The VenaPro Vascular Therapy System is intended to be an easy to use portable system, prescribed by a physician, for use in the home or clinical setting to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (stimulating muscle contractions). This device can be used to: - Aid in the prevention of DVT; - Enhance blood circulation; - Diminish post-operative pain and swelling; - Reduce wound healing time; - Aid in the treatment and healing of stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs. The unit can also be used as an aid in the prophylaxis for DVT by persons expecting to be stationary for long periods of time.
Device Story
VenaPro Vascular Therapy System is a portable, battery-powered, prescription-only device for home or clinical use. It consists of a control unit with an air pump and solenoid valve, permanently attached to a single-use inflatable PVC cuff encased in polyester fabric. The device operates by applying cyclic, intermittent pneumatic pressure (50 mmHg) to the lower limb to stimulate blood flow and muscle contractions. A 60-second cycle includes inflation followed by a rest period for deflation. The user operates the device via a single tactile switch; status is indicated by a tri-color LED. Internal software and a pressure switch monitor cuff pressure. Audible and visual alarms alert the user to leaks, low pressure, or low battery. The device aids in DVT prevention, circulation enhancement, and wound healing. It is intended for patients needing venous return support, including those stationary for long periods.
Clinical Evidence
Bench testing only. No clinical data provided. Verification included electrical safety (ANSI AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), home healthcare environment requirements (IEC 60601-1-11), and battery safety (IEC 62133-2). Software verification was conducted per FDA guidance.
Technological Characteristics
Portable, battery-powered (3.7V) pneumatic compression device. Components: PVC air bladder, polyester blend fabric cover, micro-pump, solenoid valve, microprocessor controller. Dimensions: 131x66x37mm; Weight: 0.23kg. Operation: Cyclic pneumatic pressure (50 mmHg, 60s cycle). Connectivity: Includes port for battery charger/AC adapter and unused port for future data reporting. Software: Microprocessor-controlled. Sterilization: Clean, non-sterile.
Indications for Use
Indicated for patients requiring venous return stimulation to prevent DVT, enhance circulation, reduce post-op pain/swelling, accelerate wound healing, or treat venous/arterial/diabetic ulcers, stasis dermatitis, chronic venous insufficiency, and lower limb edema. Also indicated for DVT prophylaxis in stationary individuals. Contraindicated for patients with suspected/active/untreated DVT, ischemic vascular disease, severe arteriosclerosis, pulmonary edema, severe congestive heart failure, thrombophlebitis, active infection, or where increased venous/lymphatic return is undesirable. Not for use on legs with vein ligation, gangrene, dermatitis, open wounds, recent skin grafts, massive edema, extreme deformity, neuropathy, or insensitive extremities.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 4, 2019
Innovamed Health, LLC % Bill Dai Manager Jkh Usa, LLC 20505 Valley Blvd. Suite 108 Walnut, California 91789
Re: K193020
Trade/Device Name: VenaPro Vascular Therapy System Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible limb sleeve Regulatory Class: Class II Product Code: JOW Dated: October 26, 2019 Received: October 30, 2019
Dear Bill Dai:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193020
Device Name VenaPro Vascular Therapy System
#### Indications for Use (Describe)
The VenaPro Vascular Therapy System is intended to be an easy to use portable system, prescribed by a physician, for use in the home or clinical setting to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (stimulating muscle contractions). This device can be used to:
- · Aid in the prevention of DVT;
- · Enhance blood circulation;
- · Diminish post-operative pain and swelling;
- · Reduce wound healing time;
- · Aid in the treatment and healing of stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs.
