Niobe® Magnetic Navigation System (MNS) with Navigant Workstation (NWS) and Cardiodri

K192775 · Stereotaxis, Inc. · PJB · Oct 30, 2019 · Cardiovascular

Device Facts

Record IDK192775
Device NameNiobe® Magnetic Navigation System (MNS) with Navigant Workstation (NWS) and Cardiodri
ApplicantStereotaxis, Inc.
Product CodePJB · Cardiovascular
Decision DateOct 30, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.5700
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K192775 · Oct 30, 2019Niobe® Magnetic Navigation System (MNS) with Navigant Workstation (NWS) and CardiodriStereotaxis, Inc.Clinical study datasets (ATTRAC, ATTRAC II, HEART, VERSATILE); Published clinical literatureClinical data from 511 patients across four previously submitted studies were used to assess major complication rates, acute/90-day success rates, and fluoroscopy times to support the safety and efficacy of the Niobe System. Additionally, literature data (8,000+ patients) was used to compare complication rates against manual procedures.Cardiac ablation; Magnetic navigation; Clinical study summary; Safety and efficacy; Literature review

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
ATTRAC, ATTRAC II, HEART, and VERSATILE studies; Prospective clinical studies; Follow-up/Duration: 7-day major adverse event assessment; 90-day success ratesPatients undergoing cardiac ablation using the Niobe Magnetic Navigation System; Sample Size: 511Manual comparison group (for literature data)Major complication rate, acute success rate, 90-day success rate, fluoroscopy time

Indications for Use

The Niobe System is intended to navigate compatible magnetic devices through tissue to designated target sites in the right and left heart and coronary vasculature, neurovascular and peripheral vasculature by orienting the device tip in a desired direction. The Cardiodrive Catheter Advancement System (CAS) is intended to automatically advance and retract compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart when used in conjunction with a Stereotaxis MNS. The Cardiodrive system is not intended to advance the EP mapping and ablation catheters through the coronary vasculature or the coronary sinus. The Cardiodrive system is not intended to advance or retract non-compatible catheters and/or other non-compatible devices into the neurovasculature.

Device Story

Niobe Magnetic Navigation System (MNS) is an interventional workstation for intravascular navigation of magnetically-adapted devices. System uses computer-controlled permanent magnets to orient device tips and remotely advance/retract compatible EP mapping and ablation catheters. Used in cardiac catheter labs; operated by physicians. Navigant Workstation (NWS) provides user interface for physician interaction; software determines magnetic field direction based on inputs. System integrates with digital fluoroscopy and mapping systems (e.g., AcQMap). Benefits include remote navigation of catheters to target sites in heart and vasculature. Output is physical orientation and movement of catheter tip, visualized by physician via fluoroscopy/mapping systems to guide clinical decision-making during ablation procedures.

Clinical Evidence

No new clinical trials conducted for this submission. Safety and efficacy supported by historical clinical data from 511 patients across four prospective studies (ATTRAC, ATTRAC II, HEART, VERSATILE). Reported major adverse event rate 3.9%. Acute success rate 95.0%; 90-day success rate 96.5%. Literature review of >8,000 patients showed 0.72% major complication rate for MNS vs 2.1% for manual control.

Technological Characteristics

Permanent magnet-based navigation system; generates 0.08T or 0.1T magnetic fields. Navigant Workstation software controls magnet orientation. Compatible with specific EP mapping/ablation catheters. Non-sterile system; magnet pods covered by drapes, cleaned with hospital-grade germicides. Software developed per ISO 62304 and ISO 14971. Connectivity via Open API for third-party mapping systems and digital fluoroscopy integration.

Indications for Use

Indicated for navigation of compatible magnetic devices through tissue to target sites in right/left heart, coronary, neurovascular, and peripheral vasculature. Cardiodrive indicated for automatic advancement/retraction of compatible magnetic EP mapping and ablation catheters in the heart. Contraindicated for advancement of EP catheters through coronary vasculature/sinus and for advancement/retraction of non-compatible devices into neurovasculature.

