K192771 · C4 Imaging, LLC · NEU · Oct 21, 2020 · General, Plastic Surgery
Device Facts
Record ID
K192771
Device Name
C4 Fiducial Marker
Applicant
C4 Imaging, LLC
Product Code
NEU · General, Plastic Surgery
Decision Date
Oct 21, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Indications for Use
The C4 Fiducial Marker is indicated to be implanted into the body in situations where the location of specific anatomy, normal or diseased, needs to be marked for a future medical procedure. The device can be visualized using MRI, CT or x-ray, and provides a reference from which other procedures can be guided.
Device Story
C4 Fiducial Marker is an implantable radiographic marker used to identify specific anatomy for future medical procedures. Device consists of a sealed PEEK polymer tube containing a solution of cobalt chloride and N-Acetylcysteine, with zirconium oxide sealed within the shell. It is implanted into the body by a physician. The marker is visualized using MRI, CT, or x-ray imaging modalities. By providing a stable reference point, it assists clinicians in guiding subsequent medical procedures, potentially improving accuracy and patient outcomes. It is a passive device; no software or active processing is involved.
Clinical Evidence
Bench testing only. Performance testing included structural and functional integrity (negative/positive pressure, axial/lateral load), MRI, CT, and x-ray imaging in tissue-equivalent phantoms, biocompatibility assessment per ISO 10993, and endotoxin testing.
Technological Characteristics
Materials: PEEK polymer shell, cobalt chloride (up to 1%), N-Acetylcysteine (up to 2%), zirconium oxide. Dimensions: 5.5-10.0 mm length, 1.0 mm (+/- 0.2 mm) diameter. Passive implantable marker. Visualization: MRI, CT, x-ray. Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for patients requiring anatomical marking of normal or diseased tissue for future medical procedures; implantable for visualization via MRI, CT, or x-ray to guide subsequent interventions.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
October 21, 2020
C4 Imaging LLC Andrew Bright President and CEO 196 West Ashland Street Doylestown, Pennsylvania 18901
Re: K192771
Trade/Device Name: C4 Fiducial Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: September 26, 2019 Received: September 30, 2019
Dear Mr. Bright:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192771
Device Name C4 Fiducial Marker
Indications for Use (Describe)
The C4 Fiducial Marker is indicated to be implanted into the body in situations where the location of specific anatomy, normal or diseased, needs to be marked for a future medical procedure. The device can be visualized using MRI, CT or x-ray, and provides a reference from which other procedures can be guided.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <svg height="12" width="12"> <polygon points="0,0 12,0 12,12 0,12" style="fill:none;stroke:black;stroke-width:1"></polygon> <polyline points="0,0 12,12" style="fill:none;stroke:black;stroke-width:1"></polyline> <polyline points="12,0 0,12" style="fill:none;stroke:black;stroke-width:1"></polyline> </svg> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | |
| <div> <span> <svg height="12" width="12"> <polygon points="0,0 12,0 12,12 0,12" style="fill:none;stroke:black;stroke-width:1"></polygon> </svg> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[as required by section 807.92(c)]
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
## General Information
| Submitted by: | C4 Imaging, LLC<br>196 West Ashland Street<br>Suite 109<br>Doylestown, PA 18901 USA<br>Phone: +1 (609) 933-5895<br>Email: abright@c4imaging.com<br><br>Contact Person: Andrew Bright |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | October 20, 2019 |
# Device Name
| Trade Name: | C4 Fiducial Marker |
|-----------------|--------------------------|
| Common Name(s): | Anatomical tissue marker |
# Classification
| Regulation: | 21 CFR §878.4300 |
|----------------------|-----------------------------------|
| Class: | Class II |
| Product Code: | NEU |
| Classification name: | Marker, Radiographic, Implantable |
{4}------------------------------------------------
## Predicate Devices
| Mixed media marker | K102506 | Cortex Manufacturing |
|-------------------------|---------|--------------------------------|
| BiomarC Fiducial Marker | K063193 | Carbon Medical<br>Technologies |
#### Device Description
The C4 Fiducial Marker consists of a sealed polyether ether ketone (PEEK) polymer tube containing a solution of up to 1% cobalt chloride and up to 2% N-Acetylcysteine, The ratio of cobalt chloride to N-Acetylcysteine is 1:2. Zirconium oxide is sealed within the polymer shell of the device. The length of the polymer capsule is between 5.5 mm and 10.0 mm, and the diameter is 1.0 mm (+/- 0.2 mm).
#### Indication
The C4 Fiducial Marker is indicated to be implanted into the body in situations where the location of specific anatomy, normal or diseased, needs to be marked for a future medical procedure. The device can be visualized using MRI, CT or x-ray, and provides a reference from which other procedures can be guided.
## Comparison to Predicate Devices
The indication and technological characteristics of the C4 Fiducial Marker are substantially equivalent to the Mixed Media Marker (Cortex Manufacturing -K102506) and the BiomarC Fiducial Marker (Carbon Medical Technologies -K063193). Like the Mixed Media Marker, the C4 Fiducial Marker is comprised of a sealed polyether ether ketone (PEEK) capsule that encloses an internal marker for MRI. CT and X-ray imaging. In both cases the PEEK capsule is the only patient contacting material. Like the BiomarC Fiducial, the C4 Fiducial Marker has a sealed biocompatible shell that encloses a zirconium oxide marker.
The C4 Fiducial Marker was also compared to a reference device. the C4 Imaging MRI Marker NS (K171487), which has similar technological characteristics. The C4 Fiducial Marker and the MRI Marker NS both comprise a PEEK shell enclosing a cobalt chloride N-acetyl cysteine solution.
A risk evaluation, based on a Failure Mode Effect Analysis (FEMA), determined that technological differences were not considered to have an impact on the safety or effectiveness of the C4 Fiducial Marker. Adoption evaluation of biocompatibility, MR compatibility, sterilization, structural integrity, packaging and shelf life data during the risk assessment confirmed the C4 Fiducial Marker is substantially equivalent to the predicate devices and reference device.
{5}------------------------------------------------
# Performance Testing
The C4 Fiducial Marker was subject to structural and functional integrity testing involving negative and positive pressure, as well as axial and lateral load testing. These integrity tests were followed by MRI, CT and x-ray imaging demonstrating functional performance. Separate MRI, CT and x-ray imaging tests were performed in tissue equivalent phantoms. A biocompatibility (ISO 10993) assessment and endotoxin tests were performed on the device.
### Conclusion
The C4 Fiducial Marker is substantially equivalent to the Mixed Media Marker (K102506) and to the BiomarC Fiducial Marker (K063193). There are no differences that would affect the substantial equivalence of the proposed device
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.