Attain Select II + SureValve delivery catheter system, Attain Command + SureValve Left Heart Delivery System, Attain Comman + SureValve Guide Catheters for Left Heart Delivery, C315 Delivery Catheter
K192712 · Medtronic, Inc. · DQY · Aug 3, 2020 · Cardiovascular
Device Facts
Record ID
K192712
Device Name
Attain Select II + SureValve delivery catheter system, Attain Command + SureValve Left Heart Delivery System, Attain Comman + SureValve Guide Catheters for Left Heart Delivery, C315 Delivery Catheter
Applicant
Medtronic, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Aug 3, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The left-heart delivery system is intended for introducing transvenous devices and leads into vessels of the left heart via the coronary sinus. The Attain Select™ II + SureValve™ delivery catheter system is indicated for the delivery of contrast medium and transvenous devices to the coronary sinus and left-heart venous anatomy. The delivery catheter system is indicated for use with outer guide catheters. The C315 is indicated for the introduction of various types of pacing or defibrillator leads and catheters.
Device Story
Catheter systems facilitate delivery of pacing/defibrillator leads, catheters, and contrast media into coronary sinus and left-heart venous anatomy. Systems include guide catheters, dilators, and hemostatic valves (SureValve™). C315 is a slittable catheter for lead introduction. Attain Select™ II and Attain Command™ systems function as telescoping catheters to provide sub-selection capabilities. Used in clinical settings by physicians under fluoroscopic imaging (distal tips are radiopaque). Devices are introduced percutaneously; catheters guide leads to target cardiac veins. Benefits include precise lead placement for cardiac rhythm management. Tip material updated to include hydrolysis inhibitor stabilizer and hindered amine light stabilizer.
Clinical Evidence
No clinical data generated. Substantial equivalence supported by non-clinical bench testing, including mechanical, functional, sterilization, and biocompatibility evaluations.
Technological Characteristics
Catheters constructed of Polyether Block Amide and Polyethylene. Distal tips contain Tungsten Carbide Pebax with hydrolysis inhibitor and hindered amine light stabilizer. Features include integrated hemostatic valves (SureValve™), radiopaque tips, and slittable designs (C315). Sterilization via Ethylene Oxide (EtO). Biocompatibility per ISO 10993-1 for external communicating devices with limited blood contact (<24 hours).
Indications for Use
Indicated for patients requiring transvenous device or lead placement into the coronary sinus or left-heart venous anatomy, or general pacing/defibrillator lead introduction. No specific age or gender contraindications provided.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
August 3, 2020
Medtronic, Inc. Nicola Reidy Regulatory Affairs Specialist Parkmore Business Park West Galway, Ireland
Re: K192712
Trade/Device Name: Attain Select™ II + SureValve™ Delivery Catheter System, Attain Command™ + SureValve™ Left Heart Delivery System, Attain Command™ + SureValve™ Guide Catheters for Left Heart Delivery, C315 Delivery Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: July 2, 2020 Received: July 6, 2020
Dear Nicola Reidy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hetal Odobasic Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K192712
#### Device Name
Attain CommandTM + SureValve™ Left Heart Delivery System; Attain Command™ + SureValve™ Guide Catheters for Left Heart Delivery
Indications for Use (Describe)
The left-heart delivery system is intended for introducing transvenous devices and leads into vessels of the left heart via the coronary sinus.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K192712
Device Name
Attain Select™ II + SureValve™ Delivery Catheter System
Indications for Use (Describe)
The Attain Select™ II + SureValve™ delivery catheter system is indicated for the delivery of contrast medium and transvenous devices to the coronary sinus and left-heart venous anatomy. The delivery catheter system is indicated for use with outer guide catheters.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
[ ] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
EF PSC Publishing Services (301) 443-6740
{4}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K192712
Device Name C315 Delivery Catheter
Indications for Use (Describe)
The C315 is indicated for the introduction of various types of pacing or defibrillator leads and catheters.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}------------------------------------------------
# 510(k) Summary per 21 CRF 807.92
### Date Prepared:
26 September 2019
### Applicant:
Medtronic Ireland Parkmore Business Park West Galway Ireland
### Official Correspondent:
Nicola Reidy Regulatory Affairs Specialist Medtronic Ireland Parkmore Business Park West Galway Ireland Phone: (+353) 91 750772 Fax: (+353) 91 708672 Email: nicola.m.reidy@medtronic.com
### Common Name:
- . C315 Delivery Catheter
- Attain Select™ II + SureValve™ delivery catheter system
- Attain Command™ + SureValve™ Product families (Left Heart Delivery System ● & Guide Catheters for Left Heart Delivery)
### Device Classification:
II
### Regulation Number:
21 CFR 870.1250
### Classification Name:
Percutaneous Catheter
Product Code:
DQY
{6}------------------------------------------------
### Device Description:
The Medtronic C315 Delivery Catheter contains one catheter and one dilator constructed of Polyether Block Amide and Polyethylene respectively. It is designed to aid in the introduction of various types of pacing or defibrillator leads and catheters. There are seven models in the Medtronic C315 Delivery Catheter product family, all of which have the same inner and outer diameter (5.4Fr and 7.0Fr respectively). The models differ in useable length, which varies from 20cm to 43cm. Proximally, the C315 is equipped with a hemostatic valve, and the distal tip is radiopaque to facilitate imaging under fluoroscopy. The C315 is designed to be slittable, thereby allowing its removal after device placement. A variety of curves are available to accommodate various anatomies and different lead locations.
