← Product Code [LLZ](/productcode/LLZ) · K192743

# VisionX 2.4 (K192743)

_Durr Dental SE · LLZ · Oct 31, 2019 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K192743

## Device Facts

- **Applicant:** Durr Dental SE
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Oct 31, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The software is intended for the viewing and diagnosis of image data in relation to dental issues. Its proper use is documented in the operating instructions of the corresponding image-generating systems. Image-generating systems that can be used with the software include optical video cameras, image plate scamers, extraoral X-ray devices, intraoral scanners and TWAIN compatible image sources. The software must only be used by authorized healthcare professionals in dental areas for the following tasks: - Filter optimization of the display of 2D and 3D images for improved diagnosis - Acquisition, storage, management, display, analysis, editing and supporting diagnosis of digital/digitized 2D and 3D images and videos - Forwarding of images and additional data to external software (third-party software) The software is not intended for mammography use.

## Device Story

VisionX 2.4 is standalone medical imaging software for dental practices. It receives input from optical video cameras, image plate scanners, extraoral X-ray devices, intraoral scanners, and TWAIN-compatible sources. The software enables acquisition, storage, management, display, analysis, and editing of 2D and 3D images/videos. It provides filter optimization for display to assist clinicians in diagnosis. Operated by authorized dental professionals on standard PC-compatible hardware, the software facilitates image forwarding to third-party applications. It supports clinical decision-making by providing enhanced visualization and diagnostic tools for dental pathologies. Benefits include improved diagnostic efficiency and streamlined management of digital dental imaging workflows.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including software verification, functional cross-check testing, and compliance with IEC 62304. A clinical evaluation was performed via literature review and practical tests in dental practices to confirm suitability for intended use.

## Technological Characteristics

Software-based medical imaging system; runs on standard PC-compatible hardware (Windows 8.1, 10, Server 2016/2019). DICOM compliant. Supports 2D/3D image processing, filter optimization, and TWAIN integration. Developed per IEC 62304. Cybersecurity controls implemented per FDA guidance.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- DBSWIN and VistaEasy Imaging Software ([K190629](/device/K190629.md))

## Reference Devices

- ProVecta 3D Prime ([K181432](/device/K181432.md))
- ScanX Touch / Duo Touch ([K191623](/device/K191623.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The words "U.S. FOOD & DRUG" are in a larger font than the word "ADMINISTRATION".

October 31, 2019

Durr Dental SE % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, FLORIDA 34114

Re: K192743

Trade/Device Name: VisionX 2.4 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: September 26, 2019 Received: September 30, 2019

Dear Mr. Daniel Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192743

Device Name VisionX 2.4

### Indications for Use (Describe)

The software is intended for the viewing and diagnosis of image data in relation to dental issues. Its proper use is documented in the operating instructions of the corresponding image-generating systems. Image-generating systems that can be used with the software include optical video cameras, image plate scamers, extraoral X-ray devices, intraoral scanners and TWAIN compatible image sources. The software must only be used by authorized healthcare professionals in dental areas for the following tasks:

- Filter optimization of the display of 2D and 3D images for improved diagnosis

- Acquisition, storage, management, display, analysis, editing and supporting diagnosis of digital/digitized 2D and 3D images and videos

- Forwarding of images and additional data to external software (third-party software)

The software is not intended for mammography use.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary, DÜRR DENTAL SE,

# Device Name: VisionX 2.4

# K192743

This 510(k) is being submitted in accordance with the requirements of 21 CFR §807.92.

#### 1. Date Summary Prepared:

September 19, 2019

#### 2. Submitter's Identification:

DÜRR DENTAL SE (Registered as DUERR DENTAL SE) Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Deutschland/Germany Phone: + 49 (0) 7142 70 5-0 Fax: + 49 (0) 7142 705-500 E-Mail info@duerr.de Internet: www.duerrdental.com

Responsible Individual Contact/Prepared by: Oliver Lange, Director of Quality Management DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Deutschland/Germany Phone: 49-7142-705-190 Email: oliver.lange@duerrdental.com