Regulatory Classification

Identification

A steerable cardiac ablation catheter remote control system is a prescription device that is external to the body and interacts with the manual handle of a steerable cardiac ablation catheter to remotely control the advancement, retraction, rotation, and deflection of a compatible, steerable ablation catheter used for the treatment of cardiac arrhythmias in the right side of the heart. The device allows reversion to manual control of the steerable cardiac ablation catheter without withdrawal of the catheter and interruption of the procedure.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical mechanical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance testing must be performed: (i) Mechanical performance of the system (without catheter connected); (ii) Mechanical performance of the system with compatible catheters connected to verify that the system does not impact catheter function or performance. Assessments must include the following: (A) Side-by-side remote control and manual comparisons of catheter manipulation (including all ranges of motion of catheter deflection and tip curl) for all compatible catheters; must include testing for worst-case conditions, and (B) Evaluation of the accuracy and function of all device control safety features; and (iii) Simulated-use testing in a bench anatomic model or animal model. (2) Non-clinical electrical testing must include validation of electromagnetic compatibility (EMC), electrical safety, thermal safety, and electrical system performance. The following performance testing must be performed: (i) Electrical performance of the system with compatible catheters connected to verify that the system does not impact catheter function or performance. Assessments must include the following: (A) Side-by-side remote control and manual comparisons of catheter manipulation (including all ranges of motion of catheter deflection and tip curl) for all compatible catheters; must include testing for worst-case conditions, and (B) Evaluation of the accuracy and function of all device control safety features; and (ii) Electrical safety between the device and ablation catheter system and with other electrical equipment expected in the catheter lab or operating room. (3) In vivo testing must demonstrate that the device performs as intended under anticipated conditions of use, including an assessment of the system impact on the functionality and performance of compatible catheters, and documentation of the adverse event profile associated with clinical use. Evidence must be submitted to address the following: (i) Manipulation and Positioning: Ability to manipulate compatible catheters to pre-specified cardiac locations and confirm proper anatomic placement and tissue contact, in accordance with the system indications for use and the compatible catheter indications for use; (ii) Safety: Assess device-related complication rate and major procedural complication rate (regardless of device relatedness) in comparison to literature and/or a manual comparison group for compatible ablation catheters to support the indications for use; (iii) Efficacy: Assess ablation success in comparison to literature and/or a manual comparison group for compatible ablation catheters to support the indications for use; and (iv) User assessment of device remote controls and safety features. (4) Post-market surveillance (PMS) must be conducted and completed in accordance with FDA agreed upon PMS protocol. (5) A training program must be included with sufficient educational elements that, upon completion of the training program, the clinician and supporting staff can: (i) Identify the safe environments for device use, (ii) Use all safety features of device, and (iii) Operate the device in simulated or actual use environments representative of indicated environments and use for the indication of compatible catheters. (6) Performance data must demonstrate the sterility of the sterile disposable components of the system. (7) Performance data must support shelf life by demonstrating continued sterility of the device (of the sterile disposable components), package integrity, and device functionality over the requested shelf life. (8) Labeling must include the following: (i) Appropriate instructions, warnings, cautions, limitations, and information related to the intended patient population, compatible ablation catheters, and the device safeguards for the safe use of the device; (ii) Specific instructions and the clinical training needed for the safe use of the device, which includes: (A) Instructions on assembling the device in all available configurations, including installation and removal of compatible catheters; (B) Instructions and explanation of all controls, inputs, and outputs; (C) Instructions on all available modes or states of the device; (D) Instructions on all safety features of the device; and (E) Validated methods and instructions for reprocessing/disinfecting any reusable components; (iii) A detailed summary of the mechanical compatibility testing including: (A) A table with a complete list of compatible catheters tested (manufacturer trade name and model number), and (B) A table