← Product Code [MJN](/productcode/MJN) · K192786

# Gatekeeper Balloon Catheter (K192786)

_Arch Catheter, LLC · MJN · Apr 25, 2020 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K192786

## Device Facts

- **Applicant:** Arch Catheter, LLC
- **Product Code:** [MJN](/productcode/MJN.md)
- **Decision Date:** Apr 25, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Gatekeeper Balloon Catheter is indicated for temporary occlusion of vessels in the peripheral vasculature.

## Device Story

The Gatekeeper Balloon Catheter is an over-the-wire (OTW) triple-lumen catheter featuring two compliant polyurethane balloons (proximal 15mm, distal 10mm) spaced 2 cm apart. It is used by clinicians in the peripheral vasculature to achieve temporary vessel occlusion. The device is operated via extension lines with stopcocks for balloon inflation/deflation; a third lumen accommodates a 0.014" guidewire. Radiopaque marker bands facilitate precise placement under imaging. The catheter is compatible with 6Fr introducer sheaths. By occluding vessels, the device assists in controlling blood flow or isolating vascular segments, potentially benefiting patients during peripheral vascular procedures. It is a single-use, sterile device.

## Clinical Evidence

Bench testing only. Performance verified against ISO 10555-1 and ISO 10555-4 standards. Testing included visual/dimensional inspection, leakage, compatibility (luer, guidewire, sheath), balloon compliance, inflation characteristics, radiopacity, shipping/distribution, vessel occlusion, balloon fatigue, burst/leak volume, fragmentation, tensile strength, shelf life, package integrity, environmental conditioning, kink, torque, and simulated use. Biocompatibility testing conducted per ISO 10993-1.

## Technological Characteristics

Triple-lumen OTW catheter; polyurethane balloons; 15mm (proximal) and 10mm (distal) diameters; 2cm balloon spacing; radiopaque marker bands; 0.014" guidewire compatibility; 6Fr introducer sheath compatibility. Complies with ISO 10555-1 and ISO 10555-4. Sterile, single-use.

## Regulatory Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

## Predicate Devices

- FloControl Catheter ([K090970](/device/K090970.md))
- Stent Graft Balloon Catheter ([K091624](/device/K091624.md))
- Q50 PLUS Stent Graft Balloon ([K120381](/device/K120381.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

April 25, 2020

Arch Catheter, LLC % Mary Vater 510(k) Consultant Medical Device Academy 345 Lincoln Hill Rd. Shrewsbury, Vermont 05738

Re: K192786

Trade/Device Name: Gatekeeper Balloon Catheter Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: MJN Dated: September 27, 2019 Received: September 30, 2019

Dear Mary Vater:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Carmen Gacchina Johnson Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192786

Device Name Gatekeeper Balloon Catheter

Indications for Use (Describe)

The Gatekeeper Balloon Catheter is indicated for temporary occlusion of vessels in the peripheral vasculature.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Catheter

# 510(k) SUMMARY K192786

1. SUBMITTER Arch Catheter, LLC 621 Grand Harbor Boulevard Ninety Six, SC, 29666 USA Tel: 1.864.680.2568 Fax: 1.864.725.7910

Contact Person: Michael Zhadkevich, MD, PhD Date Prepared: Sept. 27, 2019

| II. DEVICE                   |                             |
|------------------------------|-----------------------------|
| Name of Device:              | Gatekeeper Balloon Catheter |
| Common Name:                 | Occlusion balloon catheter  |
| Classification Name:         | Vascular Clamp              |
| Regulation:                  | 21 CFR § 870.4450           |
| Regulatory Class:            | Class II                    |
| Product Classification Code: | MJN                         |

| III. | PREDICATE DEVICE |
|------|------------------|
|------|------------------|

| Primary Predicate Manufacturer:<br>Primary Predicate Trade Name:<br>Primary Predicate 510(k):       | CoAxia, Inc.<br>FloControl Catheter<br>K090970                  |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Secondary Predicate Manufacturer:<br>Secondary Predicate Trade Name:<br>Secondary Predicate 510(k): | Via Biomedical, Inc.<br>Stent Graft Balloon Catheter<br>K091624 |
| Secondary Predicate Manufacturer:<br>Secondary Predicate Trade Name:<br>Secondary Predicate 510(k): | QXMedical, LLC<br>Q50 PLUS Stent Graft Balloon<br>K120381       |

No reference devices were used in this submission.

