K192770 · Aea Srl · FMF · Dec 17, 2020 · General Hospital
Device Facts
Record ID
K192770
Device Name
S dispensing line
Applicant
Aea Srl
Product Code
FMF · General Hospital
Decision Date
Dec 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The S Dispensing Line is intended for transfer of liquid drug from bag to syringes through the APOTECAsyringe automated system. The device is used inside an automatic system APOTEC Asyringe that tightens the tube with the syringe, while the connection with the bag is done manually. The transfer of liquid takes place by depression through the mechanical action on the piston of the syringe of a manipulator equipped with a gripping device.
Device Story
S Dispensing Line is a single-use, non-vented infusion set for transferring liquid drugs from bags to syringes. It is designed for use exclusively within the APOTECAsyringe automated compounding system. The device consists of a vented spike, PVC tubing, a check valve, and connectors. During operation, the spike is manually connected to a drug bag, while the automated system manipulates the syringe piston to create depression, drawing liquid through the line. A check valve prevents backflow. The device ensures secure connection to the syringe via a pressure connector. It is intended for use in clinical or pharmacy settings where automated compounding is performed. By automating the transfer process, the device reduces manual handling, potentially improving accuracy and sterility during drug preparation.
Clinical Evidence
No clinical data. Bench testing only. Performance testing confirmed connection integrity, absence of leakage, and durability after 100 usages. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity, and hemocompatibility, all complying with ISO 10993 standards. Sterilization validation performed per ISO 11135.
Technological Characteristics
Non-vented infusion set. Materials: ABS (spike, connector), LDPE (cap), Polyethylene (vent), PVC D.E.H.P. Free (tubing, elbow), Silicone (check valve), Acrylic (check valve), Polycarbonate (luer slip). Complies with ISO 10993 series. Sterilized via ethylene oxide. Single-use, blister-packaged. Mechanical operation via automated system vacuum/piston manipulation.
Indications for Use
Indicated for the transfer of liquid drugs from a bag to syringes within the APOTECAsyringe automated system. No specific patient population, age, or disease state is defined.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
December 17, 2020
AEA srl Michele Mengoni Quality & Regulatory via Fiume 16 Angeli di Rosora, 60030 It
Re: K192770
Trade/Device Name: S dispensing line Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: LHI Dated: December 10, 2020 Received: December 16, 2020
Dear Michele Mengoni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan Stevens Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192770
Device Name S Dispensing Line
#### Indications for Use (Describe)
The S Dispensing Line is intended for transfer of liquid drug from bag to syringes through the APOTECAsyringe automated system. The device is used inside an automatic system APOTEC Asyringe that tightens the tube with the syringe, while the connection with the bag is done manually. The transfer of liquid takes place by depression through the mechanical action on the piston of the syringe of a manipulator equipped with a gripping device.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "LOCCIONI" in a bold, sans-serif font. The letters are all capitalized and are a dark red color. The background is white, which makes the text stand out.
# 510(k) Summary
Thefollowing510(k)summaryhasbeenpreparedpursuanttorequirementsspecifiedin21 CFR807.92.
# I SUBMITTER
AEA SRL ViaFiume, 16AngelidiRosa60030AnconaItaly Phone: 0039-0731-816689 Contact Person: Michele Mengoni Date prepared: December 10th , 2020
## II DEVICE
| TradeNames: | S dispensing line |
|-----------------------|----------------------------------------------------------|
| CommonorUsualNames: | Intravascular administration set |
| Classification Names: | Set, IV Fluid Transfer |
| Regulation Numbers: | Sdispensingline: 880.5440-Intravascularadministrationset |
| Regulatory Class: | II |
| ProductCodes: | S dispensing line: LHI |
| PanelIdentification: | General Hospital |
## III PREDICATE DEVICES
| S dispensing line | Predicate device | Predicate device |
|-------------------|---------------------|----------------------------------|
| Trade Name: | Rapid-Fill™ Tubeset | Exacta-Mix<br>Administration Set |
| 510(k) number: | K022523 | K002705 |
# IV DEVICE DESCRIPTION
Sdispensinglineisadevice for the transferofiquids, with the purpose of preparation of in drugs. The device is intended to be used with APOTECAsyringe automatic compoundingsystem.
Thedeviceissingleuse, sterilizedbyEOandsingle-packagedinablister. The device is not manufactured with natural rubber latex.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "LOCCIONI" in a bold, sans-serif font. The color of the text is a dark red. The letters are evenly spaced and appear to be a logo or brand name.
Medical grade plastics are used, according to ISO 10993 series standards.
The device has the same intended use and the same technological characteristics as its predicate devices.
Sdispensingline is a non-vented infusion setused as a connecting part be rock syringe without hypodermic needle. The tip in contact with the syringe must perfectly by means of a pressure connector (not by screwind). in ordertoavoidloss ofmedication; the connectoris to be assembled atthe endof the line.
The short line enables the transfer of drug from a bag to a luer lock syringe through the APOTECAsyringe automated system.
