infusion System is intended to administer infusions of various therapeutic solutions into the peripheral vasculature of a patient.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K193082
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date Prepared:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 316-3845<br>(801) 826-4174<br>Mr. Jesse Nelson<br>August 7, 2020<br>1721504 |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel: | Fountain ValveTip Infusion Catheter<br>Infusion Catheter<br>Catheter, Continuous Flush<br>II<br>KRA<br>870.1210<br>Cardiovascular |
| Predicate<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | Fountain Infusion Catheter and Occluding<br>Guide Wire<br>Catheter, Continuous Flush<br>K974067<br>Merit Medical Systems, Inc. |
| Device Description | The Fountain ValveTip Infusion Catheter is intended to administer infusions of various therapeutic solutions into peripheral vasculature of a patient. The Fountain ValveTip Infusion Catheter has infusion holes at the distal end. These infusion holes allow for passage of thrombolytic fluid. The infusion segment of the catheter is indicated by two radiopaque marker bands. These marker bands can provide the user the means for positioning the infusion length at a desired location with the blood vessel.<br><br>The Fountain ValveTip Infusion Catheter uses a valve at the distal tip to seal the catheter to facilitate flow through the catheter and spray distribution through the infusion holes. The valve allows the catheter to track over a 0.035" guide wire to facilitate placement of the catheter. The Fountain ValveTip Catheter is available in 4 and 5 French including usable lengths from 45 cm to 135 cm and infusion segments length from 5cm to 50 cm. The usable length and Infusion lengths are the same as the predicate device configuration and will have same functional infusing characteristics.<br><br>This device is provided with the following Merit device which is already marketed in the US (Class II Exempt):<br>• AccessPLUS™ Hemostasis Valve | |
| Indications for Use | The Fountain ValveTip Infusion System is intended to administer infusions of various therapeutic solutions into the peripheral vasculature of a patient.<br><br><i>Note: There is no change in the Indications for Use Statement from the predicate to the subject device.</i> | |
| Comparison to Predicate Device | The subject device is substantially equivalent to the predicate device based on identical indications for use statement, and same basic performance and safety profile, principle of operation, fundamental design principles, materials and manufacturing technology. The primary reason for submitting this 510(k) is the addition of a valve on the distal tip of the catheter predicate device under 510(k), K974067 and eliminate the use of the occluding wire. | |
{4}------------------------------------------------
{5}------------------------------------------------
No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices. Performance testing of the subject Fountain ValveTip Infusion Catheter was conducted based on the risk analysis and based on the requirements of the following international standard:
- ISO 11135-1:2014, Sterilization Of Health-care Products Ethylene . Oxide - Requirements For The Development, Validation And Routine Control Of A Sterilization Process For Medical Devices
- ISO 10993-1:2009. Biological Evaluation of Medical Devices Part ● 1: Evaluation and Testing within a risk management process, and FDA quidance Required Biocompatibility Training and Toxicology Profiles for Evaluation of Medical Devices, May 1, 1995
- ISO 10993-3:2014, Biological Evaluation of Medical Devices -● Part 3: Tests for Genotoxicity, Carcinogenicity and Reproductive Toxicitv
- ISO 10993-4:2017. Biological evaluation of medical devices Part . 4: Selection of tests for interaction with blood
- . ISO 10993-5:2009, Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- . ISO 10993-7:2008, Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals
## Performance Data
.
ta
- ISO 10993-10:2010, Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization
- . ISO 10993-11:2017. Biological evaluation of medical devices -Part 11: Tests for systemic toxicity
- . ISO 10993-12, 2012 Biological Evaluation of Medical Devices -Part 12: Sample preparation and reference materials
- BS EN ISO 11607-1:2006 Packaging for Terminally Sterilized . Medical Devices. Part 1: Requirements for materials, sterile barrier systems, and packaging systems.
- ASTM F 2096-11 Standard Test Method for Detecting Gross Leaks ● in Medical Packaging by Internal Pressurization (Bubble Test)
- . ASTM F756:2017 Standard practice for Assessment of Hemolytic Properties of Materials
- . ANSI AAMI ST72:2011, Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing
- USP 41,NF 36 (USP), GP <151>, Pyrogen Test (2018)
- ASTM F 1929-15 Standard Test Method for Detecting Seal Leaks ● in porous Medical Packaging by Dye Penetration
- . ASTM D 4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems
- ASTM F640-12 Standard Test Methods for determining ● Radiopacity for medical use
Product Specific Standards:
- . ISO 10555-1:2013(E) - Intravascular Catheters - Sterile And Single-Use Catheters - Part 1: General requirements
{6}------------------------------------------------
- ISO 10555-3:2013 Intravascular catheters Sterile and single-● use catheters - Part 3: Central venous catheters
- ISO 594-1:1996(E) Conical Fittings with 6% (Luer) Taper for . Syringes, Needles and Certain Other Medical Equipment - Part 1:General Requirements
- ISO 594-2:1998(E) Conical Fittings with 6% (Luer) Taper for ● Syringes, Needles and Certain Other Medical Equipment - Part 2: Lock Fittings
## Biocompatibility Testing
The biocompatibility evaluation for the Fountain ValveTip Infusion Catheter was conducted in accordance with the ISO 10993-1 and FDA guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (CDRH's biocompatibility as recognized by FDA). The battery of testing included the following tests:
- Cytotoxicity
- Sensitization ●
- Irritation ●
- Acute Systemic Toxicity ●
- Pyrogenicity ●
- Genotoxicity ●
- Hemolysis ●
- Thrombogenicity
#### Safety & Performance Tests cont.
- Complement Activation ●
- Subacute/Subchronic Toxicity ●
- . Implantation
The Fountain ValveTip Infusion Catheter is considered as an externally communicating device with circulating blood contact for a prolonged 24 hours - 30 days.
## Performance Testing-Bench
## Bench Testing (Safety & Performance Tests)
- French size
- Usable length ●
- Surface ●
- Compatible Guide Wire ●
- Valve- shall not damage quide wire ●
- The guidewire shall not damage the valve ●
- Valve penetration and drag force ●
- Valve leak test ●
- Kink test
●
{7}------------------------------------------------
● .
- ISO-594-2, Gauge test ● ISO-594-2, Liquid leakage test ● ISO-594-2, Air leakage test ● ISO-594-2, Separation force ● ISO-594-2, Unscrewing Torque ● ISO-594-2, Ease of assembly ● ● ISO-594-2, Resisting to overriding ISO-594-2, Stress Cracking ● Design Validation Simulated use testing ● Penetration force ● Drag force . ● Trackability Tip stiffness . Leakage out of tip . The results of the testing demonstrated that the subject Fountain ValveTip Infusion Catheter met the predetermined acceptance criteria. Based on the indications for use, design, and performance testing, the Summary of Fountain ValveTip Infusion Catheter meets the requirements that are Substantial considered essential for its intended use and is substantially equivalent Equivalence to the predicate device, the Fountain Infusion Catheter and Occluding Guide Wire, K974067. manufactured by Merit Medical Systems, Inc.
Radio-detectability
Tensile force, tip attachment test
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.