K192583 · Rohrer Aesthetics, LLC · GEX · Jan 17, 2020 · General, Plastic Surgery
Device Facts
Record ID
K192583
Device Name
PicoLazer Laser System
Applicant
Rohrer Aesthetics, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PicoLazer laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery. 1064 nm The 1064 nm wavelength of the PicoLazer laser system is indicated for: - treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI) • tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI) 532 nm - The 532 nm wavelength of the PicoLazer laser system is indicated for: - treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III - tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III
Device Story
PicoLazer is a dual-wavelength (532 nm and 1064 nm) Nd:YAG laser system delivering 450-picosecond pulses. System components include a console, articulated arm, laser handpiece, footswitch, and remote interlock. Operated by clinicians in dermatology, general, or plastic surgery settings. The device emits laser energy to target pigmented lesions or tattoo inks; the short pulse duration (picosecond) facilitates photomechanical disruption of the target. Healthcare providers use the system to perform surgical and aesthetic procedures. Benefits include effective removal of tattoos and treatment of benign pigmented lesions across specified skin types.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), IEC 60601-2-22 (laser safety), and IEC 60825-1 (laser product safety).
Technological Characteristics
Dual-wavelength (532/1064 nm) Nd:YAG laser. Pulse duration: 450 ps. Max energy: 500 mJ (1064 nm), 250 mJ (532 nm). Peak power: 1.3 GW (1064 nm), 0.8 GW (532 nm). Spot size: up to 10 mm. Repetition rate: 1-10 Hz. Energy source: electrical (100-120 VAC or 200-240 VAC).
Indications for Use
Indicated for treatment of benign pigmented lesions and tattoo removal. 1064 nm: benign pigmented lesions (Fitzpatrick I-VI); dark/multicolored tattoos (Fitzpatrick I-VI). 532 nm: benign pigmented lesions (Fitzpatrick I-III); lighter colored tattoos including red/yellow (Fitzpatrick I-III).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.
Rohrer Aesthetics, LLC Mark Rohrer President 105 Citation Court Homewood, Alabama 35209
Re: K192583
Trade/Device Name: PicoLazer Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: December 17, 2019 Received: December 18, 2019
Dear Mark Rohrer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Mavadia-Shukla Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192583
Device Name
PicoLazer
#### Indications for Use (Describe)
The PicoLazer laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.
1064 nm
The 1064 nm wavelength of the PicoLazer laser system is indicated for:
- · treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI)
• tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)
532 nm
- The 532 nm wavelength of the PicoLazer laser system is indicated for:
- · treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III
- · tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III
Type of Use (Select one or both, as applicable)
| Restricted Use (Part 21 CFR 201 Subpart D) | <span style="font-size: 20px;"><b> ☑ </b></span> |
|---------------------------------------------|--------------------------------------------------------|
| Over-The-Counter Use (21 CFR 201 Subpart G) | <span style="font-size: 20px;"><b> ☐ </b></span> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) Summary PicoLazer Multi-wavelength Laser K192583
This 510(K) Summary of safety and effectiveness for the PicoLazer Multi-wavelength Laser is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.
| Applicant: | Rohrer Aesthetics, LLC |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Rohrer Aesthetics, LLC<br>105 Citation Court<br>Birmingham, AL 35209 |
| Contact Person: | Mr. Mark Rohrer |
| Telephone: | 205-356-1172 - phone<br>mrohrer@rohreraesthetics.com |
| Preparation Date: | September 15, 2019 |
| Device Trade Name: | PicoLazer Laser System |
| Common Name: | Surgical Powered Lasers and Delivery Devices/Hand piece<br>Accessories |
| Regulation Name: | Laser surgical instrument for use in general and<br>plastic surgery and dermatology |
| Regulation Number: | 21 CFR 878.4810 (Product Code: GEX) |
| Legally Marketed Predicate Device1: | PicoCare Family manufactured by Won Tech Co Ltd |
| 510(K) number: | K181272 |
| Legally Marketed Predicate Device 2: | PicoWay Laser System manufactured by Syneron Candela<br>Corporation |
| 510(K) number: | K170597 |
| Regulatory Class: | Class II Prescription Use |
| Description of the PicoLazer Multi-<br>wavelength Laser: | The Rohrer Aesthetics, LLC PicoLazer™ laser system is a<br>dual-wavelength (532 and 1064 nm) Nd:YAG laser system<br>that offers a 450-picosecond pulse duration that produces a<br>maximum energy of 500mJ at 1064nm and 250mJ at<br>532nm.<br><br>The PicoLazer laser system consists of a system console,<br>an articulated arm, a laser handpiece, a footswitch, and a<br>remote interlock plug. Other components necessary for<br>operation, such as power cables, are also included. |
