K192607 · Viasonix , Ltd. · IRP · Dec 19, 2019 · Physical Medicine
Device Facts
Record ID
K192607
Device Name
LiteWalk
Applicant
Viasonix , Ltd.
Product Code
IRP · Physical Medicine
Decision Date
Dec 19, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LiteWalk is intended for home and hospital use for temporary increase in local blood circulation in the lower extremities and temporary relief of minor muscle aches and pains where applied
Device Story
LiteWalk is a pneumatic sequential foot compression device for temporary improvement of lower limb blood circulation and relief of minor muscle aches. The system comprises a pneumatic unit (pump, valves, pressure sensor, tubing) powered by a 12V DC supply. It connects to a pair of inflatable sandals, each featuring four chambers positioned for the heel, arch, metatarsals, and toes. The device operates by sequentially inflating these chambers from heel to toes in 2-3 cycles per minute. Used in home or hospital settings, the device is operated by the patient or clinician. The user initiates therapy via a power button; the device provides relief through mechanical compression. Benefits include increased local blood flow and pain relief. The device is non-invasive and intended for OTC use.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, verification of specifications, and software validation.
Indicated for home and hospital use for temporary increase in local blood circulation in the lower extremities and temporary relief of minor muscle aches and pains. Contraindicated in patients with congestive heart failure, deep vein thrombosis (DVT), blood clots, hemorrhagic disease, coagulation problems, thrombophlebitis, or pulmonary embolism.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
December 23, 2019
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue font. The HHS logo is a symbol of an eagle.
Viasonix Ltd. Dan Manor CEO 10 Hamelacha Street Raananna, IL 4366105
Re: K192607 Trade/Device Name: LiteWalk Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: Class II Product Code: IRP Dated: September 16, 2019 Received: September 20, 2019
Dear Dan Manor:
This letter corrects our substantially equivalent letter of December 19, 2019.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Heather L. Dean -S
for: Vivek Pinto, Ph.D. Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K192607
Device Name LiteWalk
Indications for Use (Describe)
The LiteWalk is intended for home and hospital use for temporary increase in local blood circulation in the lower extremities and temporary relief of minor muscle aches and pains where applied .
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## SECTION 5 - 510(K) SUMMARY
#### 5.1 ADMINISTRATIVE INFORMATION
| Date: | 10-September-2019 |
|------------------------------------------------|-------------------------------------------------------------------------------------------|
| Submitter: | Viasonix Ltd.<br>10 Hamelacha Street<br>Raanana , ISRAEL 4366105<br>Phone : 972-9-7441692 |
| Official Correspondent: | Dan Manor, CEO |
| Trade Name: | LiteWalk |
| Classification Name:<br>Classification Number: | Powered inflatable tube massager<br>21 CFR 890.5650 |
| Product Code:<br>Device Class: | IRP<br>Class II |
| Predicate Devices: | PPCID, K000655 |
#### 5.2 Device Description
The LiteWalk is a pneumatic sequential foot compression device, designed primarily for temporary improvement in blood flow circulation in the lower limb. The system is based on a pneumatic unit that includes a pump, valves, pressure sensor and tubing, and is fed by an external power supply of 12V DC.
The unit is connected through 2 sets of 4-tubes to 2 inflatable sandals. Each sandal has 4 inflatable pressure chambers, located to match the foot anatomy: heel, arch, metatarsals, and toes. The sandals come in a pair, for right and left feet, and in 3 sizes: small, medium, large. The operation is based on sequential inflation of the 4 chambers, from heel to toes.
#### 5.3 Intended Use and Indications for Use
The LiteWalk is Intended for home and hospital use for temporary increase in local blood circulation in the lower extremities and temporary relief of minor muscle aches and pains where applied
{4}------------------------------------------------
#### 5.4 CONTRAINDICATIONS
The LiteWalk is not intended for use if any of the following medical conditions exist. Consult with a physician whenever there is doubt whether it is clinically safe to use the LiteWalk.
• Presumptive evidence of congestive heart failure where increase in blood returned to the heart could be detrimental
- Pre-existing deep vein thrombosis (DVT) or blood clots
- Existing hemorrhagic disease or coagulation problems
- Thrombophlebitis
- Pulmonary embolism
#### 5.5 Summary of Technical Characteristics
The LiteWalk device is similar to the predicate device cited above [section 5.1] intended for powered inflatable tube massager applications.
The technological characteristics, e.g., overall design, materials, mechanism of action, mode of operation, performance characteristics, the intended use, use environment and target patient population of the LiteWalk device is substantially equivalent to the predicate device cited above.