Mr. Joseph Latkowski Director of Quality and Regulatory Air Techniques, Inc. 1295 Walt Whitman Road Melville, NY 11747, USA U.S. Phone: 516-214-5574 E-Mail: Joseph.Latkowski@airtechniques.com

#### 3. Device:

Trade /Proprietary Name: VisionX 2.4 Device: Medical Imaging Software Regulation Description: Picture archiving and communications system. Regulation Medical Specialty: Radiology Review Panel Radiology Product Code LLZ 892.2050 Regulation Number Device Class 2

{4}------------------------------------------------

#### 4. Predicate Device:

Legally Marketed Predicate Device Information:

| 510(k) Number:                | K190629                                      |
|-------------------------------|----------------------------------------------|
| Manufacturer:                 | DÜRR DENTAL SE                               |
| Trade /Proprietary Name       | DBSWIN and VistaEasy Imaging Software        |
| Device:                       | Medical Imaging Software                     |
| Regulation Description:       | Picture archiving and communications system. |
| Regulation Medical Specialty: | Radiology                                    |
| Review Panel                  | Radiology                                    |
| Product Code                  | LLZ                                          |
| Regulation Number             | 892.2050                                     |
| Device Class                  | 2                                            |

#### 5. Device Description:

VisionX 2.4 imaging software is an image management system that allows dentists to acquire, display, edit, view, store, print, and distribute medical images. VisionX 2.4 software runs on user provided PC-compatible computers and utilizes previously cleared digital image capture devices for image acquisition. This software was cleared in K181432 as part of the x-Ray system ProVecta 3D Prime. With this submission VisionX will be established as standalone software. Additionally, new hardware was integrated: Support of the ScanX Touch / Duo Touch (K191623)

#### 6. Indications for Use

The software is intended for the viewing and diagnosis of image data in relation to dental issues. Its proper use is documented in the operating instructions of the corresponding imagegenerating systems. Image-generating systems that can be used with the software include optical video cameras, digital X-ray cameras, image plate scanners, extraoral X-ray devices, intraoral scanners and TWAIN compatible image sources.

The software must only be used by authorized healthcare professionals in dental areas for the following tasks:

- Filter optimization of the display of 2D and 3D images for improved diagnosis

- Acquisition, storage, management, display, analysis, editing and supporting diagnosis of digital/digitized 2D and 3D images and videos

- Forwarding of images and additional data to external software (third-party software) The software is not intended for mammography use.

