K192710 · Pega Medical, Inc. · HSB · Jun 24, 2020 · Orthopedic
Device Facts
Record ID
K192710
Device Name
The Simple Locking Intramedullary (Slim) System
Applicant
Pega Medical, Inc.
Product Code
HSB · Orthopedic
Decision Date
Jun 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Simple Locking IntraMedullary (SLIM) System is intended as a temporary implant for alignment, stabilization and fixation of long bones that have been surgically prepared (osteotomy) for correction of deformities, or have sustained fractures due to trauma or disease. This includes: - Femur and tibia in the pediatric population (child and adolescent), and small-stature adults such as patient with small intramedullary canals affected by skeletal dysplasias, osteogenesis imperfecta or other bone diseases. - Humerus, ulna, fibula in all patient populations
Device Story
Intramedullary fixation system for long bone stabilization; consists of solid shaft rods with bevelled points, conical threads for epiphyseal/cortical anchorage, and internal hexagonal drive/mechanical threads for insertion/retrieval. Includes proximal/distal locking holes for pinning and 'Bullet' screws for smaller diameter shafts. Used in surgical settings by orthopedic surgeons. Provides mechanical alignment and fixation for fractures or osteotomies; reduces migration risk via conical threads. Benefits patients with small intramedullary canals or bone diseases by providing stable, customizable fixation.
Clinical Evidence
No clinical testing provided. Substantial equivalence based on bench testing, including static and fatigue four-point bending, pull-out/push-out strength, and surgical technique validation on animal models.
Technological Characteristics
Materials: Medical grade Stainless Steel (SS316L, ASTM F138). Design: Solid shaft, bevelled point, conical thread, hexagonal drive. Dimensions: 2.0-6.4 mm diameter, 80-400 mm length. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for pediatric and small-stature adult patients (e.g., skeletal dysplasia, osteogenesis imperfecta) requiring alignment, stabilization, or fixation of femur and tibia, and all patient populations for humerus, ulna, and fibula fractures or osteotomies.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
The Simple Locking IntraMedullary (SLIM) System (K143355)
Stainless Steel Elastic Intramedullary Nail System (K081452)
Submission Summary (Full Text)
{0}------------------------------------------------
June 24, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Pega Medical Inc. Ariel R. Dujovne President and CEO 1111 Highway Chomedey Laval, Quebec, Canada H7W 5J8
Re: K192710
Trade/Device Name: The Simple Locking IntraMedullary (Slim) System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: May 21, 2020 Received: May 26, 2020
Dear Ariel Dujovne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Expiration Date: 06/30/2020
See PRA Statement below.
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K192710
Device Name
The Simple Locking IntraMedullary (SLIM) System
### Indications for Use (Describe)
The Simple Locking IntraMedullary (SLIM) System is intended as a temporary implant for alignment, stabilization and fixation of long bones that have been surgically prepared (osteotomy) for correction of deformities, or have sustained fractures due to trauma or disease. This includes:
- Femur and tibia in the pediation (child and adolescent), and small-stature adults such as patient with small intramedullary canals affected by skeletal dysplasias, osteogenesis imperfecta or other bone diseases.
. Humerus, ulna, fibula in all patient populations
Type of Use (Select one or both, as applicable)
| * | Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
Applicant : Pega Medical Inc. 1111 Highway Chomedey Laval, Quebec, Canada, H7W 5J8 Phone : 1-877-739-5175 Fax: +1 450 233-6358 Contact Person : Ariel R. Dujovne Proprietarv Name : The Simple Locking IntraMedullary (SLIM) System Common Name : SLIM Rod Device Classification : Class II Panel: Orthopedic Classification Name : 21 CFR-888.3020 Intramedullary fixation rod Device Product Code : HSB Establishment Registration Number : 9048931
#### Intended Use:
The Simple Locking IntraMedullary (SLIM) System is intended as a temporary implant for alignment, stabilization and fixation of long bones that have been surgically prepared (osteotomy) for correction of deformities, or have sustained fractures due to trauma or disease. This includes:
- · Femur and tibia in the pediatric population (child and adolescent), and small-stature adults such as patient with small intramedullary canals affected by skeletal dysplasias, osteogenesis imperfecta or other bone diseases.