{5}------------------------------------------------
# 5.5.1 Summary table of Comparison
| Specification | LiteWalk | PPCID | Differences discussion |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | Proposed Device | K000655 | NA |
| Manufacturer | VIASONIX LTD. | Medical Dynamics LLC | NA |
| Product Regulation | 21CFR890.5650 - Powered<br>inflatable tube massager | 21CFR890.5650 - Powered<br>inflatable tube massager | Identical |
| Product Code | IRP - Powered inflatable tube<br>massager | IRP - Powered inflatable tube<br>massager | Identical |
| Indications for use | Intended for home and<br>hospital use for temporary<br>increase in local blood<br>circulation in the lower<br>extremities and temporary<br>relief of minor muscle aches<br>and pains where applied | Intended for home and<br>hospital use for temporary<br>increase in local blood<br>circulation in the lower<br>extremities and temporary<br>relief of minor muscle aches<br>and pains where applied | Identical |
| Contraindications | Use of the LiteWalk is<br>contraindicated in patients<br>for whom:<br>• Presumptive evidence of<br>congestive heart failure<br>where increase in blood<br>returned to the heart<br>could be detrimental<br>• Pre-existing deep vein<br>thrombosis (DVT) or<br>blood clots<br>• Existing hemorrhagic<br>disease or coagulation<br>problems<br>• Thrombophlebitis<br>• Pulmonary embolism | Use of the PPCID is<br>contraindicated in patients for<br>whom an increase in<br>blood returned to the heart<br>could be detrimental,<br>including those with<br>presumptive evidence of<br>congestive heart failure, pre-<br>existing deep vein thrombosis,<br>thrombophelbitis, existing<br>hemorrhagic disease,<br>coagulation problems, or<br>pulmonary embolism | Identical |
| Environment<br>of Use | Hospital and Home care | Hospital and Home care | Identical |
| Rx OR OTC | OTC | OTC | Identical |
| Mode of<br>compression | Sequential | Sequential<br>*Not publicly available | Similar to the predicate |
| Use Time [min] | Recommended 30 min | 30 min or more<br>*Not publicly available | Similar to the predicate |
| Sequential cycles<br>time | 2-3 cycles per minute | 2-3 cycles per minute<br>*Not publicly available | Similar to the predicate |
| Modes | 1 mode : Sequential<br>inflation/deflation of 4<br>chambers from heel to toes | 1 mode : Sequential<br>inflation/deflation of 4<br>chambers from heel to toes<br>*Not publicly available | Similar to the predicate |
| Specification | LiteWalk | PPCID | Differences discussion |
| Max Pressure | 160 mmHg | 80-200 mmHg<br>*Not publicly available | Similar to the predicate.<br>Within the predicate pressure range |
| Types of Garments | Foot Sandal | Foot Sandal<br>*Not publicly available | Similar to the predicate.<br>Foot Sandal type is supported by the<br>predicate device |
| Garment sizes | 3 sizes: small, medium, large | 3 sizes: small, medium, large<br>*Not publicly available | Similar to the predicate |
| Patient contact<br>materials | Polyester [PVB0123, Colored<br>Fabric] | Not known<br>*Not publicly available | Similar to the predicate.<br>Both sharing a fabric for the Sandals.<br>LiteWalk Polyester skin contact material<br>comply with ISO 10993 |
| Patient contact | Non conductive | Non conductive<br>*Not publicly available | Similar to the predicate |
| Number of chambers | 4 | 4<br>*Not publicly available | Similar to the predicate |
| Line Voltage /<br>Frequency | 100-240 Vac / 50-60Hz<br>12V , 2.0A | 115 Vac / 60 Hz, 7W<br>*Not publicly available | Similar to the predicate |
| Power Consumption | 8 W | 7W<br>*Not publicly available | Similar to the predicate |
| Safety feature | Power button on main unit<br>allows user to stop therapy<br>session at any time. | Power button on main unit<br>allows user to stop therapy<br>session at any time.<br>*Not publicly available | Similar to the predicate |
| User Interface | No | No<br>*Not publicly available | Similar to the predicate |
| Indicator | Green led | Led<br>*Not publicly available | Similar to the predicate |
| Technology | Compressor and valve system<br>which sequentially inflates<br>inflatable chambers | Compressor and air distributor<br>system which sequentially<br>inflates inflatable chambers<br>*Not publicly available | Both sharing the same functionality.<br>The valve allows better control over<br>inflation and allows quick deflation.<br>This slight change doesn't impact<br>product safety and performance. |
| Standards | IEC 60601-1<br>IEC60601-1-2<br>IEC60601-1-11<br>ISO 10993-1 | IEC 60601-1<br>IEC60601-1-2<br>ISO 10993-1<br>*Not publicly available | Similar to the predicate.<br>The LiteWalk meets all latest standards<br>edition |
{6}------------------------------------------------
{7}------------------------------------------------
#### 5.6 UTILIZATION OF STANDARDS AND GUIDANCE'S:
The LiteWalk meets the following standards:
- 1. IEC 60601-1:2005+A1:2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- 2. IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- 3. IEC 60601-1-11: 2015 Medical electrical equipment Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- 4. IEC 62366-1: 2015 Medical devices Part 1: Application of usability engineering to medical devices [Including CORRIGENDUM 1 (2016)]
- 5. IEC 60601-1-6: 2013 [Edition 3.1] Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
{8}------------------------------------------------
#### 5.7 Summary of Non-Clinical Performance Testing
Summary of Non-Clinical Tests:
The LiteWalk device have been thoroughly tested through verification of specifications and validation, including software validation.
#### 5.8 Summary of Clinical Performance Data
No clinical study was conducted to support this application.
#### 5.9 Conclusions
Based on its underlying technology and bench tests performed, the LiteWalk is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.