InMode System with the Morpheus8 (Fractora) Applicators
K192695 · Inmode MD , Ltd. · GEI · Dec 27, 2019 · General, Plastic Surgery
Device Facts
Record ID
K192695
Device Name
InMode System with the Morpheus8 (Fractora) Applicators
Applicant
Inmode MD , Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 27, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The InMode System with the Morpheus8 (Fractora) Applicators is intended for use in Dermatological and General Surgical procedures for Electrocoagulation and Homeostasis. At higher energy levels greater than 62 ml/pin, use of the Morpheus8 (Fractora) Applicators is limited to Skin Types I-V.
Device Story
InMode System is a computerized, RF-based electrosurgical device for dermatological electrocoagulation and hemostasis. System comprises an AC/DC power supply, RF generator, controller, and LCD touch screen. Operators (clinicians) use a foot switch to activate energy delivery through disposable, single-use fractional RF electrode pin tips (12, 24, 40, or T-pin configurations). Device delivers RF energy to skin via multi-electrode pin arrays. System monitors RF parameters to enhance safety and minimize side effects. Output is used by clinicians to perform surgical procedures; clinical benefit is achieved through controlled tissue coagulation. Device is intended for clinical use.
Clinical Evidence
No clinical data. Bench testing only. Performance testing compared RF output (pulse duration, energy per pin, total energy, peak-to-peak voltage) of new tip heads against predicate device across various resistance loads, demonstrating equivalent performance.
Indicated for dermatological and general surgical procedures requiring electrocoagulation and hemostasis. Use at energy levels >62 mJ/pin is restricted to patients with Skin Types I-IV.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
InMode System with the Fractora3D/3D-90 Applicators (K180189)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 27, 2019
Inmode MD Ltd. % Amit Goren Regulatory Manager A. Stein- Regulatory Affairs Consulting Ltd. 20 Hata'as Str., Suite 102 Kfar Saba, 4442520 Il
Re: K192695
Trade/Device Name: InMode System with the Morpheus8 (Fractora) Applicators Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: November 25, 2019 Received: November 27, 2019
Dear Amit Goren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
Device Name
InMode System with the Morpheus8 (Fractora) Applicators
Indications for Use (Describe)
The InMode System with the Morpheus8 (Fractora) Applicators is intended for use in Dermatological and General Surgical procedures for Electrocoagulation and Homeostasis.
At higher energy levels greater than 62 ml/pin, use of the Morpheus8 (Fractora) Applicators is limited to Skin Types I-V.
Type of Use (Select one or both, as applicable)
| > Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
# INMODE RF SYSTEM
# 510(k) Number K192695
## Applicant Name:
| Company Name: | InMode Ltd |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Tabor Building, Shaar Yokneam<br>POB 44, Yokneam 20692<br>Israel<br>Tel: +972-4-9097470<br>Fax: +972-4-9097471<br>E-mail: amit@asteinrac.com |
| Contact Person: | Amit Goren |
| Official Correspondent: | Amit Goren |
| Company Name: | A. Stein - Regulatory Affairs Consulting Ltd. |
| Address: | 20 Hata'as Str., Suite 102<br>Kfar Saba 4442520 Israel<br>Tel: +972-9-7670002<br>Fax: +972-9-7668534<br>E-mail: amit@asteinrac.com |
| Date Prepared: | November 25, 2019 |
| Trade Name: | InMode System with the Morpheus8 (Fractora) Applicators |
| Classification Name: | CFR Classification section 878.4400; (Product code GEI) |
| Classification: | Class II Medical Device |
## Predicate Device:
The subject device is substantially equivalent to the following predicate device:
| Device | Manufacturer | 510(k) No. |
|--------------------------------------------------------|----------------|------------|
| InMode System with the<br>Fractora3D/3D-90 Applicators | InMode MD Ltd. | K180189 |
{4}------------------------------------------------
## Device Description:
The InMode System with the Morpheus8 (Fractora) Applicators is a computerized, programmed, RF technology-based device intended for dermatological applications which requires skin electrocoagulation and hemostasis.
