K192662 · CapsoVision, Inc. · NEZ · Feb 14, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K192662
Device Name
CapsoCam Plus (SV-3) Capsule Endoscopy System
Applicant
CapsoVision, Inc.
Product Code
NEZ · Gastroenterology, Urology
Decision Date
Feb 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1300
Device Class
Class 2
Indications for Use
The CapsoCam ® Plus video capsule system is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.
Device Story
Single-use ingestible video capsule; acquires/stores video images in on-board memory during transit through GI tract via natural peristalsis. Patient uses retrieval kit to collect excreted capsule; physician downloads/reviews images via CapsoView software or CapsoCloud web-based application. System includes capsule, data access system, retrieval kit, and software for image review/reporting. Used in clinical settings; operated by physicians. Cloud-based software manages procedures, reviews, and analyzes images. Output allows visualization of small bowel mucosa to assist in detecting abnormalities, aiding clinical diagnosis.
Clinical Evidence
Bench testing only. Performance testing conducted per FDA recognized consensus standards. Included in-vivo testing for image comparison between proposed and predicate software and image substantial equivalence analysis. No additional biocompatibility, EMC, or electrical safety testing performed as materials and design remain unchanged from predicate.
Technological Characteristics
Single-use ingestible capsule; telemetric imaging system. Components: capsule, data access system, retrieval kit, CapsoView software, CapsoCloud software. Connectivity: cloud-based software application (CapsoCloud) and local software (CapsoView). Software-based image review and reporting. Materials and sterilization processes identical to predicate (K183192).
Indications for Use
Indicated for visualization of small bowel mucosa in adults; used as a tool for detection of small bowel abnormalities.
Regulatory Classification
Identification
An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 14, 2020
CapsoVision, Inc. Azimun Jamal Sr. Director of OA/RA 18805 Cox Avenue Suite 250 Saratoga, CA 95070
Re: K192662
Trade/Device Name: CapsoCam® Plus (SV-3) Capsule Endoscopy System Regulation Number: 21 CFR 876.1300 Regulation Name: Ingestible telemetric gastrointestinal capsule imaging system Regulatory Class: II Product Code: NEZ Dated: October 10, 2019 Received: October 11, 2019
Dear Azimun Jamal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Shanil Haugen, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192662
Device Name
CapsoCam® Plus (SV-3) Capsule Endoscopy System
Indications for Use (Describe)
The CapsoCam ® Plus video capsule system is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;"> <b> <span style="font-family: Symbol;">☑</span> </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size: 10pt;"> <b>□</b> </span> Over The Counter Use (21 CFR 801 Subpart </div> |
|X | Prescription Use (Part 21 CFR 801 Subpart D)
|_ | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K192662 Page 1 of 5
Image /page/3/Picture/1 description: The image contains the logo for CapsoVision. The logo features a blue and green swirl graphic above the company name, which is written in a modern, sans-serif font. Below the company name is the tagline "empowering through innovation" in a smaller, handwritten-style font.
# 5. 510(k) Summary:
# Date Prepared: September 24, 2019
| Manufacturer: | CapsoVision, Inc.<br>18805 Cox Ave, Suite 250<br>Saratoga, CA 95070<br>Phone: (408) 624-1488<br>Fax: (408) 370-4795 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Azimun Jamal<br>Director, Quality Assurance & Regulatory Affairs<br>CapsoVision, Inc.<br>Phone: (408) 866-6358<br>E-mail: azimun.jamal@capsovision.com |
| Device Information: | |
| Classification: | Class II |
| Trade Name: | CapsoCam® Plus (SV-3) Capsule Endoscopy System |
| Common Name: | System Imaging, Gastrointestinal Wireless Capsule |
| Classification Name: | Ingestible telemetric gastrointestinal capsule imaging system<br>(21 CFR § 876.1300) |
| Product Code: | NEZ |
| Predicate Device: | CapsoCam® Plus (SV-3) System, 510(k) # K183192, cleared on 04/19/19 |
## Indications for Use:
The CapsoCam® Plus video capsule system is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel
# Device Description
CapsoCam® Plus (SV-3) capsule is a single-use. ingestible video capsule that acquires and stores video images in on-board memory while moving through the gastrointestinal tract, propelled by natural peristalsis. The patient retrieves the provided retrieval kit and returns it to the physician who downloads and reviews the images on a computer. The capsule is typically excreted within 3 to 30 hours after swallowing.