The unit can also be used as an aid in the prophylaxis for DVT by persons expecting to be stationary for long periods of time.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 1. Submitter's Information
Submitter: Innovamed Health, LLC Address: 270 West 500 North North Salt Lake, UT 84054 Contact Person: Luke Zeutzius Email: Luke@InnovamedHealth.com Date of Preparation: 09/30/2019
## 2. Subject Device
Trade/Device Name: VenaPro Vascular Therapy System Common Name: Compressible Limb Sleeve Device Regulation Medical Specialty: Neurology Review Panel: Cardiovascular Product Code: JOW Regulation Number: 21 CFR 870.5800 Device Class: II Use: Prescription
## 3. Predicate device
Trade Name: VenaPro Vascular Therapy System 510(k) Number: K133274 Clearance Date: February 12, 2014 Submitter: Innovamed Health, LLC
## 4. Description of Subject Device
The VenaPro Vascular Therapy System is a lightweight, portable, rechargeable battery powered prescriptive device. It is intended to be used in the home or clinical setting by or under the direction of a medical professional to help stimulate blood flow as an aid in the prevention of deep yein thrombosis (DVT).
The system utilizes pneumatically controlled, single chamber cuffs actuated by an electronically controlled air pump unit and solenoid valve. All pump, battery and control components are protectively housed in a plastic case that is permanently attached to a single use inflatable cuff. A single tactile touch control switch, tri-color LED for ON, LOW BATTERY, CHARGING and CHARGE COMPLETED indication, and a blue LED (for indicating a leak or low pressure alarm) provides for user interface. There is also a port for connecting the battery charger/AC adapter plug, and a currently unused port for future use in usage data reporting.
The leg wrap (cuff) component consists of a Polyvinyl Chloride (PVC) air bladder encased inside a soft, non-woven medical fabric (a Polyester blended medical fabric) or equal , which is adhered to the PVC air bladder. The units are supplied clean, non-sterile, packaged in pairs (1 left and 1 right side).
In operation, the user simply turns the power ON via the single button I/O control switch. A single user "cuff" containing air bladders is permanently connected to the unit. The control unit then fills the cuff to a pre-determined pressure (50 mmHg). Cuff pressure is monitored by an internal pressure switch and system software. Once the pressure reaches the proper level, the pump is turned OFF for an approximately 50 second "rest" period, and the cuff deflates to ambient pressure through a valve inside
{4}------------------------------------------------
the plastic case. After the "rest" period, the cycle repeats until the unit is turned off.
## 5. Indications for Use
### Prescription Use:
The VenaPro Vascular Therapy System is intended to be an easy to use portable system, prescribed by a physician, for use in the home or clinical setting to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (stimulating muscle contractions). This device can be used to:
- Aid in the prevention of DVT; ●
- Enhance blood circulation;
- Diminish post-operative pain and swelling:
- Reduce wound healing time; ●
- Aid in the treatment and healing of stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ● ulcers, chronic venous insufficiency and reduction of edema in the lower limbs.
The unit can also be used as an aid in the prophylaxis for DVT by persons expecting to be stationary for long periods of time.
### 6. Summary of Substantial Equivalence
The following comparison Table 1 summarizes the comparison between the modified subject device and the original predicate device, indicating the intended use and technical characteristics of the subject device are substantially equivalent to those of the predicate device.
