K192936 · Sedecal., Sa. · IZL · Nov 15, 2019 · Radiology
Device Facts
Record ID
K192936
Device Name
Soltus 500
Applicant
Sedecal., Sa.
Product Code
IZL · Radiology
Decision Date
Nov 15, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography
Device Story
Mobile digital X-ray system; battery-powered; motorized movement. Inputs: X-ray exposure parameters; patient data via barcode scanner. Operation: Physician/technician positions unit; captures images using Canon flat panel detectors; processes via embedded control computer. Features: Distributed Antenna System (DAS) for optimized Wi-Fi communication with detectors; Enhanced Work Flow (EWF) via secondary embedded computer for RIS/PACS connectivity. Outputs: Diagnostic radiographic images displayed on integrated touch screen; DAP dosimetry data. Used in clinical settings for bedside/point-of-care imaging. Healthcare providers use images for diagnostic decision-making; benefits include improved mobility and workflow efficiency in hospital environments.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and verification of system performance. Digital receptor panels previously cleared by FDA.
Indicated for adult and pediatric patients requiring diagnostic radiographic imaging of the skull, spine, chest, abdomen, and extremities. Contraindicated for mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 15, 2019
Sedecal., SA. % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES, FLORIDA 34114
Re: K192936
Trade/Device Name: Soltus 500 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: Class II Product Code: IZL, MOB Dated: October 17, 2019 Received: October 18, 2019
Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192936
Device Name Soltus 500
#### Indications for Use (Describe)
Intended for use by a qualified/trained doctor or technician on both adult and pediatic subjects for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary: 510(k) Number K192936
Image /page/3/Picture/2 description: The image is a teal-colored logo with the word "SEDECAL" in white letters. To the left of the word is a white icon of a person sitting at a desk. The logo is enclosed in a rounded rectangle with a white border.
SEDECAL SA C/ Pelaya, 9 - 13, Pol. Ind. Río de Janeiro 28110 Algete, Madrid España (Spain) Tel .- +34 91 6280544 Fax.- +34 91 6280574 mlgomez@sedecal.com Date Prepared: July 10, 2019 Contact: Mª Luisa Gómez de Agüero, Quality and Regulatory Manager
- 1) Identification of the Device: Trade/Device Name: Soltus 500 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB.
- Common/Usual Name: Digital Mobile Diagnostic X-Ray System
- 2) Equivalent legally marketed device: K192011 Trade/Device Name: Battery Mobile X-ray Unit PhoeniX Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
- 3) Reference devices: Digital Flat Panel Detectors (FDA 510K Cleared) Trade/Device Name: Digital Radiography CXDI-401C Wireless (K133693), Digital Radiography CXDI-701C Wireless (K131106), Digital Radiography CXDI-801C Wireless (K131106), Digital Radiography CXDI-710C Wireless (K170332), Digital Radiography CXDI-710C Wireless (K170332), Digital Radiography CXDI-410C Wireless (K171270), CXDI Control Software (K133693, K131106, K170332, K171270) Regulation Number: 21 CFR 892. 1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB
- 4) Indications for Use: Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
{4}------------------------------------------------
## 5) Description of the Device:
The Soltus 500 Mobile Digital X-Ray System, Model 10501, ("Mobile X-Ray System") is the same as the predicate mobile PhoeniX with 2 additional features, (i) Distributed Antenna System (DAS), and (ii) Enhanced Work Flow (EWF).
The Mobile X-Ray System has motorized movement and full battery operation. It contains a touch screen that operates as a control console. The Mobile X-Ray System supports various Canon flat panel detectors (Digital Radiography CXDI) supplied with the unit. (See (3) reference devices, above).
The Mobile X-Ray System additionally supports
- Ethernet communication
- Wireless Wi-Fi communications (with digital detectors and remote hand switch).
- Video Output to display EWF (Enhanced Work Flow) computer on Mobile x-ray unit monitor
- USB ports.
- Infrared data communication for registering the Wireless Detector
- Barcode scanner to register patient information. ●
- Dosimetry for Dose Area Product (DAP).
- Card reader to control access to the unit
New features:
- i) Distributed Antenna System (DAS) The Distributed Antenna system improves upon the antenna that is in the predicate device. It consists of a series of antennas to optimize the Wi-Fi communication between the wireless flat Panel detector and the wireless access point.
