Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint, Susceptibility System with Lefamulin in the dilution range of 0.008-16ug/mL
Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint, Susceptibility System with Lefamulin in the dilution range of 0.008-16ug/mL
Applicant
Thermo Fisher Scientific
Product Code
JWY · Microbiology
Decision Date
Jan 8, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Indications for Use
The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint, Susceptibility System is an in vitro diagnostic product for clinical use. It is intended for the determination of susceptibility to antimicrobial agents. The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint, Susceptibility System is indicated for the determination of susceptibility to antimicrobial agents by broth microdilution methods. The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint, Susceptibility System with Lefamulin in the dilution range of 0.008-16 µg/mL is indicated for the determination of susceptibility to Lefamulin against Haemophilus influenzae and Streptococcus pneumoniae.
Device Story
Sensititre system performs quantitative antimicrobial susceptibility testing (AST) via broth microdilution. Multi-well plastic microtiter plates contain dried, stabilized antimicrobial agents (Lefamulin) in double dilutions. Plates are inoculated with bacterial suspensions using the Sensititre AIM autoinoculator, sealed, and incubated for 20-24 hours. Growth is detected via visual inspection using the VIZION digital viewing device or automated fluorescence detection using the OptiRead system. OptiRead monitors surface enzyme activity by detecting fluorescent products generated from fluorogenic substrates. The MIC is the lowest concentration inhibiting bacterial growth. Results assist clinicians in determining appropriate antimicrobial therapy. System is intended for laboratory use by trained personnel.
Clinical Evidence
Bench testing only. Performance evaluated using 200 clinical and 50 challenge S. pneumoniae isolates, and 393 clinical and 50 challenge H. influenzae isolates. Compared against CLSI broth microdilution reference method. For S. pneumoniae (VIZION), EA was 99.2% and CA 98.4%. For S. pneumoniae (OptiRead), EA was 98.8% and CA 98.8%. For H. influenzae (VIZION), EA was 98.9% and CA 94.8%. Adjusted major error rate for H. influenzae was 0.2%. Adjusted very major error rate for S. pneumoniae was 0%.
Indicated for the determination of antimicrobial susceptibility of Haemophilus influenzae and Streptococcus pneumoniae to the antibiotic Lefamulin (dilution range 0.008-16 µg/mL) using broth microdilution methods in a clinical laboratory setting.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Predicate Devices
Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint, Susceptibility System (K190275)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K193024
B Applicant
Thermo Fisher Scientific
C Proprietary and Established Names
Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint, Susceptibility System with Lefamulin in the dilution range of 0.008-16 µg/mL
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JWY, LRG, LTT | Class II | 21 CFR 866.1640 - Antimicrobial Susceptibility Test Powder | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain a substantial equivalence determination for the addition of Lefamulin to the Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint, Susceptibility System with Lefamulin in the dilution range of 0.008 – 16 µg/mL
B Measurand:
Lefamulin in the dilution range of 0.008 – 16 µg/mL
C Type of Test:
Quantitative Antimicrobial Susceptibility Test (AST), growth-based detection
## III Intended Use/Indications for Use:
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K193024 - Page 2 of 11
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The Sensititre 20 - 24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System is an in vitro diagnostic product for clinical susceptibility testing of H. influenzae, Streptococcus pneumoniae, and Streptococcus spp.
This 510(k) is for Lefamulin in the dilution range of 0.008 - 16 ug/mL for testing Streptococcus spp. and Haemophilus influenzae on the Sensititre 20 - 24 hour MIC panel.
Lefamulin has been shown to be active both clinically and in vitro against the following organisms according to the FDA drug label:
Streptococcus pneumoniae
Haemophilus influenzae
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
The performance of lefamulin with Haemophilus influenzae and Streptococcus pneumoniae was evaluated using the AIM autoinoculator only. The use of an alternative inoculation system when testing lefamulin has not been evaluated.
The reading of lefamulin results for S. pneumoniae was performed using the AutoReader (OptiRead) and VIZION reading methods. The reading of lefamulin results for Haemophilus influenzae was only performed by VIZION method. The use of alternative reading methods when testing lefamulin has not been evaluated.
The ability of the Sensititre system to detect resistance to Lefamulin in the following species is unknown because non-susceptible strains were not available at the time of comparative testing: S. pneumoniae and H. influenzae. Isolates yielding Lefamulin MIC results suggestive of a non-susceptible interpretive category should be submitted to a reference laboratory.
