The Wireless/Wired FDR D-EVO III flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO III is not intended for mammography, fluoroscopy, tomography, and angiography applications.
Device Story
Portable digital flat panel detector system (DR-ID1800) captures x-ray exposures from standard radiographic systems; digitizes signals for display. Supports wireless or wired/tethered communication. Used in clinical environments (bucky or free-cassette) by radiology staff. Employs film-based TFT substrate and GOS scintillator for indirect x-ray conversion. Output is digital radiographic image for clinician review; aids diagnostic decision-making. Benefits include improved detector weight and handling compared to predicate; maintains equivalent image quality.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical performance testing, including image quality evaluation, software testing, and adherence to voluntary standards (AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 62304, IEC 62366-1, IEC 62494-1).
Indicated for general projection radiographic imaging of human anatomy, including pediatric and neonatal patients, in environments where conventional film/screen or CR systems are used. Contraindicated for mammography, fluoroscopy, tomography, and angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
FUJIFILM FDR D-EVO II Flat Panel Detector System (DR-ID 1200) (K142003)
Submission Summary (Full Text)
{0}------------------------------------------------
November 12, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG' in blue, and 'ADMINISTRATION' in a smaller font size below.
FUJIFILM Corporation % Ms. Kamila Sak Regulatory Affairs Specialist FUJIFILM Medical Systems U.S.A, Inc. 81 Hartwell Avenue, Suite 300 LEXINGTON MA 02421
## Re: K192932
Trade/Device Name: FDR D-EVO III flat panel detector system Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: October 16, 2019 Received: October 17, 2019
Dear Ms. Sak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-
{1}------------------------------------------------
Page 2
combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K192932
Device Name
FDR D-EVO III flat panel detector system
#### Indications for Use (Describe)
The Wireless/Wired FDR D-EVO III flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO III is not intended for mammography, fluoroscopy, tomography, and angiography applications.
| Type of Use (Select one or both, as applicable) | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | |
| | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | |
| Food and Druq Administration<br>Office of Chief Information Officer<br>PRAStaff(@fda.hhs.gov | Department of Health and Human Services<br>Paperwork Reduction Act (PRA) Staff | | |
| | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number." | | |
==============================================================================================================================================================================
{3}------------------------------------------------
# 510(k) Summary K192932
#### Date Prepared: November 05, 2019
#### Submitter's Information:
FUJIFILM Corporation 798 Miyanodai Kaisei-Machi Ashigarakami-Gun, Kanagawa, 258-8538, Japan FDA Establishment Registration Number: 3001722928
#### Contact Person:
Kamila Sak Specialist, Regulatory Affairs Telephone: (347) 577-2309 Email: kamila.sak@fujifilm.com
#### Device Name and Classification:
| Product Name: | FUJIFILM FDR D-EVO III Flat Panel Detector System |
|-----------------------|---------------------------------------------------|
| Model Number: | DR-ID 1800 |
| Classification Name: | Stationary x-ray system |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR 892.1680 |
| Device Class: | Class II |
| Product Code: | MQB |
### Predicate Device:
| Product Name | FUJIFILM FDR D-EVO II Flat Panel Detector System<br>(DR-ID 1200) |
|----------------------|------------------------------------------------------------------|
| 510(k) Number | K142003 |
| Classification Name | Stationary x-ray system |
| Classification Panel | Radiology |
| CFR Section | 21 CFR 892.1680 |
| Device Class | Class II |
| Product Code | MQB |
The subject device FDR D-EVO III Flat Panel Detector System (DR-ID1800) is essentially a modified version of legally marketed FDR D-EVO II Flat Panel Detector System (DR-
{4}------------------------------------------------
ID1200). The predicate device DR-ID 1200 had received 510(k) clearance via K142003 on October 21, 2014, and was documented internally several times after that.
### Indications for Use:
The Wireless/Wired FDR D-EVO III flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO III is not intended for mammography, fluoroscopy, tomography, and angiography applications.