with detailed test results, including type of test, acceptance criteria, and test results ( *i.e.,* pass for meeting acceptance criteria);(iv) A detailed summary of the in vivo testing including: (A) A table with a complete list of compatible catheters used during testing (manufacturer trade name and model number); (B) Adverse events encountered pertinent to use of the device under use conditions; (C) A detailed summary of the device- and procedure-related complications; and (D) A summary of study outcomes and endpoints. Information pertinent to the fluoroscopy times/exposure for the procedure, patient, and operator fluoroscopic exposure; (v) Other labeling items: (A) A detailed summary of pertinent non-clinical testing information: EMC, mechanical, electrical, and sterilization of device and components; (B) A detailed summary of the device technical parameters; and (C) An expiration date/shelf life and storage conditions for the sterile accessories; and (vi) When available, and according to the timeframe included in the PMS protocol agreed upon with FDA, provide a detailed summary of the PMS data including: (A) Updates to the labeling to accurately reflect outcomes or necessary modifications based upon data collected during the PMS experience, and (B) Inclusion of results and adverse events associated with utilization of the device during the PMS.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. October 30, 2019 Stereotaxis, Inc. Kenneth Lock Sr. Director of Clinical Affairs 4320 Forest Park Avenue, Suite 100 St. Louis, Missouri 63108 Re: K192775 Trade/Device Name: Niobe@ Magnetic Navigation System (MNS) with Navigant Workstation (NWS) and Cardiodrive Regulation Number: 21 CFR 870.5700 Regulation Name: Steerable Cardiac Ablation Catheter Remote Control System Regulatory Class: Class II Product Code: PJB, NDO Dated: September 27, 2019 Received: September 30, 2019 Dear Kenneth Lock: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Nicole Gillette Acting Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192775 #### Device Name Niobe® Magnetic Navigation System (MNS) with Navigation Workstation (NWS) and Cardiodrive #### Indications for Use (Describe) The Niobe System is intended to navigate compatible magnetic devices through tissue to designated target sites in the right and left heart and coronary vascular and peripheral vasculature by orienting the device tip in a desired direction. The Cardiodrive Catheter Advancement System (CAS) is intended to automatically advance and retract compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart when used in conjunction with a Stereotaxis MNS. The Cardiodrive system is not intended to advance the EP mapping and ablation catheters through the coronary vasculature or the coronary sinus. The Cardiodrive system is not intended to advance or retract non-compatible catheters and/or other non-compatible devices into the neurovasculature. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------| | <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### SUBMITTER | Date Prepared: | October 30, 2019 | |-----------------|---------------------------------------------------------------------------------------------------| | Submitter: | Stereotaxis, Inc.<br>4320 Forest Park Avenue<br>St. Louis, MO 63108 USA | | Contact Person: | Kenneth H. Lock<br>Sr. Director, Clinical Affairs<br>Phone: (314) 678-6123<br>Fax: (763) 493-7849 | ### DEVICE Name of Device: Niobe® Magnetic Navigation System (MNS) with Navigant™ Workstation (NWS) and Cardiodrive® Common or Usual Name: Steerable cardiac ablation catheter remote control system. Classification Name: Catheter remote control system Regulatory Class: Class II Regulation Number: 21 CFR 870.5700 Product Code: PJB FDA Panel: Cardiology PREDICATE DEVICE | Predicate | 510(k) | Regulation | Device Name | Manufacturer | |-----------|---------|------------|----------------------------------------------------------------------------------------------------|-------------------| | Primary | K183027 | 870.5700 | Niobe® Magnetic Navigation<br>System (MNS) with Navigant™<br>Workstation (NWS) and<br>Cardiodrive® | Stereotaxis, Inc. | # DEVICE DESCRIPTION The Stereotaxis Niobe Magnetic Navigation System (MNS) with Navigant Workstation (NWS) and Cardiodrive (a.k.a. the Niobe System) is an interventional workstation for the intravascular navigation of appropriately equipped. magnetically adapted, devices (e.g., catheters or guidewires) through tissue to designated target sites. The Niobe System uses computer-controlled permanent magnets for orientating the tip of a magnetic device and employs magnetic fields only to orient or steer the tip of a magnetic device and remotely advance and retract only compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart. The Niobe System incorporates software that determines the direction the magnetic field should be applied based on physician interaction with the user interface devices. ### INDICATIONS FOR USE The Niobe System is intended to navigate compatible magnetic devices through tissue to designated target sites in the right and left heart and coronary vasculature, neurovascular and peripheral vasculature by orienting the