### IV. DEVICE DESCRIPTION

The Gatekeeper Balloon Catheter is an over the wire (OTW) triple-lumen catheter with 2 compliant proximal and distal polyurethane balloons having a maximum diameter of 15 mm and 10 mm respectively. The 2 balloon lumens are connected to extension lines with stopcocks and are used to inflate and deflate the proximal and distal balloons. The third lumen is a guidewire lumen. The distance between the proximal and distal balloons is 2 cm. Two (2) radiopaque marker bands are located within the balloons at each end (30mm apart at the proximal balloon and 15mm apart at the distal balloon) to facilitate balloon placement prior to inflation. The catheter can accommodate a 0.014" diameter (or smaller) guidewire and is compatible with 6Fr (or larger) introducer sheath. The device is a single use, sterile device.

{4}------------------------------------------------

#### V. INDICATIONS FOR USE

The Gatekeeper Balloon Catheter is indicated for temporary occlusion of vessels in the peripheral vasculature.

### VI. SUMMARY OF SUBSTANTIAL EQUIVALENCE

Compared to the predicate devices of the CoAxia FloControl Catheter (K090970), Via Biomedical Stent Graft Balloon Catheter (K091624), and QXMEDICAL Q50 PLUS Stent Graft Balloon Catheter (K120381). the subject Gatekeeper Balloon Catheter has the same intended use and mechanism of action for temporary occlusion of vessels. The indication statements for the predicate devices describe additional functions for expanding vascular prostheses, controlling blood flow, or occlusion of large vessels; but all devices are intended for temporary occlusion of vessels. The subject device has different technological characteristics of catheter design and materials that do not raise different questions of safety and effectiveness and were evaluated with non-clinical testing.

The subject device is designed to fulfill the requirements of the following recognized standards:

- 6-301 ISO 10555-1 Second edition 2013-06-15 Intravascular catheters Sterile and single-use intravascular catheters - Part 1: General requirements
- 6-322 ISO 10555-4 Second edition 2013-06-15 Intravascular catheters -- Sterile and single-use catheters -- | Part 4: Balloon dilatation catheters
- 6-408 ISO 10555-1 Second edition 2013-06-15 Intravascular catheters - Sterile and single-use intravascular catheters - Part 1: General requirements [Including AMENDMENT 1 (2017)]

The following benchtop performance tests were conducted to demonstrate that the Gatekeeper Balloon Catheter performance complies with the standards.

- Visual Inspection
- -Dimensional Inspection
- Freedom from Leakage ー
- Luer Syringe Compatibility —
- Guidewire Compatibility
- Introducer Sheath Compatibility ー
- Balloon Compliance (Volume v. Diameter)
- -Inflation Time
- Balloon Inflation Characteristics —
- -Radiopacity
- -Shipping/ Distribution Testing
- -Vessel Occlusion
- Balloon Fatigue
- -Burst or Leak Volume
- Freedom from Fragmentation
- -Tensile Strength (Hub to Shaft)
- -Tensile Strength (Tip to Shaft)
- -Tensile Strength (Extension Tube)
- Shelf Life Testing ー
- Package Integrity
- -Environmental Conditioning
- Kink Testing —
- -Torque Testing
- -Simulated Use

Biocompatibility testing was conducted in accordance with ISO 10993-1.

{5}------------------------------------------------

## VII. CONCLUSIONS

The Gatekeeper Balloon Catheter has the same intended use as the predicate devices. The technological differences do not raise different questions of safety and effectiveness. Based on the results of non-clinical testing, the data supports the Gatekeeper Balloon Catheter is substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K192786](https://fda-staging.innolitics.com/device/K192786)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