Due to the lack of an intermediate category and observed trending towards higher dilutions for Lefamulin, testing of H. influenzae have resulted in major discrepancies for isolates that are otherwise within essential agreement for the reference method. Testing should be repeated using an alternative testing/reference method prior to reporting results for H. influenzae when the Sensititre with the VIZION is 4 µg/mL.
D Special Instrument Requirements:
Sensititre AIM for device inoculation
{2}
Sensititre VIZION or OptiRead for plate reading (results for Haemophilus influenzae are interpreted using VIZION only)
## IV Device/System Characteristics:
### A Device Description:
Sensititre MIC Susceptibility MIC panels are multi-well microtiter plates, dosed with dried, stabilized antimicrobials. It is a miniaturized version of the classic broth dilution method and can provide both qualitative and quantitative susceptibility results. After inoculation, plates are sealed with an adhesive seal, incubated at 34 – 36 °C for 20 – 24 hours and examined for bacterial growth.
Antimicrobial susceptibility test results can be determined by reading growth using the digital viewing device (VIZION) or automatically on an autoreader (OptiRead) using fluorescence.
### B Principle of Operation:
The Sensititre 20 - 24 hour Haemophilus influenzae/Streptococcus pneumoniae system includes multi-well plastic microtiter plates that contain doubled dilutions of antibacterial agents. Each plate includes antimicrobial agents at appropriate dilutions. Results can be read by the digital device, VIZION, or by use of an automated reader, OptiRead (Streptococcus spp. only).
The VIZION allows the panel image to be displayed on a touch screen directly from a video camera and allows the user to visually determine MIC results. The Sensititre OptiRead utilizes fluorescence technology to read the microbroth dilution plates after 20 to 24 hours incubation. The technology involves the detection of bacterial growth by monitoring the activity of specific surface enzymes produced by the test organism. Growth is determined by generating a fluorescent product from a fluorogenic substrate. The substrate is prepared by conjugating a fluorescent compound to the specific enzyme substrates with a bond which prevents fluorescence. The enzymatic action of the bacterial surface enzymes on the substrate cleaves the bond releasing fluorescence. The amount of fluorescence detected is directly related to bacterial growth. The MIC is determined by observing the lowest dilution of antimicrobial agent that inhibits growth of the organism. The substrate can be added to the inoculum broth which is dispensed into the test plate at the same time as the test organism, or, the plates can be prepared with the substrate already added to each micro-well.
### V Substantial Equivalence Information:
#### A Predicate Device Name(s):
Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Omadacycline in the dilution range of 0.008 - 32 ug/mL
#### B Predicate 510(k) Number(s):
K183324
K193024 - Page 3 of 11
{3}
# C Comparison with Predicate(s):
Table 1: Comparison with the Predicate
| Device & Predicate Device(s): | Device K193024 | Predicate K183324 |
| --- | --- | --- |
| Device Trade Name | Sensititre Haemophilus/Streptococcus pneumoniae (HP) MIC Plates with Lefamulin | Sensititre Haemophilus/Streptococcus pneumoniae (HP) MIC Plates with Omadacycline |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae, and Streptococcus species. | Same |
| Test Panel | 96 well plate is dosed with selected antimicrobial agents and substrate for the fluorescent reads, then dried. The bacterial suspension in the appropriate broth is used to rehydrate the plate. | Same |
| Test Organism | Haemophilus influenzae, Streptococcus pneumoniae, and Streptococcus spp. | Same |
| Reading Methods for Streptococcus spp. | Results can be read using the following methods:1) Automatically with the OptiRead (fluorescent substrate technology)2) On the VIZION (digital viewing device) | Same |
| Reading Methods for Haemophilus | Results can be read using the VIZION (digital viewing device) only | Same |
| Incubation | 20-24 hours, 35 ± 1°C | Same |
| Inoculation Media | CAMHBT + LHB (Streptococcus spp.), HTM broth (H. influenzae) | Same |
| General Device Characteristic Differences | | |
| Antimicrobial Agent | Lefamulin | Omadacycline |
K193024 - Page 4 of 11
{4}
K193024 - Page 5 of 11
| Concentration Range | 0.008 – 16 μg/mL | 0.008 – 32 μg/mL |
| --- | --- | --- |
## VI Standards/Guidance Documents Referenced:
Guidance for Industry and FDA - Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems – August 28, 2009.
CLSI. Performance Standards for Antimicrobial Susceptibility Testing. 29th ed. CLSI supplement M100. Wayne, PA: Clinical and Laboratory Standards Institute; 2019.