The device platform is basically constituted on the same system platform as FDA cleared for the proposed predicate device; The InMode System with the Fractora3D/3D-90 Applicators (K180189). The device applicators employ fractional RF multi-electrode technology for procedures requiring electrocoagulation and hemostasis. The Morphues8 (Fractora) Applicators are designed to deliver radiofrequency energy to the skin in a fractional manner, via an array of multielectrode pins.
The Device provides enhanced safety while minimizing possible side effects by monitoring RF parameters. The InMode System with the Morpheus8 (Fractora) Applicators consists of an AC/DC power supply unit, RF generator, controller and user interface including LCD touch screen. The Morpheus8 (Fractora) Applicators are connected to the console via a cable and a foot switch activates the energy delivery to the applicator. The subject device applicator comprises a disposable, single use, fractional RF electrode pins tip head.
The InMode System with the Morpheus8 (Fractora) Applicators comprises the following applicator tip heads:
- Morpheus8 24 Pin Applicator (Fractora3D) tip (FDA Cleared in K180189)
- Morpheus8 40 Pin treatment tip (New tip) .
- . Morpheus8 12 Pin treatment tip (New tip)
- . Morpheus8 T Pin treatment tip (New tip)
### Device Specifications:
| Main Line Frequency (nominal) | 50 - 60 Hz |
|----------------------------------------|----------------------|
| Input Voltage (nominal) | 100 - 240 VAC |
| Electrosurgical Unit dimensions (inch) | 18.2"W x18.2"D x40"H |
| Platform weight (lb.) | 70.5 |
| Applicator weight (lb.) | 1.26 |
| RF Max Output Power (Watt) | 65 |
| RF Output Frequency (MHz) | $1\pm 2%$ |
{5}------------------------------------------------
## Intended Use/Indication for Use:
The InMode System with the Morpheus8 (Fractora) Applicators is intended for use in Dermatologic and General Surgical procedures for Electrocoagulation and Hemostasis.
At higher energy levels greater than 62 mJ/pin, use of the Morrpheus8 (Fractora) Applicator is limited to Skin Types I-IV.
## Performance Standards:
The InMode System with the Morpheus8 (Fractora) Applicators complies with the following FDA recognized consensus standards:
- AAMI/ANSI 60601-1 (2012), Medical Electrical Equipment Part 1: General . Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, Mod).
- IEC 60601-1-2 (Edition 3.0, 2007), Medical Electrical Equipment Part 1-2: . General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.
- IEC 60601-2-2 Edition 5.0 2009-02, Medical Electrical Equipment Part 2-2: . Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories.
## Non-Clinical (Bench) Performance Data:
The InMode System with the Morpheus8 (Fractora) Applicators was tested for its applicator tip RF output performance specifications. The device utilizing the Morphues8 12, 24, 40 pin and T tip heads preset to different RF energy levels, was tested under different resistance loads with the following parameters measured and compared to the results obtained in similar testing conducted on the predicate device: pulse duration, energy per pin, total energy, and pick to pick voltage. The results demonstrate equivalent measurements obtained for the subject and predicate devices.
## Animal Performance Data / Histology Data:
Not Applicable
## Clinical Performance Data:
Not Applicable
{6}------------------------------------------------
## Biocompatibility
All of the modified device materials which come in direct contact with the patient skin are biocompatible and identical to the materials used in the predicate device manufacturing.
