← Product Code [OLO](/productcode/OLO) · K192911

# Brainlab Compatible K2M Navigation Instruments (K192911)

_K2m, Inc. · OLO · Nov 14, 2019 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K192911

## Device Facts

- **Applicant:** K2m, Inc.
- **Product Code:** [OLO](/productcode/OLO.md)
- **Decision Date:** Nov 14, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

Brainlab Compatible K2M Navigation Instruments are intended to be used in the preparation and placement of K2M pedicle screws (DENALI, MESA, EVEREST, YUKON) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Brainlab Navigation system, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. The Brainlab Compatible K2M Navigation Instruments are not intended for navigation of occipital screws.

## Device Story

Manual surgical instruments designed to interface with Brainlab Navigation system; used for preparation and placement of K2M pedicle screws (DENALI, MESA, EVEREST, YUKON) during spinal surgery. Instruments assist surgeons in locating anatomical structures via stereotactic navigation using CT/MR models, fluoroscopy, or digitized landmarks. Subject device features modified connection mechanism geometry and material to improve user experience. Used in OR by surgeons; provides visual guidance for screw placement; enhances precision in spinal procedures.

## Clinical Evidence

No clinical data. Bench testing only, including risk assessment, sterilization, biocompatibility, and rotational testing.

## Technological Characteristics

Manual surgical instruments; modified connection mechanism geometry and material. Designed for integration with Brainlab Navigation system. Sterilization required.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Brainlab Compatible K2M Navigation Instruments ([K181890](/device/K181890.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 14, 2019

K2M, Inc. % Megan Callanan Regulatory Affairs Associate Stryker 2 Pearl Court Allendale, New Jersey 07401

Re: K192911

Trade/Device Name: Brainlab Compatible K2M Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: October 11, 2019 Received: October 15, 2019

Dear Megan Callanan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For: Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192911

#### Device Name

Brainlab Compatible K2M Navigation Instruments

### Indications for Use (Describe)

Brainlab Compatible K2M Navigation Instruments are intended to be used in the preparation and placement of K2M pedicle screws (DENALI, MESA, EVEREST, YUKON) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Brainlab Navigation system, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. The Brainlab Compatible K2M Navigation Instruments are not intended for navigation of occipital screws.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span style="text-decoration: underline;"><b></b></span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b></b></span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)             |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) Summary: Brainlab Compatible K2M Navigation Instruments |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                                                     | K2M, Inc.<br>600 Hope Pkwy SE<br>Leesburg, VA 20175                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Contact Person :                                               | Name: Megan Callanan<br>Phone: (551)262-2429<br>Email: megan.callanan1@stryker.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Date Prepared:                                                 | 11/14/2019                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Trade Name:                                                    | Brainlab Compatible K2M Navigation Instruments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Common Name:                                                   | Navigation Instruments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Proposed Class:                                                | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification Name:                                           | Orthopedic / Orthopedic Stereotaxic Instruments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Regulation Number:                                             | 21 CFR 882.4560                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Product Code:                                                  | OLO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Predicate Device:                                              | Primary Predicate: Brainlab Compatible K2M Navigation Instruments (K181890)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Description:                                            | Brainlab Compatible K2M Navigation Instruments are manual surgical instruments<br>intended be used when implanting previously cleared components of MESA, DENALI,<br>EVEREST and YUKON Spinal Systems.<br><br>These instruments are designed to interface with the Brainlab Navigation system<br>when used for navigation during spinal surgery. The subject instruments feature a<br>modification of the connection mechanism geometry and material to enhance user<br>experience.                                                                                                                                                                                                                                                                                                                                                             |
| Indications for use:                                           | Brainlab Compatible K2M Navigation Instruments are intended to be used in the<br>preparation and placement of K2M pedicle screws (DENALI, MESA, EVEREST, YUKON)<br>during spinal surgery to assist the surgeon in precisely locating anatomical structures<br>in either open or minimally invasive procedures. These instruments are designed for<br>use with the Brainlab Navigation system, which is indicated for any medical condition<br>in which the use of stereotactic surgery may be appropriate, and where reference to a<br>rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified<br>relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the<br>anatomy. The Brainlab Compatible K2M Navigation Instruments are not intended for<br>navigation of occipital screws. |
| Summary of the Technological<br>Characteristics                | Brainlab Compatible K2M Navigation Instruments were compared to predicate<br>devices and the design features, materials and indications were the same or similar<br>to the previously cleared devices. The subject instruments feature a modification of<br>the connection mechanism geometry and material to enhance user experience.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Non-clinical Performance<br>Evaluation                         | A risk assessment including sterilization, biocompatibility, and additional rotational<br>testing was conducted to confirm that the subject Brainlab Compatible K2M<br>Navigation Instruments do not introduce new issues of safety or effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Conclusion                                                     | There are no significant differences between the Brainlab Compatible K2M Navigation<br>Instruments and other legally marketed systems. It is substantially equivalent to these<br>other devices in design, function, material and intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

---

**Source:** [https://fda-staging.innolitics.com/device/K192911](https://fda-staging.innolitics.com/device/K192911)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).