K192785 · Optiscan Biomedical Corporation · LZF · Feb 5, 2020 · General Hospital
Device Facts
Record ID
K192785
Device Name
OptiScanner® 5000 Glucose Monitoring System
Applicant
Optiscan Biomedical Corporation
Product Code
LZF · General Hospital
Decision Date
Feb 5, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Indications for Use
The OptiScanner® 5000 Glucose Monitoring System is intended for in-vitro diagnostic use.
Device Story
Automated, bedside glucose monitoring system for surgical intensive care unit (SICU) patients. System comprises instrument, transport cart, and sterile, single-use disposable cartridge. Instrument houses pump, spectrometer, and user interface; cart provides mobility and power. Cartridge connects to patient's central venous catheter (CVC), peripherally inserted central catheter (PICC), or multi-lumen access catheter (MAC) to draw 3 mL venous blood. System centrifuges sample, measures plasma glucose concentration via reagent-free spectrophotometry, and returns remaining blood with saline flush. Operates on 15-minute sampling frequency. Provides glucose trends and patterns to clinicians; alerts for high/low glucose, air in line, or occlusion. Benefits include automated, continuous glucose monitoring without enzymatic reagents, reducing manual sampling burden and supporting dysglycemia management.
Clinical Evidence
No new clinical data provided. Performance characteristics, including analytical accuracy, precision, linearity, and clinical supportive data, were previously established and reviewed under K162042. Current submission relies on bench testing to verify mechanical compatibility of the device with specific PICC models.
Technological Characteristics
Spectrophotometric glucose quantification; reagent-free. System includes instrument, cart, and sterile disposable cartridge (single-patient use, 72-hour duration). Connectivity: standalone with 72-hour data storage. Power: AC with 3-hour backup battery. Sampling: 0.17 mL analyzed, 3 mL total draw, 2.5 mL saline flush return. Hematocrit range: 15-60%. Glucose range: 40-400 mg/dL.
Indications for Use
Indicated for persons age 18+ in surgical intensive care units for detecting trends and tracking patterns in glucose levels; determining dysglycemia. Contraindicated for patients <18 years, pregnant or nursing women, and patients undergoing MRI or CT.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
OptiScanner® 5000 Glucose Monitoring System (K162042)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K192785
B Applicant
Optiscan Biomedical Corporation
C Proprietary and Established Names
OptiScanner® 5000 Glucose Monitoring System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LZF | Class II | 21 CFR 880.5725 - Infusion Pump | General Hospital |
| PYV | Class II | 21 CFR 862.1345 - Glucose test system | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
The purpose of this submission is to add selected peripherally inserted central catheters (PICCs) to the list of catheters which are compatible with the device.
B Measurand:
Plasma glucose from central venous catheter blood draw
C Type of Test:
Quantitative, mid-infrared (MIR) spectrophotometric assay
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The OptiScanner® 5000 Glucose Monitoring System is an automated, bedside glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) in the surgical intensive care unit. The system collects a venous whole blood sample via connection to a central venous catheter, centrifuges the sample, and measures the plasma glucose concentration. It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysglycemia. The OptiScanner® 5000 Glucose Monitoring System is for in vitro diagnostic use.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Contraindications:
- Patients that are <18 years of age
- Women who are pregnant or nursing
- Patients undergoing Magnetic Resonance Imaging (MRI) or Computerized Tomography (CT)
Warnings:
- The OptiScanner disposable cartridge should be used only by medical personnel having specific training in, and understanding of, the techniques associated with such devices and procedures.
- The disposable cartridge is for single patient use only. Do not re-sterilize. Do not re-use. Re-use of the cartridge may pose infection to the patient due to non-sterile condition and cross contamination of blood.
- Do not use the cartridge if the packaging is received open or damaged.
- The OptiScanner has not been tested for use in the operating room.
- The OptiScanner should be connected to the most proximal port when using a Central Venous Catheter (this includes peripherally inserted central catheters). If the most proximal port is not available, use next most proximal port for OptiScanner primary connection.- Do not infuse anything above the OptiScanner primary connection.
- Do not rest containers of liquids on top of the OptiScanner enclosure.
- Avoid infusion of any glucose containing solutions adjacent to the OptiScanner port. Infuse these solutions in the most distal port available, as clinical research has shown glucose administered through the distal port does not affect OptiScanner accuracy.
- Be advised of possible complications associated with central venous catheterization. These include but are not limited to: vessel wall perforation, cardiac tamponade, air embolism, catheter embolism, thrombosis, bacteremia, septicemia.
