K192764 · Neuroone, Inc. · GYC · Nov 26, 2019 · Neurology
Device Facts
Record ID
K192764
Device Name
NeuroOne Cortical Electrode
Applicant
Neuroone, Inc.
Product Code
GYC · Neurology
Decision Date
Nov 26, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NeuroOne Cortical Electrodes are intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals on the surface of the brain.
Device Story
NeuroOne Cortical Electrode is a sterile, single-use electrical conductor placed on the brain surface for <30 days. Device consists of platinum contacts mounted on thin polyimide film strips or grids. It functions as an interface between the brain surface and external, commercially available recording, monitoring, or stimulation instrumentation. Used in intraoperative and neurological monitoring settings by clinicians. Device records/monitors electrical activity or delivers stimulation to the brain. Benefits include temporary monitoring/stimulation capabilities for neurological assessment. Output is raw electrical signal data processed by external equipment; clinicians use this data to inform clinical decision-making regarding brain function.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included design verification (dimensional measurements, durability, mapping/stimulation, reliability), electrical testing per IEC 60601, biocompatibility per ISO 10993-1 and ASTM F756, packaging/shelf life per ASTM standards (D4169-16, F1980-16, 2096-11, F88-15), and sterilization validation per ISO 11135.
Technological Characteristics
Sterile, single-use cortical electrode. Contacts: Platinum. Film: Polyimide. Thickness: 0.0127 mm. Configuration: 2-16 contacts per electrode. Sensing/Actuation: Electrical signal recording and stimulation. Connectivity: Passive interface to external instrumentation. Sterilization: ISO 11135 validated.
Indications for Use
Indicated for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals on the surface of the brain.
Regulatory Classification
Identification
A cortical electrode is an electrode which is temporarily placed on the surface of the brain for stimulating the brain or recording the brain's electrical activity.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
NeuroOne, Inc. % Debra Kridner President, Debra J Kridner LLC, Regulatory Consultant Debra J Kridner 10901 Red Circle Dr., Suite 150 Minnetonka. Minnesota 55343
Re: K192764
Trade/Device Name: NeuroOne Cortical Electrode Regulation Number: 21 CFR 882.1310 Regulation Name: Cortical Electrode Regulatory Class: Class II Product Code: GYC Dated: September 28, 2019 Received: September 30, 2019
Dear Debra Kridner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192764
Device Name
NeuroOne Cortical Electrode
Indications for Use (Describe)
The NeuroOne Cortical Electrodes are intended for temporary (less than 30 days) use with recording, and stimulation equipment for the recording, and stimulation of electrical signals on the surface of the brain.
Type of Use (Select one or both, as applicable)
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "Neuroone" in a stylized font. The "Neuro" part of the word is in gray, while the "one" part is in blue. There is a circular graphic between the two words that is divided into four sections, colored red, light pink, light blue, and dark blue. The logo appears to be for a company or organization called Neuroone.
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| 510(k) Number: | |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | September 28, 2019 |
| Applicant: | NeuroOne, Inc.<br>10901 Red Circle Dr., Suite 150<br>Minnetonka, MN 55343<br>Email: info@NeuroOneinc.com<br>Website: www.NeuroOneinc.com |
| Contact Person: | Debra Kridner<br>President, Debra J Kridner Consulting, LLC<br>10901 Red Circle Dr., Suite 150<br>Minnetonka, MN 55343<br>Email: info@NeuroOneinc.com |
| SUBJECT DEVICE | |
| Trade Name: | NeuroOne Cortical Electrode |
| Common Name: | Cortical Electrode |
|----------------------|--------------------------------------|
| Classification Name: | Cortical electrode (21 CFR§882.1310) |
| Device Class: | Class 2 |
| Product Code | GYC |
#### DEVICE DESCRIPTION (For the Device Subject to This 510(k) Notification)
The NeuroOne Cortical Electrode is a sterile, single use electrical conductor that is temporarily placed (< 30 days) on the surface of the brain. The electrodes are designed to record and/or monitor the brain's electrical activity as well as stimulate the brain. It consists of a series of contacts mounted in thin plastic strips, or in a square/rectangular grid to cover larger surface areas. The device is connected to recording, monitoring, or stimulation/response instrumentation currently commercially available for use with other cortical electrodes, including the predicate device.
There are Similarities and Differences between the subject and predicate devices. The NeuroOne and PMT electrodes have the identical indications for use and intended use. The devices both have platinum contacts (PMT also has platinum/iridium contacts as an option). Both devices have the identical principles of operation. PMT uses silicone and NeuroOne uses polyimide film both are biocompatible for their intended use. Overall the configurations, dimensions/surface area of the electrodes are similar: with the NeuroOne electrode having a slightly smaller footprint. The NeuroOne device has undergone electrical, bench, and biocompatibility testing to demonstrate the differences in material and size of the contacts/electrodes do not raise questions of safety or efficacy.
#### INDICATIONS FOR USE (For the Device Subject to This 510(k) Notification)
The NeuroOne Cortical Electrodes are intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals on the surface of the brain.
