K192720 · Covidien · GDW · Mar 13, 2020 · General, Plastic Surgery
Device Facts
Record ID
K192720
Device Name
GIA Stapler with Tri-Staple Technology
Applicant
Covidien
Product Code
GDW · General, Plastic Surgery
Decision Date
Mar 13, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The GIA™ stapler with Tri-Staple™ technology has applications in abdominal and thoracic surgical procedures for resection, transection and creation of anastomosis.
Device Story
Manual, single-use surgical stapler; used in open abdominal and thoracic surgery. Device places two triple staggered rows of titanium staples while simultaneously cutting/dividing tissue between rows. Features height-progressive staple technology (purple cartridge: 3.0/3.5/4.0 mm; black cartridge: 4.0/4.5/5.0 mm) to accommodate varying tissue thicknesses. Operated by surgeon via single-handle squeeze mechanism; firing knob slides forward to deploy staples and actuate knife blade. Provides audible feedback upon firing. Reloadable up to 7 times (8 total firings). Benefits include consistent staple formation and tissue management during resection/anastomosis.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical performance testing: in-vitro (staple formation, firing force, multi-fire evaluation), ex-vivo (burst pressure, pneumostasis, staple line evaluation), and in-vivo (safety, efficacy, tissue abrasion, hemostasis, chronic survival). Human factors/usability testing conducted per IEC 62366. Biocompatibility per ISO 10993-1. Sterilization validated per ISO 11135.
Technological Characteristics
Manual linear stapler; titanium staples (ASTM F67 Grade I); stainless steel knife and anvil. 80mm staple line length. Two cartridge types: purple (3.0/3.5/4.0mm) and black (4.0/4.5/5.0mm). Single-handle squeeze operation. Sterile (ethylene oxide). 5-year shelf life.
Indications for Use
Indicated for patients undergoing abdominal and thoracic surgical procedures requiring resection, transection, and creation of anastomosis.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
March 13, 2020
Covidien Frank Gianelli Senior Regulatory Affairs Specialist 60 Middletown Avenue North Haven, Connecticut 06473
Re: K192720
Trade/Device Name: GIA Stapler with Tri-Staple Technology Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW, GAG Dated: September 13, 2019 Received: September 27, 2019
Dear Frank Gianelli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K192720
Device Name
GIATM Stapler with Tri-Staple™ Technology
Indications for Use (Describe)
The GIA™ stapler with Tri-Staple™ technology has applications in abdominal and thoracic surgical procedures for resection, transection and creation of anastomosis.
| Type of Use (Select one or both, as applicable) | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|----------|---------------------------------------------|
| <span> <table style="display: inline-table;"><tr><td><div> <table style="border-collapse: collapse;"><tr><td style="border: 1px solid black; padding: 5px;"><b> </b></td></tr></table> </div></td></tr></table> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <div> <table style="border-collapse: collapse;"><tr><td style="border: 1px solid black; padding: 5px;"><b> </b></td></tr></table> </div> | <b> </b> | Over-The-Counter Use (21 CFR 801 Subpart C) |
| <div> <table style="border-collapse: collapse;"><tr><td style="border: 1px solid black; padding: 5px;"><b> </b></td></tr></table> </div> | <b> </b> | | |
| <b> </b> | | | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | | | |
| Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br> | PRAStaff@fda.hhs.gov | | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of | | | |
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the word "COVIDIEN" in a sans-serif font. To the left of the word is a blue square with a white square in the center. A light blue line runs vertically through the white square. The word "COVIDIEN" is in a dark blue color.
K192720 Page 1 of 5
# 510(k) Summary
### Submission Date: Sep 27, 2019
### K192720
### SUBMITTER INFORMATION:
Sarah Tang Covidien Rooms 501, 502, 601, 602, No.3 building No.2388 Chen Hang Road Min Hang District, Shanghai, 201114, China
#### Contact:
Frank Gianelli (U.S. Agent) on behalf of Sarah Tang Covidien 60 Middletown Avenue North Haven, CT 06473, USA Senior Requlatory Affairs Specialist Telephone: (203) 492-5352 Fax: frank.j.gianelli@medtronic.com Email:
#### Subject Device:
Proprietary/Trade GIA™ Stapler with Tri-Staple™ Technology Name: GIA80MTS, GIA80XTS, GIA80MTC, GIA80XTC Model Numbers: Staple, Implantable Classification Name: 21 CFR 878.4750 Requlations Number: GDW Product Codes: Class II Device Class: General and Plastic Surgery Review Panel: Surgical Staple Common Name:
#### Predicate Device:
Proprietary/Trade Name: GIATM Stapler with DST™ Technology 510(k) Number: K111825 Classification Name: Staple, Implantable Regulations Number: 21 CFR 878.4750 Product Codes: GDW Device Class: Class II Review Panel: Common Name:
General and Plastic Surgery Surgical Staple
#### Device Description:
The subject GIA™ stapler with Tri-Staple™ technology places two triple staggered rows of titanium staples and simultaneously cuts and divides tissue between these two triple rows. The GIA™ stapler and cartridges with Tri-Staple™ technology are available in 80 mm staple line length and two staple sizes to accommodate various tissue thicknesses: medium/thick and extra thick. Staplers with medium/thick staple size (purple cartridge) deploy three height-progressive
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Covidien. The logo consists of a blue square with a white cross in the center, followed by the word "COVIDIEN" in blue, sans-serif font. The logo is simple and modern, and the colors are clean and professional.