The Attain Command™ + SureValve™ Left Heart Delivery System and Attain Command™ + SureValve™ Guide Catheters for Left Heart Delivery are intended for introducing transvenous devices and leads into vessels of the left heart via the coronary sinus. The Attain Command™ + SureValve™ Left-Heart Delivery System kits each contain two outer guide catheters with an integrated hemostasis valve (SureValve™), up to two valve tools, one dilator, one guide wire, and one slitter. The Attain Command™ + Sure Valve™ Left-Heart Delivery System is available in two models:
- Attain Command™ + SureValve™ 6250VC Left Heart Delivery System .
- Attain Command™ + SureValve™ 6250VS Left Heart Delivery System
With the exception of the two guide catheters, all system components packaged in each kit are identical. Each guide catheter model is different with respect to the guide catheter shape and length.
The Attain Command™ + SureValve™ Guide Catheters for Left Heart Delivery individual packs each contain one guide catheter with an integrated hemostasis valve (SureValve™), up to two valve tools, and one dilator. The Attain Command™ + SureValve™ Guide Catheters for Left Heart Delivery are available in 12 models:
- . Attain Command™ + SureValve™ 6250V-45S Guide Catheter for Left Heart Delivery
- Attain Command™ + SureValve™ 6250V-50S Guide Catheter for Left Heart Delivery ●
- Attain Command™ + SureValve™ 6250V-57S Guide Catheter for Left Heart Delivery ●
- Attain Command™ + SureValve™ 6250V-AM Guide Catheter for Left Heart Delivery ●
- Attain Command™ + SureValve™ 6250V-EH Guide Catheter for Left Heart Delivery .
- Attain Command™ + SureValve™ 6250V-EHXL Guide Catheter for Left Heart ● Delivery
- Attain Command™ + SureValve™ 6250V-MB2 Guide Catheter for Left Heart ● Delivery
- Attain Command™ + SureValve™ 6250V-MB2X Guide Catheter for Left Heart ● Delivery
- Attain Command™ + SureValve™ 6250V-MP Guide Catheter for Left Heart Delivery ●
{7}------------------------------------------------
- Attain Command™ + SureValve™ 6250V-MPR Guide Catheter for Left Heart ● Delivery
- Attain Command™ + SureValve™ 6250V-MPX Guide Catheter for Left Heart ● Delivery
- Attain Command™ + SureValve™ 6250V-3D Guide Catheter for Left Heart Delivery ●
Each model is different with respect to the guide catheter shape and length and dilator length.
The Attain Select™ II + SureValve™ delivery catheter system is designed to facilitate leftheart lead delivery to a desired cardiac vein. The Attain Select™ II + SureValve™ delivery catheter system is indicated for the delivery of contrast medium and transvenous devices to the coronary sinus and left-heart venous anatomy. The delivery catheter system is indicated for use with outer guide catheters. Together, the catheters function as a telescoping system that can provide additional sub selecting capabilities.
