K192552 · Chongqing Peninsula Medical Technology Co., Ltd. · OAP · Apr 17, 2020 · Physical Medicine
Device Facts
Record ID
K192552
Device Name
Irradiation Cosmetic Device
Applicant
Chongqing Peninsula Medical Technology Co., Ltd.
Product Code
OAP · Physical Medicine
Decision Date
Apr 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Irradiation Cosmetic Device (Model: HairPro Plus) is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications of I - II and males with androgenetic alopecia who have Norwood-Hamilton Classifications of IIa - V and for both, Fitzpatrick Classification of Skin Phototypes I to IV.
Device Story
Irradiation Cosmetic Device (HairPro Plus) is a wearable cap containing 272 laser diodes (650nm) for scalp treatment. Device operates via diode lasers to stimulate hair growth without manual movement. Intended for home use by patients. Features include 30-minute automatic timer; built-in IR sensor detects if cap is worn, suspending laser output and shutting down after 10 minutes if removed. Indicator light blinks based on laser frequency. Healthcare providers use device to treat androgenetic alopecia; patient benefits include non-invasive hair growth promotion.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and performance evaluated via IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-1-11, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (sensitization/irritation).
Technological Characteristics
Laser-based hair growth device; 272 laser diodes; 650nm wavelength; 5mW per diode; Class 3R laser product. Dimensions: 199mm x 179mm x 88mm. Weight: 1100g. Biocompatible materials per ISO 10993-5/10. Electrical safety per IEC 60601-1/1-2/1-11. Powered by internal electronics with automatic 30-minute timer and IR-based safety shut-off.
Indications for Use
Indicated for females with androgenetic alopecia (Ludwig-Savin I-II) and males with androgenetic alopecia (Norwood-Hamilton IIa-V), Fitzpatrick Skin Phototypes I-IV, to promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and consists of the letters "FDA" stacked on top of the words "U.S. FOOD & DRUG ADMINISTRATION".
Chongqing Peninsula Medical Technology Co., Ltd. % Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District Guangzhou, 51006 China
## Re: K192552
Trade/Device Name: Irradiation Cosmetic Device Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: February 28, 2020 Received: March 11, 2020
### Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192552
Device Name Irradiation Cosmetic Device
### Indications for Use (Describe)
The Irradiation Cosmetic Device (Model: HairPro Plus) is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications of I - II and males with androgenetic alopecia who have Norwood-Hamilton Classifications of IIa - V and for both, Fitzpatrick Classification of Skin Phototypes I to IV
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
### 1. Submitter Information
Sponsor Name: Chongqing Peninsula Medical Technology Co., Ltd. Address: No.2-4, 2F Block I, Biological and Biochemical Park, 70-1&70-2# Ke Yuan 4th Road, Jiulongpo District, Chongqing, China Contact name: Zhang Sudi (RA) Tel: +86-150 1257 8550 Fax: 023-6861 1963 E-mail: zsd@peninsulalaser.com
### Application Correspondent:
Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China Tel: +86 20 8266 2446 Email: regulatory@glomed-info.com
#### Subject Device Information 2.
| Type of 510(k): | Traditional |
|----------------------|-----------------------------|
| Trade Name: | Irradiation Cosmetic Device |
| Model Name: | HairPro Plus |
| Classification Name: | Infrared Lamp |
| Review Panel: | General & Plastic Surgery |
| Product Code: | OAP |
| Regulation Number: | 890.5500 |
| Regulation Class: | 2 |
#### Predicate Device (1) Information 3.
Sponsor: Chongqing Peninsula Medical Technology Co., Ltd.
Classification Name: Infrared Lamp
{4}------------------------------------------------
Sponsor: Chongqing Peninsula Medical Technology Co., Ltd Subject Device: Irradiation Cosmetic Device, Model: HairPro Plus File No.: 510(k) summary (V1.0)
Trade Name: Irradiation Aesthetic Device Model name: HairPro 510(K) Number: K171835 Review Panel: General & Plastic Surgery Product Code: OAP Requlation Number: 890.5500 Requlation Class: 2
### Predicate Device (2) Information:
Sponsor: Capillus LLC Classification Name: Infrared Lamp Trade Name: Capillus272 510(K) Number: K153618 Review Panel: General & Plastic Surgery Product Code: OAP Requlation Number: 890.5500 Requlation Class: 2
#### 4. Device Description
Irradiation Cosmetic Device (Model: HairPro Plus) consists of laser diodes that are spread throughout the cap. The device uses diode lasers to cover the entire area of the head that is normally covered with hair, and this unique design allows the treatment of the entire scalp without manual movement. In the process of the cap working, the indicator blinks red based on the working frequency of the laser diodes. Built-in timing function, the cap can record the time per treatment, and the cap can stop working automatically after each 30 minutes treatment. When the built-in IR detected the cap is not worn on the head, the laser light output will be automatically suspended immediately and the cap will automatically shut down after 10 minutes.
