icobrain ctp is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" computer or a virtual platform, such as VM ware, and can be used to perform image processing, and communication of computed tomography (CT) perfusion scans of the brain. Data and images are acquired through DICOM-compliant imaging devices. icobrain ctp provides both analysis and communication capabilities for dynamic imaging datasets that are acquired with CT Perfusion imaging protocols. Analysis includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. Results of image processing which include CT perfusion parameter maps generated from a raw CTP scan are exported in the standard DICOM format and may be viewed on existing radiological imaging viewers.
Device Story
icobrain ctp is image processing software for brain CT perfusion (CTP) scans; inputs are DICOM-compliant dynamic imaging datasets. Software performs pre-processing (loading scans, extracting tags), calculates perfusion parameters (CBF, CBV, MTT, Tmax), and identifies Tmax and CBF abnormalities. Outputs include an electronic report (PDF/DICOM) containing abnormality volumes, mismatch volume/ratio, time series data, and perfusion parameter maps. Used in clinical settings by physicians/technicians; results viewed on standard radiological viewers to assist in stroke assessment and clinical decision-making. Benefits include automated quantification of tissue perfusion to aid diagnosis.
Clinical Evidence
Performance validated via digital phantom (simulated tracer kinetic theory) and clinical stroke patient datasets. Accuracy metrics (correlation, p90 absolute difference, mean relative difference) for perfusion parameters (CBV, CBF, MTT) and abnormality volumes (Tmax, CBF) met pre-established acceptance criteria based on literature. Digital phantom correlations >0.90. Reproducibility tested via test-retest with simulated motion. Software verification and validation confirmed design requirements met.
Technological Characteristics
Software-based image processing; runs on off-the-shelf hardware or virtual platforms (e.g., VMware). Inputs/outputs via DICOM. Calculates perfusion parameters (CBF, CBV, MTT, Tmax) using tracer kinetic theory. Standalone software package; no hardware components.
Indications for Use
Indicated for trained professionals, including physicians and medical technicians, to perform image processing, analysis, and communication of brain CT perfusion scans in patients, including stroke patients.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
February 28, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Icometrix NV % Dirk Smeets VP Clinical Applications Kolonel Begaultlaan 1b/12 3012 Leuven BELGIUM
Re: K192962
Trade/Device Name: icobrain-ctp Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: January 24, 2020 Received: January 27, 2020
Dear Dirk Smeets:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192962
Device Name icobrain ctp
#### Indications for Use (Describe)
icobrain ctp is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" computer or a virtual platform, such as VM ware, and can be used to perform image processing, and communication of computed tomography (CT) perfusion scans of the brain. Data and images are acquired through DICOM-compliant imaging devices.
icobrain ctp provides both analysis and communication capabilities for dynamic imaging datasets that are acquired with CT Perfusion imaging protocols. Analysis includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. Results of image processing which include CT perfusion parameter maps generated from a raw CTP scan are exported in the standard DICOM format and may be viewed on existing radiological imaging viewers.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K192962
- 5.1 Submitter
- 5.2 Device
- 5.3 Predicate Device
- 5.4 Device Description
- 5.5 Indications for Use
- 5.6 Comparison with predicate device
- 5.7 Performance testing
### 5. I Submitter
| Name: | icometrix NV |
|-------------------|-----------------------------------------------------|
| Address: | Kolonel Begaultlaan 1b/12<br>3012 Leuven<br>Belgium |
| Contact Person: | Dirk Smeets |
| Telephone number: | +32 16 369 000 |
| Fax Number: | N.A. |
| E-mail: | dirk.smeets@icometrix.com |
| Date Prepared: | 28 Feb 2020 |
{4}------------------------------------------------
## 5.2 Device
| Device Trade Name: | icobrain ctp |
|-----------------------|----------------------------------------|
| Common Name | Medical Image Processing Software |
| Classification Name | System, Image processing, Radiological |
| Number | 892.2050 |
| Product Code: | LLZ |
| Classification Panel: | Radiology |
# Predicate Device 5.3
| Item | Description |
|-----------------------------|-----------------|
| 510(k) Number | K180161 |
| Device Name | Viz CTP |
| Original Applicant | Viz.ai, Inc. |
| Regulation Number | 21 CFR 892.2050 |
| Classification Product Code | LLZ |
| 510k Review Panel | Radiology |
#### 5.4 Device Description
The following flowchart illustrates the overall architecture of icobrain ctp.
Image /page/4/Figure/6 description: This image is a flowchart that shows the steps involved in image processing. The first step is to input images in DICOM format. The next steps are preprocessing, image processing, and output generation (report/images). The final step is to generate a final report in PDF or DICOM format and output images in DICOM format.
{5}------------------------------------------------
The input images are CT perfusion images. During the pre-processing, each scan is loaded from the DICOM format: the image data and relevant dicom tags are extracted. The image processing block calculates the perfusion parameters and the volumes of the Tmax abnormality (defined as tissue with delayed arrival) and the CBF abnormality (defined as tissue with delayed arrival and critically decreased cerebral blood flow). Finally, the computed measurements are summarized into an electronic report. Optionally if requested, Tmax and CBF abnormalities segmentations are overlaid on the input images and image volumes of perfusion parameters maps are sent.