device tip in a desired direction. {4}------------------------------------------------ The Cardiodrive Catheter Advancement System (CAS) is intended to automatically advance and retract compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart when used in conjunction with a Stereotaxis MNS. The Cardiodrive system is not intended to advance the EP mapping and ablation catheters through the coronary vasculature or the coronary sinus. The Cardiodrive system is not intended to advance or retract non-compatible catheters and/or other noncompatible devices into the neurovasculature. # COMPARISON OF TECHNOLOGICAL CHARACTERISTICS The Niobe System has the same primary technological characteristics as the predicate device, including, but not limited to the following: Primary Predicate Device: - Remotely controlling PMA class III ablation catheter ● Additional Reference Devices - Primary Design Elements (magnets, positioners, software) ● - . Principle of Operation (magnetic navigation) - Performance Parameters (field modes and positions) - User Interface Devices (controllers, keyboard, display monitors) ● - . Companion Devices and Systems (digital fluoroscopy systems, mapping systems) # CHANGES MADE TO THE PROPOSED DEVICE Software modifications in the proposed device with respect to the predicate device: - Onen Mapping Application Programming Interface (Open API) that allows physicians to use their . preferred mapping system with the Niobe MNS. AcQMap® Mapping System from Acutus Medical is now compatible with the Niobe System - . Companion digital fluoroscopy system from Omega Imaging LLC was added to the list of compatible fluoroscopy systems - . User interface changes - Improved some informational messages O - Better sizing of registration image O - Optimize window layout for quad screen O - O Provide import of pre-procedure surface images - Added an ablation history informational graph O Any minor differences in technological characteristics between the proposed and reference Niobe System have demonstrated continued performance and do not raise new issues of safety or effectiveness. ### NON-CLINICAL STUDIES The Niobe System underwent verification and validation activities to ensure that the device continues to meet established performance criteria. Evaluations included: {5}------------------------------------------------ - . Biocompatibility Testing The Niobe System does not contain patient-contacting components. Biocompatibility testing was not required to support of this submission. - Electromagnetic Compatibility (EMC) ● Since the changes in the subject device are to software only, Electromagnetic compatibility testing (EMC) was not completed. - Software Verification and Validation ● Stereotaxis performed software verification and validation activities under design controls in accordance with ISO 62304 and ISO 14971 for software utilized by the Niobe System device to ensure that the software is safe and appropriate risk control measures are effective. ## Performance Testing For the changes for the subject device, system verification testing were conducted to assess the performance of the device following software modifications. The testing demonstrated that the Niobe System met pre-determined requirements and continued to perform as intended. ## Special Controls Testing Summary The following updates the additional testing that was conducted to meet the special controls requirements. Special Controls testing was leveraged from the predicate device since the software modifications had no effect on the performance of the Niobe System. In Special Control 1) iii, in the table below, animal testing was conducted to test the integration of the AcQMap system with the Niobe System. | Special Controls | How Special Control Has Been Met | | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | 1) Non-clinical mechanical performance<br>testing must demonstrate that the device<br>performs as intended under anticipated<br>conditions of use. The following<br>performance testing must be performed:<br>i. Mechanical performance of the system<br>(without catheter connected); | Using the proposed device, Magnetic field performance testing<br>without the catheter connected uses a THM-7025 Hall Effect tesla<br>meter to measure the applied field strength and accuracy in the<br>cardinal direction and at 0.08T and 0.1T (which represent worst<br>case positions). | | | ii. Mechanical performance of the system<br>with compatible catheters connected to<br>verify that the system does not impact<br>catheter function or performance.<br>Assessments must include the<br>following:<br>(A) Side-by-side remote control and<br>manual comparisons of catheter<br>manipulation (including all ranges<br>of motion of catheter deflection and<br>tip curl) for all compatible<br>catheters; must include testing for<br>worst-case conditions, and | Manual Control Performance Testing<br>Performance testing data for the reference Niobe device included<br>deflection testing, accuracy testing with target phantom and<br>anatomical position testing with a plastic heart model. The results<br>of this testing provide mechanical performance data for the<br>reference device.