- ii) Enhanced Work Flow (EWF)
The Enhance Work Flow functionality allows the user to connect to the hospitals RIS and PACS system via a second computer embedded in the Mobile X-Ray Unit
### 6) Substantial Equivalence Chart
| Characteristic | Predicate: K192011, Trade/Device Name:<br>Battery Mobile X-ray Unit PhoeniX | Soltus 500 |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Indications for<br>Use: | Intended for use by a qualified/trained doctor<br>or technician on both adult and pediatric<br>subjects for taking diagnostic radiographic<br>exposures of the skull, spinal column, chest,<br>abdomen, extremities, and other body parts.<br>Applications can be performed with the<br>patient sitting, standing, or lying in the prone<br>or supine position. Not for mammography | SAME |
| Configuration | Mobile System with digital x-ray panel and<br>image acquisition computer | SAME |
| Characteristic | Predicate: K192011, Trade/Device Name:<br>Battery Mobile X-ray Unit Phoenix | Soltus 500 |
| X-ray<br>Generator(s) | kW rating: 20 kW, 32 kW, 40 kW and 50 kW.<br>kV range: from 40 kV to 150 kV in 1 kV steps.<br>mA range: from 10 mA to 630 mA / 640 mA /<br>650 mA. | SAME |
| Collimator | Ralco R108 DHHS | SAME |
| Photos | Phoenix® Mobile<br>Image: Phoenix Mobile X-ray | Soltus 500<br>Image: Soltus 500 X-ray |
| Digital X-ray<br>Panel Supplied | CANON CXDI-401C Wireless<br>CANON CXDI-701C Wireless<br>CANON CXDI-801C Wireless<br>CANON CXDI-710C Wireless<br>CANON CXDI-810C Wireless<br>CANON CXDI-410C Wireless | SAME |
| Software | Canon control software CXDI-NE | SAME |
| Panel Interface | Ethernet or Wi-Fi wireless | SAME |
| Meets US<br>Performance<br>Standard | YES 21 CFR 1020.30 | SAME |
| Computer | One control computer | One computer with two motherboards<br>(control PC, Enhanced Work Flow (EWF) |
| Power Source | Input transformer with 7 input voltage taps<br>(100 / 110 / 120 /, 127 / 220 / 230 / 240 V~).<br>AC 20 amp and Batteries | SAME |
| Wireless<br>Antennas | One single internal antenna | Distributed Antenna System (DAS) |
{5}------------------------------------------------
{6}------------------------------------------------
- 7) The technological characteristics, including design, materials, composition, and energy source, are substantially the same, so there are no issues impacting safety and effectiveness. Safety and Effectiveness, comparison to predicate device. The results of bench testing indicates that the new device is as safe and effective as the predicate device. Proper system operation is fully verified upon installation. We verified that the modified combination of components worked properly and produced diagnostic quality images as good as our predicate generator/panel combination.
- 8) Summary of non-clinical testing: Systems covering all generator/panel combinations were assembled and tested and found to be operating properly. In addition to the testing performed on the predicate, the modifications received bench testing and safety testing was updated. Firmware was validated according to the FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005. Because the system uses Wi-Fi and Ethernet, we observed the recommendations contained in the FDA Guidance Document: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014. The digital panel software employed was already reviewed by FDA in K190368. The Soltus 500 Battery Mobile X-Ray Units have been tested to be in compliance with the following
- a) IEC 60601-1:2005+A1:2012 (Edition 3.1)
- b) IEC 60601-1-2:2014 (Edition 4.0)
International Standards:
- c) IEC 60601-1-3:2008+A1:2013 (Edition 2.1)
- d) IEC 60601-2-54:2009+A1:2015 (Edition 1.1)
- e) IEC 60601-2-28:2010 (Edition 2.0)
- IEC 60601-1-6:2010 + A1:2013 (Edition 3.1) f)
- g) IEC 62304:2006 + A1:2016 (Edition 1.1)
- 9) Summary of clinical testing: Clinical testing was not required to establish substantial equivalence because all digital x-ray receptor panels have had previous FDA clearance.
- 10) Conclusion: After analyzing bench and clinical tests, it is the conclusion of Sedecal SA. that the new Soltus 500 Digital Diagnostic Mobile X-Ray Systems are as safe and effective as the predicate device, have few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.