## Substantial Equivalence:
A comparison table is provided below comparing the intended use and basic technological characteristics of the InMode System with the Morpheus8 (Fractora) Applicators to the intended use and basic technological characteristics of the predicate device.
| Technological<br>Characteristic | InMode System with the<br>Morpheus8 (Fractora)<br>Applicators<br>InMode Ltd.<br>(K192695) | InMode System with the<br>Fractora3D/3D-90<br>Applicators<br>InMode MD Ltd.<br>(K180189) |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Dimensions | 46cm W x 46cm D x<br>100cm H<br>[18.2" W x 18.2" D x<br>40" H] | 46cm W x 46cm D x<br>100cm H<br>[18.2" W x 18.2" D x<br>40" H] |
| Weight | | |
| Platform weight | 32 Kg (70.5 lbs.) | 32 Kg (70.5 lbs.) |
| Applicator weight | 0.4 Kg (1.26 lbs.)<br>Tip weight - 0.02 Kg | 0.4 Kg (1.26 lbs.)<br>Tip weight - 0.02 Kg |
| Applicator Tip Heads: | Morpheus8 (Fractora) 12 tip head Morpheus8 (Fractora) 24 tip head Morpheus8 (Fractora) 40 tip head Morpheus8 (Fractora) T tip head | Morpheus8 (Fractora3D)<br>24 tip head |
| Number of pins | 12-40 pins | 24 pins |
| Maximal Treatment<br>depth | 4.0mm<br>T tip head - 0.5mm<br>(Fixed) | 4.0mm |
| Cable Dimensions: | 250 cm | 250 cm |
| Performance | Frequency: 1 MHz | Frequency: 1 MHz |
{7}------------------------------------------------
| Technological<br>Characteristic | InMode System with the<br>Morpheus8 (Fractora)<br>Applicators<br>InMode Ltd.<br>(K192695) | InMode System with the<br>Fractora3D/3D-90<br>Applicators<br>InMode MD Ltd.<br>(K180189) |
|---------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| | Maximal RF output<br>power: 65W | Maximal RF output<br>power: 65W |
| Standards Met | IEC 60601-1<br>IEC 60601-1-2<br>ANSI AAMI 60601-2-2<br>for safety of high<br>frequency surgical<br>equipment | IEC 60601-1<br>IEC 60601-1-2<br>ANSI AAMI 60601-2-2<br>for safety of high<br>frequency surgical<br>equipment |
| Reprocessing &<br>Sterilization | For single use.<br>All applicator tip heads<br>are gamma sterilized. | For single use.<br>Non-sterilized. To be<br>sterilized prior to use. |
## Comparison Discussion:
The indications for use and technological characteristics of the modified device are substantially equivalent to the indications for use and technological characteristics of the predicate device.
The design and components of the modified device, including the console (with power supply, RF generator, controller and display panel) and the Applicator handle (with cable and connector to console) are identical to the design and components found in the predicate device. The modified device comprises the predicate device Morpheus8 (Fractora) 24 pin tip head and additional three compatible tip heads; the Morpheus8 (Fractora) 12, 40 pin and T tip heads. The performance specifications (including RF frequency, pulse duration, RF current distribution and RF energy per pin) of the modified device are comparable to those of the predicate device. The safety features and compliance with safety standards in the modified device are identical to the safety features and compliance with safety standards found in the predicate device. Patient contact materials are also identical. Any minor differences in the technological characteristics do not raise new safety or effectiveness concerns. Furthermore, the modified device underwent bench performance testing, including software validation testing, electrical and mechanical safety testing to IEC 60601-1 and electromagnetic compatibility testing according to IEC 60601-1-2 and IEC 60601-2-2. Sterilization methods applied on the device applicator tip heads were validated in accordance with ISO 11137-1.
{8}------------------------------------------------
The performance tests demonstrated that the minor differences in the device design and specifications meet the system requirements and do not raise new safety or effectiveness concerns.
Consequently, it can be concluded that the InMode System with the Morpheus8 (Fractora) Applicators is substantially equivalent to the predicate InMode System with the Fractora3D/3D-90 Applicators, FDA cleared in 510(k) K180189, and therefore, may be legally marketed in the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.