### K192720 Page 2 of 5
# 510(k) Summary
rows of 3.0 mm, 3.5 mm and 4.0 mm titanium staples on either side of the tissue cut line. Staplers with extra thick staple size (black cartridge) deploy three height progressive rows of 4.0 mm, 4.5 mm and 5.0 mm titanium staples on either side of the cut line. Each GIA™ stapler with Tri-Staple™ technology may be reloaded with GIA™ Cartridge with Tri-Staple™ Technology up to 7 times for a total of 8 firings per instrument.
The subject device GIA™ Stapler with Tri-Staple™ Technology is manufactured with the same patient contact materials that are utilized within the predicate device (K111825).
The subject device GIA™ stapler™ technology has the same principle operation as the predicate device. The GIA™ Stapler with Tri-Staple™ Technology is a manual single-use device. It is provided sterile (ethylene oxide) with 5-year shelf life, and intended for multiple use during a single procedure, which is the same as the predicate device (K111825).
#### Indications for Use:
The GIA™ stapler with Tri-Staple™ technology has applications in abdominal and thoracic surgical procedures for resection, transection and creation of anastomosis.
#### Technological Characteristics:
The subject new device GIA™ stapler with Tri-Staple™ technology is substantially equivalent to the predicate K111825 GIA™ Stapler with DST™ Technology regarding the fundamental stapling technologies employed, intended use and indications for use. Both are single-use manual linear staplers.
The major technical difference between the subject device GIA™ staple™ technology and predicate device GIA™ Stapler with DST™ Technology is the new proposed device is equipped with Tri-Staple™ Technology.
The subject device GIA™ stapler with Tri-Staple™ technology places two triple staggered row of height progressive titanium staples, hence the name "Tri-Staple". The is available in 80mm length. The staplers are offered in 2 cartridge sizes, the purple cartridge deploying three height-progressive rows of 3.0 mm, 3.5 mm and 4.0 mm titanium staples is for medium/thick tissue thickness range and the black cartridge deploying three height progressive rows of 4.0 mm, 4.5 mm and 5.0 mm titanium staples is for extra thick tissue thickness range.
The predicate GIA™ Stapler with DST™ Technology deploy two double staggered same-height rows of titanium staples. The length of the staple line is 80mm. The staplers are offered in 2 cartridge sizes, blue cartridge deploying 2 rows of 3.8mm titanium staples is for medium/thick tissue thickness range, green cartridge deploying 2 rows of 4.8mm titanium staples is for extra thick tissue thickness range.
### Substantial Equivalent:
The subject devices have the same intended use and indications for use as the predicate devices.
They are similar in fundamental scientific technology in that they are all sterile, single used, handheld, manual surqical instruments equipped with titanium staples intended to have applications in abdominal and thoracic surgical procedures for resection, transection and creation of anastomosis. The subject and predicate device are similar in design, materials and are sterilized via ethylene oxide.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Covidien. The logo consists of a blue square with a white cross in the center, and the word "COVIDIEN" in blue text to the right of the square. The text is in a sans-serif font and is slightly smaller than the square.
# K192720 Page 3 of 5
# 510(k) Summary
The below table further summarizes the similarities and differences between the subject and predicate device.
