← Product Code [KNS](/productcode/KNS) · K193039

# Single Use Balloon Dilator V (with knife) (K193039)

_Olympus Medical Systems Corp. · KNS · Dec 17, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K193039

## Device Facts

- **Applicant:** Olympus Medical Systems Corp.
- **Product Code:** [KNS](/productcode/KNS.md)
- **Decision Date:** Dec 17, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

These instruments have been designed to be used with an Olympus endoscope for papillotomy and for dilating the major papilla to retrieve biliary stones.

## Device Story

Single Use Balloon Dilator V (with knife) is an endoscopic accessory used for papillotomy and biliary stone retrieval. Device consists of a balloon and a papillotomy knife. Operated by a physician during endoscopic procedures; device is inserted through a compatible Olympus endoscope. Principle of operation: high-frequency current applied to a cutting wire for tissue incision; inflation device attached to a balloon port injects fluid to dilate the papilla. V-marking on the insertion portion allows for visual confirmation of insertion length relative to the endoscope biopsy valve. Hook on the handle assists in manipulation. Device facilitates biliary stone retrieval by dilating the papilla and performing sphincterotomy, potentially reducing the need for more invasive surgical interventions.

## Clinical Evidence

Bench testing only. Performance data included electrical safety (ANSI/AAMI ES 60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), sterilization/shelf-life (ASTM F1980-16), and biocompatibility (ISO 10993-1). Bench performance testing verified dimensions, endoscopic compatibility, knife operation, papillotomy resistance, tensile strength, balloon deflation time, balloon burst strength, balloon fatigue, and package integrity.

## Technological Characteristics

4-lumen catheter with C-channel, integrated balloon, and papillotomy knife. Distal tip diameter 1.5mm; working length 1950mm. Balloon diameters: 8mm, 10mm, 12mm. Maximum pressure 3.5 bar. Pre-curved distal end. Materials evaluated per ISO 10993-1 for limited mucosal contact. Sterilized device. No software or electronic algorithm components.

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- Single Use Balloon Dilator V (with knife) ([K150142](/device/K150142.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 17, 2019

Olympus Medical Systems Corp. Sheri Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610

Re: K193039

> Trade/Device Name: Single Use Balloon Dilator V (with knife) Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: KNS, FGE Dated: October 30, 2019 Received: October 31, 2019

Dear Sheri Musgnung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Daniel G. Walter, Jr. Assistant Director THT3A3: Obesity and Hepatobiliary Devices DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

K193039

Device Name Single Use Balloon Dilator V (with Knife)

Indications for Use (Describe)

These instruments have been designed to be used with an Olympus endoscope for papillotomy and for dilating the major papilla to retrieve biliary stones.

Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly stylized, with a modern and clean font. Underneath the word "OLYMPUS", there is a horizontal yellow bar that spans the length of the word. The overall design is simple and corporate, likely representing the logo of the Olympus Corporation.

October 30, 2019

K193039 Page 1 of 5

# Section 5

# 510(k) Summary

#### 5.1 GENERAL INFORMATION

- 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
- Sheri L. Musgnung Contact Person: Olympus Corporation of the America 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 Fax: 484-896-7128 Email: sheri.musgnung@olympus.com
- Manufacturing site: Aomori Olympus Co., Ltd. 2-248-1 Okkonoki, Kuroishi-shi, Aomori 036-0357, Japan

#### 5.2 DEVICE IDENTIFICATION

| ■ Device Name       | Single Use Balloon Dilator V (with knife)                                             |
|---------------------|---------------------------------------------------------------------------------------|
| ■ Model Name        | BD-VC431Q-1240-20<br>BD-VC431Q-1240-25<br>BD-VC431Q-1240-30                           |
| ■ Common Name       | Single Use Balloon Dilator (with knife)                                               |
| ■ Regulation Number | 876.4300<br>876.5010                                                                  |
| ■ Regulation Name   | Endoscopic electrosurgical unit and accessories.<br>Biliary catheter and accessories. |
| ■ Regulatory Class  | II                                                                                    |

Section 5 510(k) Summary Page 1 of 5

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are evenly spaced and appear to be a sans-serif font. A thin, horizontal, yellow line is located underneath the word. The registered trademark symbol is located in the upper right corner of the image.

- Product Code KNS FGE
■ Classification Panel Gastroenterology/Urology

#### 5.3 PREDICATE DEVICE

#### Predicate device

| Device name                               | 510(k) Submitter              | 510(k) No. |
|-------------------------------------------|-------------------------------|------------|
| Single Use Balloon Dilator V (with knife) | OLYMPUS MEDICAL SYSTEMS CORP. | K150142    |

#### 5.4 DEVICE DESCRIPTION

#### ■ General Description of the subject device

The subject device consists of the balloon and the papillotomy knife for endoscopic sphincterotomy.

The only difference from the predicate device which has been cleared under K150142 is the additional balloon diameter offerings

The intended use of the subject device is identical to the predicate device.

#### Principle of Operation

There are no differences in composition and basic principle compared with the predicate.

The subject device used in combination with compatible Olympus endoscopes, A-Cord (power cable), electrosurgical unit and compatible inflation device and the subject device is inserted endoscopically into the body cavity, and a tissue is cut by applying a high frequency current while the cutting wire is tightened.

In addition, an inflation device is attached to a balloon port, and the inflation fluid is injected by utilizing an inflation device to dilate the papilla.