- · Humerus, ulna, fibula in all patient populations
#### Description:
The Simple Locking Intra-Medullary (SLM) system consists of intramedullary fixation devices for use in long bones. The solid shaft, bevelled point, and pre-determined or blank length options is designed for easy insertion in the medullar canal. Anchorage of the device is achieved through a conical thread to obtain a stable fixation in the epiphyses or cortical bone, which aims to reduce the risk of migration. Internal features, such as a hexagonal drive and an internal mechanical thread in the device, are designed for capture and guidance during insertion and retrieval. Additional proximal and distal locking holes allow pinning to ensure fixation with locking pegs when required, and the Bullet screws allow fixation of smaller diameter shafts that cannot be cross-pinned due to their size. The SLIM implants are manufactured in medical grade Stainless Steel (SS316L, ASTM F138). The rods are available in seven diameters: 2.0.2.6.3.2. 4.0. 4.8. 5.6 and 6.4 mm, from 80mm up to 400 mm in length allows end user to customize the length of the rod.
{4}------------------------------------------------
### Basis for substantial equivalent:
The Simple Locking IntraMedullary (SLM) System is claimed to be substantially equivalent in design, indicated use and function to the following devices:
Primary predicate:
- 0 The Simple Locking IntraMedullary (SLIM) System, K143355, Pega Medical Inc.
References:
- 0 GAP Endo-Exo Medullary System, K160545, Pega Medical Inc.
- 0 Stainless Steel Elastic Intramedullary Nail System, K081452, Synthes (USA)
## Summary of Technologies:
The technological characteristics of the Simple Locking IntraMedullary (SLIM) System are the same or similar to the predicate devices.
## Non-clinical Performance Data:
No additional testing was considered necessary for the SLIM Blanks, pins or pegs.
- SLIM blanks have the identical diameters and material as previously cleared SLIM rods; . therefore, they will have the same mechanical resistance.
- o The added pins and pegs have the same or larger diameter and the same material than previously cleared pins, therefore they will have the same or higher mechanical resistance.
As per the findings of the bench testing and the performance of the data supports the use of this product as substantially equivalent to the predicate device; the anticipated benefits of such a system clearly outweigh the possible residual risks.
Additional testing was conducted for the addition of the Bullet screw to the system. In four-point static and fatigue testing, the Bullet screws were found equivalent for their intended uses. In comparison to the previously cleared locking pins, it was considered that the locking pegs perform equivalently than simple locking pin that is bent against the cortex rather than threaded into the cortex in terms of torsion, drive torque and axial pull-out.
A list of the tests carried for this application can be found below:
- Static four-point bending bench testing;
- Fatigue four-point bending bench testing;
- Pull-out & Push-out strength;
- Surgical technique validation on animal model;
{5}------------------------------------------------
- Evaluation of a Shaft to Bullet assembly and instrumentation: migration resistance of the ● assembly under static loading:
- o Evaluation of a Shaft to Bullet assembly and instrumentation: migration resistance of the assembly under cyclic loading (fatigue);
- Static four-point bending bench test of the Bullet;
- Fatigue four-point bending bench test of the Bullet;
- Surgical technique validation of the SLIM and Bullet; .
- . Comparative loading in physiological relevant setting of the SLIM and Bullet against predicate;
- . Functional testing of the cutting operation of the Intro-OP Implant Cutter (ICON) Prototype on a SLIM blank.
### Clinical Performance Data:
No clinical testing is provided as a basis for substantial equivalence.
### Conclusion:
Based on the similarities in the intended use, design, materials, manufacturing methods, and packaging, The Simple Locking IntraMedullary (SLIM) System has been established as substantially equivalent to the previously cleared predicate device. Furthermore, mechanical evaluation results demonstrate that the proposed SLIM System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.