The DANNIK Disposable Monopolar Laparoscopic Instrument have application in a variety of minimally invasive procedures to facilitate grasping, mobilization, dissection and transection of tissue.
Device Story
Disposable monopolar laparoscopic instrument for minimally invasive surgery; used by surgeons to grasp, mobilize, dissect, and transect tissue. Device features handle and 360-degree rotation knob (ABS thermoplastic), 5mm diameter stainless steel shaft with high-density polyethylene insulation, and RF post for electrosurgical coagulation when connected to an external generator and ground electrode. Operated manually by surgeon in clinical/OR settings. Provides mechanical tissue manipulation and electrical coagulation; benefits include precise tissue handling and hemostasis during laparoscopic procedures.
Clinical Evidence
No clinical data. Bench testing only, including performance and functionality evaluations: thermal effects on tissue, connection strength, clamping performance, cable connection, mechanical open/close performance, hardness, and corrosion resistance.
Technological Characteristics
5mm diameter, 33cm length. Materials: ABS thermoplastic (handle/knob), medical-grade stainless steel (shaft), high-density polyethylene (insulation). Energy: Monopolar RF electrosurgery via external generator. Sterilization: Ethylene Oxide (ETO) per ISO 11135-1. Biocompatibility: ISO 10993 compliant. Manual operation.
Indications for Use
Indicated for patients undergoing minimally invasive surgical procedures requiring tissue grasping, mobilization, dissection, and transection.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 4, 2020
DANNIK Olga Haberland Regulatory Compliance 941 W Morse Blvd., Suite 100 Winter Park, Florida 32789
Re: K193019
Trade/Device Name: DANNIK Disposable Monopolar Laparoscopic Instrument Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: February 18, 2020 Received: February 24, 2020
Dear Olga Haberland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193019
Device Name
DANNIK Disposable Monopolar Laparoscopic Instrument
Indications for Use (Describe)
The DANNIK Disposable Monopolar Laparoscopic Instrument have application in a variety of minimally invasive procedures to facilitate grasping, mobilization, dissection and transection of tissue.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
| X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) /Summary
1. Contact Information DANNIK 941 West Morse Blvd. Suite #100 Winter Park, Florida 32789 Phone: (407) 745-1698 Olga Haberland, Regulatory Compliance January 30, 2020
- 2. Device Name
- Trade Name DANNIK Disposable Monopolar Laparoscopic . Instrument
- Common Name: Disposable Monopolar Laparoscopic Instrument .
- Classification Name : Electrosurgical, Cutting & Coagulation Device . and Accessories
- Classification: Class II General and Plastic Surgery Devices GEI, 21 . CFR878.4400
Performance Standards: Devices are manufactured according to Good Manufacturing Practices (G.M.P.), Association for Advancement of Medical Instrumentation (A.AM.I.) and American Society for Testing and Materials (A.S.T.M.) requirements and applicable Harmonized Standards ISO 13485.
- 3. Substantially Equivalent Device
- Legally Marketed (unmodified Devices): .
- Ethicon Endo-Surgery Endopath Endoscopic Instrument FDA 510K . (K984240)
005 1
{4}------------------------------------------------
# K193019
- 4. Device Description
The DANNIK Disposable Monopolar Laparoscopic Instrument consists of Handle which activates the instruments jaws and scissor blades, Rotator knob that provides 360 degrees rotation , both handle and rotation knob are made of ABS thermoplastic, and a Shaft made from Medical grade Stainless Steel and insulated with high-density polyethylene PE. 5mm in diameter with a working length of 33cm. RF Post for electrosurgery when attached to an approved electrosurgical generator to provide coagulation of tissue when used with appropriate ground electrode.
This device is packaged and sterilized for single use only. Do NOT re-use, reprocess or resterilize. Discard after use with care.
Biocompatibility Conforms to ISO 10993 Prescription Only Yes Sterilized by ETO ( Ethelyn Oxide )
- 5. Intended Use
The DANNIK Disposable Monopolar Laparoscopic Instrument have application in a variety of minimally invasive procedures to facilitate grasping, mobilization, dissection and transection of tissue.
- 6. Technological Characteristics of the Subject Device Compared to the Predicate Device
Comparison of Technological Characteristics: DANNIK Disposable Monopolar Laparoscopic Instrument is substantially equivalent to the predicate device.
There are no new technologies being added to this device from the predicate, in terms of finished device functions. The device has the same intended use and application as the predicate device.
{5}------------------------------------------------
| Device | Dannik Disposable Monopolar<br>Laparoscopic Instrument | Ethicon Endo-Surgery Endopath<br>Endoscopic Instrument FDA 510K<br>(K984240) |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | The DANNIK Disposable<br>Monopolar Laparoscopic<br>Instrument have application in a<br>variety of minimally invasive<br>procedures to facilitate grasping,<br>mobilization, dissection and<br>transection of tissue. | Same |
| Product Picture | Image: Dannik Disposable Monopolar Laparoscopic Instrument | Image: Ethicon Endo-Surgery Endopath Endoscopic Instrument |
| Device Description | The DANNIK Disposable Monopolar<br>Laparoscopic Instrument consists of<br>Handle which activates the<br>instruments jaws and scissor blades,<br>Rotator knob that provides 360<br>degrees rotation to shaft; both<br>handle and rotation knob are made<br>of ABS thermoplastic, and a Shaft<br>5mm in diameter with a working<br>length of 33cm made from Medical<br>grade Stainless Steel and insulated<br>with high-density polyethylene PE.<br>5mm in diameter with a working<br>length of 33cm. RF Post for<br>electrosurgery when attached to an<br>approved electrosurgical generator<br>to provide coagulation of tissue<br>when used with appropriate ground<br>electrode. | Same/Equivalent<br>"The Instruments have a rotating<br>insulated shaft with a diameter of<br>3mm, 5mm or 10mm. The Rotation<br>knob located on the handle rotates<br>the shaft 260 degrees in either<br>direction. The ring handles are<br>compressed and released to activate<br>the instrument jaws or scissor<br>blades. Each of the curved scissors<br>and dissectors has a monopolar<br>cautery connector that extends<br>from the top of the handle. The<br>connector is used for electrosurgery<br>when properly attached to a<br>standard cautery cable and proper<br>generator |
| Classification | Class II General and Plastic Surgery<br>Devices GEI, 21 CFR878.4400 | Same |
| Diameter | 5mm x 33cm | Same ( also offers additional diameters) |
| Materials | Non-Patient Contact:<br>• ABS Thermoplastic-<br>Patient Contact:<br>• Stainless Steel<br>• PE (Polyethylene) | Unknown |
| Sterilization | Ethylene Oxide ( ETO)<br>I.S.O 11135-1 | Unknown |
| Prescription Only | Yes | Yes |
| Biocompatibility | Conforms to ISO 10993 | Unknown |
| | | |
{6}------------------------------------------------
## 7. Non-Clinical Tests
DANNIK Disposable Monopolar Laparoscopic Instrument has been evaluated by Design Engineers, through performance and functionality evaluations that include Thermal effects on Tissue Connection strength, clamping performance , cable connection, open and close performance, Hardness , corrosion resistance and outward.
- 8. Clinical Tests
No Clinical trials performed on the DANNIK Disposable Monopolar Laparoscopic Instrument
- 9. Conclusions
The subject device has equivalent indications for use as the predicate device. There are no new technologies being added to this device from the predicate, in terms of finished device functions. The device has the same intended use and application as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.