<br>Remote Control Performance Testing<br>The current version (proposed device) of the Niobe MNS<br>Substantial Equivalence Application Testing demonstrates<br>equivalent mechanical performance between the subject device<br>and reference device. | | | Special Controls | How Special Control Has Been Met | | | (B) Evaluation of the accuracy and<br>function of all device control safety<br>features; and | Testing of the four primary safety controls related to the physical<br>motion of the system was conducted on the proposed device.<br>These concern physical movement of the magnet positioners,<br>movement of the covers toward the patient, continuous<br>advancement of the catheter, and motion e-stops. All four safety<br>controls passed the test. | | | iii.<br>Simulated-use testing in a bench<br>anatomic model or animal model. | Four non-GLP animal validation studies were conducted and<br>tested the following using iterations of software after the<br>reference device was cleared:<br>Navigant 4.4 features and ablation lesion assessment<br>• Sheath visualization<br>• Zone mapping<br>Navigant 5.0.1 Validation Study<br>• Auto NaviLine ./ single step NaviLine<br>Navigant 5.0.2 Validation Study<br>• NaviLine<br>• Targeting<br>• Point and line ablation<br>Navigant 5.0.3 Validation Study<br>• Targeting / Electrode Targeting<br>• NaviLine<br>• Point and Line Ablation<br>Navigant 5.0.5<br>• Integration of the Acutus AcQMap system with the Niobe<br>System<br>• Acutus Workflow integration | | | Special Controls | How Special Control Has Been Met | | | 2.<br>Non-clinical electrical testing must<br>include validation of electromagnetic<br>compatibility (EMC), electrical safety,<br>thermal safety, and electrical system<br>performance. The following performance<br>testing must be performed:<br>i. | Manual Control Performance Testing<br>EMC and Electrical safety testing showing conformance with IEC<br>60601 were performed by TuV for the reference device. | | | i.<br>Electrical performance of the system<br>with compatible catheters connected<br>to verify that the system does not<br>impact catheter function or<br>performance. Assessments must<br>include the following:<br>A. Side-by-side remote control and<br>manual comparisons of catheter<br>manipulation (including all<br>ranges of motion of catheter<br>deflection and tip curl) for all<br>compatible catheters; must<br>include testing for worst-case<br>conditions, and | Remote Control Performance Testing<br>EMC testing for the proposed device hardware demonstrating<br>conformance with IEC 60601 was performed by TuV. | | | | All tests passed. | | | B. Evaluation of the accuracy and<br>function of all device control<br>safety features; and | Electrical safety testing on the proposed device demonstrating<br>compliance with IEC 60601, was performed by TuV. | | | ii.<br>Electrical safety between the device<br>and ablation catheter system and with<br>other electrical equipment expected in<br>the catheter lab or operating room. | The proposed system has been tested for compatibility with<br>specific x-ray, ablation generators, mapping systems. In addition,<br>electrical isolation and emissions testing have been performed by<br>TuV.<br>All tests passed. | | | 3.<br>In vivo testing must demonstrate that the<br>device performs as intended under<br>anticipated conditions of use, including an<br>assessment of the system impact on the<br>functionality and performance of<br>compatible catheters, and documentation<br>of the adverse event profile associated<br>with clinical use. Evidence must be<br>submitted to address the following:<br>i.<br>Manipulation and Positioning:<br>Ability to manipulate compatible<br>catheters to pre-specified cardiac<br>locations and conform proper<br>anatomic placement and tissue<br>contact, in accordance with the<br>system indications for use and<br>compatible catheter indications for<br>use. | The VERSATILE Study used to support K141530 as well as in<br>vivo animal testing provided in K071029 for the Cardiodrive<br>system, demonstrated the ability to manipulate compatible<br>catheters to pre-specified cardiac locations. | | | Special Controls | How Special Control Has Been Met | | | ii. Safety: Assess device-related complication rate and major procedural complication rate (regardless of device relatedness) in comparison to literature and/or manual comparison group for compatible ablation catheters to support the indications for use; | Clinical data to support the safety of the Niobe Magnetic Navigation System, which includes Niobe, Navigant Software, the Cardiodrive catheter advancement system and a compatible ablation catheter, was reviewed.<br>The following is a summary of data from clinical studies sponsored by Stereotaxis that have been submitted to the FDA in three separate submissions:<br>• P050029 Helios II ablation catheter (ATTRAC, ATTRAC II and HEART)<br>• K071029 Cardiodrive with Ablation Catheter (ATTRAC II)<br>• K140804 V-CAS (VERSATILE)<br>A total of 511patients from 4 studies were enrolled using the Niobe Magnetic Navigation System, which included the Niobe MNS, Cardiodrive, a Magnetic Ablation Catheter, and the Navigant Software.