SensiCare Neoprene Synthetic Polychloroprene Surgical Glove (Tested for Use with Chemotherapy Drugs)

K192820 · Medline Industries, Inc. · KGO · Feb 28, 2020 · General, Plastic Surgery

Device Facts

Record IDK192820
Device NameSensiCare Neoprene Synthetic Polychloroprene Surgical Glove (Tested for Use with Chemotherapy Drugs)
ApplicantMedline Industries, Inc.
Product CodeKGO · General, Plastic Surgery
Decision DateFeb 28, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4460
Device ClassClass 1

Indications for Use

This surgeon's glove is a disposable device made of synthetic polychloroprene intended to be worn by surgeons and/or operating room personnel to protect a surgical wound from contamination. These gloves are tested for use with chemotherapy drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Story

Disposable synthetic polychloroprene surgical glove; worn by surgeons and operating room personnel to provide barrier protection for surgical wounds. Tested for permeation resistance against various chemotherapy drugs per ASTM D6978-05. Provides protection during handling of hazardous chemotherapy agents; breakthrough times for specific drugs provided in labeling. Used in clinical/surgical settings.

Clinical Evidence

Bench testing only. Device performance evaluated via ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Breakthrough detection times reported for 24 specific chemotherapy agents.

Technological Characteristics

Material: Synthetic polychloroprene. Form factor: Disposable surgical glove. Standards: ASTM D6978-05 for chemotherapy drug permeation. Non-powdered. Class I, reserved.

Indications for Use

Indicated for use by surgeons and operating room personnel as a protective barrier against contamination of surgical wounds. Tested for use with specific chemotherapy drugs.

Regulatory Classification

Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. February 28, 2020 Medline Industries, Inc. Matt Clausen Senior Regulatory Affairs Specialist Three Lakes Drive Northfield, Illinois 60093 ### Re: K192820 Trade/Device Name: SensiCare Neoprene Synthetic Polychloroprene Surgical Glove (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO, LZC Dated: January 13, 2020 Received: January 15, 2020 ## Dear Matt Clausen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192820 Device Name SensiCare® Neoprene Synthetic Polychloroprene Surgical Glove (Tested for Use with Chemotherapy Drugs) Indications for Use (Describe) This surgeon's glove is a disposable device made of synthetic polychloroprene intended to be worn by surgeons and/or operating room personnel to protect a surgical wound from contamination. These gloves are tested for use with chemotherapy drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. The following chemicals have been tested with these gloves. Chemotherapy Drug Permeation (Breakthrough Detection Time) in Minutes. Bleomycin 15 mg/ml >240 minutes Busulfan 6 mg/ml >240 minutes Carboplatin 10.0 mg/ml >240 minutes Carmustine (BCNU) 3.3 mg/ml 36.3 (36.3, 37.4, 38.4) minutes Cisplatin 1.0 mg/ml >240 minutes Cyclophosphamide (Cytoxan) 20 mg/ml >240 minutes Dacarbazine (DTIC) 10.0 mg/ml >240 minutes Doxorubicin Hydrochloride 2.0 mg/ml >240 minutes Epirubicin (Ellence) 2 mg/ml >240 minutes Etoposide (Toposar) 20.0 mg/ml >240 minutes Fludarabine 25.0 mg/ml >240 minutes Fluorouracil 50.0 mg/ml >240 minutes Idarubicin 1.0 mg/ml >240 minutes Ifosfamide 50.0 mg/ml >240 minutes Mechlorethamine HCI 1.0 mg/ml >240 minutes Melphalan 5 mg/ml >240 minutes Methotrexate 25 mg/ml >240 minutes Mitomycin C 0.5 mg/ml >240 minutes Mitoxantrone 2.0 mg/ml >240 minutes Paclitaxel (Taxol) 6.0 mg/ml >240 minutes Paraplatin 10 mg/ml >240 minutes Rituximab 10 mg/ml >240 minutes Thiotepa 10.0 mg/ml 49.2 (56.0. 49.2, 78.5) minutes Vincristine Sulfate 1.0 mg/ml >240 minutes CAUTION: Testing showed a minimum breakthrough time of 36.3 minutes with Carmustine and 49.2 minutes for Thiotepa. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) |X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...