The insertion portion of this product has a V-marking, and the relative insertion length of this product into the endoscope can be confirmed by the positional relationship between the V-marking and the biopsy valve of the endoscope.

The Hook provided on the handle is for attaching the subject device to an endoscope or an inflation device, which assists manipulation of the subject device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. A thin, horizontal, yellow line is underneath the word. The registered trademark symbol is located to the right of the letter "S".

K193039 Page 3 of 5

#### 5.5 INDICATIONS FOR USE

These instruments have been designed to be used with an Olympus endoscope for papillotomy and for dilating the major papilla to retrieve biliary stones.

## 5.6 COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEIVCE

The Single Use Balloon Dilator V (with knife) has the same technological characteristics and design as the predicate device except for the following new feature:

- । smaller balloon diameters
All other technological characteristics of both the subject and predicate device are identical.

Validation from non-clinical testing demonstrated that these technological features do not raise any new issues of safety or effectiveness of the subject device.

| Comparison Table                                                                |                                                                                             |                                                                                              |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
|                                                                                 | Subject Device                                                                              | Predicate Device                                                                             |
| Item                                                                            | BD-VC431Q-1240-20/25/30                                                                     | BD-VC431Q-1840-20/25/30                                                                      |
| Distal tip diameter (mm)                                                        | ø1.5 mm                                                                                     | ø1.5 mm                                                                                      |
| Tip length (mm)                                                                 | 7 mm                                                                                        | 7 mm                                                                                         |
| Knife length (mm)                                                               | 20 mm / 25 mm / 30 mm                                                                       | 20 mm / 25 mm / 30 mm                                                                        |
| Maximum insertion portion<br>diameter (mm)<br>(before inflating the<br>balloon) | ø 3.65 mm                                                                                   | ø 3.65 mm                                                                                    |
| Working length (mm)                                                             | 1950 mm                                                                                     | 1950 mm                                                                                      |
| Balloon diameter (mm (Fr))<br>/ Pressure (bar (atm))                            | ø 8 mm(24Fr) / 0.5bar(0.5atm)<br>ø 10mm(30Fr) / 2bar(2atm)<br>ø 12mm(36Fr) / 3.5bar(3.5atm) | ø 12 mm(36Fr) / 0.5bar(0.5atm)<br>ø 15mm(45Fr) / 2bar(2atm)<br>ø 18mm(54Fr) / 3.5bar(3.5atm) |
| Balloon length (mm)                                                             | 40mm                                                                                        | 40mm                                                                                         |
| Maximum pressure (bar<br>(atm))                                                 | 3.5 bar (3.5 atm)                                                                           | 3.5 bar (3.5 atm)                                                                            |
| Shape of the distal end                                                         | Pre-curved                                                                                  | Pre-curved                                                                                   |
| Knife wire diameter (mm)                                                        | ø 0.2 mm                                                                                    | ø 0.2 mm                                                                                     |
| Coated portion                                                                  | Provided                                                                                    | Provided                                                                                     |
| Lumen type                                                                      | 4-Lumen with C-channel                                                                      | 4-Lumen with C-channel                                                                       |

Comparison Table

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, dark blue letters. The letters are closely spaced and appear to be a sans-serif font. A thin, horizontal gold bar is located beneath the word "OLYMPUS", adding a subtle visual element to the logo. The registered trademark symbol is located to the right of the letter "S".

K193039 Page 4 of 5

#### 5.7 PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

#### 1) Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Single Use Balloon Dilator V (with knife). The system complies with the ANSVAAMI ES 60601-1:2005/(R)2012 and A1:2012 and IEC 60601-2-18:2009 standards for safety and the IEC 60601-1-2:2014 standards for EMC.

#### 2) Sterilization/Shelf-life testing

Sterilization/shelf-life testing for the Single Use Balloon Dilator V (with knife) was conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile".

Accelerated aging test for the Single Use Balloon Dilator V (with knife) was conducted in accordance with ASTM F1980-16(2016), the standard guide for accelerated aging of sterile barrier systems for medical devices. The three years real-time aging test will be performed to demonstrate longer stability and support the results of the accelerated aging test.

#### 3) Biocompatibility testing

Biocompatibility testing for the Single Use Balloon Dilator V (with knife) were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff. Use of International Standard ISO 10993-1:2009. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".

#### Contact classification:

The Single Use Balloon Dilator V (with knife) is considered a surface-contacting device in contact with mucosal membranes. The contact duration is limited exposure (i.e. contact is up to 24 hours).

#### 4) Performance testing - Bench

Bench testing for the Single Use Balloon Dilator V (with knife) as listed below was conducted to ensure that the subject device performs as intended and meet design specifications.

Dimensional verification Endoscopic compatibility

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are spaced closely together, creating a strong visual impact. Below the word, there is a horizontal yellow line that spans the width of the word. The overall design is simple and clean, with a focus on the brand name.

Traditional 510(k) Notification Single Use Balloon Dilator V (with knife) BD-VC431Q-1240

Knife operation Papillotomy resistance Tensile strength Balloon deflation time Balloon burst strength Balloon fatigue Package integrity

#### 5) Risk analysis

Risk analysis for the Single Use Balloon Dilator V (with knife) was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.

#### 5.8 CONCLUSIONS

Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate device, the Single Use Balloon Dilator V (with knife) raise no new issue of safety and effectiveness and are substantially equivalent to the predicate device in terms of safety, effectiveness, and performance.

---

**Source:** [https://fda-staging.innolitics.com/device/K193039](https://fda-staging.innolitics.com/device/K193039)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