{4}------------------------------------------------
# 510(K) Summary PicoLazer Multi-wavelength Laser
| Indication for Use: | The PicoLazer system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery. |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 1064nm:<br>The 1064nm wavelength of the PicoLazer laser system is indicated for: |
| | <ul><li>treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-VI.</li><li>tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)</li></ul> |
| | 532nm:<br>The 532nm wavelength of the PicoLazer laser system is indicated for: |
| | <ul><li>treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III</li><li>tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III</li></ul> |
| Performance Data: | The following performance data was provided in support of the substantial equivalence determination: |
| | IEC 60601-1 Test for Medical Electrical equipment was performed for General Requirements for basic safety and essential performance; |
| | IEC 60601-1-2 Test for Medical Equipment for General Requirements for basic safety and essential performance: electromagnetic compatibility |
| | IEC 60601-2-22 Medical Electrical Equipment-Part 2-22:<br>Requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser Equipment |
| | IEC 60825-1 Safety of laser products - Part 1: Equipment classification, and requirements [Including: Technical Corrigendum 1 (2008), Interpretation Sheet 1 (2007), Interpretation Sheet 2 (2007)] |
| Results of Clinical Study: | A human clinical study was not required as the device is substantially equivalent to the predicate devices. |
{5}------------------------------------------------
# 510(K) Summary PicoLazer Multi-wavelength Laser
Technical Specification Comparison:
| Characteristic | PicoLazer<br>(K192583) | | PicoWay Laser<br>System (K170597) | | PicoCare Family<br>(K181272) | |
|-------------------------------|------------------------|-------|-----------------------------------|--------|------------------------------|-------|
| Wavelength (nm) | 1064nm | 532nm | 1064nm | 532nm | 1064nm | 532nm |
| Laser Type | Nd:YAG | | Nd:YAG | | Nd:YAG | |
| Max Energy (mJ) | 500mJ | 250mJ | 800mJ | 300mJ | 600mJ | 300mJ |
| Peak Power | 1.3GW | 0.8GW | 0.9GW | 0.53GW | 1.33GW | 0.8GW |
| Spot Size | Max 10mm | | Max 10mm | | Max 10mm | |
| Pulse Duration | 450ps | 375ps | 450ps | 375ps | 450ps | |
| Pulse Repetition<br>rate (Hz) | 1~10Hz | | 1~10Hz | | 1~10Hz | |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Physical Specification:
| aracteristic | PicoLazer<br>(K192583) | PicoWay Laser<br>System (K170597) | PicoCare Family<br>(K181272) |
|------------------|----------------------------------------|-----------------------------------|----------------------------------------|
| Console Weight | 176lbs | 275lbs | 198lbs |
| Console Size | 25(W)x26(D)x48(H)in | 18(W)x27(D)x42(H)in | 18(W)x37(D)x36(H)in |
| Voltage, Current | 100-120 VAC, 20 A<br>200-240 VAC, 15 A | 200-240 VAC, 30A | 100-120 VAC, 20 A<br>220-230 VAC, 15 A |
| Frequency | 50/60Hz | 50/60Hz | 50/60Hz |
Indication for Use Comparison:
| | | PicoLazer<br>(K192583) | PicoWay Laser System<br>(K170597) | PicoCare Family (K181272) |
|--------------|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| 1064nm Laser | | Treatment of benign<br>pigmented lesions on<br>patients with Fitzpatrick<br>skin types I-VI | Treatment of benign pigmented<br>lesions on patients with<br>Fitzpatrick skin types I-IV | Tattoo removal for dark<br>colored tattoo inks and for<br>multicolored tattoos<br>containing dark colored<br>tattoo inks on patients with all |
| | | Tattoo removal for dark<br>colored tattoo inks and<br>for multicolored tattoos<br>containing dark colored<br>tattoo inks on patients<br>with all skin types<br>(Fitzpatrick I-VI) | Removal of tattoos for all skin<br>types (Fitzpatrick I-VI) to treat the<br>following colors: black, brown,<br>green, blue, and purple. | skin types (Fitzpatrick I-VI) |
| 532nm Laser | | Treatment of benign<br>pigmented lesions on<br>patients with Fitzpatrick<br>skin types I-III | Treatment of benign pigmented<br>lesions on patients with<br>Fitzpatrick skin types I-IV | Tattoo removal for lighter<br>colored tattoo inks, including<br>red and yellow inks, on<br>patients with Fitzpatrick skin<br>types I-III |
| | | Tattoo removal for<br>lighter colored tattoo<br>inks, including red and<br>yellow inks, on patients<br>with Fitzpatrick skin<br>types I-III | Removal of tattoos for Fitzpatrick<br>skin types I-III to treat the<br>following colors: red, yellow, and<br>orange. | |
{6}------------------------------------------------
## 510(K) Summary PicoLazer Multi-wavelength Laser
Conclusion:
The PicoLaser Laser System's intended use, indications for use and technical specifications are substantially equivalent to the PicoWay Laser System and the PicoCare Family of Lasers.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.