← Product Code [LNH](/productcode/LNH) · K193021

# Vantage Orian 1.5T, MRT-1550, V6.0 (K193021)

_Canon Medical Systems Corporation · LNH · Jun 3, 2020 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K193021

## Device Facts

- **Applicant:** Canon Medical Systems Corporation
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Jun 3, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Vantage Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: ·Proton density (PD) (also called hydrogen density) - ·Spin-lattice relaxation time (T1) - ·Spin-spin relaxation time (T2) - ·Flow dynamics - ·Chemical Shift Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

## Device Story

Vantage Orian 1.5T is a magnetic resonance imaging (MRI) system; utilizes 1.5T magnet with Pianissimo noise reduction technology. Inputs: proton density, T1/T2 relaxation times, flow dynamics, chemical shift data. Transforms inputs via gradient coils and pulse sequences into cross-sectional images. Features Compressed SPEEDER (compressed sensing + parallel imaging) for accelerated scanning and T2 mapping using pre-contrast pulses. Operated by trained clinicians in clinical settings. Output: diagnostic images for physician interpretation to aid diagnosis. Benefits: shorter scan times or higher resolution; improved clinical utility. System includes XGO gradient options and Windows 10 software environment.

## Clinical Evidence

No clinical trials; bench testing and volunteer imaging only. Performance of Compressed SPEEDER assessed via phantoms and volunteers for SNR, unfolding error artifacts, and phase encode direction performance. Representative images reviewed by American Board Certified Radiologists for image degradation, diagnostic performance, lesion conspicuity, and clinical utility. T2 mapping evaluated via phantom and volunteer data. All testing met predetermined acceptance criteria.

## Technological Characteristics

1.5T MRI system; 1.4m short magnet; Pianissimo noise reduction; max FOV 55x50cm. Standard or XGO gradient systems. Connectivity: Windows 10 OS. Standards: IEC 60601-2-33, IEC 62304, NEMA MS 1-5. Software: Moderate level of concern.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Vantage Orian 1.5T, MRT-1550, V4.5 ([K182282](/device/K182282.md))

## Reference Devices

- Vantage Galan 3T, MRT-3020, V6.0 ([K192506](/device/K192506.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Canon Medical Systems Corporation % Ms. Janine Reyes Manager, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive TUSTIN CA 92780

### Re: K193021

Trade/Device Name: Vantage Orian 1.5T, MRT-1550, V6.0 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: April 27, 2020 Received: April 28, 2020

Dear Ms. Reyes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

June 3, 2020

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

#### K193021

Device Name Vantage Orian 1.5T, MRT-1550, V6.0

#### Indications for Use (Describe)

Vantage Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

·Proton density (PD) (also called hydrogen density)

- ·Spin-lattice relaxation time (T1)
- ·Spin-spin relaxation time (T2)
- ·Flow dynamics
- ·Chemical Shift

Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

Type of Use (Select one or both, as applicable)

| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D)    </div> |
|-----------------------------------------------------------------------------|
| <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)    </div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/3 description: The image shows the logo for Canon Medical Systems USA, Inc. The word "Canon" is in red, and the rest of the text is in black. The text is aligned to the left and is in a sans-serif font. The logo is simple and professional.

**Made For life**

## 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92

## 1. CLASSIFICATION and DEVICE NAME

| Classification Name:    | Magnetic Resonance Diagnostic Device |
|-------------------------|--------------------------------------|
| Regulation Number:      | 90-LNH (Per 21 CFR § 892.1000)       |
| Trade Proprietary Name: | Vantage Orian 1.5T, MRT-1550, V6.0   |
| Model Number:           | MRT-1550                             |

## 2. SUBMITTER'S NAME

Canon Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550

## 3. OFFICIAL CORRESPONDENT

Naofumi Watanabe Senior Manager, Regulatory Affairs and Vigilance Canon Medical Systems Corporation

## 4. CONTACT PERSON, U.S. AGENT and ADDRESS

### Contact Person

Janine F. Reyes Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 669-7853 Fax: (714) 730-1310 E-mail: jfreyes@us.medical.canon

## Official Correspondent/U.S. Agent

Paul Biggins Senior Director, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-7808 Fax: (714) 730-1310 E-mail: pbiggins@us.medical.canon

{4}------------------------------------------------

## DICAL SYSTEMS USA, INC.

*Made For life*

#### 5. MANUFACTURING SITE

Canon Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan

#### ESTABLISHMENT REGISTRATION 6.

9614698

#### 7. DATE PREPARED

October 29th, 2019

#### 8. DEVICE NAME

Vantage Oran 1.5T, MRT-1550, V6.0

#### 9. TRADE NAME

Vantage Oran 1.5T, MRT-1550, V6.0

#### 10. CLASSIFICATION NAME

Magnetic Resonance Diagnostic Device (MRDD)

