Electric Wheelchair [Models: JRWD6010/JRWD6012]

K192739 · Jerry Medical Instrument (Shanghai) Co., Ltd. · ITI · Jul 13, 2021 · Physical Medicine

Device Facts

Record IDK192739
Device NameElectric Wheelchair [Models: JRWD6010/JRWD6012]
ApplicantJerry Medical Instrument (Shanghai) Co., Ltd.
Product CodeITI · Physical Medicine
Decision DateJul 13, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position.

Device Story

Motor-driven indoor wheelchair; battery-powered; DC motor-driven. User operates via joystick on control panel to manage speed and direction; pushing joystick increases speed; releasing joystick stops device and engages electromagnetic brakes. Consists of foldable aluminum frame, seat, backrest, armrests, two rear driving wheels with hub motors, and two front casters. Used by disabled or elderly individuals for indoor mobility. Controller processes user input signals to adjust motor output. Provides mobility and independence to users with limited walking ability.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including static/dynamic stability, brake effectiveness, energy consumption, dimensions, speed/acceleration, climatic tests, obstacle climbing, and electromagnetic compatibility per ISO 7176 standards. Biocompatibility verified per ISO 10993.

Technological Characteristics

Frame: A6061-T4 aluminum alloy. Power: 24V DC motor (180W), 24V 20Ah Li-ion battery. Control: Joystick-based electronic controller. Connectivity: None. Sterilization: N/A. Software: Moderate level of concern. Standards: ISO 7176 series (stability, brakes, dimensions, etc.), IEC 60601-1-2 (EMC), IEC 62133 (battery safety).

Indications for Use

Indicated for disabled or elderly persons limited to a seated position requiring indoor mobility assistance.