CLSI. Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically. 11th ed. CLSI standard M07. Wayne, PA: Clinical and Laboratory Standards Institute; 2018.
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
1. **Precision/Reproducibility:**
A reproducibility study was performed at four sites using a panel comprised of 10 isolates of *Streptococcus* spp.: *S. pneumoniae* (three isolates), *S. pyogenes* (one isolate), *S. agalactiae* (two isolates), *S. anginosus* (two isolates), *S. mitis* (one isolate), and *S. salivarius* (one isolate) and a panel comprised of 10 strains of *H. influenzae*. Of the *Streptococcus* spp., five non-indicated species (7 isolates) were tested but this was considered acceptable due to similarities among the species and all isolates provided MIC values that are on-scale. All isolates were tested in triplicate over three days with each read method (i.e., VIZION and OptiRead for *Streptococcus* spp., and VIZION only for *H. influenzae*). The Sensititre AIM inoculator was used for plate inoculation. The mode MIC value was determined and the reproducibility was calculated based on MIC values falling within ±1 dilution of the mode MIC value. Reproducibility was greater than 95% for both read methods for *Streptococcus* spp. and VIZION for the *H. influenzae* panel; results were considered to be acceptable.
2. **Linearity:**
Not applicable
3. **Analytical Specificity/Interference:**
Not applicable
4. **Assay Reportable Range:**
Not applicable
5. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):**
{5}
Quality controls strains recommended by CLSI were tested with Lefamulin at four sites. The QC organisms tested were S. pneumoniae ATCC 49619 and H. influenzae ATCC 49247. The QC strains were tested a minimum of 20 times per site and read using the OptiRead (S. pneumoniae only) and VIZION (both S. pneumoniae and Haemophilus influenzae). The results demonstrate that the Sensititre Lefamulin MIC quality controls results were within the recommended range $\geq 95\%$ of the time (Table 2).
Table 2. Quality Control Results for Sensititre 18 – 24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Lefamulin with the VIZION and OptiRead Methods.
| QC Organism | Lefamulin Range (μg/mL) | Concentration (μg/mL) | Reference | Read Method | |
| --- | --- | --- | --- | --- | --- |
| | | | | VIZION | OptiRead* |
| S. pneumoniae ATCC 49619 | 0.06 – 0.5 | 0.03 | 0 | 0 | 0 |
| | | 0.06 | 12 | 4 | 6 |
| | | 0.12 | 40 | 37 | 50 |
| | | 0.25 | 36 | 50 | 35 |
| | | 0.5 | 0 | 0 | 0 |
| | | 1 | 0 | 0 | 0 |
| | | | | | |
| H. influenzae ATCC 49247 | 0.5 - 2 | 0.25 | 1 | 0 | N/A |
| | | 0.5 | 12 | 10 | N/A |
| | | 1 | 66 | 43 | N/A |
| | | 2 | 0 | 34 | N/A |
| | | 4 | 0 | 0 | N/A |
*QC results for H. influenzae were interpreted only using the VIZION.
Inoculum Density. Inoculum density checks were performed a sufficient number of times; all organism suspensions were in the acceptable range.
Purity checks. Purity checks were performed on all isolates following plate inoculation. Only results from pure cultures were evaluated.
Growth Failures. All isolates tested showed growth in the Sensititre panels.
6. Detection Limit:
Not applicable
7. Assay Cut-Off:
Not applicable
B Comparison Studies:
1. Method Comparison with Predicate Device:
K193024 - Page 6 of 11
{6}
Results obtained with Sensititre 20 - 24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Lefamulin were compared to results obtained with the CLSI broth microdilution reference panel. Drug dilutions were prepared using fresh Mueller Hinton broth with lysed horse blood (LHB) for Streptococcus spp. and Haemophilus Test Medium (HTM) broth for H. influenzae isolates as indicated in CLSI M07, 11th ed. Clinical testing was performed at four sites, three of which were in the U.S. A total of 200 clinical S. pneumoniae isolates were tested and read by both VIZION and OptiRead. In addition, a total of 393 H. influenzae isolates were tested of which results were read using the digital viewing device (VIZION) only. All of the clinical isolates tested were contemporary isolates (i.e., tested within 6 months of isolates). During the course of the clinical trial, all Sensititre dried MIC panels were inoculated using the Sensititre Autoinoculator (AIM) and the same panel was read on both the VIZION and the OptiRead for S. pneumoniae isolates in a blinded manner. The sponsor added the following limitations to the device labeling to reflect these inoculation and read methods:
The performance of lefamulin with Haemophilus influenzae and Streptococcus pneumoniae was evaluated using the AIM autoinoculator only. The use of an alternative inoculation system when testing lefamulin has not been evaluated.