#### 11. CLASSIFICATION PANEL

Radiology

#### 12. DEVICE CLASSIFICATION

Class II (per 21 CFR 892.1000, Magnetic Resonance Diagnostic Device)

#### 13. PRODUCT CODE

90-LNH

#### 14. PREDICATE DEVICE

Predicate Device: Vantage Orian 1.5T, MRT-1550, V4.5 (K182282 Reference Device: Vantage Galan 3T, MRT-3020, V6.0 (K192506)

| TABLE NO. 1. Predicate Device |                                 |                                 |                                 |
|-------------------------------|---------------------------------|---------------------------------|---------------------------------|
| System                        | Subject Device                  | Predicate Device                | Reference Device                |
| Marketed<br>By                | Canon Medical Systems USA, Inc. | Canon Medical Systems USA, Inc. | Canon Medical Systems USA, Inc. |
| 510(k)<br>Number              | This Submission                 | K182282                         | K192506                         |
| Clearance<br>Date             |                                 | October 19th, 2018              | March 3rd, 2020                 |

TABLE No. 1: Predicate Device

2441 Michelle Drive, Tustin, CA 92780 PHONE: 800-421-1968 https://us.medical.canon

{5}------------------------------------------------

## EDICAL SYSTEMS USA, INC.

*Made For life*

## 15. REASON FOR SUBMISSION

Modification of a cleared device

## 16. SUBMISSION TYPE

Traditional 510(k) Premarket Notification

## 17. DEVICE DESCRIPTION

The Vantage Orian (Model MRT-1550) is a 1.5 Tesla Magnetic Resonance Imaging (MRI) System. The Vantage Orian uses 1.4 m short and 3.8 tons light weight magnet. It includes the Pianissimo™ technology (scan noise reduction technology). The design of the gradient coil and the WB coil of the Vantage Orian 1.5T provides the maximum field of view of 55 x 50 cm. The Model MRT-1550/AC, AD, AG, AH includes the standard gradient system and Model MRT-1550/AK, AL, AO, AP includes the XGO gradient system.

This system is based upon the technology and materials of previously marketed Canon Medical Systems MRI systems and is intended to acquire and display cross-sectional transaxial, coronal, sagittal, and oblique images of anatomic structures of the head or body. The Vantage Orian MRI System is comparable to the current 1.5T Vantage Orian MRI System (K182282), cleared October 19th, 2018 with the following modifications.

## 18. SUMMARY OF CHANGE(S)

This submission is to report the following software changes:

- Compressed SPEEDER (Compressed Sensing)*:
	- . This application allows accelerated fast scan of the brain, cervical spine, thoracic spine, lumbar spine, shoulder, knee, foot, ankle, elbow and wrist. Compressed SPEEDER combines parallel imaging by using multi-sensitivity map and compressed sensing.
	- . Compressed SPEEDER feature allows acceleration factors for shorter scan times or higher resolution in FSE2D imaging based upon the principle of Compressed Sensing in combination with parallel imaging.
- T2 Map Using Pre-Contrast Pulses: ECG gating or peripheral pulse gating is used in scanning with FFE2D sequences, and different Pre-contrast pulses are used to obtain multiple TEeff images.
- Windows 10: Windows 10 has been applied to V6.0 Software.

*Note: Compressed SPEEDER should not be used for contrast-agent-enhanced imaging.

{6}------------------------------------------------

## L SYSTEMS USA, INC.

*Made For life*

## 19. SAFETY PARAMETERS

| Item                                                                | Subject Device:<br>Vantage Orian 1.5T,<br>MRT-1550, V6.0                                          | Predicate Device:<br>Vantage Orian 1.5T,<br>MRT-1550, V4.5                                | Notes |
|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-------|
| Static field strength                                               | 1.5T                                                                                              | 1.5T                                                                                      | Same  |
| Operational Modes                                                   | Normal and 1st Operating Mode                                                                     | Normal and 1st Operating Mode                                                             | Same  |
| i. Safety parameter display                                         | SAR, dB/dt                                                                                        | SAR, dB/dt                                                                                | Same  |
| ii. Operating mode access<br>requirements                           | Allows screen access to 1st level<br>operating mode                                               | Allows screen access to 1st level<br>operating mode                                       | Same  |
| Maximum SAR                                                         | 4W/kg for whole body (1st operating<br>mode specified in IEC 60601-2-33:<br>2010+A1:2013+A2:2015) | 4W/kg for whole body (1st<br>operating mode specified in IEC<br>60601-2-33: 2010+A1:2013) | Same  |
| Maximum dB/dt                                                       | 1st operating mode specified in IEC<br>60601-2-33: 2010+A1:2013+A2:2015                           | 1st operating mode specified in<br>IEC 60601-2-33: 2010+A1:2013                           | Same  |
| Potential emergency<br>condition and means<br>provided for shutdown | Shutdown by Emergency Ramp Down<br>Unit for collision hazard for<br>ferromagnetic objects         | Shutdown by Emergency Ramp<br>Down Unit for collision hazard for<br>ferromagnetic objects | Same  |