| | Subject Device | Predicate Device | Equivalence |
|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | N/A | K133274 | |
| Submitter | Innovamed Health, LLC | Innovamed Health, LLC | Identical |
| Device Name/Model | VenaPro Vascular Therapy System | VenaPro Vascular Therapy System | Identical |
| Intended Use | Prescription Use: | Prescription Use: | Identical |
| | The VenaPro Vascular Therapy<br>System is intended to be an easy to use<br>portable system, prescribed by a<br>physician, for use in the home or<br>clinical setting to help prevent the<br>onset of DVT in patients by<br>stimulating blood flow in the<br>extremities (stimulating muscle<br>contractions). This device can be used<br>to:<br><br>• Aid in the prevention of<br>DVT;<br>• Enhance blood circulation;<br>• Diminish post-operative pain<br>and swelling;<br>• Reduce wound healing time;<br>• Aid in the treatment and<br>healing of stasis dermatitis,<br>venous stasis ulcers, arterial | The VenaPro Vascular Therapy<br>System is intended to be an easy to use<br>portable system, prescribed by a<br>physician, for use in the home or<br>clinical setting to help prevent the<br>onset of DVT in patients by<br>stimulating blood flow in the<br>extremities (stimulating muscle<br>contractions). This device can be used<br>to:<br><br>• Aid in the prevention of<br>DVT;<br>• Enhance blood circulation;<br>• Diminish post-operative pain<br>and swelling;<br>• Reduce wound healing time;<br>• Aid in the treatment and<br>healing of stasis dermatitis,<br>venous stasis ulcers, arterial | |
| | | | |
| | and diabetic leg ulcers,<br>chronic venous insufficiency<br>and reduction of edema in<br>the lower limbs.<br><br>The unit can also be used as an aid in<br>the prophylaxis for DVT by persons<br>expecting to be stationary for long<br>periods of time. | and diabetic leg ulcers,<br>chronic venous insufficiency<br>and reduction of edema in<br>the lower limbs.<br><br>The unit can also be used as an aid in<br>the prophylaxis for DVT by persons<br>expecting to be stationary for long<br>periods of time. | |
| Prescription or OTC | Prescription | Prescription | Identical |
| Power Source(s) | Rechargeable battery | Rechargeable battery | Identical |
| Battery Specifications | 3.7V rechargeable battery | 7.4V rechargeable battery | The predicate device uses the<br>7.4V rechargeable battery. In<br>comparison, the subject device<br>switches to the 3.7V<br>rechargeable battery. The<br>voltage difference of batteries<br>used does not change the<br>product performance or<br>parameters, which will not raise<br>any new issue of the safety or<br>effectiveness. |
| Battery Charge | Takes approximately 2.5 hours (from<br>depleted state). | Takes approximately 1.5 hours (from<br>depleted state). | The difference of charging time<br>does not change the product<br>performance or parameters,<br>which will not raise any new<br>issue of the safety or<br>effectiveness. |
| Power Supply | Input: 100 - 240 Vac, 50 - 60 Hz,<br>Output: 5 Vdc @ 1 Amp) | Input: 100 - 240 Vac, 50 - 60 Hz,<br>Output: 10 Vdc @ 1.1 Amp) | The predicate device uses the<br>7.4V rechargeable battery that is<br>charged by a 10V charger. In<br>comparison, the subject device<br>switches to the 3.7V<br>rechargeable battery, which is<br>charged by a 5V charger. The<br>voltage difference of power<br>supply used does not change the<br>product performance or<br>parameters, which will not raise<br>any new issue of the safety or<br>effectiveness. |
| Internal rechargeable<br>batteries | Yes | Yes | Identical |
| Compliance with<br>Voluntary Standards? | Yes | Yes | Identical |
| Electrical Safety<br>Mechanical Safety<br>Chemical Safety<br>Thermal Safety<br>Radiation Safety? | Yes | Yes | Identical |
| Functions and design | Aids venous return by using cyclic,<br>intermittent, pneumatic pressure<br>application (inflation followed by<br>deflation) to compress the lower<br>limb(s). | Aids venous return by using cyclic,<br>intermittent, pneumatic pressure<br>application (inflation followed by<br>deflation) to compress the lower<br>limb(s). | Identical |
| Contraindication(s) | The VenaPro Vascular Therapy<br>System MUST NOT be used to treat<br>the following conditions: Persons with<br>suspected, active or untreated: deep<br>vein thrombosis, ischemic vascular<br>disease, severe arteriosclerosis,<br>pulmonary edema | The VenaPro Vascular Therapy<br>System MUST NOT be used to treat<br>the following conditions: Persons with<br>suspected, active or untreated: deep<br>vein thrombosis, ischemic vascular<br>disease, severe arteriosclerosis,<br>pulmonary edema | Identical |