## Description of the Device:
The subject device FDR D-EVO III Flat Panel Detector System (DR-ID1800) is a portable digital detector system that interfaces with, and acquires and digitizes x-ray exposures from, standard radiographic systems. DR-ID1800 is designed to be used in any environment that would typically use a radiographic cassette for examinations of adults, pediatrics and neonates. The detector models support both wireless and wired/tethered data communication between the detector and the system. Detectors can be placed in a wall bucky for upright exams, a table bucky for recumbent exams, or removed from the bucky for non-grid or free cassette exams. The software that supports the functions of the FDR D-EVO III Flat Panel Detector System is unchanged from the predicate device, cleared under K142003.
## Comparison of Technological Characteristics:
A comparison of the technological characteristics between the subject device and predicate device is provided below:
{5}------------------------------------------------
| Comparison Item | Subject Device<br>FDR D-EVO III FPD system (DR-ID1800) | Predicate Device<br>FDR D-EVO II FPD system (DR-ID 1200) | Discussion<br>&<br>Conclusion |
|----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Wireless/Wired FDR D-EVO III flat panel<br>detector system is intended to capture for display<br>radiographic images of human anatomy. It is<br>intended for use in general projection<br>radiographic applications including pediatric and<br>neonatal exams wherever conventional<br>film/screen or CR systems may be used. The<br>FDR D-EVO III is not intended for<br>mammography, fluoroscopy, tomography, and<br>angiography applications. | The Wireless/Wired FDR D-EVO II flat panel detector<br>system is intended to capture for display radiographic<br>images of human anatomy. It is intended for use in<br>general projection radiographic applications including<br>pediatric and neonatal exams wherever conventional<br>film /screen or CR systems may be used. The FDR D-<br>EVO II is not intended for mammography, fluoroscopy,<br>tomography, and angiography applications. | Same(Except device name) |
| Detector characteristics | | | |
| Panel variation | DR-ID 1801SE (14" x 17")<br>DR-ID 1802SE (17" x 17") | DR-ID 1201SE (14" x 17"),DR-ID 1202SE (17" x 17")<br>DR-ID 1211SE (14" x 17"),DR-ID 1212SE (17" x 17")<br>DR-ID 1213SE (9" x 11") | Described as below |
| Scintillator | GOS (Gadolinium oxysulfide) | GOS (Gadolinium oxysulfide);<br>DR-ID 1201SE (14" x 17")<br>DR-ID 1202SE (17" x 17")<br>CsI(Secium Iodide);<br>DR-ID 1211SE (14" x 17")<br>DR-ID 1212SE (17" x 17")<br>DR-ID 1213SE (9" x 11") | The subject device has only<br>GOS based detector.<br>No impact on safety or<br>effectiveness of the device. |
| Dimensions<br>(Detector exterior) | DR-ID1801SE<br>: 38.4 cm x 46 cm x 1.5 cm<br>DR-ID1802SE<br>: 46 cm x 46 cm x 1.5 cm | DR-ID1201SE, DR-ID1211SE<br>: 38.4 cm x 46 cm x 1.5 cm<br>DR-ID1202SE, DR-ID1212SE<br>: 46 cm x 46 cm x 1.5 cm<br>DR-ID1213SE<br>: 26.8 cm x 32.8 cm x 1.5 cm | Same* |
| Number of Pixels | DR-ID1801SE: 2336 x 2836<br>DR-ID1802SE: 2832 x 2836 | DR-ID1201SE, DR-ID1211SE<br>: 2336 x 2836<br>DR-ID1202SE, DR-ID1212SE<br>: 2832 x 2836<br>DR-ID1213SE<br>:1920 x 1536 | Same* |
| Pixel Size | 150 μm | 150 μm | Same |
| Comparison Item | Subject Device<br>FDR D-EVO III FPD system (DR-ID1800) | Predicate Device<br>FDR D-EVO II FPD system (DR-ID 1200) | Discussion<br>&<br>Conclusion |
| X-ray Conversion | Indirect conversion (a-Si) | Indirect conversion (a-Si) | Same |
| TFT sensor<br>substrate | Film-based TFT substrate | Glass-based TFT substrate | Only the material of<br>substrate is changed, there<br>is no change in the<br>mechanism. No impact on<br>safety or effectiveness of<br>the device. |