The reading of lefamulin results for S. pneumoniae was performed using the AutoReader (OptiRead) and VIZION reading methods. The reading of lefamulin results for Haemophilus influenzae was only performed by VIZION method. The use of alternative reading methods when testing lefamulin has not been evaluated.
A total of 100 challenge isolates were tested at a single site. Species tested included S. pneumoniae (50 isolates) and H. influenzae (50 isolates).
Results for S. pneumoniae and H. influenzae were evaluated for essential agreement (EA) and category agreement (CA) when compared to the reference method. Breakpoints for both are noted on the FDA-Recognized Susceptibility Test Interpretive Criteria Website (STIC). Essential Agreement (EA) is when the Sensititre results agree exactly or within one doubling dilution of the reference broth microdilution results. Category Agreement (CA) is when the Sensititre result interpretation agrees exactly with the reference broth microdilution result interpretation.
The results from clinical and challenge testing determined with the VIZION demonstrated a combined EA of 99.2% and CA of 98.4% for S. pneumoniae (Table 3). There was one major error (1/245 =0.4%) and 3 very major errors (3/5 = 60%). Due to the lack of an intermediate interpretive category, our analysis takes into consideration MIC results that are within essential agreement and the error rate is adjusted accordingly. For lefamulin, the MIC values for all four of these categorical errors were within essential agreement of the reference method. The adjusted error rate is 0% for major errors and 0% for very major errors. When interpreted by the OptiRead (Table 4), results demonstrated a combined EA of 98.8% and CA of 98.8%. No major errors were identified, however, the same three very major errors were identified when interpreted by the OptiRead. Similarly as with the VIZION, because all three very major errors were within essential agreement, the adjusted very major error rate is 0% and therefore, acceptable.
K193024 - Page 7 of 11
{7}
The results from clinical and challenge testing determined with the VIZION demonstrated a combined EA of 98.9% and CA of 94.8% for H. influenzae (Table 3). There were 23 (23/429=5.4%) major errors and no very major errors identified. Due to the lack of an intermediate or resistant interpretive category, our analysis takes into consideration MIC results that are within essential agreement and the error rate is adjusted accordingly. For lefamulin, many of the observed major errors were otherwise within essential agreement of the reference method. The adjusted major error rate was 0.2% (1/429). However, given the high number of major errors and trending in favor of higher dilutions when compared to the reference method (see Table 5), the following was included in the labeling to instruct users to test with an alternative method if a lefamulin MIC result of 4 µg/mL is obtained:
Due to the lack of an intermediate category and observed trending towards higher dilutions for Lefamulin, testing of H. influenzae have resulted in major discrepancies for isolates that are otherwise within essential agreement for the reference method. Testing should be repeated using an alternative testing/reference method prior to reporting results for H. influenzae when the Sensititre with the VIZION is 4 µg/mL.
Table 3. Performance of S. pneumoniae and H. influenzae Clinical and Challenge Isolates, Read Using VIZION
| | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. NS | No. S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| S. pneumoniae | | | | | | | | | | | | | |
| Clinical | 200 | 198 | 99.0 | 200 | 198 | 99.0 | 198 | 99.0 | 4 | 196 | NA | 0 | 2 |
| Challenge | 50 | 50 | 100 | 50 | 50 | 100 | 48 | 96.0 | 1 | 49 | NA | 1 | 1 |
| Total | 250 | 248 | 99.2 | 250 | 248 | 99.2 | 246 | 98.4 | 5 | 245 | NA | 1 | 3¹ |
| H. influenzae | | | | | | | | | | | | | |
| Clinical | 393 | 388 | 98.7 | 386 | 381 | 98.7 | 372 | 97.7 | 13 | 380 | NA | 21 | 0 |
| Challenge | 50 | 50 | 100 | 50 | 50 | 100 | 48 | 96.0 | 1 | 49 | NA | 2 | 0 |
| Total | 443 | 438 | 98.9 | 436 | 431 | 98.9 | 420 | 94.8 | 14 | 429 | NA | 23² | 0 |
¹Adjusted very major error rate is 0%.
²Adjusted major error rate is 0.2% (1/429).