### 20. IMAGING PERFORMANCE PARAMETERS

No change from the previous predicate submission, K182282.

### 21. INDICATIONS FOR USE

Vantage Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces crosssectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

- Proton density (PD) (also called hydrogen density) ●
- . Spin-lattice relaxation time (T1)
- Spin-spin relaxation time (T2)
- Flow dynamics
- Chemical Shift

Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

{7}------------------------------------------------

# ICAL SYSTEMS USA, INC.

*Made For life*

## 22. SUMMARY OF DESIGN CONTROL ACTIVITIES

Risk Management activities for new software functionalities and pulse sequences are included in this submission. The test methods used are the same as those submitted in the previously cleared submission of the predicate device, Vantage Orian 1.5T, MRT-1550, V4.5 (K182282). A declaration of conformity with design controls is included in this submission.

#### 23. SAFETY

This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards.

This device is based upon the same technologies, materials and software as the predicate device. Risk activities were conducted in concurrence with established medical device development standards and guidance. Additionally, testing was done in accordance with applicable recognized consensus standards published by the International Electrotechnical Commission (IEC) for medical devices and the National Electrical Manufacturers Association (NEMA):

#### LIST OF APPLICABLE STANDARDS

- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012
- IEC60601-1-2 (2014)
- IEC60601-1-6 (2010), Amd.1 (2013) ●
- IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015)
- IEC60825-1 (2007)
- IEC62304 (2006), Amd.1 (2015) .
- IEC62366 (2007), Amd.1 (2014) ●
- NEMA MS 1 (2008) ●
- NEMA MS 2 (2008) ●
- NEMA MS 3 (2008)
- NEMA MS 4 (2010)
- NEMA MS 5 (2010) ●

#### 24. TESTING

Risk analysis and verification/validation testing conducted through bench testing are included in this submission which demonstrate that the system requirements have been met. Additionally, image quality testing was completed which demonstrated that the subject device meets predetermined acceptance criteria.

MR image quality metrics were performed, utilizing phantoms and volunteer images, to assess Compressed SPEEDER with regards to SNR, unfolding error artifacts and performance in all the phase encode directions. It was concluded that Compressed SPEEDER met all acceptance criteria.

Additionally, representative images, reviewed by American Board Certified Radiologists, were obtained using the subject device. Reviewers provided detailed assessments of image degradation, diagnostic performance, lesion conspicuity, and clinical utility. It was confirmed that Compressed SPEEDER images were of diagnostic quality.

T2 Map Using Pre-Contrast Pulses was evaluated utilizing phantom and volunteer images. It was concluded that T2 maps can be generated using the data acquired using pre-contrast pulses.

PHONE: 800-421-1968 2441 Michelle Drive, Tustin, CA 92780 https://us.medical.canon

{8}------------------------------------------------

Image /page/8/Picture/3 description: The image shows the logo for Canon Medical Systems USA, Inc. The word "Canon" is in red, and the words "CANON MEDICAL SYSTEMS USA, INC." are in black. The logo is simple and clean, and it is likely used on the company's website and marketing materials. The logo is a recognizable symbol of the company.

*Made For life*

Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.

#### 25. SUBSTANTIAL EQUIVALENCE

Canon Medical Systems Corporation believes that the Vantage Orian 1.5T, MRT-1550, V6.0, Magnetic Resonance Imaging (MRI) System is substantially equivalent to the previously cleared predicate device, Vantage Orian 1.5T, MRT-1550, V4.5, referenced in this submission. Canon Medical Systems Corporation believes that the changes incorporated into the Vantage Orian 1.5T, MRT-1550, V6.0 are substantially equivalent to the previously cleared predicate device.

#### 26. CONCLUSION

The modifications incorporated into the Vantage Orian 1.5T, MRT-1550, V6.0 do not change the indications for use or the intended use of the device. Based upon bench testing, volunteer clinical imaging, successful completion of software validation and application of risk management and design controls, it is concluded that the subject device is safe and effective for its intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K193021](https://fda-staging.innolitics.com/device/K193021)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).