The delivery catheter system consists of a delivery catheter with an integrated hemostasis valve (SureValve™), an inner catheter, and up to two valve tools. The Attain Select™ II + SureValve™ delivery catheter system is available in 8 models:
- Attain Select™ II + SureValve™ 6248V-90 Delivery Catheter System ●
- Attain Select™ II + SureValve™ 6248V-90S Delivery Catheter System ●
- Attain Select™ II + SureValve™ 6248V-90L Delivery Catheter System
- Attain Select™ II + SureValve™ 6248V-130 Delivery Catheter System ●
- Attain Select™ II + SureValve™ 6248V-130L Delivery Catheter System ●
- Attain Select™ II + SureValve™ 6248V-90P Delivery Catheter System ●
- Attain Select™ II + SureValve™ 6248V-90SP Delivery Catheter System ●
- Attain Select™ II + SureValve™ 6248V-130P Delivery Catheter System ●
The Attain Select™ II + SureValve™ inner catheter is identical for all configurations. Each model is different with respect to delivery catheter shape and length.
# Indications For Use:
The indications for use for each of the applicable devices is listed below;
- C315 Delivery Catheter .
The device is indicated for the introduction of various types of pacing or defibrillator leads and catheters.
- Attain Command™ + SureValve™ Left Heart Delivery System ●
- Attain Command™ & SureValve™ Guide Catheters for Left Heart Delivery: ●
{8}------------------------------------------------
The left-heart delivery system is intended for introducing transvenous devices and leads into vessels of the left heart via the coronary sinus.
- Attain Select™ II + SureValve™ Delivery Catheter System: ●
The Attain Select™ II + SureValve™ delivery catheter system is indicated for the delivery of contrast medium and transvenous devices to the coronary sinus and left-heart venous anatomy. The delivery catheter system is indicated for use with outer guide catheters.
### Substantially Equivalent Device:
The applicable device uses similar technology and has similar intended uses, function, materials and method of operation to the following predicate devices:
- C315 Delivery Catheter (K101885, cleared on September 9, 2010) .
- Attain Select™ II + SureValve™ delivery catheter system (K123153, cleared on April ● 9, 2013)
- Attain Command™ + SureValve™ Product families (Left Heart Delivery System and ● Guide Catheters for Left Heart Delivery) (K123153, cleared on April 9, 2013)
### Summary of Technological Differences to the Predicate Device:
The difference to the predicate device is the change in the tip material.
| | Current Tip Material | Proposed Tip Material |
|-----------|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Materials | Tungsten Carbide Pebax<br>material | Tungsten Carbide Pebax material with addition<br>of hydrolysis inhibitor stabilizer and a hindered<br>amine light stabilizer. |
### Summary of Non-Clinical Data:
Device integrity testing was performed to support the equivalency of the tip material change of the applicable devices device to the predicate devices. Testing included mechanical, functional, sterilization and biocompatibility testing. The devices listed below met all specified design and performance requirements.
- C315 Delivery Catheter (K101885, cleared on September 9, 2010) ●
- Attain Select™ II + SureValve™ delivery catheter system (K123153, cleared on ● April 9, 2013)
- Attain Command™ + SureValve™ Product families (Left Heart Delivery System and ● Guide Catheters for Left Heart Delivery) (K123153, cleared on April 9, 2013)
# Biocompatibility Information:
The biocompatibility evaluation completed for the devices listed below verifies that the applicable devices are biocompatible.
- C315 Delivery Catheter
- . Attain Select™ II + SureValve™ delivery catheter system
{9}------------------------------------------------
- Attain Command™ + SureValve™ Product families (Left Heart Delivery System and ● Guide Catheters for Left Heart Delivery)
The testing which supports the biocompatibility is consistent with International Standard ISO 10993-1: "Biological Evaluation of Medical devices- Part 1: Evaluation and Testing." When classified according to this standard, the catheter and dilator included in the applicable devices are external communicating devices with limited exposure (<24 hours) to circulating blood.
### Summary of Clinical Data:
Clinical data was not generated. This section is not applicable.
### Sterilization Validation:
The applicable devices will be sterilized using a validated Ethylene Oxide (EtO) sterilization process.
### Conclusion:
Through the data and information presented, Medtronic Ireland considers the
- . C315 Delivery Catheter
- Attain Select™ II + SureValve™ delivery catheter system ●
- Attain Command™ + SureValve™ Product families (Left Heart Delivery System and ● Guide Catheters for Left Heart Delivery)
device with the tip material change to be substantially equivalent to legally marketed predicate devices:
- . C315 Delivery Catheter
- Attain Select™ II + SureValve™ delivery catheter system ●
- Attain Command™ + SureValve™ Product families (Left Heart Delivery System and ● Guide Catheters for Left Heart Delivery)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.