Table 1. Comparison between the subject device and the predicate device
{5}------------------------------------------------
{6}------------------------------------------------
| | severe congestive heart failure, thrombophlebitis, or an active infection.<br>On the legs where cuffs would interfere with the following conditions: vein ligation, gangrene, dermatitis, open wounds, a recent skin graft, massive edema or extreme deformity of the leg.<br>On any neuropathy.<br>On extremities that are insensitive to pain.<br>Where increased venous or lymphatic return is undesirable. | severe congestive heart failure, thrombophlebitis, or an active infection.<br>On the legs where cuffs would interfere with the following conditions: vein ligation, gangrene, dermatitis, open wounds, a recent skin graft, massive edema or extreme deformity of the leg.<br>On any neuropathy.<br>On extremities that are insensitive to pain.<br>Where increased venous or lymphatic return is undesirable. | | | | | |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|----------------------|--|
| Target Population /<br>Intended Users | Patients who need venous return. | Patients who need venous return. | Identical | | | | |
| Where Used | Home, Hospital, Surgery Center, Altitude travel, areas of limited mobility | Home, Hospital, Surgery Center, Altitude travel, areas of limited mobility | Identical | | | | |
| Application | Non-invasive / external | Non-invasive / external | Identical | | | | |
| Portability | Portable, ambulant | Portable, ambulant | Identical | | | | |
| Basis of operation | Aids venous return by using cyclic, intermittent, pneumatic pressure application (inflation followed by deflation) to compress the lower limb(s). | Aids venous return by using cyclic, intermittent, pneumatic pressure application (inflation followed by deflation) to compress the lower limb(s). | Identical | | | | |
| Anatomical Site /<br>Location of treatment<br>application | Lower limb(s) (Calf) | Lower limb(s) (Calf) | Identical | | | | |
| System management | Electronic, microprocessor controlled | Electronic, microprocessor controlled | Identical | | | | |
| Pressure Source | Micro pump controlled by electronic processor | Micro pump controlled by electronic processor | Identical | | | | |
| Operating Modes | Default mode one | Default mode one | Identical | | | | |
| Working Pressure | Default mode is preset at 50 mmHg | Default mode is preset at 50 mmHg | Identical | | | | |
| Cycle Time | 60 seconds | 60 seconds | Identical | | | | |
| System diagnostics | Audible and visual alarms prompt recognition of system faults | Audible and visual alarms prompt recognition of system faults | Identical | | | | |
| Modes | 1 Modality | 1 Modality | Identical | | | | |
| Air delivery from pump<br>to cuff bladder | Via flexible plastic (PVC) tube(s) connected directly to the air bladder. | Via flexible plastic (PVC) tube(s) connected directly to the air bladder. | Identical | | | | |
| Sterility | Clean / non-sterile | Clean / non-sterile | Identical | | | | |
| Leg cuff usage | Single Patient Use | Single Patient Use | Identical | | | | |
| Material Used | | Single bladder PVC chambers encased in a covering of soft, nonlatex, non-woven medical fabric (a Polyester blend) or equivalent medical material for increased patient comfort and | for increased patient comfort and<br>biocompatibility compliance. | Single bladder PVC chambers encased in a covering of soft, nonlatex, non-woven medical fabric (a Polyester blend) or equivalent medical material for increased patient comfort and | for increased patient comfort and<br>biocompatibility compliance. | Identical or similar | |
| | | Glue | | | | | |
| Fastening between the<br>plastic case and the<br>fabric wrap | Snap and screw | Glue | The plastic case and the fabric<br>wrap of the predicate device are<br>glued together, while the plastic<br>case and the fabric wrap of the<br>subject device are snapped and<br>screwed together. The patients<br>can not reach the snap and | | | | |
{7}------------------------------------------------
| | | | screws that are hidden in the<br>subject device. Therefore, this<br>difference will not raise any new |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| | | | issue of safety and effectiveness. |
| Biocompatibility | Biocompatible | Biocompatible | Identical |
| Software | Moderate | Moderate | Identical |
| Dimensions | 131x66x37mm | 131x66x37mm | Identical of similar |
| Weight Approx.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.