← Product Code [OCX](/productcode/OCX) · K193036

# C2 CryoBalloon EndoGrip (K193036)

_Pentax Medical, A Division of Pentax of America, Inc. · OCX · Nov 22, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K193036

## Device Facts

- **Applicant:** Pentax Medical, A Division of Pentax of America, Inc.
- **Product Code:** [OCX](/productcode/OCX.md)
- **Decision Date:** Nov 22, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The C2 CryoBalloon™ EndoGrip is intended to provide access for endoscopic device passage and exchange throughout the endoscopic procedure.

## Device Story

Accessory for therapeutic endoscopes/catheters; anchors catheters relative to endoscope. Comprises clip base, rubber grip, compression spring. Physician inserts catheter through endoscope; positions EndoGrip against biopsy valve via spring-clip mechanism. Allows catheter placement/repositioning without full withdrawal/reinsertion. Used in hospitals, surgery centers, clinics. Benefits patient by facilitating efficient device exchange during endoscopic procedures.

## Clinical Evidence

Bench testing only. Included physical dimensional verification, simulated use, performance, and reliability testing (attachment/detachment, 10 endoscopic delivery cycles, compatibility, dislodgement resistance, kinking resistance), and usability testing. All tests met predetermined acceptance criteria.

## Technological Characteristics

Consists of clip base, rubber grip, and compression spring. Spring-clip mechanism for attachment to proximal biopsy channel port of endoscope. Compatible with 3.7mm accessory channel ID endoscopes or C2 Sidecar (FG-1011). Single-patient use.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- BOA Endoscopic Valve ([K152140](/device/K152140.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.

November 22, 2019

PENTAX Medical, A Division of PENTAX of America, Inc. Gurvinder Nanda Senior Director of Regulatory Affairs 303 Convention Way, Suite 1 Redwood City, California 94063

Re: K193036

Trade/Device Name: C2 CryoBalloon EndoGrip Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OCX Dated: October 30, 2019 Received: October 31, 2019

Dear Gurvinder Nanda:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193036

Device Name C2 CryoBalloon™ EndoGrip

Indications for Use (Describe)

The C2 CryoBalloon™ EndoGrip is intended to provide access for endoscopic device passage and exchange throughout the endoscopic procedure.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### PENTAX Medical C2 CryoBalloon™ EndoGrip (K193036)

Image /page/3/Picture/1 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold letters, and it is positioned above the word "MEDICAL", which is in gray, bold letters. There is a gray line separating the two words.

# 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92.

#### I. SUBMITTER

PENTAX Medical 303 Convention Way, Suite 1 Redwood City, CA 94063

Phone: 650-316-8601 Fax: 650-316-8601

Gurvinder Singh Nanda, Ph.D. Contact Person: Senior Director of Regulatory Affairs

Date Prepared: October 30, 2019

#### II. DEVICE

Name of Device: C2 CryoBalloon™ EndoGrip Common Name: EndoGrip Classification Name: Endoscopic and Accessories Classification Code: 21 CFR §876.1500 Regulatory Class: Class II Product Code: OCX

#### III. PREDICATE DEVICE

BOA Endoscopic Valve (K152140)

This predicate device has not been subject to a design-related recall.

#### IV. DEVICE DESCRIPTION

The C2 CryoBalloon™ EndoGrip is an accessory product that is used in conjunction with therapeutic endoscopes and therapeutic catheters (the C2 CryoBalloon™ Catheter of the C2 CryoBalloon™ Ablation System). The C2 CryoBalloon™ EndoGrip consists of a Clip Base, Rubber Grip, and Compression Spring. This is supplied non-sterile. The EndoGrip is used to anchor catheters in place relative to the scope. In a clinical application, once the endoscope is delivered into the gastric anatomy and positioned at its

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif font. Below it is a gray line, and below that is the word "MEDICAL" in gray, bold, sans-serif font. 

desired location the physician may insert the catheter through the endoscope. When the catheter is located properly in its axial position, the EndoGrip is positioned against the biopsy valve by depressing the clip mechanism and sitting the catheter on the rubber grips. The revised design allows for placing the EndoGrip after the CryoBalloon catheter is in place in the scope and for repositioning without having to completely withdraw and reinsert the catheter. The C2 CryoBalloon™ EndoGrip is designed for one-time, continuous application use (single patient) in conjunction with a therapeutic endoscope (3.7mm accessory channel ID) or a C2 Sidecar (FG -1011), which serves as an external channel for therapeutic devices when a therapeutic endoscope is unavailable for use.

#### V. INDICATIONS FOR USE

The C2 CryoBalloon™ EndoGrip is intended to provide access for endoscopic device passage and exchange throughout the endoscopic procedure.

### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject C2 CrvoBalloon™ EndoGrip has similar technological characteristics to the legally marketed predicate C2 BOA Endoscopic Valve (K152140). The subject and predicate devices are based on the following same technological elements:

- Attached against the proximal biopsy channel port of an endoscopic device to ● provide endoscopic device passage and exchange
- . Used in hospitals, sub-acute care institutions, surgery centers, or doctor's office where endoscopic procedures may be performed
- . Used in single patient (disposable) undergoing endoscopic procedures

The following technological differences exist between the modified device and the predicate device:

The modified device is secured through the use of a spring-clip mechanism, and the predicate device is secured through the use of a twist-Tuohy tightening mechanism.

#### PERFORMANCE DATA VII.

The performance testing for the C2 CryoBalloon™ EndoGrip included:

- 1. Physical characteristic evaluations for dimensional verification
- 2. Simulated use, performance, and reliability testing were performed to verify the device can be attached to and detached from endoscope, withstand 10

{5}------------------------------------------------

### PENTAX Medical C2 CryoBalloon™ EndoGrip (K193036)

Image /page/5/Picture/1 description: The image contains the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif font. A gray line is underneath the word "PENTAX". Below the gray line, the word "MEDICAL" is in gray, bold, sans-serif font.

endoscopic delivery cycles, compatible with endoscopic devices, resist dislodgement forces, and resist kinking.

- 3. Usability testing for validation of user needs
All the tests passed and met the predetermined acceptance criteria.

# VIII. CONCLUSION

The subject C2 CryoBalloon™ EndoGrip has the same clinical attributes, technological characteristics and expected performance as the legally marketed predicate, C2 BOA Endoscopic valve (K152140). The performance data demonstrate that the subject C2 CryoBalloon™ EndoGrip performs as intended in the specified use conditions and performs comparably to the legally marketed predicate device that is currently marketed for the same intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K193036](https://fda-staging.innolitics.com/device/K193036)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com