Regulatory Classification

Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 13, 2021 Jerry Medical Instrument (Shanghai) Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, 200120 China Re: K192739 Trade/Device Name: Electric Wheelchair (Models: JRWD6010 and JRWD6012) Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: May 31, 2021 Received: June 16, 2021 Dear Boyle Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192739 Device Name Electric Wheelchair (Models: JRWD6010 and JRWD6012) Indications for Use (Describe) The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary # (K192739) This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92. ### 1.0 Submitter's information Name: JERRY MEDICAL INSTRUMENT (SHANGHAI) CO., LTD. Address: Building 12, No. 615 Fengdeng Rd, Malu Town, Jiading District, Shanghai 201801, China Tel: 86-13817397985 Fax: 86-21-59517526 Contact: Jianguo Chen Date of Preparation: May 31,2021 #### Designated Submission Correspondent Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn #### 2.0 Device information Trade name: Electric Wheelchair Powered Wheelchair Common name: Classification name: Powered Wheelchair Model(s): JRWD6010, JRWD6012 ### 3.0 Classification Production code: ITI Regulation number: 21 CFR 890.3860 Classification: Class II Panel: Physical Medicine #### 4.0 Predicate device information Manufacturer: Nanjing Jin Bai He Medical Apparatus Co., Ltd. Device: Powered Wheelchair DYW30A(D09) {4}------------------------------------------------ #### 510(k) number: K170787 #### 5.0 Indication for Use Statement The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position. #### 6.0 Device description The proposed device, Electric wheelchair, mainly powered by battery, motivated by DC motor, driven by user controlling joystick and adjusting speed. The electric wheelchair consists of two foldable armrests, a backrest, a seat cushion, a foldable frame, two rear driving wheels with hub motor/electromagnetic brake assemblies, two pivoting casters, a Li-ion batteries, an off-board battery charger, a control panel, and an electric motor controller. The electric wheelchair is intended to provide mobility to a person with a disability or an older adult limited to a sitting position. Folding design: This foldable electric wheelchair has main frame, rear frame, backrest frame, seat frame, front wheel frame and battery frame. Release the locking device of the backrest frame and the rear frame, push the backrest frame downward, drive the seat frame to rotate toward the backrest frame with the rotating mechanism, the two armrest frames rotate towards the direction of the seat frame. At the same time two front wheel frames move to the direction of the seat frame. When the back cushion is in contact with the seat cushion, the folding is completed. Joystick: This controls the speed and direction of the wheelchair . Push the joystick in the direction you wish to go. The further you push it , the faster the speed . Releasing the joystick stops the wheelchair and automatically applies the brakes. Controller: The controller includes a power switch, a speed adjustment button, a horn button and a direction joystick. The operation interface controller receives the operation signal and transmits it to the main board of the controller body, and sends a control signal to the electric wheelchair to adjust and operate the electric wheelchair. Wheel and frame connection method: First, the left and right frames are connected to the motor respectively, and then the motor shaft is connected to the wheel. The motor rotates to drive the wheels to rotate to drive the wheelchair. Frame design: 1. The tube of this product is A6061-T4 aluminum, and the frame tube mostly is D22.2*2.0t aluminum tube. (The main frame of JRWD6012 is 80*25*2.0t aluminum tube, and the main frame of JRWD6010 is 40*22.2*2.0t flat oval tube). 2. The operation mode of the folding mechanism is to remove the locking device by someone else and manually push the backrest frame to complete the folding operation under non-riding state. The electric wheelchair has two models: JRWD6010, JRWD6012. {5}------------------------------------------------ Model JRWD6010: 8 inch front wheel and 10 inch rear tire. Modle JRWD6012: 8 inch front wheel and 12inch rear tire. The motor of electric wheelchair is DC24V 180W; the battery is 24V 20AH, Li-ion battery; the charger is 24V/6A. Max. loading can not be over than 100Kgs. The following surfaces are re commended NOT to operate on: Sand surface Wet or icy surface Road maintenance hole metal cover Do not use on stairs Do not use escalators. Use the elevator. Too steep incline over 10 degrees. Obstacle climbing ability: 1.97" Do not use outdoors #### 7.0 Summarv of Non-Clinical Testing Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: ISO 7176-1 :2014 Wheelchairs - Part 1: Determination of static stability ISO 7176-2 :2017 Wheelchairs - Part 2: Determination of dynamic stability of electrically powered wheelchairs ISO 7176-3 : 2012 Wheelchairs - Part 3: Determination of effectiveness of brakes ISO 7176-4 : 2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range ISO 7176-5 : 2008 Wheelchairs - Part 5: Determination of overall dimensions, mass and maneuvering space ISO 7176-6: 2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs ISO 7176-7 : 1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions ISO 7176-8 : 2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths ISO 7176-9: 2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs ISO 7176-10 : 2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs ISO 7176-11 : 2012 Wheelchairs - Part 11: Test dummies ISO 7176-13 : 1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces ISO 7176-14 : 2008 Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods ISO 7176-15: 1996 Wheelchairs - Part 15: Requirements for information disclosure, {6}------------------------------------------------ #### documentation and labeling ISO 7176-16:2012 Wheelchairs -- Part 16: Resistance to ignition of postural support devices ISO 7176-21: 