CapsoCam® Plus (SV-3) capsule endoscope is a single-use ingestible capsule system for diagnostic visualization of the adult small bowel. The system consists of the following accessories:
- CapsoCam® Plus (SV-3) capsules, which are intended for visualization of the small bowel mucosa in adults and used as a tool in the detection of abnormalities of the small bowel.
- CapsoView (CVV) software program, which is used to download and view CapsoCam® Plus ● (SV-3) capsule images and to generate capsule endoscopy reports.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for CapsoVision. The logo features a blue and green swirl design above the company name. Below the company name is the tagline "empowering through innovation" in a cursive font.
K192662 Page 2 of 5
- CapsoAccess (CDAS) Capsule Data Access System, which accesses data from the ● CapsoCam® Plus (SV-3) capsules.
- CapsoRetrieve (CVR1) Capsule Retrieval Kit, which is used for the collection, storage, and ● transportation of the excreted CapsoCam® Plus (SV-3) capsules.
- CapsoCloud (CLD), which is a cloud-based software application used to manage procedures ● and to review and analyze images from the CapsoCam® Plus (SV-3) capsule.
# Software Description and System Requirements
# CapsoView CVV:
Capso View is Capso Vision's proprietary software program used to download and view CapsoCam images and to generate capsule endoscopy reports.
| CPU | Minimum Required | Recommended |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| | Intel Core i3, 4th generation (for image reviewing and reporting only)<br>Intel Core i5, 4th generation (4+ threads for downloading, reviewing and reporting) | Intel Core i7 Quad-Core, 6th generation (8+ threads for optimal downloading performance) |
| Memory | Minimum Required | Recommended |
| | 4GB | 8GB |
| Operating System | Minimum Required | Recommended |
| | PC: Windows 7 (Service Pack 1) (64-bit)<br>Mac: OS X El Capitan | PC: Windows 10 (64-bit)<br>Mac: macOS Sierra or macOS High Sierra |
| Display | Minimum Required | Recommended |
| | Display Resolution of: 1366x768 for Windows 1280x800 for Mac<br>For low-resolution monitors, it is not recommended to set Scaling over 100%. | Display Resolution: 1920x1080 Panel Type: IPS (In-Plane Switching) Color Gamut: 100% sRGB with Delta-E <2 (For maximum color accuracy) |
| Additional Software | PDF Reader (e.g. Adobe Reader) | |
CapsoView® can be installed on a computer that meets the following requirements:
# CapsoCloud CLD
CapsoCloud is a cloud-based software application used to manage procedures and to review and analyze images from the CapsoCam® Plus capsule endoscopy system.
CapsoCloud is a web-based interface that can be accessed on a device that meets the following requirements:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for CapsoVision. The logo features a circular graphic with blue and green swirls at the top. Below the graphic is the company name, "CapsoVision," with the first letter capitalized and the rest in lowercase. Underneath the company name is the tagline "empowering through innovation" in a smaller, handwritten-style font.