<br>The following table describes the major complication rates reported in these trials. | | | | Study | RMN Patients | | | ATTRAC | 182 | | | ATTRAC II | 80 | | | HEART Study | 129 | | | VERSATILE | 120 | | | Total | 511 | {6}------------------------------------------------ {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ | Special Controls | How Special Control Has Been Met | | | | | | | | | | | | | | | | | | | | | | | | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--| | iii. Efficacy: Assess ablation success in comparison to literature and/or a manual comparison group for compatible ablation catheters to support the indications for use; and | Data to support the ablation success of the Niobe Magnetic Navigation System which includes Niobe, Navigant Software, the Cardiodrive catheter advancement system and a compatible ablation catheter was reviewed.<br>The following is a summary of data from clinical studies sponsored by Stereotaxis that have been submitted to the FDA in three separate submissions:<br>P050029 Helios II ablation catheter (ATTRAC ,ATTRAC II and HEART study) K071029 Cardiodrive with Ablation Catheter (ATTRAC) K140804 V-CAS (VERSATILE) A total of 511 patients from 4 studies were enrolled using the Niobe Magnetic Navigation System, which included the Niobe MNS, Cardiodrive, a Magnetic Ablation Catheter, and the Navigant software. The ATTRAC study series and the HEART Study used the Helios ablation catheter, and the Versatile study used the BWI RMT Thermocool catheter. The following table describes the acute and 90-day success rates reported in these trials.<br>Study RMN Patients Acute Success RMN 90 day success RMN ATTRAC 182 175/182 (96.2%) 145/147 (98.6%) ATTRAC II 80 71/75 (94.7%) 51/54 (94.4%) HEART Study 129 108/121 (89.3) 82/87 (94.3%) VERSATILE 120 119/120 (99.2%) Not reported Total 511 473/498 (95.0%) 278/288 (96.5%) | | | | | | | | | | | | | | | | | | | | | | | | | | iv. User assessment of device remote controls and safety features. | The Navigant system was used to steer the catheter during the evaluations performed for Navistar RMT, Celsius RMT and Helios II. Performance impressions were documented on physician feedback forms which are included in the reports. | | | | | | | | | | | | | | | | | | | | | | | | | | 4. Post-market surveillance (PMS) must be conducted and completed in accordance with FDA agreed upon PMS protocol. | Given the amount of clinical data from submitted studies and peer reviewed publications, a post market surveillance study is not warranted.<br>The following summarizes the overwhelming evidence available to support the safety and efficacy of the MNS system: Four studies submitted to the FDA demonstrated a major adverse event rate of 3.9%. Data from these four studies supported 510(k) clearances for K071029, and K140804 and PMA approval in P050029. Acute success and 90 success rates were 95.0% and | | | | | | | | | | | | | | | | | | | | | | | | | {10}------------------------------------------------ | Special Controls | How Special Control Has Been Met | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | 96.5% respectively in those studies More than 8,000 patients were reported in the literature using the MNS System with a major complication rate of 0.72% compared to a manual rate of 2.1% Acute success rates and long-term success rates were similar in both the MNS and manual groups Stereotaxis proposes to continue monitoring the safety and efficacy of the Niobe System through the literature and Post Marketing Surveillance program. | | 5. A training program must be included with sufficient educational elements that, upon completion of the training program, the clinical and supporting staff can:<br>i. Identify the safe environments for device use,<br>ii. Use all safety features of the device, and<br>iii. Operate the device in simulated or actual use environments representative of indicated environments and use for the indication of compatible catheters. | Representatives from the company train the physician and staff in the use of the Niobe Magnetic Navigation system.<br><br>The Niobe ES User's Manual provides detailed operating instructions on the system and the navigation software.<br><br>This information is reviewed with the physician and staff during simulation sessions and phantom training with the actual system | | 6. Performance data must demonstrate the sterility of the sterile disposable components of the system | The Niobe System is not provided in sterile form nor is it required to be sterilized prior to use, therefore, sterilization information is not applicable. The Niobe System does not include sterile disposable components as part of the system.<br><br>Cardiodrive includes a single-use disposable (QuikCAS) to interface with the compatible ablation catheter. The sterile components underwent sterilization testing. All sterilization testing resulted in a PASS. | | 7. Performance data must support shelf life by demonstrating continued sterility of the device (of the sterile disposable components), package integrity, and device functionality over the requested shelf life. | The Niobe System is not provided in sterile form nor is it required to be sterilized prior to use, therefore, shelf life information, including package integrity and function of the device over the stated shelf life are not applicable.