| Features | Subject Device (K192720) | Predicate Device (K111825) |
|--------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | GIA™ Stapler with Tri-Staple™ Technology | GIA™ Staplers with DST™ Technology |
| Manufacturer | Covidien | Covidien |
| Constructional | Image: GIA Stapler with Tri-Staple Technology | Image: GIA Staplers with DST Technology |
| Indications for Use | The GIA™ stapler with Tri-Staple™ technology has applications in abdominal and thoracic surgical procedures for resection, transection and creation of anastomosis. | The GIA™ staplers with DST™ technology and the knifeless stapler have applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis. |
| Operation Method | Manual | Manual |
| Anatomical Site | Alimentary tract and Thoracic | Alimentary tract and Thoracic |
| Surgical Approach | Open surgery | Open surgery |
| Method of Operation | The instruments are activated by sliding the firing knob forward to a complete stop and Immediately after staple formation, the knife blade resects the excess tissue, creating a linear anastomosis. | The instruments are activated by sliding the firing knob forward to a complete stop and Immediately after staple formation, the knife blade resects the excess tissue, creating a linear anastomosis. |
| Product Codes | Stapler with Cartridge:<br>GIA80MTS, GIA80XTS<br>Cartridge:<br>GIA80MTC, GIA80XTC | Stapler with Cartridge:<br>GIA8038S, GIA8048S<br>Cartridge:<br>GIA8038L, GIA8048L |
| Staple Rows | 3 staggered rows of staples on either side of the tissue cut line with different staple height in each staple row | 2 staggered rows of staples on either side of the tissue cut line with same staple height in each staple row |
| Instrument Handle Type | Single-handle squeeze | Single-handle squeeze |
| Staple Cartridge Configuration | 2 triple rows staples, step-faced cartridge with different staple size in each staple row | 2 double rows staples, flat-faced with same staple size in each staple row |
| Cartridge Color | Purple, Black | Blue, Green |
| Staple Size (open leg height) | Purple cartridge: 3.0mm, 3.5mm 4.0mm<br>Black cartridge: 4.0mm, 4.5mm 5.0mm | Blue cartridge: 3.8mm<br>Green cartridge: 4.8mm |
| Anvil | 2 triple staggered rows of anvil pocket design | 2 double staggered rows of anvil pocket design |
| Staple Line Length | 80mm | 80mm |
| Staple Material | Titanium per ASTM F67 Grade I | Titanium per ASTM F67 Grade I |
| Features | Subject Device (K192720) | Predicate Device (K111825) |
| Identification of<br>Materials of<br>Implant (staple)<br>and tissue cutting<br>component (knife) | Staple: Titanium per ASTM F67 Grade I<br>Knife: Stainless Steel<br>Anvil: Stainless Steel | Staple: Titanium per ASTM F67 Grade I<br>Knife: Stainless Steel<br>Anvil: Stainless Steel |
| Biocompatibility | Evaluated per ISO 10993-1 series and<br>FDA 2016 biocompatibility guidance | Evaluated per ISO 10993-1 series |
| Audible Feedback | Yes | Yes |
| Knife | Yes.<br>without knifeless option | Yes<br>with knifeless option |
| Single Use | Yes | Yes |
| Disposable | Yes | Yes |
| Sterile | Ethylene oxide | Ethylene oxide |
| Shelf Life | 5 years | 5 years |
| | GIA™ Stapler with Tri-Staple™ Technology | GIA™ Staplers with DST™ Technology |
GIA™ Stapler with Tri-Staple™ Technology
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Covidien. The logo consists of a blue square with a white cross in the center, and the word "COVIDIEN" in blue letters to the right of the square. The logo is simple and modern, and the colors are clean and professional.
### K192720 Page 4 of 5
# 510(k) Summary
Tests performed to evaluate and compare technological and performance characteristics: Non-clinical performance data - the following testing has been performed to
demonstrate substantial equivalence to the predicate devices.
- 1. Performance Test In-Vitro
- Visual inspection on product, packaging and instruction for use .
- Staple formation on test media ●
- . Firing force test
- . Multi-Fire Evaluation
- 2. Performance Test Ex-Vivo
- Burst Pressure ●
- . Pneumostasis
- Across Staple Line Evaluation ●
- 3. Performance Test In-Vivo
- In vivo safety and efficacy .
- Tissue Abrasion
- Hemostasis test .
- Staple formation on tissues .
- Chronic survival testing .
Above device testing is completed per FDA Guidance " Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submission" issued on April 26, 2019 on the finial finished sterilized devices that have been subject to all manufacturing processes.
- 4. Performance Test Human Factors/Usability per IEC 62366 and FDA Guidance 'Applying Human Factors and Usability Engineering to Medical Devices"
- Human factors evaluation was conducted on the subject device and the predicate . device. Based on the results of those studies, the subject device has been found to be
GIA™ Stapler with Tri-Staple™ Technology
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Covidien logo. The logo consists of a blue square with a white cross in the center, followed by the word "COVIDIEN" in blue, sans-serif font. The logo is simple and modern, and the colors are clean and professional.
### K192720 Page 5 of 5
# 510(k) Summary
safe and effective for the intended users, uses and use environments.
- 5. Biocompatibility testing has been conducted to meet the requirements of ISO 10993-1 and FDA Guidance 'Use of international Standard ISO 10993-1" issued June 16, 2016 on the final manufactured, packaged and sterilized medical device that have been subject to all manufacturing processes.
- . Cytotoxicity test
- . Sensitization
- Intracutaneous irritation .
- Acute system toxicity
- . Pyrogenicity
- 6. Sterilization assessment per ISO 11135
- Overkill method used for validation .
- ETO residual test .
- 7. Stability/Shelf-life studies
- Accelerated aging test for 5 years shelf life .
- . Product and packaging functional test
Clinical performance data - No clinical study is deemed necessary since the substantial equivalence has been sufficiently demonstrated by non-clinical studies.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.