#### 5. Intended Use
The Irradiation Cosmetic Device (Model: HairPro Plus) is indicated to promote hair growth in females with androqenetic alopecia who have Ludwig-Savin Classifications of I - II and males with androgenetic alopecia who have Norwood-Hamilton Classifications of Ila – V and for both, Fitzpatrick Classification of Skin Phototypes I to IV.
#### 6. Test Summary
{5}------------------------------------------------
| Sponsor: | Chongqing Peninsula Medical Technology Co., Ltd |
|-----------------|--------------------------------------------------|
| Subject Device: | Irradiation Cosmetic Device, Model: HairPro Plus |
| File No.: | 510(k) summary (V1.0) |
Irradiation Cosmetic Device (Model: HairPro Plus) has been evaluated the safety and performance by lab bench testing according to the following standards:
| Standards No. | Standard Title | Version | Date |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|------------|
| IEC 60601-1 | Medical Electrical Equipment -<br>Part 1: General Requirements<br>for Safety | 2005+A1:2012 | 07/09/2014 |
| IEC 60601-1-2 | Medical electrical equipment -<br>Part 1-2: General requirements<br>for basic safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br>compatibility - Requirements<br>and tests | 2014 | 09/17/2018 |
| IEC 60825-1 | Safety of laser products - Part<br>1:<br>Equipment classification and<br>requirements | Second<br>edition:03/2007 | 07/09/2014 |
| IEC60601-1-11 | Medical Electrical Equipment -<br>Part 1-11: General<br>Requirements for Basic Safety<br>and Essential Performance -<br>Collateral Standard:<br>Requirements for Medical<br>Electrical Equipment and<br>Medical Electrical Systems<br>Used | 2015-1 | 06/27/2016 |
| ISO 10993-5 (Cytotoxicity) | Biological evaluation of<br>medical devices - Part 5: Tests<br>for In Vitro cytotoxicity | 2009/(R)2014 | 12/23/2016 |
| ISO 10993-10<br>(Sensitization and<br>Irritation) | Biological evaluation of<br>medical devices - Part 10:<br>Tests for irritation and skin<br>sensitization | 2010/(R)2014 | 07/26/2016 |
### Comparison to Predicate Device 7.
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness.
| Elements of Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Verdict |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Company | Chongqing Peninsula Medical Technology Co., Ltd. | Chongqing Peninsula Medical Technology Co., Ltd. | Capillus LLC | SE |
| Trade Name | Irradiation Cosmetic Device<br>(model:HairPro Plus) | Irradiation Aesthetic Device<br>HairPro | Capillus272 Pro,<br>Capillus272 OfficePro,<br>Capillus82,<br>Capillus202 | SE |
| Elements<br>of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Verdict |
| Classification<br>Name | Infrared Lamp | Infrared Lamp | Infrared Lamp | SE |
| 510(k) Number | Applying | K171835 | K160285 | -- |
| Product Code | OAP | OAP | OAP | SE |
| Intended Use /<br>Indications for<br>Use | The Irradiation<br>Cosmetic Device<br>(Model: HairPro Plus)<br>is indicated to promote<br>hair growth in females<br>with androgenetic<br>alopecia who have<br>Ludwig-Savin<br>Classifications of I - II<br>and males with<br>androgenetic alopecia<br>who have Norwood-<br>Hamilton<br>Classifications of Ila -<br>V and for both,<br>Fitzpatrick<br>Classification of Skin<br>Phototypes I to IV. | The Irradiation<br>Cosmetic Device<br>(Model: HairPro) is<br>indicated to promote<br>hair growth in females<br>with androgenetic<br>alopecia who have<br>Ludwig-Savin<br>Classifications of I - II<br>and males with<br>androgenetic alopecia<br>who have Norwood-<br>Hamilton<br>Classifications of Ila -<br>V and for both,<br>Fitzpatrick<br>Classification of Skin<br>Phototypes I to IV. | The Capillus272<br>Pro, Capillus272<br>OfficePro.<br>Capillus82, and<br>Capillus202 are<br>intended for the<br>promotion of hair<br>growth in females<br>with androgenic<br>alopecia who have<br>Ludwig-Savin<br>Classifications I- II,<br>and in males with<br>androgenic alopecia<br>who have Norwood<br>Hamilton<br>Classifications Ila-<br>V ; and both<br>genders having<br>Fitzpatrick<br>Classification of<br>Skin Phototypes I to<br>IV. | SE |