Table 4. Performance of S. pneumoniae Clinical and Challenge Isolates, Read Using OptiRead
| | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. NS | No. S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| S. pneumoniae | | | | | | | | | | | | | |
| Clinical | 200 | 197 | 98.5 | 200 | 197 | 98.5 | 198 | 99.0 | 4 | 196 | NA | 0 | 2 |
| Challenge | 50 | 50 | 100 | 50 | 50 | 100 | 49 | 98.0 | 1 | 49 | NA | 0 | 1 |
| Total | 250 | 247 | 98.8 | 250 | 247 | 98.8 | 247 | 98.8 | 5 | 245 | NA | 0 | 3 |
¹Adjusted very major error rate is 0%.
EA – Essential agreement
CA – Category agreement
Eval – Evaluable isolates
NS – Non-susceptible isolates
NA – Not Applicable due to lack of intermediate category
maj – Major errors
vmj – Very major errors
min – Minor errors
S – Susceptible
To address the testing and reporting of non-indicated species the following comment was added to the Precautions section of the device labeling:
The safety and efficacy of antimicrobial drugs for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-
K193024 - Page 8 of 11
{8}
controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labelling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.
Due to the lack of intermediate or resistant categories and availability of insufficient number of non-susceptible isolates tested in the comparison study, the sponsor also included the following to instruct users to use an alternative method of testing when non-susceptible isolates of S. pneumoniae and H. influenzae are encountered:
The ability of the Sensititre system to detect resistance to Lefamulin in the following species is unknown because non-susceptible strains were not available at the time of comparative testing: S. pneumoniae and H. influenzae. Isolates yielding Lefamulin MIC results suggestive of a non-susceptible interpretive category should be submitted to a reference laboratory.
## MIC Trending
An analysis of trending was calculated using the combined clinical and challenge data for each organism group. This trending calculation takes into account MIC values that are determined to be one or more doubling dilutions lower or higher compared to the reference method irrespective of whether the device MIC values are on-scale or not. Trending results are shown in Table 5. Results for S. pneumoniae were stratified by both reading methods and H. influenzae was evaluated using the VIZION only to determine if particular trends were observed. The acceptable percent difference between higher and lower dilution readings is <30%. Based on the results observed, a trend was identified for H. influenzae when interpreted by the VIZION which trended towards higher dilutions when compared to the reference method. Given this, the following was included in the package insert as a footnote:
Lefamulin MIC values tended to be in exact agreement or at least one doubling dilution higher when testing H. influenzae with the VIZION reading method compared to the CLSI reference broth microdilution.
Table 5: Lefamulin Trending Analysis for S. pneumoniae and H. influenzae, Read by VIZION and OptiRead (S. pneumoniae only)
| Organism (Read Method) | Total Evaluable for Trending | ≥ 1 Dilution lower No. (%) | Exact No. (%) | ≥ 1 Dilution Higher No. (%) | Percent Difference (CI) | Trending Noted |
| --- | --- | --- | --- | --- | --- | --- |
| S. pneumoniae (VIZION) | 250 | 67 (26.8) | 157 (62.8) | 26 (10.4) | -16.4 (-23.0 to -9.7) | No |
| S. pneumoniae (OptiRead) | 250 | 80 (32.0) | 154 (61.6) | 16 (6.4) | -25.6 (-32.1 to -19.0) | No |
| H. influenzae (VIZION) | 438 | 10 (2.3) | 191 (43.6) | 237 (54.1) | 51.8 (46.8 – 56.6) | Yes |
K193024 - Page 9 of 11
{9}
2. Matrix Comparison:
Not applicable
# C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
# D Clinical Cut-Off:
Not applicable
# E Expected Values/Reference Range:
The FDA-identified susceptibility interpretive criteria for lefamulin are as listed in Table 6.
Table 6: FDA-Recognized Interpretive Criteria ${}^{a}$ for Lefamulin $\left( {\mu \mathrm{g}/\mathrm{{mL}}}\right)$
| | Susceptible (S) | Intermediate (I) | Resistant (R) |
| --- | --- | --- | --- |
| Streptococcus pneumoniae | ≤0.5 | - | - |
| Haemophilus influenzae | ≤2 | - | - |
a FDA STIC Webpage
# VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
# IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage
(https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm). The protocol outlined the specific procedures and acceptance criteria that ThermoFisher intends to use to evaluate the Sensititre 20-24 hour Haemophilus influenzae/Streptococcus
K193024 - Page 10 of 11
{10}
pneumoniae MIC or Breakpoint, Susceptibility System with Lefamulin in the dilution range of 0.008-16 ug/ml when revised breakpoints for lefamulin are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, ThermoFisher will update the lefamulin device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
K193024 - Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.