2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers ISO 7176-22 : 2014 Wheelchairs - Part 22: Set-up procedures IEC 60601-1-2:2015 Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests IEC 62133:2012 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications #### Biocompatibility of patient-contacting parts #### Statement for Biocompatibility Certification The materials of the Joystick knob, Joystick Gaiter, Enclosure Mouldings and Keypad of new\/Si electric wheelchair controller & Joystick of the Electric Wheelchair, model JRWD6010, JRWD6012, manufactured by JERRY MEDICAL INSTRUMENT (SHANGHAI) CO., LTD. are identical to the materials of newVSi electric wheelchair controller & Joystick of the Y207 Electric Wheelchair, Model Y207, JIANGSU INTCO MEDICAL PRODUCTS CO., LTD, K202482, clearance date 03/18/2021, in formulation, processing, and geometry, and no other chemicals have been added. The patient-contacting materials of the new\/Si electric wheelchair controller & Joystick of the Electric Wheelchair, JRWD6010, JRWD6012, have the same nature of tissue contact and contact duration (e.g., surface device category, intact skin contact, less than 24-hour duration) as the Y207 Electric Wheelchair, ModelY207. Other patient-contacting material are carried out biocompatibility assessment in accordance with IS0 10993-1: 2018, including : Cytotoxicity per ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity Irritation and Skin Sensitization per ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. #### 8.0 Summarv of Clinical Testing No clinical study implemented for the electric wheelchair. ### 9.0 Technological Characteristic Comparison Table {7}------------------------------------------------ | Item | Proposed device | Predicated device | Remark | |------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------| | Product Code | ITI | ITI | Same | | Regulation No. | 21 CFR 890.3860 | 21 CFR 890.3860 | Same | | Class | II | II | Same | | Product name | Electric Wheelchair | Powered Wheelchair | - | | | | DYW30A(D09) | | | 510(k) No. | K192739 | K170787 | - | | Models | JRWD6010 JRWD6012 | DYW30A(D09) | - | | Intended Use | The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position. | The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position. | Same | | Use environment | Indoor use | Indoor use | Same | | Patient<br>Population | The electric wheelchair is intended to provide mobility to a person with a disability or an older adult limited to a sitting position | The electric wheelchair is intended to provide mobility to a person with a disability or an older adult limited to a sitting position | Same | | Product structure | Consists of two foldable armrests, a backrest, a seat cushion, a foldable frame, two rear driving wheels with hub motor/ electromagnetic brake assemblies, two pivoting casters, a Li-ion batteries, an off-board battery charger, a control panel, and an electric motor controller. | Two foldable armrests, a seat belt, a backrest, a seat cushion, a foldable frame, two rear driving wheels with hub motor/ electromagnetic brake assemblies, two pivoting casters, two Li-ion batteries, an off-board battery charger, a control panel, and an electric motor controller. | Similar | | Driving system | Direct drive on the rear wheels | Direct drive on the rear wheels | Same | | Number of<br>wheels | 4 | 4 | Same | | Main frame<br>material | Aluminium alloy | Aluminium alloy | Same | | Motor | DC24V* 180W*2pcs | 24 VDC *250W * 2 pcs | | | Battery | DC 24V 20Ah Lithium-ion | Lithium-ion, ITP2406 | Minor differences in the | | | 1 pcs | 6 Ah x 24 VDC x 2 pcs | dimensions will not | | | High Power Technology<br>Inc. | High Power Technology<br>Inc. | impact the safety and | | | | | effectiveness of the | | Battery charger | HP0180WL2 | HP0060W(L2) | substantial equivalence. | | | Input: 100-240 VAC | Input: 100-240 VAC | | | | Output: DC 24V, 6 Amp | Output: DC 24V, 2 Amp | | # Table1-General Comparison {8}------------------------------------------------ # Table2 Performance Comparison | Item | Proposed Device | | Predicate Device | Remark | |----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------| | | JRWD6010 | JRWD6012 | | | | Dimensions | 38.1"x24.0"x37.0" | 39.3"x23.6"x37.0" | 37.4" x 22.6" x 36.2" | Minor differences in the<br>dimensions will not<br>impact the safety and<br>effectiveness of the<br>substantial equivalence. | | Weight, w/<br>Battery | 58.4 lbs. /26.5kg | 58.2 lbs. /26.4kg | 51.8" lbs. / 23.5 kg | The difference will<br>not raise any new safety<br>and effectiveness<br>concerns. | | Frame design | Foldable/<br>The device consists of a foldable<br>and non-rigid type of power<br>wheelchair base with rear drive and<br>2 casters in the front and two<br>anti-tippers in the rear. | Foldable/<br>The device consists of a foldable<br>and non-rigid type of power<br>wheelchair base with rear drive and<br>2 casters in the front and two<br>anti-tippers in the rear. | Foldable/<br>The device consists of a<br>foldable and non-rigid<br>type of power wheelchair<br>base with rear drive and 2<br>casters in the front and<br>two<br>anti-tippers in the rear. | Same | | Folding<br>mechanism | A foldable seat frames<br>(The backrest could be folded to<br>seat) | A foldable seat frames<br>(The backrest could be folded to<br>seat) | A foldable seat frames<br>(The backrest could be<br>folded to seat) | Same | | Front<br>wheel(inch) | 8<br>(PU solid tire) | 8 | 7<br>(PU solid tire) | Larger sizes of front<br>wheels bring steadier<br>pivoting function than<br>predicate device. | | Rear tire (inch) | 10<br>(PU solid tire) | 12<br>(Pneumatic tire) | 12.5<br>(PU solid tire) | Smaller sizes of rear wheels,<br>The difference will not raise<br>any new safety and<br>effectiveness concerns. | | Cruising<br>Range(km) | | 20 | 18 | This difference is due to<br>the less weight for the<br>subject device.