| DQE (RQA5, 1<br>lp/mm, 1mR) -<br>detector alone,<br>without tabletop | 33%<br>Measurement tolerance (±10%) | 31%<br>Measurement tolerance (±10%) | Slightly improved DQE by<br>using Film-based TFT<br>substrate<br>This difference of DQE<br>does not impact clinical<br>Image Quality. |
| MTF (RQA5, 1<br>lp/mm) | Two options are available by setting.<br>-Normal 60%<br>-High 75% | Two options are available by setting.<br>-Normal 60%<br>-High 75% | Same |
| Appearance<br>characteristics | - Adopt more rounded shape<br>- Changed alighnment of LEDs/Operation button<br>that is easier to use | - Rounded corners and curved edges are adopted.<br>- LEDs/Operation button is placed. | This difference of detector<br>appearance does not<br>impact on safety or<br>effectiveness of the device. |
| Detector Weight | DR-ID1801SE:<br>: Approx. 1.8kg (without battery)<br>DR-ID1802SE<br>: Approx. 2.1kg (without battery) | DR-ID1201SE<br>: Approx. 2.3kg (without battery)<br>DR-ID1202SE<br>: Approx. 2.9kg (without battery)<br>DR-ID1211SE<br>: Approx. 2.4kg (without battery)<br>DR-ID1212SE<br>: Approx. 3.0kg (without battery)<br>DR-ID1213SE<br>: Approx. 1.3kg (without battery) | This difference of detector<br>weight does not impact on<br>safety or effectiveness of<br>the device. |
| Comparison Item | Subject Device<br>FDR D-EVO III FPD system (DR-ID1800) | Predicate Device<br>FDR D-EVO II FPD system (DR-ID 1200) | Discussion &<br>Conclusion |
| Software Charactisctic | | | |
| Workstation | FDX Console Version 13.0 | FDX Console Version 8.0 and above (Latest version:<br>12.1) | This difference of the<br>workstation is only to add<br>this subject detector as the<br>connectable Flat panel<br>detector options. No impact<br>on safety or effectiveness of<br>the device. |
{6}------------------------------------------------
{7}------------------------------------------------
*Note: Csl scintillator has not been adopted in DR-ID 1800 yet. Then, we discuss and conclude the substantial equivalence by excluding Csl scintillator.
{8}------------------------------------------------
#### Substantial Equivalence:
Both the subject device and predicate device have the same Indication for Use. There are some differences, but the differences are minor and do not affect fundamental scientific technology, principles of operation, safety and effectiveness, and image quality. Therefore, the subject device FDR D-EVO III Flat Panel Detector System (DR-ID1800) can be considered to be substantially equivalent to the predicate device FDR D-EVO II Flat Panel Detector System (DR-ID1200) limited to the flat panel detector with GOS scintillator.
## Summary Of Studies:
Non-clinical Performance Data: FDR D-EVO III Flat Panel Detector System (DR-ID1800) conforms to the voluntary standards such as AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 62304, IEC 62366-1, IEC 62494-1 and DICOM. In addition, the FDA's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices (September 1, 2016) was followed to describe the detector characteristics, and Radio Frequency Wireless Technology in Medical Devices (August 14, 2013) was followed to test wireless features. As required by the risk analysis, necessary verification and validation activities were performed including software testing, and the results were satisfactory. Furthermore, the image quality evaluation confirmed that the image quality of the FDR D-EVO III Flat Panel Detector System (DR-ID1800) is substantially equivalent to that of the predicate device.
Clinical Performance Data: No clinical study has been performed. The substantial equivalence has been demonstrated by non-clinical studies.
#### Conclusion:
This Special 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.