{4}------------------------------------------------
# INDICATIONS FOR USE AND DEVICE CLASSIFICATION COMPARISON
The following tables provide a side-by-side comparison of the Cortical Electrode to the predicate device to support this pre-market notification.
| Intended Use/Indications for Use Comparison | | |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Predicate Device<br>K082474 |
| Intended Use | A cortical electrode is an electrode which is<br>temporarily placed on the surface of the brain<br>for stimulating the brain or recording the<br>brain's electrical activity. | A cortical electrode is an electrode which is<br>temporarily placed on the surface of the brain<br>for stimulating the brain or recording the<br>brain's electrical activity. |
| Indications for<br>Use | The NeuroOne Cortical Electrodes | Catalog Numbers 21 10-XXX: The Subdural<br>Cortical Electrodes (Dual Side Macro<br>Electrodes, Dual Side Micro Electrodes,<br>Dual Side Macro-Micro Electrodes, Dual<br>Sided Hemispherical Macro Electrodes,<br>Dual Sided Hemispherical Micro Electrodes,<br>Dual Side Hemispherical Macro-Micro<br>Electrodes, Macro Grids, Macro Strip,<br>Macro-Micro Grids, Macro-Micro Strips,<br>Micro Grids, and Micro Strips.) |
| | are intended for temporary (less than 30 days)<br>use with recording, monitoring, and<br>stimulation equipment for the recording,<br>monitoring, and stimulation of electrical<br>signals on the surface of the brain. | are intended for temporary (<30 day)<br>use with recording, monitoring, and<br>stimulation equipment for the recording,<br>monitoring and stimulation of electrical<br>signals on the surface of the brain. |
| Device Classification Comparison Subject Device | | |
|-------------------------------------------------|-----------------------------|---------------------------------|
| | Subject Device | Predicate Device<br>K082474 |
| Trade/Device Name: | NeuroOne Cortical Electrode | PMT Subdural Cortical Electrode |
| Common Name: | Cortical electrode | Cortical electrode |
| Classification Number: | 21 CFR 882.1310 | 21 CFR 882.1310 |
| Classification Name: | Cortical electrode | Cortical electrode |
| Device Class: | Class 2 | Class 2 |
| Product Code: | GYC | GYC |
| Technological and Performance Characteristics Comparison | | |
|----------------------------------------------------------|----------------|-----------------------------|
| | Subject Device | Predicate Device |
| Manufacturer | NeuroOne, Inc. | PMT |
| Number of Contacts per Electrode | 2 - 16 | 2 - 64 |
| Thickness of Contact | 0.0127 mm | 0.5 - 1.01mm |
| Contact Diameter | 3 mm | 2 - 4.5mm |
| Contact Material(s) | Platinum | Platinum or stainless steel |
| Surface Area of Electrode | 1x2 469mm² | 1x2 570mm² |
| Electrode Film Material | Polyimide | Silicone |
| Electrode Film Thickness | 0.0127 mm | 0.5 - 1.01 mm |
| Thickness of Tail | 0.0127 mm | 1.40mm |
{5}------------------------------------------------
| Technological and Performance Characteristics Comparison | | |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|------------------|
| | Subject Device | Predicate Device |
| Single patient use, Disposable | Yes | Same |
| Provided Sterile | Yes | Yes |
| Environment of Use | Intraoperative and Neurological<br>monitoring locations on surface of<br>brain | Same |
| Duration of Use | Less than 30 days | Same |
| Principles of Operation | Cortical electrodes are placed on<br>the surface of the brain to record,<br>monitor and stimulate brain<br>activity | Same |
### SUMMARY OF TESTING and PERFORMANCE STANDARDS
Performance testing was conducted to evaluate and characterize the performance of the device to support a determination of substantial equivalence to the predicate device. A comparison was made against the predicate where data was available. The NeuroOne device has undergone electrical, bench, and biocompatibility testing to demonstrate any differences do not raise questions of safety or efficacy.
- . Design Verification Testing
- o Dimensional Measurements
- O Durability
- Mapping/Stimulation O
- Reliability o
- Electrical to IEC 60601 O
- Biocompatibility Testing to ISO 10993-1 and ASTM F756
- . Packaging/Shelf Life to ASTM D4169-16, ASTM F1980-16, ASTM 2096-11, ASTM F88-15
- Sterilization Validation to ISO 11135 ●
- Labeling to ISO 15223-1 .
The Indications for Use, Intended Use, Fundamental Scientific Technology and Principles of Operation for the Cortical Electrode are the same as those described for the predicate device. The Cortical Electrode is found to have a safety and effectiveness profile that is the same as the predicate device and is determined by NeuroOne to be substantially equivalent.
In summary, the Cortical Electrode has the following similarities to the predicate device which has previously received 510(k) clearance:
- Has the same indications for use
- Has the same intended use
.
- Used in the same anatomical site
- Uses the same technological characteristics
#### ● Uses the same principles of operation
- Uses the same sterilization methodology
- Biocompatible for its intended use
- SUBSTANTIAL EQUIVALENCE CONCLUSIONS
Through the comparison of technological and performance characteristics, the subject device is determined to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.