<br>There is a larger cruising<br>range for the subject device. | | Obstacle | | 50 | | The larger height in the | | climbing(mm) | | | obstacle climbing will not<br>impact the safety and<br>effectiveness of the subject<br>device. | | | Max. Speed<br>(km/h) | 6 | 6 | Same | | | Static stability<br>forward | 21.8° | | Both of the devices are<br>evaluated according to<br>standard ISO 7176-1:2014, | | | Static stability<br>rearward | 19° | Not publicly available | so the different static stability | | | Static stability<br>sideways | 19.2° | | will not impact the safety<br>and effectiveness | | | Max. loading<br>(kg) | 220lbs(100kg) | 264lbs (120kg) | Less loading weight means<br>more convenient for the<br>transportation | | | Maximum safe<br>operational<br>incline | 10 degrees | 8 degrees | Larger safe operational<br>incline of subject bring more<br>convenient for the use<br>environment | | | Min. Turning<br>radium | 1820mm | 833 mm | The difference in the turning<br>radius will bring more<br>convenience when it<br>turns.The difference will not<br>raise any new safety and<br>effectiveness concerns. | | | Maximum<br>obstacle<br>climbing | 1.97" (50mm) | 1.36" (34.5 mm) | The larger height in the<br>obstacle climbing will<br>not impact the safety<br>and effectiveness of the<br>subject device. | | | Minimum<br>braking<br>distance | 1m | 1 m | Same | | | Max Speed<br>Forwards | 3.75 mph (6 km/h) | 3.75 mph (6 km/h) | Same | | | Max. Speed<br>Backward | 2.80 mph (4.5 km/h) | 1.86 mph (3.0 km/h) | The devices are evaluated<br>according to standard ISO<br>7176-6:2018, so the different<br>will not impact the safety<br>and effectiveness | | | Controller | PG Drives Technology Ltd., newVSI | Changzhou Billon<br>Electronic Appliance<br>Co.,Ltd., WS-1 | Different<br>Although different controller<br>is used, both the control | | | | | | system, including the joystick | | | | | | controller,<br>the | | | | | | electromagnetic brakes and | | | | | | the user interface are similar. | | | | | | The joystick controls the | | | | | | directions and speed of | | | | | | movement, and when the | | | | | | joystick is released, the | | | | | | powered wheelchair will slow | | | | | | down to stop and the brakes | | | | | | will automatically re-engage. | | | | | | The controller also provides | | | | | | the battery status displaying | | | | | | and abnormal condition | | | | | | displaying. Both of the control | | | | | | systems are evaluated | | | | | | according to standard ISO | | | | | | 7176-14:2008 and software | | | | | | validation requirement and | | | | | | there are no new safety and | | | | | | effectiveness concerns due | | | | | | to the difference. | | | Speed control<br>method | Joystick control method | Joystick control method | Same | | {9}------------------------------------------------ {10}------------------------------------------------ # Table3 Safety Comparison | Item | Proposed Device | Predicate Device | Remark | | |---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|------| | Main materials | Frame: Aluminium alloy;<br>Wheel, Armrest: PU;<br>Backrest: oxford cloth | Frame: Aluminium alloy;<br>Wheel, Armrest: PU;<br>Backrest: PVC Vinyl | Biocompatibility<br>evaluation has<br>been carried out<br>per ISO 10993-1. | | | Materials<br>contacting<br>user | Armrest: PU;<br>Backrest: oxford cloth<br>Seat: oxford cloth<br>newVSi electric wheelchair controller:<br>Joystick knob: Santoprene<br>101-80;<br>Joystick Gaiter: Silicone 3032 (50%) &<br>5031 (50%)<br>Enclosure Moulding(s): ABS/PC<br>Wonderloy PC-540<br>Keypad: Silicone keypad<br>coatings TC-2407<br>& CH-6330 | Armrest: PU;<br>Seat: PVC Vinyl<br>Backrest: PVC Vinyl<br>Safety belt: PVC Vinyl<br>Joystick: PVC Vinyl | There are no new<br>safety and<br>effectiveness<br>concerns due to the<br>difference. | | | | Biocompatibilit | Comply with ISO 10993-1, FDA | Comply with ISO 10993-1, FDA | Same | {11}------------------------------------------------ | y of materials contacting user | Guidance, Tests included | Guidance, Tests included | | |----------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|------| | | Cytotoxicity (ISO 10993-5:2009),<br>Sensitization and Intracutaneous<br>Reactivity (ISO 10993-10:2010) | Cytotoxicity (ISO 10993-5:2009),<br>Sensitization and Intracutaneous<br>Reactivity (ISO 10993-10:2010) | | | Label and<br>Labeling | Conforms to FDA Regulatory<br>Requirements | Conforms to FDA Regulatory<br>Requirements | Same | | Level of<br>Concern of the<br>Software | Moderate | Moderate | Same | #### Summary of substantial equivalence discussion: The Electric Wheelchair model JRWD6010 and JRWD6012 complied with the requirements of ISO 7176-1:2014, ISO 7176-2:2017, ISO 7176-3:2012, ISO 7176-4:2008, ISO 7176-5:2008, ISO 7176-6:2001, ISO 7176-7:1998, ISO 7176-8:2014. ISO 7176-9:2009. ISO 7176-10:2008. ISO 7176-11:2008. ાડભ 7176-13:1989, ISO 7176-14:2008, ISO 7176-15:1996, ISO 7176-16:2012, ISO 7176-21:2009, ISO 7176-22:2014, IEC 60601-1-2: 2014, IEC 62133:2012, ISO 10993-1:2018, ISO10993-5:2009, ISO 10993-10:2010. The intended uses for both devices are the same. Mainframes of two devices are folded by way of front and rear close, and frame materials all meet the Tensile Strength. Yield Load, and Elongation tests. The design principles of the controller and Driving system are the same, and both meet the requirements of the ISO 7176-14:2008. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3:2012. Maximum obstacle climbing and Maximum safe operational incline are slightly different while such differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-2:2001, ISO 7176-10:2008. The biocompatibility of the Predicate device and Subject device meet the requirements of the ISO 10993-5:2009 & ISO 10993-10:2010. The flame retardant test of the seat cushion/back cushion and armrest of both subject device and predicate device is carried out according to the ISO 7176-16 test. Therefore, both devices are assured to be under the same safety level. In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. #### Conclusion 10.0 {12}------------------------------------------------ The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device in K170787 and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...