K192785 - Page 2 of 8
{2}
- Due to the risk of further blood loss, do not use the OptiScanner in patients with low hematocrit (less than 15%).
- The OptiScanner should not be used concomitantly with the intravenous administration of high dose ascorbate (IVC) for the treatment of patients with cancer.
- Use caution when OptiScanner readings are below 60 and above 300 mg/dL as these concentrations have not been studied in the intended use population.
- Use caution when OptiScanner is reading in the hypoglycemic range (<70 mg/dL). The comparator measurement check may not be sufficient to indicate instances of severe hypoglycemia, especially when the Low Glucose Alarm is set to 80 mg/dL.
- Though line occlusions were not studied in the trial, it should be noted that the catheters utilized with the OptiScanner may become occluded. In the event of a line occlusion, follow hospital protocol to clear the line.
- Certain substances should be used with caution with the OptiScanner. Refer to the Interfering Substance section in Chapter 7 for more information.
- Not all substances which could potentially interfere with OptiScanner measurement accuracy have been identified. Following the instructions regarding the daily reference check is imperative to screen for potential offsets.
- No additives of any kind are to be passed through or injected into the saline bag or saline flush line.
- Do not use the OptiScanner in patients that are being administered intravenous immunoglobulin therapies (IVIG). These therapies may generate false glucose measurements. Examples of IVIG therapies are Gamimune N, HepaGam B, Octagam, Vaccinia Immune Globulin, and WinRho SDF Liquid.
- Do not use the OptiScanner in patients that have used peritoneal dialysis solutions any time in the past week. These solutions may generate false glucose measurements.
- Do not inject heparin or any other substances into the saline bag.
- If moving the OptiScanner without disconnecting from the patient, always maintain the connection between the venous line and the OptiScanner patient line.
- Do not use stopcocks. Ensure catheter lines are not pinched off by clamps, hemostats or physically kinked.
- Use of stopcocks or catheters with valves (for example, Groshong catheters) that potentially alter, restrict or impede flow should never be used with the OptiScanner.
- Only use normal, 0.9% saline with the OptiScanner. Do not use saline with dextrose, ringers, or any other substance or additive to connect to the OptiScanner.
- The disposable cartridge must be primed before being connected to the patient. Do NOT connect to patient until priming is complete.
- Be sure there is no air in the catheter lumen when connecting the OptiScanner to the patient. If present, air could be returned to the patient. Ensure catheter is primed per manufacturer's instructions for use.
D Special Instrument Requirements:
OptiScanner® 5000 Instrument
K192785 - Page 3 of 8
{3}
K192785 - Page 4 of 8
# IV Device/System Characteristics:
# A Device Description:
The OptiScanner® 5000 Glucose Monitoring System is comprised of the OptiScanner® 5000 Instrument, a transportation cart, disposable OptiScanner® cartridges (sold separately), and a barcode scanner with USB connectivity. The OptiScanner® 5000 instrument houses the mechanisms that automatically draw and return blood to the patient, analyzes the sample and calculates the glucose value. The disposable cartridges are sterile, single use only, and designed for use on a single patient for up to 72 hours.
# B Principle of Operation:
The OptiScanner® 5000 Glucose Monitoring System samples and measure blood glucose levels every 15 minutes. Samples ( $\sim$ 3 mL) are drawn from the patients central venous catheter (CVC), multi-lumen access catheter, or peripherally inserted central catheter (PICC). Of the withdrawn blood, $0.17\mathrm{mL}$ is retained for measurement and the remaining blood is returned to the patient. The retained volume is heparinized to prevent clotting while it is processed and is centrifuged to separate the red blood cells from the plasma. The plasma is evaluated and the glucose quantified by the spectrophotometer by analyzing mid-infrared (MIR) light ( $7\mathrm{um} - 10\mathrm{um}$ wavelength) absorption spectra through an array of optical filters. The OptiScanner® 5000 Glucose Monitoring System is designed to display glucose values every 15 minutes along with a trend line with settable high and low alarms.
C Instrument Description Information:
| Modes of Operation | Yes | No |
| --- | --- | --- |
| Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐ |
| Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☑ | ☐ |
| Software | | |
| FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types. | ☑ | ☐ |
1. Instrument Name:
OptiScanner® 5000 Glucose Monitoring System
2. Specimen Identification:
The patient ID can be entered into the OptiScanner® 5000 Glucose Monitoring System manually or by using the barcode scanner included with the system. The patient ID is displayed on the device screen along with the glucose measurement and glucose trend information.
{4}
K192785 - Page 5 of 8
3. Specimen Sampling and Handling:
The venous blood drawn from the patient for measurement is immediately centrifuged in the disposable test cartridge to plasma and analyzed by the device.