<br><br>Cardiodrive includes a single-use disposable (QuikCAS) to interface with the compatible ablation catheter. The sterile components underwent shelf-life /packaging and sterilization testing. All sterilization and packaging testing resulted in a PASS, and each component was validated for a shelf life of 3 years. | | Special Controls | How Special Control Has Been Met | | 8. Labeling must include:<br>i. Appropriate instructions, warnings, cautions, limitations, and information related to the intended patient population, compatible ablation catheters, and the device safeguards for the device;<br>ii. Specific instructions and clinical training needed for the safe use of the device, which includes: | Niobe ES User's Manual includes:<br>• Compatible Catheters<br>• Indications for Use<br>• Warnings<br>• Safety Controls<br>The Niobe System does not require assembly of the device prior to use. The Niobe System is considered permanent equipment installed in a Stereotaxis Magnetic Navigation Lab. Stereotaxis performs all installation activities for each Niobe System; therefore, assembly and installation instructions to the physician user are not required. | | A. Instructions on assembling the device in all available configurations, including installation and removal of compatible catheters; | With regard to use of magnetically-compatible devices with the Niobe System, assembly and installation of these devices to/with the Niobe System is not required. Magnetically-compatible catheters interface with the Cardiodrive CAS device which advances and retracts the device and instructions related to the catheter-CAS interface are provided in the Cardiodrive instructions for use. The Niobe System only applies a magnetic field to orient the distal tip of a magnetically-compatible device; therefore, the manipulation of the distal end of the device is by indirect (non-physical) means for which assembly or installation of other devices is not required. | | B. Instructions on all available modes or states of the devices;<br>C. Instructions and explanation of all controls, inputs, and outputs<br>D. Instructions on all safety features of the device and<br>E. Validated methods and instructions for reprocessing/disinfecting any reusable components | The Niobe System User Guide includes an explanation and instructions for the following, controls, modes, and states of the device:<br>• Procedure room components<br>• System positions<br>• Tableside Magnet Controller<br>• System power up<br>• Cover Force Sensor<br>• Navigation Position Assistance<br>• Software Basic Information<br>• Cardiodrive CAS user interface<br>• Activation codes<br>The Niobe System is a medical device that does not have any reusable components that reprocessing or disinfection prior to use. The magnet pods are covered by drapes during the procedure and may be wiped clean using hospital-grade EPA-registered germicide solutions following each procedure. Cleaning instructions are provided in the "Cleaning the Niobe System" section of the Niobe System User Guide). | | Special Controls | How Special Control Has Been Met | | iii.<br>A detailed summary of the<br>mechanical compatibility testing<br>including:<br>A. A table with a complete list of<br>compatible catheters tested<br>(manufacturer trade name and<br>model number), and | - Biosense Webster Navistar RMT<br>- Biosense Webster Navistar RMT Thermocool<br>- Biosense Webster Celsius RMT<br>- Biosense Webster Celsius RMT Thermocool<br>- Stereotaxis Helios II | | B. A table with detailed test results,<br>including type of test, acceptance<br>criteria, and test results (i.e., pass<br>for meeting acceptance criteria); | Mechanical performance testing of compatible catheter is<br>included in the Niobe Users Guide. | | iv.<br>A detailed summary of the in vivo<br>testing including:<br>A. A table with a complete list of<br>compatible catheters used during<br>testing (manufacturer trade name<br>and model number); | - Biosense Webster Navistar RMT<br>- Biosense Webster Navistar RMT Thermocool<br>- Biosense Webster Celsius RMT<br>- Biosense Webster Celsius RMT Thermocool<br>- Stereotaxis Helios II | | B. Adverse events encountered<br>pertinent to use of the device<br>under use conditions; | Data from four clinical studies involving 511 patients who<br>underwent catheter ablation using the Stereotaxis Magnetic<br>Navigation System (MNS) are summarized. Collectively, the data<br>in these studies demonstrates the safety and effectiveness of the<br>Magnetic Navigation System for catheter ablation.<br><br>Study Design:<br>All four studies were prospective in nature and included safety<br>endpoints. Data from all four studies were used to support<br>regulatory approvals. Evaluation of 7 day major adverse events<br>for safety is reported.<br><br>Study Results:<br>Adverse Events:<br>The overall 7 day major complication rate for all four studies was<br>20/511 (3.9%) Major adverse events that occurred within 7 days<br>post procedure included:<br>- 1 cardiac tamponade related to right sided<br>catheter<br>- 1 cardiac tamponade related to the transseptal<br>puncture…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...