The line is intended for the connection of the spike perforator; a check valve prevents the flow back towards thebag. Components and materials are reported in the following table.
| COMPONENT | MATERIAL |
|----------------------------------|----------------------------------------------------------------|
| Vented spike with protective cap | ABS (vented spike), LDPE (protective cap), Polyethylene (vent) |
| Tubing | PVC D.E.H.P. Free |
| Check valve | Silicone (internal part), Acrylic (upper and lower parts) |
| Straight connector | ABS |
| Elbow connector | PVC D.E.H.P. Free |
| Female luer slip connector | Polycarbonate |
## V INTENDED USE AND INDICATIONS FOR USE
The S dispensing line is intended for transfer of liquid drug from a bag to syringes through the APOTECAsyringe automated system.
The device is used inside an automatic system APOTECAsyringe that tightens the tube with the connection with the bagis done manually. The transfer of liquid takes place by depression through the piston of the syringe of a manipulator equipped with a gripping device.
## VI comparison of technological characteristics with The Predicate devices
S dispensing line has the intended use and the same technological characteristics of its predicate devices:
- . The S dispensing line has the same intended use of the predicated devices (K022523 and K002705).
- The S dispensing line and the predicate devices (K022523 and K002705) component materials comply with ISO 10993 ● as applicable to the intended use of the device.
- The S dispensing line hasvery similar design and materials of predicated devices (K022523 and K002705). .
The differences between the predicate devices and the subject devices include:
- The proximal end connector of the S dispensing line has been appropriately designed to be managed by APOTECAsyringe . compoundingsystemandatthesametimetofitwiththeluerlockofthesyringe used inside the automaticsystem.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "LOCCIONI" in a bold, sans-serif font. The letters are all capitalized and are a dark red color. The word appears to be a logo or brand name.
AEA srl Via Fiume 16,60030 Angelidi Rosora, AN (Italy) R.I. - C.F. - P. Iva 00686250424 info@loccioni.com
- The sterilization method for the proposed device S dispensing line is ethylene oxide whereas the sterilization method . for the predicate device is gamma irradiation.
The differences between proposed device and predicate devices do not raise any issues of safety and effectiveness.
#### VII. PERFORMANCE DATA
#### Non-Clinical Testing
AEA srl has performed non-clinicalion testing based on the risk analysis conducted, as summarized in the following tables. The favourable outcome of this testing demonstrates that the AEA proposed device performs in an equivalent manner to the predicate devices.
| Performance Characteristic | Acceptance Criteria/Standard | |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-----------------------------------------|
| Performance Testing Ref.<br>Section 018<br>"Performance Testing-<br>Bench" | Right connection/absence of<br>ruptures between syringe tip and<br>connector after 100 usages | Right connection, no ruptures |
| | Absence of leakage during filling after<br>100 usages | No leakage |
| | Right disconnection/absence of<br>ruptures between syringe tip and<br>connector after 100 usages | Right disconnection, no ruptures |
| | Absenceofspillingafterthe<br>disconnection | No spilling |
| | Performance testing after real aging<br>(ongoing) | No change in performances after 3 years |
| | Test for particulate contamination | ISO 8536-4:2010 |
| | Test for leakage | ISO 8536-4:2010 |
| | Test for tensile strength | ISO 8536-4:2010 |
| | Test for the closure-piercing device | ISO 8536-4:2010 |
| | Test for transparency | ISO 8536-4:2010 |
| Biocompatibility Testing Ref.<br>Section 015<br>"Biocompatibility" | Cytotoxicity MEM | ISO 10993-5:2009, ISO 10993-12:2012 |
| | Elution test | |
| | Sensitization | ISO 10993-10:2010, ISO 10993-12:2012 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "LOCCIONI" in a bold, sans-serif font. The letters are all capitalized and are a dark red color. The background is white, which makes the text stand out.
# AEA srl Via Fiume 16,60030 Angelidi Rosora, AN (Italy) R.I. - C.F. - P. Iva 00686250424 info@loccioni.com
| Performance Characteristic | | Acceptance Criteria/Standard |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| | Guinea Pig Sensitization Test | |
| | Irritation or Intracutaneous Reactivity<br>Rabbit Intracutaneous reactivity | ISO 10993-10:2010, ISO 10993-12:2012 |
| | Acute Systemic Toxicity | ISO 10993-11:2017, ISO 10993-12:2012 |
| | Material-Mediated Pyrogenicity<br>Pyrogen Test (USP Rabbit Test) | USP 41-NF36:2018 <151> Pyrogen Test<br>(USP Rabbit Test) |
| | Hemocompatibility<br>Haemolysis test indirect contact | ISO 10993-4:2017, ISO 10993-12:2012 |
| | Particulate matters testing | USP 788 |
| | Sterilization Validation | ISO 11135:2014 |
| Sterilization and Shelf Life<br>Ref. Section 014<br>"Sterilization and Shelf Life" | Packaging Validation | Effective microbiological barrier, product sterility and<br>integrity preservation |
| | Shelf life - 3 years | Sterility and product integrity maintained over the<br>entire shelf life |
| | | Sterility and product integrity maintained over 3 year<br>shelf life |
| | | 3 year Real Time Study (Ongoing)<br>Accelerated Ageing Study completed |
## Clinical Testing
Clinical testing was not required for this submission.
#### Substantial Equivalence
The S dispensing line is substantially equivalent to the predicate devices in intended use, principles of operation, technology, design, materials and performance.
## VII CONCLUSION
Results of performance and biocompatibility testing conducted on the proposed devices demonstrate that the is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.