4. Calibration:
The OptiScanner® 5000 requires no calibration by the user.
5. Quality Control:
There is no quality control material for use with the OptiScanner® 5000 Glucose Monitoring System.
V Substantial Equivalence Information:
A Predicate Device Name(s):
OptiScanner 5000 Glucose Monitoring System
B Predicate 510(k) Number(s):
K162042
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K192785 | K162042 |
| --- | --- | --- |
| Device Trade Name | OptiScanner® 5000 Glucose Monitoring System | Same |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Intended for hospital bedside glucose monitoring for detecting trends and tracking patterns in glucose | Same |
| Intended use population | Patients in the surgical intensive care unit (SICU) | Same |
| General Device Characteristic Differences | | |
| Point of access for sample | Central, multi lumen, or peripheral venous catheter | Central or multi-lumen venous catheter |
{5}
VI Standards/Guidance Documents Referenced:
Not applicable.
VII Performance Characteristics (if/when applicable):
The device in this submission is different from the predicate device in that it is intended for use in the peripherally inserted central catheters described in Table 1 under section VII.F, below. There are no changes to the physical device, and aside from claimed use with the catheters described below the device claims are unchanged from those of the predicate device. Clinical and nonclinical data as reviewed in the clearance of the predicate device, k162042, also support the clearance of this device. For more information regarding the information supporting the performance of this device, see the decision summary for clearance of k162042.
A Analytical Performance:
1. Precision/Reproducibility:
Previously established in k162042.
2. Linearity:
Previously established in k162042.
3. Analytical Specificity/Interference:
Previously established in k162042.
The sponsor has included the following in the labeling:
- Not all substances which could potentially interfere with OptiScanner® measurement accuracy have been identified. Following the instructions regarding the daily reference check is imperative to screen for potential offsets. The OptiScanner® should not be used during the first 12 hours following the administration of a D-Xylose absorption test.
- Do not use the OptiScanner® in patients who are being treated with Sodium Thiosulfate.
- Do not use the OptiScanner® when glycerol is being infused intravenously for the treatment of brain edema.
- Substances containing maltose, or substances that can be metabolized into maltose, may generate falsely high blood glucose readings on the OptiScanner®.
Do not use the OptiScanner® in patients that have are receiving any of these medications.
4. Assay Reportable Range:
The OptiScanner® 5000 will display results between 40 mg/dL and 400 mg/dL glucose.
K192785 - Page 6 of 8
{6}
The reportable range for this device was established in k162042.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
**Traceability:**
The clinical study performed to support the predicate device, k162042, used YSI 2300 as the comparator method.
**Stability:**
The disposable cartridges and accessories (plasma separation chamber, OptiScanner Heparin Syringe and Adapter) are packaged together and are provided as sterile. The other components of the system (the instrument, a transportation cart, and a barcode scanner) are provided as non-sterile.
The one year shelf-life of the disposable cartridges and accessories is stated by the sponsor to be up to one year when stored at 5°C - 35°C (41°F - 95°F).
The sterility and shelf life of the OptiScanner® 5000 were previously reviewed in k162042.
6. Detection Limit:
See linearity, described above in Section VII.A.2.
7. Assay Cut-Off:
Not applicable.
8. Accuracy (Instrument):
Previously established in k162042.
9. Carry-Over:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
See section VII.C.3 below.
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
1. Clinical Sensitivity:
K192785 - Page 7 of 8
{7}
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Previously established in k162042.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
Not applicable.
F Other Supportive Instrument Performance Characteristics Data:
Mechanical Compatibility:
Selected PICCs were assessed for compatibility with the OptiScanner® 5000 Glucose Monitoring System by conducting bench testing that assessed whether samples could be drawn through the catheter to the instrument according to device specifications. Results from these studies support the use of the following PICCs with the OptiScanner® 5000 System, which are listed in the user manual:
Table 1. PICC compatibility with the OptiScanner® 5000 Glucose Monitoring System
| Brand | Model | Length (cm) | French Size |
| --- | --- | --- | --- |
| Bard | 4134115 | 65 | 4 |
| Bard | 5A5002057 | 55 | 4 |
| Bard | 5A5001307 | 55 | 5 |
| Bard | 5A5000574 | 55 | 5 |
| Arrow | CA-05052-HP | 50 | 4 |
| Arrow | CA-05041-HP | 50 | 5 |
| Arrow | PR-45541-HPHNM | 55 | 4.5 |
| Arrow | PR-45552-HPHNM | 55 | 5.5 |
| Arrow | PR-45563-HPHNM | 55 | 6 |
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K192785 - Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.