#### Output of icobrain ctp:
#### Report:
- volume of brain tissue with elevated time to maximum residual function (Tmax abnormality)
- volume of brain tissue with elevated time to maximum residual function and decreased cerebral blood flow (CBF abnormality)
- mismatch volume (difference between Tmax abnormality and CBF abnormality volumes)
- ratio between Tmax abnormality volume and CBF abnormality volume
- · the time series of the arterial input function, venous output function and the averaged time density curves in the parenchymal tissue
- · the time points/image volumes of the CTP that were excluded because of artefacts (if any)
- temporal mean of the input CT input images with corresponding perfusion parameter maps (time to maximum residual function, cerebral blood flow, cerebral blood volume, mean transit time) at various slices
#### 5.5 Indications for Use
icobrain ctp is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" computer or a virtual platform, such as VMware, and can be used to perform image processing, analysis, and communication of computed tomography (CT) perfusion scans of the brain. Data and images are acquired through DICOM-compliant imaging devices.
icobrain ctp provides both analysis and communication capabilities for dynamic imaging datasets that are acquired with CT Perfusion imaging protocols. Analysis includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. Results of image processing which include CT perfusion parameter maps generated from a raw CTP scan are exported in the standard DICOM format and may be viewed on existing radiological imaging viewers.
{6}------------------------------------------------
# 5.6 Comparison with predicate device
| | Device to market | Proposed predicate device | Proposed Reference device |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Trade Name | icobrain ctp | Viz CTP | RAPID |
| Common<br>Name | Medical Image Processing Software | Picture archiving and communications system | Picture archiving and communications<br>system |
| 510(k)<br>Number | K192962 | K180161 | K121447 |
| Manufacturer | icometrix NV | Viz.ai, Inc. | iSchemaView, Inc. |
| | Kolonel Begaultlaan Ib / 12<br>3012 Leuven<br>BELGIUM | 855 El Camino Real Suite 13A-252<br>Palo Alto, CA 94301 | 323 Olmsted Road<br>Stanford, CA 94305 |
| Regulation<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | 21 CFR 892.2050 |
| Device<br>Classification<br>Name | System, Image processing,<br>Radiological | System, Image processing, Radiological | System, Image processing, Radiological |
| Product<br>Code | LLZ | LLZ | LLZ |
| Regulatory<br>Class | II | II | II |
| Classification<br>Panel | Radiology | Radiology | Radiology |
| Indications<br>for use | icobrain ctp is an image processing<br>software package to be used by<br>trained professionals, including but<br>not limited to physicians and<br>medical technicians. The software<br>runs on a standard "off-the-shelf"<br>computer or a virtual platform,<br>such as VMware, and can be used<br>to perform image processing,<br>analysis, and communication of<br>computed tomography (CT)<br>perfusion scans of the brain. Data<br>and images are acquired through<br>DICOM-compliant imaging devices.<br><br>icobrain ctp provides both analysis<br>and communication capabilities for<br>dynamic imaging datasets that are<br>acquired with CT Perfusion imaging<br>protocols. Analysis includes<br>calculation of parameters related to<br>tissue flow (perfusion) and tissue<br>blood volume. Results of image<br>processing which include CT<br>perfusion parameter maps<br>generated from a raw CTP scan are<br>exported in the standard DICOM<br>format and may be viewed on<br>existing radiological imaging viewers. | Viz CTP is an image processing software<br>package to be used by trained professionals,<br>including but not limited to physicians and<br>medical technicians. The software runs on a<br>standard "off-the-shelf" computer or a virtual<br>platform, such as VMware, and can be used<br>to perform image processing, analysis, and<br>communication of computed tomography<br>(CT) perfusion scans of the brain. Data and<br>images are acquired through DICOM-<br>compliant imaging devices.<br><br>Viz CTP provides both analysis and<br>communication capabilities for dynamic<br>imaging datasets that are acquired with CT<br>Perfusion imaging protocols. Analysis<br>includes calculation of parameters related to<br>tissue flow (perfusion) and tissue blood<br>volume. Results of image processing which<br>include CT perfusion parameter maps<br>generated from a raw CTP scan are<br>exported in the standard DICOM format<br>and may be viewed on existing radiological<br>imaging viewers. | iSchemaView's RAPID is an image<br>processing software package to be used<br>by trained professionals, including but not<br>limited to physicians and medical<br>technicians. The software runs on a<br>standard "off-the-shelf" computer or a<br>virtual platform, such as VMware, and can<br>be used to perform image viewing,<br>processing and analysis of brain images.<br>Data and images are acquired through<br>DICOM compliant imaging devices.<br><br>iSchemaView's RAPID provides both<br>viewing and analysis capabilities for<br>functional and dynamic imaging datasets<br>acquired with CT Perfusion and MRI<br>including a Diffusion Weighted MRI<br>(DWI) Module and a Dynamic Analysis<br>Module (dynamic contrast enhanced<br>imaging data for MRI and CT). |
| Basic PACS<br>functions | Yes | Yes | Yes |
| Computer<br>platform | Standard "off-the-shelf" computer<br>or a virtual platform. | Same | Same |
| DICOM<br>compliance | Yes | Yes | Yes |
| Functional<br>overview | icobrain ctp is a software package<br>that provides for the quantification<br>and visualization of the perfusion of<br>tissue based on dynamic contrast<br>enhanced CT images. | Same | Automatic analysis for functional and<br>dynamic imaging datasets acquired with<br>CT Perfusion and MRI including a<br>Diffusion Weighted MRI (DWI) Module<br>and a Dynamic Analysis Module (dynamic<br>contrast enhanced imaging data for MRI<br>and CT). |
| Data<br>Acquisition | Acquires medical image data from<br>DICOM compliant imaging devices<br>and modalities. | Same | Same |
| Data/Image<br>Types | Computed Tomography (CT) | Same | CT and MR scanner |
| Acquisition and Modalities Features | | | |
| CT | CT Perfusion | Yes | Yes |
| Computed Parameter Maps | | | |
| Perfusion CT | Cerebral Blood Flow (CBF) | Yes | Yes |
| | Cerebral Blood Volume (CBV) | Yes | Yes |
| | Mean Transit Time (MTT) | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.