# SYSTEM REQUIREMENTS
## 1. Computer
| CPU | Recommended |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| | Intel Core i7 Quad-Core, 6th generation |
| Memory | Recommended |
| | 8GB |
| Operating System | Recommended |
| | PC: Windows 10 (64-bit) |
| | Mac: macOS Sierra or macOS High Sierra |
| Display | Recommended |
| | Display Resolution: 1920x1200 Panel Type: IPS (In-Plane<br>Switching) Color Gamut: 100% sRGB with Delta-E <2 (For<br>maximum color accuracy) |
| Network | Recommended |
| | 25 Mbps, 100ms latency |
| Chrome | Supported Versions |
| | 75.0.3770 - 79.0.3945 |
| Additional Software | PDF Reader (e.g. Adobe Reader) |
## 2. Mobile device (iPhone, Android phone, iPad, Android tablet)
## a. Procedure management and shipping label generation
| Processor | Recommended |
|------------|------------------------------------------------------|
| | iPhone: A9<br>Android Phone: Qualcomm Snapdragon 425 |
| Memory | Recommended |
| | 2GB |
| iOS | Supported Versions |
| | 9.0 and above |
| Android OS | Supported Versions |
| | 4.1.x and above |
#### b. Video Streaming (iPad)
| Memory | Recommended |
|--------|-------------|
|--------|-------------|
#### CONFIDENTIAL
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for CapsoVision. The logo features a circular graphic with swirling blue and green colors at the top left. Below the graphic is the company name, "CapsoVision," written in a modern, handwritten-style font. Underneath the company name is the tagline "empowering through innovation" in a smaller, cursive font.
| | 1.4GB |
|------------|-------------------------------------------------|
| iOS | Supported Versions |
| | 9.0 and above |
| iPad Model | Recommended |
| | iPad 6 2018, iPad air 2019, iPad Pro all models |
## Comparison of the Technological Characteristic with the Predicate Device:
The characteristics of the modified CapsoCam® Plus (SV-3) Capsule Endoscopy System is substantially equivalent to the predicate device based on the indications for use and technological characteristics.
The CapsoCam® Plus (SV-3) capsule endoscopy system that supports the Capsule endoscopy procedure consists of the following:
- CapsoCam® Plus (SV-3) capsules, which is intended for visualization of the small bowel ● mucosa in adults and used as a tool in the detection of abnormalities of the small bowel.
- CapsoView (CVV) software program, which is used to download and view CapsoCam® Plus ● (SV-3) images and to generate capsule endoscopy reports.
- CapsoAccess (CDAS) Capsule Data Access System, which accesses data from the ● CapsoCam® Plus (SV-3) capsules.
- CapsoRetrieve (CVR1) Capsule Retrieval Kit, which is used for the collection, storage, and transportation of the excreted CapsoCam® Plus (SV-3) capsules.
The modified CapsoCam® Plus (SV-3) Capsule Endoscopy System has the following similarities to the predicate device:
- Has the same intended use and Indications for Use ●
- Uses the same operating principle ●
- Incorporates the same capsule design ●
- Incorporates the same materials
- Has the same shelf life
- Is packaged and cleaned using the same material and processes
The modified CapsoCam® Plus (SV-3) Capsule Endoscopy System is different from the predicate system in that it includes a new cloud-based software application to manage procedures and review and analyze images from the CapsoCam capsule.
## Performance Data
The CapsoCam® Plus (SV-3) Capsule Endoscope System performance testing was conducted per the appropriate FDA Recognized Consensus Standards and required bench testing. The modified CapsoCam® Plus (SV-3) Capsule Endoscopy System was tested for the CapsoCloud Software application related changes including in-vivo testing for the image comparison from the proposed and the predicate software and Image Substantial Equivalence Analysis. No additional testing was performed since no other changes were made to the subject device.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for CapsoVision. The logo features a stylized blue and green swirl above the company name, which is written in a modern, sans-serif font. Below the company name is the tagline "empowering through innovation" in a smaller, handwritten-style font.
K192662 Page 5 of 5
The result of this testing concludes that the material and technological characteristics of the CapsoCam® Plus (SV-3) Capsule Endoscopy System do not raise different questions of safety or effectiveness when compared to the predicate device.
Note: Since the material is the same as the cleared device CapsoCam® Plus (SV-3) Capsule Endoscopy System under K183192, no additional testing was performed for the Biocompatibility, EMC or electrical Safety test for the subject device.
## Conclusions:
The subject and predicate device share the same intended use, Indications for Use, and similar technological characteristics. The differences in technological characteristics do not impact safety or effectiveness and verification and validation data demonstrate that the subject device is as safe and effective as the predicate device. The subject device is substantially equivalent to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.