| Waveform | Visible red laser | Visible red laser | Visible red laser | SE |
| Wavelength | 650nm±5nm | 650nm±5nm | 650 | SE |
| Amounts of Laser<br>Lamp | 272 | 81 | Capillus272 Pro:<br>272<br>Capillus202: 202<br>Capillus82: 82 | SE |
| Energy of per<br>Laser Lamp | 5mW ±10% | 5mW ±10% | <5mW | SE |
| Classification<br>according to<br>IEC60825-1 | Class 3R | Class 3R | Class 3R | SE |
| Treatment Time | Each Treatment: 30<br>min<br>Total Treatment: 3<br>times per week | Each Treatment: 30<br>min<br>Total Treatment: 3<br>times per week | Each Treatment: 30<br>min<br>Total Treatment:<br>every other day, for<br>17 weeks. | SE |
| Treatment Area | 495.37 cm²<br>Mathematically Max.<br>derived | 202.3 cm²<br>Mathematically Max.<br>derived | Capillus272: 495.37<br>cm2<br>Capillus202: 449.51<br>cm2<br>Capillus82: 194.42<br>cm2 | SE |
| Elements<br>of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Verdict |
| | | | Mathematically<br>Max. derived | |
| Irradiance (power<br>per area) | 2.7454 mW/cm²<br>Mathematically Max.<br>derived | 2.2022 mW/cm²<br>Mathematically Max.<br>derived | Capillus272: 2.7454<br>mW/cm²<br>Capillus202: 2.2469<br>mW/cm²<br>Capillus82: 2.1088<br>mW/cm²<br>Mathematically<br>Max. derived | SE |
| Fluence | 4.9417 J/cm²<br>Mathematically Max.<br>derived | 3.9639 J/cm²<br>Mathematically Max.<br>derived | Capillus272: 4.9417<br>J/cm²<br>Capillus202: 4.044<br>J/cm²<br>Capillus82: 3.7920<br>J/cm²<br>Mathematically<br>Max. derived | SE |
| Dimension | 199mm*179mm*88m<br>m (L x W x H) | 199mm*179mm*88m<br>m (L x W x H) | -- | SE |
| Life Expectancy | 5 years | 5 years | -- | SE |
| Weight | 1100g | 1000g | -- | SE |
| Environment for<br>Operation | Temperature:<br>5°C~30°C<br>Humidity: ≤ 80%<br>Atmosphere range:<br>700hPa-1060hPa | Temperature:<br>5°C~30°C<br>Humidity: ≤ 80%<br>Atmosphere range:<br>700hPa-1060hPa | -- | SE |
| Environment for<br>Storage | Temperature:<br>0°C~50°C<br>Humidity: ≤ 85%<br>Atmosphere range:<br>50kPa-110k | Temperature:<br>0°C~50°C<br>Humidity: ≤ 85%<br>Atmosphere range:<br>50kPa-110k | -- | SE |
| Safety Feature | Complied with IEC<br>60601-1 and IEC<br>60601-1-2 | Complied with IEC<br>60601-1 and IEC<br>60601-1-2 | Complied with IEC<br>60601-1 and IEC<br>60601-1-2 | SE |
| Biocompatibility<br>Feature | All patient contacting<br>materials are complied<br>with ISO 10993-5, ISO<br>10993-10 | All patient contacting<br>materials are complied<br>with ISO 10993-5, ISO<br>10993-10 | All patient<br>contacting materials<br>are complied with<br>ISO 10993-5, ISO<br>10993-10 | SE |
{6}------------------------------------------------
Chongqing Peninsula Medical Technology Co., Ltd Sponsor: Subject Device: Irradiation Cosmetic Device, Model: HairPro Plus File No.: 510(k) summary (V1.0)
{7}------------------------------------------------
Sponsor: Chongqing Peninsula Medical Technology Co., Ltd Subject Device: Irradiation Cosmetic Device, Model: HairPro Plus File No.: 510(k) summary (V1.0)
### Final Conclusion
The subject device Irradiation Cosmetic Device (Model: HairPro Plus) has all features of the predicate device. The few differences do not affect the safety and effectiveness of the subject device. Thus, the subject device is substantially equivalent to the predicate device.
#### 8. Summary Prepared Date 11 April 2020
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.