### 7. Summary of the technological characteristics of the device compared to the predicate device:

{5}------------------------------------------------

## Comparison of the Technological Characteristics

| Descriptive<br>Information:                              | Predicate Device:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Subject Device:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                          | DBSWIN and VistaEasy Imaging Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | VisionX 2.4                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                          | K190629                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Indications for<br>Use                                   | DBSWIN and VistaEasy imaging software<br>are intended for use by qualified dental<br>professionals for windows-based<br>diagnostics. The software is a diagnostic<br>aide for licensed radiologists, dentists<br>and clinicians, who perform the actual<br>diagnosis based on their training,<br>qualification, and clinical experience.<br>DBSWIN and VistaEasy are clinical<br>software applications that receive<br>images and data from various imaging<br>sources (i.e., radiography devices and<br>digital video capture devices) that are<br>manufactured and distributed by DÜRR<br>Dental and Air Techniques. It is intended<br>to acquire, display, edit (i.e., resize,<br>adjust contrast, etc.) and distribute<br>images using standard PC hardware. In<br>addition, DBSWIN enables the<br>acquisition of still images from 3rd party<br>TWAIN compliant imaging devices (e.g.,<br>generic image devices such as scanners)<br>and the storage and printing of clinical<br>exam data, while VistaEasy distributes<br>the acquired<br>images to 3rd party TWAIN compliant<br>PACS systems for storage and printing.<br><br>DBSWIN and VistaEasy software are not<br>intended for mammography use. | The software is intended for the<br>viewing and diagnosis of image data in<br>relation to dental issues. Its proper use<br>is documented in the operating<br>instructions of the corresponding<br>image-generating systems. Image-<br>generating systems that can be used<br>with the software include optical video<br>cameras, digital X-ray cameras, image<br>plate scanners, extraoral X-ray devices,<br>intraoral scanners and TWAIN<br>compatible image sources.<br>The software must only be used by<br>authorized healthcare professionals in<br>dental areas for the following tasks:<br>- Filter optimization of the display of<br>2D and 3D images for improved<br>diagnosis<br>- Acquisition, storage, management,<br>display, analysis, editing and<br>supporting diagnosis of<br>digital/digitized 2D and 3D images<br>and videos<br>- Forwarding of images and<br>additional data to external<br>software (third-party software)<br><br>The software is not intended for<br>mammography use. |
| Patient<br>Management                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Image<br>Management                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| X-ray<br>(i.e., Phosphor<br>Plate, Digital<br>Panoramic) | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Laser<br>Fluorescence<br>Caries<br>Detection Aid         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Video                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Photos                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Documents                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Import                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Display Images                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Save/Store Images                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Produce Reports                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Print/Export Images                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Brightness                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Contrast                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Colorize                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Crop                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Rotate                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Zoom In/Out                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Invert                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Sharpen                                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Measure                                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Annotate                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Run on standard PC compatible computers                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Operating Systems                                        | Microsoft Windows 7, 32-bit (from Home to Premium)<br>Microsoft Windows 7, 64-bit (from Home to Premium)<br>Microsoft Windows 8.1, 64-bit<br>Microsoft Windows 10, 64-bit<br>Microsoft Windows Server 2012<br>Microsoft Windows Server 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Microsoft Windows 7, 64-bit<br>Microsoft Windows 8.1, 64-bit<br>Microsoft Windows 10, 64-bit<br>Microsoft Windows Server 2016<br>Microsoft Windows Server 2019<br>Same but since Windows 7 is no longer supported by Microsoft we no longer recommend using this OS..                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Supported Devices                                        | Supported Device Families:<br>• ScanX<br>• ProVecta S-Pan<br>• CamX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Supported Device Families:<br>• ScanX<br>• ProVecta S-Pan<br>• CamX<br>Added Device Family:<br>• ProVecta 3D                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

{6}------------------------------------------------

{7}------------------------------------------------

#### 8. Discussion of Similarities and Differences:

VisionX 2.4 represents these enhancements as compared to our predicate:

- VisionX as standalone software. Currently the VisionX is cleared as part of the System of ● ProVecta 3D prime only.
- . 3D functionality: 3D images and videos for acquisition, storage, management, display, analysis, editing and supporting diagnosis of digital/digitized.
- Integration of the device family ProVecta 3D (x-ray system, cleared in K181432).
- Integration of new variants of image plate scanner of the "ScanX" device family.
- . Integration of a new interface for an intraoral scanner.

#### 9. Non-Clinical Data and Performance Testing

VisionX 2.4 was developed in compliance with the harmonized standard of IEC 62304 for medical device software life cycle requirements. We consulted this FDA guidance document during the software development: Guidance for Industry and FDA Staff- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Software testing, effectiveness, and functionality were successfully conducted and verified between VisionX 2.4 and image capture devices. VisionX 2.4 is DICOM compliant. Risk Analysis based design development and design reviews were conducted. Full functional software cross check testing was performed.

Labeling and internal procedures have been updated to assure compliance with the FDA Guidance: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices-Guidance for Industry and Food and Drug Administration Staff Document. In addition, we performed what we call a "Clinical Evaluation." This was an extensive detailed review of literature data, data from practical tests in dental practices, and safety data. It concluded that the software is suitable for an effective, efficient and safe dental use for the indications mentioned.

- 10. Clinical Data: Not required for a finding of substantial equivalence.
## 11. Conclusion:

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the similarity to the predicate device in terms of technology, performance and indications for use, DÜRR DENTAL SE concludes that the VisionX 2.4 Imaging Software is substantially equivalent to the predicate device as described herein. The device modifications to VisionX 2.4 do not alter the fundamental scientific technology of the predicate device and summary level information is adequate to assess the modifications. The verification testing demonstrates that the device continues to meet its performance specifications and the results of the testing did not raise new issues of safety or effect…

---

**Source:** [https://fda-staging.innolitics.com/device/K192743](https://fda-staging.innolitics.com/device/K192743)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com